!NOT RECOMMENDED FOR NEW DESIGNS!
• Long 5 year warranty
Features
• 2MOPP/250VAC
AC/DC Converter
• Suitable for built in Class II applications
• Wide input voltage range (85-264VAC)
• Low leakage current (8.0mm
>8.0mm
Notes:
Note5: For repeat Hi-Pot testing, reduce the time and/or the test voltage
www.recom-power.com
REV.: 8/2021
MED-3
! NOT RECOMMENDED FOR NEW DESIGNS !
RACM65
AC/DC Converter
Series
Specifications (measured at Ta= 25°C, 250VAC, full load and after warm-up)
ENVIRONMENTAL
Parameter
Condition
Operating Temperature Range
Value
refer to derating graph
-40°C to +85°C
Temperature Coefficient
±0.02%/K
Operating Altitude
5000m max.
Operating Humidity
non-condensing
5% to 95% RH
Pollution Degree
PD2
Shock
according to IEC60068-2-27
Vibration
according to IEC60068-2-6
MTBF
1494 x 103 hours
according to MIL-HDBK-217F, full load, +25°C
Derating Graph
(@ natural convection 0.1m/s)
RACM65-15,24S(/OF)
100
100
90
90
80
80
70
70
Output Load [%]
Output Load [%]
RACM65-05,12,48S(/OF)
60
50
40
30
50
40
30
20
10
60
20
115VAC
230VAC
10
0
-40 -30 -20 -10 0 10 20 30 40 50 60 70 80 90 100 110
75 85
Ambient Temperature [°C]
115VAC
230VAC
0
-40 -30 -20 -10 0 10 20 30 40 50 60 70 80 90 100 110
75 85
Ambient Temperature [°C]
SAFETY AND CERTIFICATIONS
Certificate Type (Safety)
Report / File Number
Medical Electric Equipment, General Requirements for Safety and Essential Performance
E314885
Medical Electric Equipment, General Requirements for Safety and Essential Performance
(CB Scheme)
151101302
Information Technology Equipment - General Requirements for Safety (LVD)
TW1708008-001
Information Technology Equipment - General Requirements for Safety
EAC
RU-AT.49.09571
RoHs2+
Standard
CAN/CSA-C22.2 No. 60601-1:14
ANSI/AAMI ES60601-1:2005 + A2:2010
IEC60601-1:2005 + C2:2007, 3rd Edition
EN60601-1:2006
EN60950-1:2006 + A2:2013
IEC60950-1:2005, 2nd Edition + A2:2013
TP TC 004/2011 TP TC 004/2011
RoHS-2011/65/EU + AM-2015/863
EMC Compliance (Medical)
Conditions
Standard / Criterion
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
EN60601-1-2:2015
Industrial, scientific and medical equipment - Radio frequency disturbance characteritics Limits and methods of measurement
CISPR11:2009 + A1:2010, Class B
continued on next page
www.recom-power.com
REV.: 8/2021
MED-4
! NOT RECOMMENDED FOR NEW DESIGNS !
AC/DC Converter
RACM65
Series
Specifications (measured at Ta= 25°C, 250VAC, full load and after warm-up)
EMC Compliance (Medical)
Conditions
ESD Electrostatic discharge immunity test
Air ±15kV; Contact ±8kV
Standard / Criterion
IEC61000-4-2:2008
Radiated, radio-frequency, electromagnetic field immunity test
20V/m (80-2700MHz)
27V/m (385MHz)
28V/m (450MHz)
IEC61000-4-3:2006 + A2:2010
Fast Transient and Burst Immunity
AC Power Port: ±2kV
IEC61000-4-4:2012
Surge Immunity
AC Port:
IEC61000-4-5:2014
20Vr.m.s
IEC61000-4-6:2013
50Hz, 30A/m
IEC61000-4-8:2009
Dips: >95%; 30%;
Interruptions >95%
IEC61000-4-11:2004
Immunity to conducted disturbances, induced by radio-frequency fields
Power Frequency Magnetic Field
Voltage Dips and Interruptions
L-N= ±1kV
L-GND= ±2kV
Limits of Voltage Fluctuations and Flicker
EN61000-3-3:2013
Limitations on the amount of electromagnetic intererence allowed from digital & electronic devices
47CFR FCC Part 15 Subpart B, Class B
Methods of Measurement of Radio-Noise Emissions from Low-Voltage Electrical and
Electronic Equipment in the Range of 9 kHz to 40 GHz
ANSI C63.4:2014
FCC methods of measurement of radio noise emissions from industrial, scientific, and
medical equipment
FCC OST/MP-5
EMC Compliance (Industrial)
Conditions
Electromagnetic compatibility of multimedia equipment – Emission Requirements
Information technology equipment - Immunity characteristics - Limits and methods of
measurement
ESD Electrostatic discharge immunity test
Standard / Criterion
EN55032:2015+AC:2013, Class B
EN55024:2010+A1:2015
Air ±15kV; Contact ±6kV
IEC61000-4-2:2008, Criteria A
Radiated, radio-frequency, electromagnetic field immunity test
10V/m (80-1000MHz)
20V/m (80-1000MHz)
IEC61000-4-3:2006 + A2:2010, Criteria A
Fast Transient and Burst Immunity
AC Power Port: ±4kV
IEC61000-4-4:2012, Criteria A
Surge Immunity
AC Port:
L-N= ±2kV
L-PE= ±4kV
IEC61000-4-5:2014, Criteria A
Immunity to conducted disturbances, induced by radio-frequency fields
AC Power Port 10V, 20V
IEC61000-4-6:2013, Criteria A
50Hz/60Hz, 100A/m,
1000A/m
IEC61000-4-8:2009, Criteria A
Voltage Dips and Interruptions
Dips: >95%; 60%; 30%
Interruptions >95%
IEC61000-4-11:2004, Criteria A
IEC61000-4-11:2004, Criteria B
Damped oscillatory wave immunity test
AC Port:
L-N= ±1kV
L/N-G= ±2.5kV
IEC61000-4-18:2006 + A1:2010, Criteria A
Power Frequency Magnetic Field
Limits of Voltage Fluctuations and Flicker
EN61000-3-3:2013
DIMENSION and PHYSICAL CHARACTERISTICS
Parameter
Type
Value
Material
enclosed case
PCB
aluminum
FR4, (UL94V-0)
Dimension (LxWxH)
enclosed case
open frame
91.4 x 60.5 x 33.3mm
76.2 x 50.8 x 26.5mm
Weight
enclosed case
open frame + “-ST” version
172g
137g
continued on next page
www.recom-power.com
REV.: 8/2021
MED-5
! NOT RECOMMENDED FOR NEW DESIGNS !
RACM65
AC/DC Converter
Series
Specifications (measured at Ta= 25°C, 250VAC, full load and after warm-up)
Dimension Drawing Enclosed Case (mm)
Top View
14.3
51.3
60.5
FC 41.3
CON2
CON1
Trim Adj.
logo embossed
76.2
FC 84.2
91.4
AC Input Connector (CON1)
Pin#
Terminal
Mating Housing
1 AC/L
3 AC/N
Molex KK156
(SD-2478)
Molex KK156
(09508031)
DC Output Connector (CON2)
Pin#
Terminal
Mating Housing
1,2 V3,4 V+
Molex KK156
(SD-2478)
Molex KK156
(09508041)
1.20
33.3
Side View
10.0
9.6
Bottom View
3.5
M3 screws
continued on next page
www.recom-power.com
REV.: 8/2021
MED-6
! NOT RECOMMENDED FOR NEW DESIGNS !
RACM65
AC/DC Converter
Series
Specifications (measured at Ta= 25°C, 250VAC, full load and after warm-up)
Dimension Drawing Open Frame (/OF) (mm)
Top View
76.2
10.0
44.0
50.8
ScrewACTerminal
Version
Input Connector
(CON1)„-ST“
Ø3.2
FC 69.8
51.3
14.3
51.3
Side View
Pin#
Terminal
Mating Housing
1 AC/L
3 AC/N
Molex KK156
(SD-2478)
Molex KK156
(09508031)
14.3
Connector Version
FC 44.4
CON2
CON1
Trim Adj.
DC Output Connector (CON2)
24.0
CON1
1,2 V3,4 V+
Trim Terminal
Adj.
Mating Housing
Molex KK156
(SD-2478)
Molex KK156
(09508041)
2.5
1.6
CON2
CON1
Pin#
CON2
Trim Adj.
Screw Terminal Connection “-ST”
Top View
logo embossed
logo embossed
Screw Terminal Version „-ST“
Trim Adj.
CON1
Trim Adj.
Trim Adj.
CON2
CON2
CON2
CON1
CON2
CON1
51.3
Open Frame Version
14.3
Enclosed Version
logo embossed
CON1
Screw terminal information
#
Function
AWG
1
3
1,2
3,4
Trim Adj.
VAC in (L)
VAC in (N)
-Vout
+Vout
26-16
26-16
26-16
26-16
Model
ETB30
(EK381V)
ETB30
(EK381V)
CON2
CON2
recommended tightening torque: 0.2Nm
www.recom-power.com
REV.: 8/2021
MED-7
! NOT RECOMMENDED FOR NEW DESIGNS !
AC/DC Converter
RACM65
Series
Specifications (measured at Ta= 25°C, 250VAC, full load and after warm-up)
PACKAGING INFORMATION
Parameter
Packaging Dimension (LxWxH)
Type
cardboard box
Value
enclosed case
120.0 x 80.0 x 85.0mm
open frame
111.0 x 94.0 x 51.0mm
Packaging Quantity
1pcs
Storage Temperature Range
Storage Humidity
-40°C to +85°C
non-condensing
5% to 95% RH
The product information and specifications may be subject to changes even without prior written notice.The product has been designed for various applications; its suitability lies in the responsibility of each customer. The
products are not authorized for use in safety-critical applications without RECOM’s explicit written consent. A safety-critical application is an application where a failure may reasonably be expected to endanger or cause
loss of life, inflict bodily harm or damage property. The applicant shall indemnify and hold harmless RECOM, its affiliated companies and its representatives against any damage claims in connection with the unauthorized
use of RECOM products in such safety-critical applications.
www.recom-power.com
REV.: 8/2021
MED-8