0
登录后你可以
  • 下载海量资料
  • 学习在线课程
  • 观看技术视频
  • 写文章/发帖/加入社区
创作中心
发布
  • 发文章

  • 发资料

  • 发帖

  • 提问

  • 发视频

创作活动
NIUSB-6009

NIUSB-6009

  • 厂商:

    NI

  • 封装:

  • 描述:

    NIUSB-6009 - 8 analog inputs at 12 or 14 bits,up to 48 kS/s - National Instruments Corporation

  • 数据手册
  • 价格&库存
NIUSB-6009 数据手册
Low-Cost, Bus-Powered Multifunction DAQ for USB – 12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs NI USB-6008, NI USB-6009 • 8 analog inputs at 12 or 14 bits, up to 48 kS/s • 2 analog outputs at 12 bits, software-timed • 12 TTL/CMOS digital I/O lines • 32-bit, 5 MHz counter • Digital triggering • Bus-powered • 1-year warranty Recommended Software • • • • LabVIEW LabVIEW SignalExpress LabWindows™/CVI Measurement Studio Other Compatible Software • C#, Visual Basic .NET • ANSI C/C++ Operating Systems • • • • • Windows Vista (32- and 64-bit)/XP/2000 Mac OS X1 Linux®1 Windows Mobile1 Windows CE1 Measurement Services Software (included) • NI-DAQmx driver software • Measurement & Automation Explorer configuration utility • LabVIEW SignalExpress LE 1You need to download NI-DAQmx Base for these operating systems. Product USB-6009 USB-6008 1 SE Bus USB USB Analog Inputs1 8 SE/4 DI 8 SE/4 DI Input Resolution (bits) 14 12 Max Sampling Rate (kS/s) 48 10 Input Range (V) ±1 to ±20 ±1 to ±20 Analog Outputs 2 2 Output Resolution (bits) 12 12 Output Rate (Hz) 150 150 Output Range (V) 0 to 5 0 to 5 Digital I/O Lines 12 12 32-Bit Counter 1 1 Trigger Digital Digital = single ended, DI = differential 2Software-timed Overview and Applications With recent bandwidth improvements and new innovations from National Instruments, USB has evolved into a core bus of choice for measurement applications. The NI USB-6008 and USB-6009 are lowcost entry points to NI flagship data acquisition (DAQ) devices. With plug-and-play USB connectivity, these modules are simple enough for quick measurements but versatile enough for more complex measurement applications. The USB-6008 and USB-6009 are ideal for a number of applications where low cost, small form factor, and simplicity are essential. Examples include: • Data logging – quick and easy environmental or voltage data logging • Academic lab use – student ownership of DAQ hardware for completely interactive lab-based courses (Academic pricing available. Visit ni.com/academic for details.) • OEM applications as I/O for embedded systems Recommended Software National Instruments measurement services software, built around NI-DAQmx driver software, includes intuitive application programming interfaces, configuration tools, I/O assistants, and other tools designed to reduce system setup, configuration, and development time. National Instruments recommends using the latest version of NI-DAQmx driver software for application development in NI LabVIEW, LabVIEW SignalExpress, LabWindows/CVI, and Measurement Studio software. To obtain the latest version of NI-DAQmx, visit ni.com/support/daq/versions. NI measurement services software speeds up your development with features including: • A guide to create fast and accurate measurements with no programming using the DAQ Assistant. • Automatic code generation to create your application in LabVIEW. • LabWindows/CVI; LabVIEW SignalExpress; and C#, Visual Studio .NET, ANSI C/C++, or Visual Basic using Measurement Studio. • Multithreaded streaming technology for 1,000 times performance improvements. • Automatic timing, triggering, and synchronization routing to make advanced applications easy. • More than 3,000 free software downloads available at ni.com/zone to jump-start your project. • Software configuration of all digital I/O features without hardware switches/jumpers. • Single programming interface for analog input, analog output, digital I/O, and counters on hundreds of multifunction DAQ hardware devices. M Series devices are compatible with the following versions (or later) of NI application software – LabVIEW, LabWindows/CVI, or Measurement Studio versions 7.x; and LabVIEW SignalExpress 2.x. Low-Cost, Bus-Powered Multifunction DAQ for USB – 12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs Every M Series data acquisition device also includes a copy of LabVIEW SignalExpress LE data-logging software, so you can quickly acquire, analyze, and present data without programming. The NI-DAQmx Base driver software is provided for use with Linux, Mac OS X, Windows Mobile, and Windows CE operating systems. Information for Student Ownership To supplement simulation, measurement, and automation theory courses with practical experiments, NI has developed the USB-6008 and USB-6009 student kits, which include the LabVIEW Student Edition and a ready-to-run data logger application. These kits are exclusively for students, giving them a powerful, low-cost, hands-on learning tool. Visit ni.com/academic for more details. Recommended Accessories The USB-6008 and USB-6009 have removable screw terminals for easy signal connectivity. For extra flexibility when handling multiple wiring configurations, NI offers the USB-600x Connectivity Kit, which includes two extra sets of screw terminals, extra labels, and a screwdriver. In addition, the USB-600x Prototyping Kit provides space for adding more circuitry to the inputs of the USB-6008 or USB-6009. Information for OEM Customers For information on special configurations and pricing, call (800) 813 3693 (U.S. only) or visit ni.com/oem. Go to the Ordering Information section for part numbers. NI USB DAQ for OEMs Shorten your time to market by integrating world-class National Instruments OEM measurement products into your embedded system design. Board-only versions of NI USB DAQ devices are available for OEM applications, with competitive quantity pricing and available software customization. The NI OEM Elite Program offers free 30-day trial kits for qualified customers. Visit ni.com/oem for more information. Ordering Information NI USB-60081 ........................................................................779051-01 NI USB-60091 ........................................................................779026-01 NI USB-6008 OEM ................................................................193132-02 NI USB-6009 OEM ................................................................193132-01 NI USB-6008 Student Kit1,2 ..................................................779320-22 NI USB-6009 Student Kit1,2 ..................................................779321-22 NI USB-600x Connectivity Kit ..............................................779371-01 NI USB-600x Prototyping Kit ................................................779511-01 1 Includes 2 Includes NI-DAQmx software, LabVIEW SignalExpress LE, and a USB cable. LabVIEW Student Edition. BUY NOW! For complete product specifications, pricing, and accessory information, call 800 813 3693 (U.S. only) or go to ni.com/usb. BUY ONLINE a t ni.com o r CALL 800 813 3693 (U.S.) 2 Low-Cost, Bus-Powered Multifunction DAQ for USB – 12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs Output range ....................................... Output impedance............................... Output current drive............................ Power-on state.................................... Slew rate............................................. Short-circuit current............................ 0 to +5 V 50 Ω 5 mA 0V 1 V/µs 50 mA Specifications Typical at 25 °C unless otherwise noted. Analog Input Absolute accuracy, single-ended Range ±10 Typical at 25 ˚C (mV) 14.7 Maximum (0 to 55 ˚C) (mV) 138 Digital I/O Number of channels............................ 12 total 8 (P0.) 4 (P1.) Direction control ................................. Each channel individually programmable as input or output Output driver type USB-6008........................................ Open-drain USB-6009........................................ Each channel individually programmable as push-pull or open-drain Compatibility ....................................... CMOS, TTL, LVTTL Internal pull-up resistor ...................... 4.7 kΩ to +5 V Power-on state.................................... Input (high impedance) Absolute maximum voltage range...... -0.5 to +5.8 V Digital logic levels Level Input low voltage Input high voltage Input leakage current Output low voltage (I = 8.5 mA) Output high voltage (push-pull, I = -8.5 mA) Output high voltage (open-drain, I = -0.6 mA, nominal) Output high voltage (open-drain, I = -8.5 mA, with external pull-up resistor) Min -0.3 2.0 – – 2.0 2.0 2.0 Max 0.8 5.8 50 0.8 3.5 5.0 – Units V V µA V V V V Absolute accuracy at full scale, differential1 Range ±20 ±10 ±5 ±4 ±2.5 ±2 ±1.25 ±1 Typical at 25 ˚C (mV) 14.7 7.73 4.28 3.59 2.56 2.21 1.70 1.53 Maximum (0 to 55 ˚C) (mV) 138 84.8 58.4 53.1 45.1 42.5 38.9 37.5 Number of channels............................ 8 single-ended/4 differential Type of ADC ........................................ Successive approximation ADC resolution (bits) Module USB-6008 USB-6009 Differential 12 14 Single-Ended 11 13 Maximum sampling rate (system dependent) Module USB-6008 USB-6009 Maximum Sampling Rate (kS/s) 10 48 Input range, single-ended................... ±10 V Input range, differential...................... ±20, ±10, ±5, ±4, ±2.5, ±2, ±1.25, ±1 V Maximum working voltage ................. ±10 V Overvoltage protection ....................... ±35 V FIFO buffer size ................................... 512 B Timing resolution ................................ 41.67 ns (24 MHz timebase) Timing accuracy .................................. 100 ppm of actual sample rate Input impedance ................................. 144 kΩ Trigger source...................................... Software or external digital trigger System noise....................................... 5 m Vrms (±10 V range) Counter Number of counters ............................ Resolution ........................................... Counter measurements....................... Pull-up resistor.................................... Maximum input frequency.................. Minimum high pulse width................. Minimum low pulse width.................. Input high voltage ............................... Input low voltage ................................ Power available at I/O connector +5 V output (200 mA maximum) ......... +5 V typical +4.85 V minimum +2.5 V output (1 mA maximum) .......... +2.5 V typical +2.5 V output accuracy ....................... 0.25% max Voltage reference temperature drift... 50 ppm/°C max 1 32 bits Edge counting (falling edge) 4.7 kΩ to 5 V 5 MHz 100 ns 100 ns 2.0 V 0.8 V Analog Output Absolute accuracy (no load) ............... 7 mV typical, 36.4 mV maximum at full scale Number of channels............................ 2 Type of DAC ........................................ Successive approximation DAC resolution.................................... 12 bits Maximum update rate ........................ 150 Hz, software-timed 1 Input voltages may not exceed the working voltage range. BUY ONLINE a t ni.com o r CALL 800 813 3693 (U.S.) 3 Low-Cost, Bus-Powered Multifunction DAQ for USB – 12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs Physical Characteristics If you need to clean the module, wipe it with a dry towel. Dimensions (without connectors) ....... 6.35 by 8.51 by 2.31 cm (2.50 by 3.35 by 0.91 in.) Dimensions (with connectors) ............ 8.18 by 8.51 by 2.31 cm (3.22 by 3.35 by 0.91 in.) Weight (without connectors) .............. 59 g (2.1 oz) Weight (with connectors) ................... 84 g (3 oz) I/O connectors..................................... USB series B receptacle (2) 16-position (screw-terminal) plug headers Screw-terminal wiring ........................ 16 to 28 AWG Screw-terminal torque........................ 0.22 to 0.25 N•m (2.0 to 2.2 lb•in.) Power Requirement USB (4.10 to 5.25 VDC)....................... 80 mA typical 500 mA maximum USB suspend....................................... 300 µA typical 500 µA maximum Environmental The USB-6008 and USB-6009 are intended for indoor use only. Operating environment Ambient temperature range ........... 0 to 55 °C (tested in accordance with IEC-60068-2-1 and IEC-60068-2-2) Relative humidity range ................. 10 to 90%, noncondensing (tested in accordance with IEC-60068-2-56) Storage environment Ambient temperature range ........... -40 to 85 °C (tested in accordance with IEC-60068-2-1 and IEC-60068-2-2) Relative humidity range ................. 5 to 90%, noncondensing (tested in accordance with IEC-60068-2-56) Maximum altitude............................... 2,000 m (at 25 °C ambient temperature) Pollution degree.................................. 2 Safety and Compliance Safety This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: • IEC 61010-1, EN 61010-1 • UL 61010-1, CSA 61010-1 Note: For UL and other safety certifications, refer to the product label or visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column. Electromagnetic Compatibility This product is designed to meet the requirements of the following standards of EMC for electrical equipment for measurement, control, and laboratory use: • EN 61326 EMC requirements; Minimum Immunity • EN 55011 Emissions; Group 1, Class A • CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A Note: For EMC compliance, operate this device according to product documentation. CE Compliance This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows: • 2006/95/EC; Low-Voltage Directive (safety) • 2004/108/EC; Electromagnetic Compatibility Directive (EMC) Note: Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory compliance information. To obtain the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column. Waste Electrical and Electronic Equipment (WEEE) EU Customers: At the end of their life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers and National Instruments WEEE initiatives, visit ni.com/environment/weee.htm. BUY ONLINE a t ni.com o r CALL 800 813 3693 (U.S.) 4 NI Services and Support IN TA PL SERVICE NEEDS LO Y V DE NI has the services and support to meet your needs around the globe and through the application life cycle – from planning and development through deployment and ongoing maintenance. We offer services and service levels to meet customer requirements in research, design, validation, and manufacturing. Visit ni.com/services. Local Sales and Technical Support In offices worldwide, our staff is local to the country, giving you access to engineers who speak your language. NI delivers industry-leading technical support through online knowledge bases, our applications engineers, and access to 14,000 measurement and automation professionals within NI Developer Exchange forums. Find immediate answers to your questions at ni.com/support. We also offer service programs that provide automatic upgrades to your application development environment and higher levels of technical support. Visit ni.com/ssp. MA IN Training and Certification NI training is the fastest, most certain route to productivity with our products. NI training can shorten your learning curve, save development time, and reduce maintenance costs over the application life cycle. We schedule instructor-led courses in cities worldwide, or we can hold a course at your facility. We also offer a professional certification program that identifies individuals who have high levels of skill and knowledge on using NI products. Visit ni.com/training. EL OP AN Professional Services Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than 600 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance. OEM Support We offer design-in consulting and product integration assistance if you want to use our products for OEM applications. For information about special pricing and services for OEM customers, visit ni.com/oem. P DE Hardware Services NI Factory Installation Services NI Factory Installation Services (FIS) is the fastest and easiest way to use your PXI or PXI/SCXI combination systems right out of the box. Trained NI technicians install the software and hardware and configure the system to your specifications. NI extends the standard warranty by one year on hardware components (controllers, chassis, modules) purchased with FIS. To use FIS, simply configure your system online with ni.com/pxiadvisor. Calibration Services NI recognizes the need to maintain properly calibrated devices for high-accuracy measurements. We provide manual calibration procedures, services to recalibrate your products, and automated calibration software specifically designed for use by metrology laboratories. Visit ni.com/calibration. Repair and Extended Warranty NI provides complete repair services for our products. Express repair and advance replacement services are also available. We offer extended warranties to help you meet project life-cycle requirements. Visit ni.com/services. ni.com • (800) 813 3693 National Instruments • info@ni.com *351378C-01* 351378C-01 2008-9621-301-101-D ©2008 National Instruments. All rights reserved. CVI, LabVIEW, Measurement Studio, National Instruments, National Instruments Alliance Partner, NI, ni.com, SCXI, and SignalExpress are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the United States and other countries. Linux® is the registered trademark of Linus Torvalds in the U.S. and other countries. Other product and company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from NI and has no agency, partnership, or joint-venture relationship with NI.
NIUSB-6009 价格&库存

很抱歉,暂时无法提供与“NIUSB-6009”相匹配的价格&库存,您可以联系我们找货

免费人工找货