0
登录后你可以
  • 下载海量资料
  • 学习在线课程
  • 观看技术视频
  • 写文章/发帖/加入社区
创作中心
发布
  • 发文章

  • 发资料

  • 发帖

  • 提问

  • 发视频

创作活动
778204-01

778204-01

  • 厂商:

    NI

  • 封装:

  • 描述:

    778204-01 - S Series Multifunction DAQ 12 or 16-Bit,1 to 10 MS/s, 4 Analog Inputs - National Instrum...

  • 数据手册
  • 价格&库存
778204-01 数据手册
S Series Multifunction DAQ 12 or 16-Bit, 1 to 10 MS/s, 4 Analog Inputs NI 6120, NI 6115, NI 6110, NI 6111 • 2 or 4 analog inputs; dedicated A/D converter per channel • 1 to 10 MS/s per channel maximum sample rate • Analog and digital triggering • AC or DC coupling • 8 input ranges from ±200 mV to ±42 V • 2 analog outputs at 4 MS/s single channel or 2.5 MS/s dual channel • 8 digital I/O lines (5 V TTL/CMOS) • Two 24-bit counter/timers • Measurement services that simplify configuration and measurements Recommended Software • LabVIEW 7.x or higher • LabWindows/CVI 7.x or higher • Measurement Studio 7.x or higher • Digital Waveform Editor • SignalExpress 1.x or higher Other Compatible Software • VI Logger 2.x or higher • Visual Studio .NET • Visual Basic, C/C++, and C# Measurement Services Software (included)1 • NI-DAQmx driver software • Measurement & Automation Explorer configuration utility • VI Logger Lite data-logging software 1Mac Operating Systems • Windows 2000/NT/XP • Mac OS X • Linux OS X and Linux applications must use NI-DAQmx Base driver software. Calibration Certificate Available Analog Inputs 4 4 4 2 Input Resolution (bits) 16 12 12 12 Sampling Rate (MS/s) 13 10 5 5 Input Range (V) ±0.2 to ±42 ±0.2 to ±42 ±0.2 to ±42 ±0.2 to ±42 Analog Outputs 2 2 2 2 Max Output Rate (MS/s) 41 41 41 41 Output Range (V) ±10 ±10 ±10 ±10 Digital I/O 82 82 8 8 Counter/ Timers 2, 24-bit 2, 24-bit 2, 24-bit 2, 24-bit Family NI 6120 NI 6115 NI 6110 NI 6111 14 Bus PCI, PXI PCI, PXI PCI PCI Triggers Analog, digital Analog, digital Analog, digital Analog, digital MS/s single channel; 2.5 MS/s on two channels 2Hardware-timed up to 10 MB/s 3800 kS/s with NI-DAQmx, 1 MS/s with additional download. Special conditions apply. Table 1. Channel, Speed, and Resolution Specifications Overview and Applications National Instruments S Series devices combine the latest in PC technologies to deliver simultaneous-sampling, high-bandwidth capabilities for low and high-channel-count data acquisition and control systems. Use S Series devices in a variety of applications including: • Stimulus/response applications where synchronization among highspeed analog input (AI), analog output (AO), and digital I/O is required • 42 V automotive applications • Radar, sonar, ultrasound • Transient/ballistics measurements • High-energy physics • IF digitization Features The NI high-performance S Series DAQ devices have the same basic features of a typical multifunction device – eight digital I/O lines, two counter/timers, two analog outputs, RTSI or PXI trigger bus for multidevice synchronization, and PFI lines for software-controlled routing of timing and triggering signals. S Series devices exceed the analog input performance of traditional multiplexed devices with features including: Family NI 6120 NI 6115 NI 6110 NI 6111 1Filter Onboard Memory (samples) Antialias Filters1 64 M 5-pole Bessel, 100 kHz 32 M or 64 M 3-pole Bessel 50 and 500 kHz 8,192 – 8,192 – Signal Coupling AC and DC AC and DC DC DC Input Configuration Pseudodifferential Pseudodifferential Pseudodifferential Pseudodifferential DIO Hardware-timed Hardware-timed Software-timed Software-timed can be bypassed. Table 2. Special Features S Series Multifunction DAQ 12 or 16-Bit, 1 to 10 MS/s, 4 Analog Inputs Dedicated A/D Converters per Channel – help you achieve faster sampling rates per channel, simultaneous sampling, and better dynamic accuracy for transient or frequency measurements. 200 mV to 42 V Input Ranges – configure each analog input channel to read from a very wide variety of voltage ranges. The input impedance is lower for 20 and 42 V ranges, preserving the bandwidth for signals with high frequencies and large amplitudes. AC or DC Coupling – configure analog inputs for AC or DC coupling on a per-channel basis. AC coupling removes the DC offset for applications only analyzing signals in the frequency domain. Antialias Filters (NI 6115 and NI 6120) – the antialias filters are turned on from software on a per-channel basis. NI 6115 devices have 50 or 500 kHz filters for each channel, and NI 6120 devices have 100 kHz filters for each channel. Large Onboard Memory (NI 6115 and NI 6120) – NI 6120 devices feature a 64 MS onboard memory, and NI 6115 devices feature a 64 or 32 MS onboard memory. With the extra memory, which requires no extra programming, you can capture high-bandwidth signals over long periods of time, even if multiple devices are sharing the same PCI or PXI bus. Hardware-Timed Digital I/O – the digital I/O lines can synchronize with the analog input, analog output, counter/timer I/O, or an external clock for mixed-signal measurement applications. Using the onboard FIFO, you can achieve reliable data transfers to 10 MB/s. Phase Locking (PXI only) – synchronize multiple devices to a common PXI backplane clock to preserve the phase between measurements across multiple device analog input channels. Figure 1. Sound and Vibration Toolkit NI LabVIEW Sound and Vibration Toolkit This toolkit includes a full set of frequency-domain analysis and presentation tools required for sound and vibration applications. These tools include: • Distortion, gain, phase, cross-talk, and dynamic range measurements • Octave analysis • Swept-sine analysis • Linear, exponential, and peak hold measurements • Baseband and zoom FFT, power, FFT, power spectral density (PSD) • Limit testing • Short-time Fourier transform and shock response spectrum • Weighting filters • System calibration • Waterfall/colormap display NI LabVIEW Signal Processing Toolkit This toolkit is a suite of software tools, example programs, and utilities for time-frequency analysis and digital filter design. The toolkit includes: • Wavelet and filter-bank design • Digital filter design of FIR and IIR filters • Joint time-frequency analysis (JTFA) with the Gabor spectrograph • High-frequency resolution model-based spectral estimation for small datasets Professional Analysis and Presentation Toolkits There are several analysis and presentation add-on toolkits available from National Instruments that help make you more productive when developing custom applications with your S Series devices. Turn your S Series device into a transient recorder, spectrum analyzer, or one of several other instruments with these toolkits. 2 National Instruments • Tel: (800) 813 3693 • info@ni.com • ni.com S Series Multifunction DAQ 12 or 16-Bit, 1 to 10 MS/s, 4 Analog Inputs NI Spectral Measurements Toolkit This toolkit includes all of the analysis and presentation tools to convert your S Series device into a full-featured spectrum analyzer. The toolkit includes: • Measurements such as in-band power and adjacent-channel power • 3D spectrogram • Analog modulation and demodulation • I-Q data for digital demodulation • Highly optimized FFT processing SCXI is limited to 100 kS/s for 16-bit accuracy and 333 kS/s for 12-bit accuracy when used in multiplexed mode. However, in parallel mode, the sample rate is limited by the maximum scan rate of the data acquisition device. When SCXI is used in parallel mode with S Series, all channels are simultaneously sampled and the sampling rate is limited by the S Series device, from 250 kS/s per channel to 10 MS/s per channel. Measurement Services Software National Instruments measurement services software, built around NI-DAQmx driver software, includes intuitive application programming interfaces, configuration tools, I/O assistants, and other tools designed to reduce system setup, configuration, and development time. Automatic Code Generation – DAQ Assistant is an interactive guide that helps you navigate through configuring, testing, and programming measurement tasks and automatically generates the necessary code for LabVIEW, LabWindows/CVI, or Measurement Studio. Cleaner Code Development – Basic and advanced software functions have been combined into one easy-to-use yet powerful set to help you build cleaner code and move from basic to advanced applications without replacing functions. High-Performance Driver Engine – NI-DAQmx delivers maximum I/O system throughput with a multithreaded driver. Test Panels – With the Measurement & Automation Explorer configuration utility, you can test all of your module functionality before you begin development. Scaled Channels – Easily scale your voltage data into the proper engineering units using the NI-DAQmx Measurement Ready virtual channels by choosing from a list of common sensors and signals or creating your own custom scale. Data-Logging Software – VI Logger Lite is configuration-based software designed specifically for data logging. Features include easy logging and viewing of data, data extraction to Microsoft Excel, and code generation in LabVIEW. LabVIEW Integration – All NI-DAQmx functions create the waveform data type, which carries acquired data and timing information directly into more than 400 LabVIEW built-in analysis routines for display of results in engineering units on a graph. High-Performance Data Throughput Using NI-DAQmx driver software and S Series, you can take full advantage of your PC’s throughput potential, which is useful in many applications including transient recording. Flexible Data Transfer With NI-DAQmx, you can choose where to accumulate data – on the onboard memory, PC RAM memory, or PC hard drive memory. Thus, you can balance your PC application requirements and stream-to-disk/memory capabilities. In any case, NI-DAQmx optimizes data transfers by using DMA (direct memory access) and transferring large chunks of data, while ensuring no data loss. You can also choose the format of your data, further optimizing throughput. Choices include raw (binary), scaled, or the high-level waveform data type. Using these techniques, you can achieve rates of more than 40 MS/s sustained over the PCI/PXI bus to hard drive (depending on the PC). Data Compaction With the release of NI-DAQmx 7.4 or later, data compaction is available for stream-to-disk applications. The NI-DAQmx compaction feature strips unused or unwanted bits of data from each sample taken from your S Series device before it is permanently recorded to hard disk. This is helpful because hard drives are typically the throughput bottleneck compared to the PCI/PXI bus and PC RAM memory read/write speed. Using data compression, users with standard hard drives can see up to 20 percent higher throughput when streaming-to-disk. Signal Conditioning With the release of NI-DAQmx 7.4 or later, all S Series devices are compatible with parallel-mode SCXI modules. Model SCXI-1520 SCXI-1125 SCXI-1141/2/3 Description Universal Strain Gage Input Isolation Amplifier 8-Pole Lowpass Filters Channels 8 8 8 Table 3. Most Popular Parallel-Mode SCXI Modules NI-DAQmx Base Driver NI-DAQmx Base (available at ni.com/downloads) offers Mac OS X and Linux users a programming interface similar to NI-DAQmx. It includes ready-to-use LabVIEW VIs and C function features similar to those included in NI-DAQmx driver software. National Instruments • Tel: (800) 813 3693 • info@ni.com • ni.com 3 S Series Multifunction DAQ 12 or 16-Bit, 1 to 10 MS/s, 4 Analog Inputs CH0+ CH0– CH1+ CH1– CH2+ CH2– CH3+ CH3– AI CH3 Mux AI CH2 Mux AI CH1 Mux AI CH0 Mux + CH0 Amplifier – + CH1 Amplifier – + CH2 Amplifier – + CH3 Amplifier – AntiAliasing Filter CH0 12-Bit ADC 12 CH0 Latch Data (16) AntiAliasing Filter CH1 12-Bit ADC 12 CH1 Latch Data (16) AntiAliasing Filter CH2 12-Bit ADC 12 CH2 Latch Data (16) Control ADC FIFO Data (32) Generic Mini PCI Bus Bus MITE Interface Interface Address/Data AntiAliasing Filter CH3 12-Bit ADC 12 CH3 Latch Data (16) I/O Connector Calibration Mux 2 Analog Trigger Circuitry Trigger Counter/ Timing I/O Analog Input Timing/Control AI Control EEPROM IRQ DMA Trigger Trigger Level DACs PFI / Trigger DMA/IRQ Bus Interface RTSI Bus Interface Analog Input Control DAQ-STC Bus Interface Analog Output Control EEPROM Control FPGA DIO Control DMA Interface DIO FIFO I/O Bus Interface Timing STC DIO (8) Digital I/O (8) DIO Mux DIO FIO DAQ - STC Analog Output Timing/Control Digital I/O AO Control FPGA Digital I/O (8) DAC0 Data (12) DAC FIFO Data (32) DAC1 RTSI Bus Calibration DACs Figure 2. NI 6120 and NI 6115 Hardware Block Diagram 4 National Instruments • Tel: (800) 813 3693 • info@ni.com • ni.com PXI/CompactPCI Bus S Series Multifunction DAQ 12 or 16-Bit, 1 to 10 MS/s, 4 Analog Inputs Voltage REF Calibration DACs Gain Amplifier ACH0+ ACH0ACH1+ ACH1ACH2+ ACH2ACH3+ ACH31 AI CH0 Mux AI CH1 Mux AI CH21 Mux AI CH31 Mux Not available on PCI-6111 + - CH0 12-Bit ADC CH1 12-Bit ADC CH2 12-Bit ADC CH3 12-Bit ADC ADC FIFO Generic Bus Interface MITE PCI Bus Interface + - + - Control + - Address/Data I/O Connector AI Control Data (32) Calibration Mux Trigger Level DACs Trigger PFI / Trigger EEPROM Address (5) Analog Trigger Circuitry Trigger Counter/ Timing I/O Digital I/O Analog Input Timing/Control DMA/ Interrupt Request Bus Interface RTSI Bus Interface IRQ DMA Analog Input Control DAQ-STC Bus Interface Analog Output Control EEPROM Control DMA Interface Timing DAQ - STC Analog Output Timing/Control MIO Interface I/O Bus Interface Digital I/O (8) AO Control DAC0 DAC1 DAC FIFO Data (32) Calibration DACs RTSI Bus Figure 3. NI 6110 and 6111 Hardware Block Diagram Ordering Information PCI PXI NI PCI-6120 ..................................................................778397-01 NI PCI-6115 with 32 MS ........................................................................778534-01 64 MS ........................................................................778554-01 NI PCI-6111 ..................................................................777530-01 NI PCI-6110 ..................................................................777475-01 NI PXI-6115 with 32 MS ........................................................................778194-01 64 MS ........................................................................778204-01 NI PXI-6120 ..................................................................778396-01 Includes data acquisition driver software. BUY NOW! For complete product specifications, pricing, and accessory information, call (800) 813 3693 (U.S. only) or go to ni.com/dataacquisition. PCI Bus 5 National Instruments • Tel: (800) 813 3693 • info@ni.com • ni.com S Series Multifunction DAQ 12 or 16-Bit, 1 to 10 MS/s, 4 Analog Inputs Specifications These specifications are typical for 25 °C unless otherwise noted. Timing I/O Number of channels Up/down counter/timers ........................... Frequency scaler........................................ Resolution Up/down counter/timers ........................... Frequency scaler........................................ Compatibility .................................................... Base clocks available Up/down counter/timers ........................... Frequency scaler........................................ Base clock accuracy ......................................... Maximum source frequency Up/down counter/timers ........................... Minimum source pulse duration ...................... Minimum gate pulse duration.......................... Data transfers .................................................. 2 1 24 bits 4 bits TTL/CMOS 20 MHz, 100 kHz 10 MHz, 100 kHz ±0.01% 20 MHz 10 ns, edge-detect mode 10 ns, edge-detect mode DMA (scatter-gather), interrupts, programmed I/O Analog Input Input Characteristics Number of channels NI 6110, NI 6115, NI 6120 ........................ 4 pseudodifferential NI 6111 ...................................................... 2 pseudodifferential ADC resolution NI 6110, NI 6111, NI 6115 ........................ 12 bits, 1 in 4,096 NI 6120 ...................................................... 16 bits, 1 in 65,536 ADC pipeline NI-6110, NI 6111 ....................................... 3 NI 6115 ...................................................... 4 NI 6120 ...................................................... 0 Maximum sampling rate NI 6110, NI 6111 ....................................... 5 MS/s NI 6115 ...................................................... 10 MS/s NI 6120 ...................................................... 1 MS/s1 Minimum sampling rate NI 6110, NI 6111 ....................................... 1 kS/s NI 6115 ...................................................... 20 kS/s NI 6120 ...................................................... No minimum Input signal ranges (selectable by channel).............................. ±42, ±20, ±10, ±5, ±2, ±1, ±0.5, ±0.2 V Input coupling................................................... AC or DC Max working voltage for all analog input channels Overvoltage protection (ACH+, ACH-) ............................................. ±42 V Input FIFO buffer NI 6110, NI 6111 ....................................... 8,192 samples NI 6115, NI 6120 ....................................... 16 Msamples or 32 Msamples Data transfers .................................................. DMA (scatter-gather), interrupts, programmed I/O Analog filter type NI 6115 ...................................................... 3-pole Bessel NI 6120 ...................................................... 5-pole Bessel Analog filter frequency (-3 dB) NI 6115 ...................................................... 50 and 500 kHz NI 6120 ...................................................... 100 kHz 1800 Triggers Analog Trigger Purpose Analog input .............................................. Analog output ............................................ General-purpose counter/timers ............... Source............................................................... Level Internal source, ACH ....................... External source, PFI0/TRIG1...................... Slope................................................................. Resolution NI 6110, NI 6111, NI 6115 ........................ NI 6120 ...................................................... Hysteresis......................................................... Bandwidth (-3 dB) Internal source, ACH ....................... External source, PFI0/TRIG1...................... Start and stop trigger, gate, clock Start trigger, gate, clock Source, gate All analog input channels, PFI0/TRIG1 ±Full-scale ±10 V Positive or negative; software-selectable 8 bits, 1 in 256 12 bits, 1 in 4,096 Programmable 5 MHz 5 MHz Digital Trigger Purpose Analog input .............................................. Analog output ............................................ General-purpose counter/timers ............... External Sources .............................................. Compatibility .................................................... Response .......................................................... Pulse width....................................................... Start and stop trigger, gate, clock Start trigger, gate, clock Source, gate PFI, RTSI 5 V/TTL Rising or falling edge 10 ns minimum kS/s with NI-DAQmx, 1 MS/s with additional download. Special conditions apply. Analog Output Output Characteristics Number of channels......................................... Resolution NI 6110, NI 6111, NI 6120 ........................ NI 6115 ...................................................... Max update rate 1 channel .................................................. 2 channel ................................................... Output FIFO buffer size NI 6110, NI 6111 ....................................... NI 6115, NI 6120 ....................................... Data transfers .................................................. 2 voltage 16 bits, 1 in 65,536 12 bits, 1 in 4,096 4 MS/s 2.5 MS/s 2,048 samples 16 or 32 Msamples DMA (scatter-gather), interrupts, programmed I/O PXI Trigger Bus (PXI only) Trigger lines...................................................... 7 RTSI Bus (PCI only) Trigger lines...................................................... 7 Bus Interface PXI or PCI.......................................................... Master, slave Voltage Output Ranges.............................................................. ±10 V Output coupling ................................................ DC Physical Dimensions (not including connectors) PCI ............................................................ 31.2 by 10.6 cm (12.3 by 4.2 in.) PXI ............................................................ 16 by 10 cm (6.3 by 3.9 in.) I/O connector.................................................... 68-pin male SCSI II type Digital I/O Number of channels......................................... Compatibility .................................................... Power-on state ................................................. Data transfers NI 6110, NI 6111 ....................................... NI 6115, NI 6120 ....................................... Transfer rate NI 6115, NI 6120 ....................................... Input buffer size NI 6115, NI 6120 ....................................... Output buffer size NI 6115, NI 6120 ....................................... 8 input/output TTL/CMOS Input (high-impedence) Programmed I/O DMA (scatter-gather), interrupts, programmed I/O 10 Mbytes/s 2,048 bytes 2,048 bytes 6 National Instruments • Tel: (800) 813 3693 • info@ni.com • ni.com S Series Multifunction DAQ Cables and Accessories Recommended Configurations I/O Configuration BNC shielded Front-mount shielded screw terminals (PXI only) Noise-rejecting shielded screw terminals Low-cost unshielded screw terminals Accessory BNC-2110 (777643-01) TB-2705 (778241-01) SCB-68 (776844-01) CB-68LPR (777145-02) Cable SH68-68-EP (184749-01) – SH68-68-EP (184749-01) R6868 (777643-01) SH68-68-EP Noise-Rejecting Shielded Cable SH68-68-EP Shielded Cable The SH68-68-EP is a shielded 68-conductor cable terminated with two 68-pin female 0.050 series D-type connectors. It features individually shielded analog twisted pairs for reduced cross-talk with high-speed devices. 1 m ....................................................................................................................184749-01 2 m ....................................................................................................................184749-02 R6868 Low-Cost, Ribbon Cable The R6868 is a 68-conductor flat ribbon cable terminated with two 68-pin female 0.050 series D-type connectors. 1 m ....................................................................................................................182482-01 R6868 Ribbon Cable BNC-2110 Noise-Rejecting BNC I/O Connector Block The BNC-2110 is a shielded connector block with signal-labeled BNC connectors for easy connectivity of your analog input, digital I/O, and counter/timer signals. BNC-2110 ........................................................................................................777643-01 Dimensions – 20.3 by 11.2 by 5.5 cm (8.0 by 4.4 by 2.2 in.) TB-2705 Noise-Rejecting Front-Mount I/O Connector Block BNC-2110 I/O Connector Block The TB-2705 is a screw terminal block for your PXI DAQ devices. It latches to the front of your PXI modules with locking screws and provides strain relief and easy access to your analog, digital, trigger, and counter/timer signals through screw terminals. TB-2705 ............................................................................................................778241-01 Dimensions – 8.43 by 10.41 by 2.03 cm (3.32 by 4.1 by 0.8 in.) SCB-68 Noise-Rejecting Shielded I/O Connector Block The SCB-68 is a shielded I/O connector block for rugged, very-low-noise signal termination for connecting 68-pin DAQ devices. Silk-screened component locations provide an easy addition of simple signal-conditioning circuitry for your analog input channels. It also includes a general-purpose breadboard area as well as an IC temperature sensor for cold-junction compensation in temperature measurements. SCB-68 ..............................................................................................................776844-01 Dimensions – 19.5 by 15.2 by 4.5 cm (7.7 by 6.0 by 1.8 in.) TB-2705 Terminal Block CB-68LP and CB-68LPR Low-Cost I/O Connector Blocks SCB-68 Shielded I/O Connector Block The CB-68 and CB-68LP are low-cost termination accessories with 68 screw terminals for easy connector of field I/O signals to 68-pin DAQ devices. The connector blocks include standoffs for use on a desktop or for mounting in a custom panel. The CB-68LP has a vertical-mounted 68-pin connector. The CB-68-LPR has a right-angle mounted connector, and it is used with the CA-1000. CB-68LP ............................................................................................................777145-01 Dimensions – 14.35 by 10.74 cm (5.65 by 4.23 in.) CB-68LPR ........................................................................................................777145-02 Dimensions – 7.62 by 16.19 cm (3.00 by 6.36 in.) CB-68LP and CB-68LPR I/O Connector Blocks National Instruments • Tel: (800) 813 3693 • info@ni.com • ni.com 7 NI Services and Support NI has the services and support to meet your needs around the globe and through the application life cycle – from planning and development through deployment and ongoing maintenance. We offer services and service levels to meet customer requirements in research, design, validation, and manufacturing. Visit ni.com/services. Local Sales and Technical Support In offices worldwide, our staff is local to the country, giving you access to engineers who speak your language. NI delivers industryleading technical support through online knowledge bases, our applications engineers, and access to 14,000 measurement and automation professionals within NI Developer Exchange forums. Find immediate answers to your questions at ni.com/support. We also offer service programs that provide automatic upgrades to your application development environment and higher levels of technical support. Visit ni.com/ssp. SERVICE NEEDS Training and Certification NI training is the fastest, most certain route to productivity with our products. NI training can shorten your learning curve, save development time, and reduce maintenance costs over the application life cycle. We schedule instructor-led courses in cities worldwide, or we can hold a course at your facility. We also offer a professional certification program that identifies individuals who have high levels of skill and knowledge on using NI products. Visit ni.com/training. Hardware Services NI Factory Installation Services NI Factory Installation Services (FIS) is the fastest and easiest way to use your PXI or PXI/SCXI combination systems right out of the box. Trained NI technicians install the software and hardware and configure the system to your specifications. NI extends the standard warranty by one year on hardware components (controllers, chassis, modules) purchased with FIS. To use FIS, simply configure your system online with ni.com/pxiadvisor. Professional Services Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than 600 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance. Calibration Services NI recognizes the need to maintain properly calibrated devices for high-accuracy measurements. We provide manual calibration procedures, services to recalibrate your products, and automated calibration software specifically designed for use by metrology laboratories. Visit ni.com/calibration. Repair and Extended Warranty OEM Support We offer design-in consulting and product integration assistance if you want to use our products for OEM applications. For information about special pricing and services for OEM customers, visit ni.com/oem. NI provides complete repair services for our products. Express repair and advance replacement services are also available. We offer extended warranties to help you meet project life-cycle requirements. Visit ni.com/services. ni.com • (800) 813 3693 National Instruments • info@ni.com © 2005 National Instruments Corporation. All rights reserved. CVI, DAQ-STC, LabVIEW, Measurement Studio, MITE, National Instruments, National Instruments Alliance Partner, NI, ni.com, NI-DAQ, RTSI, SCXI, and SignalExpress are trademarks of National Instruments. Other product and company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from NI and has no agency, partnership, or joint-venture relationship with NI. 2005_4961_301_101_D_6120_611x
778204-01 价格&库存

很抱歉,暂时无法提供与“778204-01”相匹配的价格&库存,您可以联系我们找货

免费人工找货