AD8233ACBZ-R7

AD8233ACBZ-R7

  • 厂商:

    AD(亚德诺)

  • 封装:

    WLCSP20_2.04X1.71MM

  • 描述:

    IC ECG FRONT END 20WLCSP

  • 数据手册
  • 价格&库存
AD8233ACBZ-R7 数据手册
FEATURES FUNCTIONAL BLOCK DIAGRAM Fully integrated, single-lead ECG front end Low quiescent supply current: 50 µA (typical) Leads on/off detection while in shutdown ( 50 kΩ RL = 50 kΩ Max Unit kHz V/µs nV/√Hz µV p-p µV p-p 7.5 10 +VS − 0.1 µA V mA kΩ kHz 0.1 IOUT VOS IB IOUT Typ 15 0.01 120 7 9 11 150 20 180 1 100 12 0.1 +VS − 0.7 +VS − 0.27 125 1.5 fAC IAC IDC 50 +IN, SDN = low −IN, SDN = low Between +IN and −IN, SDN = high 10 mV pA mA V V mV µs 100 200 250 −300 20 100 175 kHz nA p-p nA nA MΩ μs 10 100 12 kΩ pA S1 and S2 RON tS1 Filter Recovery Switch On Time tS2 Fast Restore Reset tRST 8 From either rail 100 2 mV µs +VS = 3 V +VS = 1.8 V +VS = 3 V +VS = 1.8 V +VS = 3 V +VS = 1.8 V 160 80 80 40 3 1.5 ms Rev. D | Page 4 of 32 ms µs Data Sheet Parameter LOGIC INTERFACE Input Characteristics Input Voltage (AC/DC, FR, and RLD SDN) Low High Input Voltage (SDN) Low High Output Characteristics Output Voltage Low High SYSTEM SPECIFICATIONS Quiescent Supply Current Wakeup Current Shutdown Current Peak-to-Peak Voltage Noise (RTI) AD8233 Symbol Test Conditions/Comments 2 Typ Max Unit VIL VIH 0.41 × +VS 0.45 × +VS V V VIL VIH 0.6 × + VS 0.3 × + VS V V 0.05 +VS − 0.05 V V LOD terminal RL = 100 kΩ VOL VOH 50 60 0.65 0.75 0.5 0.6 TA = 0°C to 70°C SDN = low, LOD = low TA = 0°C to 70°C SDN = low, LOD = high TA = 0°C to 70°C VDIFF = 0 V f = 0.5 Hz to 40 Hz f = 0.05 Hz to 150 Hz VDIFF = ±0.3 V f = 0.5 Hz to 40 Hz f = 0.05 Hz to 150 Hz Supply Range Specified Temperature Range Operational Temperature Range 1 Min 1.7 0 −40 Offset is referred to the input of the instrumentation amplifier inputs. In ac leads off and shutdown mode, the dc leads off comparator at the +IN pin trips the LOD pin. Rev. D | Page 5 of 32 70 1.5 1 µA µA µA µA µA µA 9 15 µV p-p µV p-p 11 21 µV p-p µV p-p V °C °C 3.5 70 +85 AD8233 Data Sheet ABSOLUTE MAXIMUM RATINGS THERMAL RESISTANCE Table 3. Parameter Supply Voltage Output Short-Circuit Current Duration Maximum Voltage, Any Terminal1 Minimum Voltage, Any Terminal1 Storage Temperature Range Operating Temperature Range Maximum Junction Temperature Electrostatic Discharge (ESD) Rating Human Body Model (HBM) Field Induced Charged Device Model (FICDM) 1 Thermal performance is directly linked to printed circuit board (PCB) design and operating environment. Careful attention to PCB thermal design is required. Rating 3.6 V Indefinite +VS + 0.3 V −0.3 V −65°C to +125°C −40°C to +85°C 140°C Table 4. Thermal Resistance Package Type CB-20-13 PCB 1S0P1 2S2P2 8 kV 1 kV Stresses at or above those listed under Absolute Maximum Ratings may cause permanent damage to the product. This is a stress rating only; functional operation of the product at these or any other conditions above those indicated in the operational section of this specification is not implied. Operation beyond the maximum operating conditions for extended periods may affect product reliability. 0 ms 108.5 101.1 47.9 46.8 θJA (°C/W) 1 ms 2 ms 89.0 82.3 87.3 87.3 43.4 42.1 43.3 42.1 θJC (°C/W) 0.6 0.6 0.7 0.7 Simulated thermal numbers per JESD51-9: 1-layer PCB (1S0P), low effective thermal conductivity test board. 2 4-layer PCB (2S2P), high effective thermal conductivity test board. 1 This level or the maximum specified supply voltage, whichever is the lesser, indicates the superior voltage limit for any terminal. If input voltages beyond the specified minimum or maximum voltages are expected, place resistors in series with the inputs to limit the current to less than 5 mA. Power (W) 0.25 1.25 0.25 1.25 ESD CAUTION Rev. D | Page 6 of 32 Data Sheet AD8233 PIN CONFIGURATIONS AND FUNCTION DESCRIPTIONS BALL A1 INDICATOR AD8233 1 2 GND +VS SDN AC/DC 3 4 5 REFIN HPSENSE HPDRIVE A FR IAOUT +IN LOD RLD SDN REFOUT RLDFB –IN OUT OPAMP– OPAMP+ RLD B C SW TOP VIEW (BALL SIDE DOWN) Not to Scale 13737-002 D Figure 2. 20-Ball WLCSP Pin Configuration Table 5. 20-Ball WLCSP Pin Function Descriptions Ball No. A1 A2 A3 A4 Mnemonic GND +VS REFIN HPSENSE A5 HPDRIVE B1 B2 SDN AC/DC B3 B4 B5 C1 C2 C3 FR IAOUT +IN LOD RLD SDN REFOUT C4 C5 D1 RLDFB −IN OUT D2 D3 D4 D5 OPAMP− OPAMP+ SW RLD Description Power Supply Ground. Power Supply Terminal. Reference Buffer Input. Use REFIN, a high impedance input terminal, to set the level of the reference buffer. High-Pass Sense Input for Instrumentation Amplifier. Connect HPSENSE to the junction of R and C that sets the corner frequency of the dc blocking circuit. High-Pass Driver Output. Connect HPDRIVE to the capacitor in the first high-pass filter. The AD8233 drives this pin to keep HPSENSE at the same level as the reference voltage. Shutdown Control Input. Drive SDN low to enter the low power shutdown mode. Leads Off Mode Control Input. Drive the AC/DC pin low for dc leads off mode. Drive the AC/DC pin high for ac leads off mode. Fast Restore Control Input. Drive FR high to enable fast recovery mode. Otherwise, drive it low. Instrumentation Amplifier Output Terminal. Instrumentation Amplifier, Positive Input. +IN is typically connected to the left arm (LA) electrode. Leads Off Detection Comparator Output. Right Leg Drive Shutdown Control Input. Drive RLD SDN low to power down the RLD amplifier. Reference Buffer Output. The instrumentation amplifier output is referenced to this potential. Use REFOUT as a virtual ground for any point in the circuit that requires a signal reference. Right Leg Drive Feedback Input. RLDFB is the feedback terminal for the right leg drive circuit. Instrumentation Amplifier, Negative Input. −IN is typically connected to the right arm (RA) electrode. Operational Amplifier Output. The fully conditioned heart rate signal is present at this output. OUT can be connected to the input of an ADC. Operational Amplifier Inverting Input. Operational Amplifier Noninverting Input. Fast Restore Switch Terminal. Connect this terminal to the output of the second high-pass filter. Right Leg Drive Output. Connect the driven electrode (typically right leg) to the RLD pin. Rev. D | Page 7 of 32 AD8233 Data Sheet TYPICAL PERFORMANCE CHARACTERISTICS +VS = 3 V, VREF = 1.5 V, VCM = 1.5 V, TA = 25°C, unless otherwise noted. All typical performance characteristics are measured for the WLCSP package. INSTRUMENTATION AMPLIFIER PERFORMANCE CHARACTERISTICS 2100 100 80 1750 INPUT BIAS CURRENT (pA) 60 UNITS 1400 1050 700 40 20 0 –20 –40 –60 350 30 0 –30 60 90 120 CMRR (µV/V) 0 40 2500 30 GAIN (dB) 3000 1500 10 500 0 –0.5 0 0.5 1.0 1.5 2.0 GAIN ERROR (%) 2.5 3.0 3.5 NO DC OFFSET 300mV OFFSET –10 1 10 100 1k 10k 100k FREQUENCY (Hz) Figure 4. Gain Error Distribution Figure 7. Gain vs. Frequency 3.5 120 3.0 100 CMRR RTI (dB) 2.5 2.0 1.5 1.0 80 60 0.5 40 –0.5 0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 OUTPUT VOLTAGE (V) Figure 5. Input Common-Mode Voltage vs. Output Voltage 20 10 100 1k 10k FREQUENCY (Hz) Figure 8. CMRR RTI vs. Frequency Rev. D | Page 8 of 32 100k 13737-008 NO DC OFFSET +300mV OFFSET –300mV OFFSET 0 13737-005 INPUT COMMON-MODE VOLTAGE (V) 2.0 20 1000 13737-004 UNITS 50 –1.0 1.5 Figure 6. Input Bias Current vs. Input Common-Mode Voltage 3500 –1.5 1.0 INPUT COMMON-MODE VOLTAGE (V) Figure 3. CMRR Distribution 0 –2.0 0.5 13737-007 –60 13737-003 –90 –100 13737-006 –80 0 –120 Data Sheet AD8233 100 90 80 PSRR RTI (dB) 70 5µV/DIV 60 50 40 30 10 1 10 100 1k 10k 100k FREQUENCY (Hz) 13737-009 200ms/DIV 0 0.1 13737-012 20 Figure 12. 0.5 Hz to 40 Hz Noise (RTI) Figure 9. PSRR RTI vs. Frequency 1.0 10k 0.9 0.8 GAIN ERROR (%) 100 0.6 0.5 0.4 0.3 0.2 10 100 1k 10k 100k FREQUENCY (Hz) Figure 10. Voltage Noise Spectral Density (RTI) 0 0 50 100 150 200 250 300 DC OFFSET (mV) Figure 13. Gain Error vs. DC Offset 22pF 470pF 1nF 5µV/DIV 1s/DIV 400µs/DIV Figure 11. 0.1 Hz to 10 Hz Noise (RTI) 50mV/DIV Figure 14. Small Signal Pulse Response Rev. D | Page 9 of 32 13737-013 1 13737-014 10 0.1 13737-010 0.1 13737-011 NOISE (nV/√Hz) 0.7 1k AD8233 Data Sheet 0.5V/DIV 70 3.0 60 2.5 50 2.0 40 1.5 30 1.0 20 0.5 10 0 0 –0.5 –10 –1.0 –40 –20 0 20 40 60 –20 100 80 13737-018 13737-015 INPUT BIAS CURRENT (IB) (nA) 3.5 400µs/DIV 80 IB IOS INPUT OFFSET CURRENT (IOS) (pA) 4.0 TEMPERATURE (°C) Figure 18. Input Bias Current (IB) and Input Offset Current (IOS) vs. Temperature Figure 15. Large Signal Pulse Response 0.5 1.5 0.4 0.3 0.2 GAIN ERROR (%) 0.5 0 –0.5 0.1 0 –0.1 –0.2 –0.3 –40°C +25°C +85°C –1.5 100 10k 1k –0.4 1M 100k LOAD (Ω) –0.5 –40 0 20 40 60 80 100 80 100 TEMPERATURE (°C) Figure 16. Output Voltage Swing vs. Load Figure 19. Gain Error vs. Temperature 10 0.4 8 0.3 6 0.2 4 CMRR (µV/V) 0.1 0 –0.1 2 0 –2 –4 –0.2 –6 –0.3 –8 –20 0 20 40 60 80 100 TEMPERATURE (°C) –10 –40 –20 0 20 40 60 TEMPERATURE (°C) Figure 20. CMRR vs. Temperature Figure 17. DC Blocking Input Offset Drift on Multiple Parts Rev. D | Page 10 of 32 13737-020 –0.4 –40 13737-017 DC BLOCKING INPUT OFFSET (mV) –20 13737-019 –1.0 13737-016 OUTPUT VOLTAGE SWING (V) 1.0 Data Sheet AD8233 OPERATIONAL AMPLIFIER PERFORMANCE CHARACTERISTICS 1000 800 UNITS 600 400 400µs/DIV –2 0 2 4 OFFSET VOLTAGE (mV) Figure 24. Large Signal Transient Response Figure 21. Offset Distribution 80 –40 60 –60 40 –80 20 –100 0 –120 –20 –140 –40 –160 –60 0.1 1 10 100 1k –180 100k 10k VOLTAGE NOISE SPECTRAL DENSITY (nV/√Hz) –20 FREQUENCY (Hz) Figure 22. Open-Loop Gain and Phase Margin vs. Frequency 10k 1k 100 10 0.1 1 10 100 1k 10k 100k FREQUENCY (Hz) Figure 25. Voltage Noise Spectral Density vs. Frequency 22pF 470pF 1nF 100µs/DIV 20mV/DIV 1s/DIV Figure 26. 0.1 Hz to 10 Hz Noise Figure 23. Small Signal Response for Various Capacitive Loads Rev. D | Page 11 of 32 13737-026 5µV/DIV 13737-023 OPEN-LOOP GAIN (dB) 100 0 PHASE MARGIN (Degrees) GAIN PHASE MARGIN 13737-022 120 13737-025 –4 0.5V/DIV 13737-021 0 13737-024 200 AD8233 Data Sheet 1.5 5µV/DIV 0.5 0 –0.5 –40°C +25°C +85°C 200ms/DIV 13737-027 –1.0 –1.5 100 10k 1k 100k 1M LOAD (Ω) Figure 29. Output Voltage Swing vs. Load Figure 27. 0.5 Hz to 40 Hz Noise 100 120 80 110 100 60 90 40 80 PSRR (dB) 20 0 –20 70 60 50 40 –40 30 –60 20 –80 0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 INPUT COMMON-MODE VOLTAGE (V) 0 0.1 1 10 100 1k 10k FREQUENCY (Hz) Figure 30. Power Supply Rejection Ratio vs. Frequency Figure 28. Input Bias Current vs. Input Common-Mode Voltage Rev. D | Page 12 of 32 100k 13737-030 10 –100 13737-028 INPUT BIAS CURRENT (pA) 13737-029 OUTPUT VOLTAGE SWING (V) 1.0 Data Sheet AD8233 INPUT BIAS CURRENT (pA) 1k 10 1 0.1 –40 –20 0 20 40 60 80 TEMPERATURE (°C) Figure 31. Load Transient Response (100 μA Load Change) Figure 33. Input Bias Current vs. Temperature 0.8 0.6 0.2 0 –0.2 –0.4 –0.6 –0.8 –40 –20 0 20 40 60 80 TEMPERATURE (°C) 100 13737-032 OFFSET (mV) 0.4 Figure 32. Offset vs. Temperature on Multiple Parts Rev. D | Page 13 of 32 100 13737-033 13737-031 50mV/DIV 100µs/DIV 100 AD8233 Data Sheet RLD AMPLIFIER PERFORMANCE CHARACTERISTICS –20 –40 80 –60 60 –80 40 –100 20 –120 0 –140 –20 –160 OPEN-LOOP GAIN (dB) 100 –40 0.1 100 10 1 1k 10k 5µV/DIV 13737-037 120 0 PHASE MARGIN (Degrees) GAIN PHASE MARGIN 1s/DIV –180 100k 13737-034 140 FREQUENCY (Hz) Figure 37. 0.1 Hz to 10 Hz Noise Figure 34. Open-Loop Gain and Phase Margin vs. Frequency 1.5 0.5 5µV/DIV 0 –40°C +25°C +85°C –1.0 –1.5 100 1k 100k 10k 200ms/DIV 1M LOAD (Ω) Figure 38. 0.5 Hz to 40 Hz Noise Figure 35. Output Voltage Swing vs. Load 12 10k AC/DC = LOW, RLD SDN = HIGH 1k 100 8 6 4 2 10 0.1 1 10 100 1k 10k FREQUENCY (Hz) 100k 0 –40 +VS = 1.8V +VS = 3V +VS = 3.5V –20 0 20 40 60 80 TEMPERATURE (°C) Figure 39. RLD Supply Current vs. Temperature Figure 36. Voltage Spectral Noise Density vs. Frequency Rev. D | Page 14 of 32 100 13737-039 RLD SUPPLY CURRENT (µA) 10 13737-036 VOLTAGE NOISE SPECTRAL DENSITY (nV/√Hz) 13737-038 –0.5 13737-035 OUTPUT VOLTAGE SWING (V) 1.0 Data Sheet AD8233 REFERENCE BUFFER PERFORMANCE CHARACTERISTICS 15 100k SOURCE SINK OUTPUT IMPEDANCE (Ω) 10k 5 0 –5 100 10 1 LOAD CURRENT (mA) 1 0.1 1 10 100 1k 10k 13737-042 0.1 13737-040 –15 0.01 1k 10 –10 100k FREQUENCY (Hz) Figure 40. Load Regulation Figure 42. Output Impedance vs. Frequency 50mV/DIV 100µs/DIV 100 10 1 0.1 –40 –20 0 20 40 60 80 TEMPERATURE (°C) Figure 41. Load Transient Response (100 μA Load Change) Figure 43. Input Bias Current vs. Temperature Rev. D | Page 15 of 32 100 13737-043 INPUT BIAS CURRENT (pA) 1k 13737-041 OUTPUT ERROR (mV) 10 AD8233 Data Sheet SYSTEM PERFORMANCE CHARACTERISTICS SDN = HIGH, AC/DC = LOW, RLD SDN = LOW 60 50 40 30 20 +VS = 1.8V +VS = 3V +VS = 3.5V 10 0 –40 –20 0 20 40 60 80 100 TEMPERATURE (°C) 13737-044 SUPPLY CURRENT (µA) 70 10k 1k 100 10 0.1 1 10 100 1k 10k 100k FREQUENCY (Hz) Figure 47. Voltage Noise Spectral Density (RTI), Measured at IAOUT Figure 44. Supply Current vs. Temperature 900 SDN = LOW, AC/DC = LOW RLD SDN = LOW, LOD = HIGH 700 600 5µV/DIV 500 400 300 +VS = 1.8V +VS = 3V +VS = 3.5V 100 0 –40 –20 0 20 40 60 80 100 TEMPERATURE (°C) 200ms/DIV 13737-048 200 13737-045 SHUTDOWN CURRENT (nA) 800 Figure 48. 0.5 Hz to 40 Hz Noise (RTI), Measured at IAOUT Figure 45. Shutdown Current vs. Temperature 1000 900 SDN = LOW, AC/DC = LOW RLD SDN = LOW, LOD = LOW 700 10µV/DIV 600 500 400 300 +VS = 1.8V +VS = 3V +VS = 3.5V 100 0 –40 –20 0 20 40 60 80 TEMPERATURE (°C) 100 2s/DIV 13737-049 200 13737-046 WAKEUP CURRENT (nA) 800 Figure 49. 0.05 Hz to 150 Hz Noise (RTI), Measured at IAOUT Figure 46. Wakeup Current vs. Temperature Rev. D | Page 16 of 32 13737-047 VOLTAGE NOISE SPECTRAL DENSITY (nV/√Hz) 80 Data Sheet AD8233 THEORY OF OPERATION The feedback of the amplifier is applied via GM2 through two separate paths: the two resistors divide the output signal to set an overall gain of 100, whereas the dc blocking amplifier integrates any deviation from the reference level. Consequently, dc offsets as large as ±300 mV across the GM1 inputs appear inverted and with the same magnitude across the inputs of GM2, all without saturating the signal of interest. ARCHITECTURE OVERVIEW The AD8233 is an integrated front end for signal conditioning of cardiac biopotentials for heart rate monitoring. It consists of a specialized instrumentation amplifier (IA), an operational amplifier (A1), a right leg drive amplifier (A2), and a midsupply reference buffer (A3). In addition, the AD8233 includes leads on or off detection circuitry and an automatic fast restore circuit that restores the signal shortly after leads are reconnected. To increase the common-mode voltage range of the instrumentation amplifier, a charge pump boosts the supply voltage for the two transconductance amplifiers. This boost in supply voltage further prevents saturation of the amplifier in the presence of large common-mode signals, such as line interference. The charge pump runs from an internal oscillator, the frequency of which is set around 500 kHz. The AD8233 contains a specialized instrumentation amplifier that amplifies the ECG signal while rejecting the electrode half cell potential on the same stage. The amplification of the ECG signal and the rejection of the electrode half cell potential are possible with an indirect current feedback architecture, which reduces size and power compared with traditional implementations. OPERATIONAL AMPLIFIER INSTRUMENTATION AMPLIFIER The general-purpose operational amplifier (A1) is a rail-to-rail device that can be used for low-pass filtering and to add additional gain. The following sections provide details and example circuits that use this amplifier. The instrumentation amplifier shown in Figure 50 is composed of two well matched transconductance amplifiers (GM1 and GM2), the dc blocking amplifier (HPA), and an integrator formed by C1 and an op amp. The transconductance amplifier, GM1, generates a current that is proportional to the voltage present at its inputs. When the feedback is satisfied, an equal voltage appears across the inputs of the transconductance amplifier, GM2, thereby matching the current generated by GM1. The difference generates an error current that is integrated across Capacitor C1. The resulting voltage appears at the output of the instrumentation amplifier. RLD AMPLIFIER The RLD amplifier inverts the common-mode signal that is present at the instrumentation amplifier inputs. When the right leg drive output current is injected into the subject, it counteracts common-mode voltage variations, thus improving the common-mode rejection of the system. +VS HPDRIVE HPSENSE IAOUT SW A2 A5 A4 B4 D4 OPAMP– D3 D2 10kΩ CHARGE PUMP +VS OPAMP+ A1 D1 OUT S2 AC/DC +IN B5 RFI FILTER –IN C5 HPA GM1 GM2 VCM AC/DC S1 R 10kΩ B3 FR +VS – 0.1V 99R B1 SDN C1 0.1V B2 AC/DC INSTRUMENTATION AMPLIFIER (IA) S1 S2 SYNC RECTIFIER RLDFB C4 SWITCH TIMING 0.7V RLD SDN C2 RLD D5 C1 LOD A2 150kΩ AC/DC +VS – 0.27V A1 GND C3 REFOUT A3 *ALL SWITCHES SHOWN IN DC LEADS OFF DETECTION POSITION AND FAST RESTORE DISABLED = REFOUT Figure 50. Simplified Schematic Diagram Rev. D | Page 17 of 32 13737-050 REFIN A3 AD8233 Data Sheet An integrator can be built by connecting a capacitor between the RLD FB and RLD terminals. A good starting point is a 1 nF capacitor, which places the crossover frequency at about 1 kHz (the frequency at which the amplifier has an inverting unity gain). This configuration results in about 26 dB of loop gain available at a frequency range from 50 Hz to 60 Hz for common-mode line rejection. Higher capacitor values reduce the crossover frequency, thereby reducing the gain that is available for rejection and, consequently, increasing the line noise. Lower capacitor values move the crossover frequency to higher frequencies, allowing increased gain. However, when using higher gain, the system can become unstable and saturate the output of the right leg amplifier. When using this amplifier to drive an electrode, place a resistor in series with the output to limit the current to be always less than 10 µA, even in fault conditions. For example, if the supply used is 3.0 V, ensure that the resistor is greater than 330 kΩ to account for component and supply variations. The reference voltage level is set at the REFIN pin. It can be set with a voltage divider or by driving the REFIN pin from some other point in the circuit (for example, from the ADC reference). The voltage is available at the REFOUT pin for the filtering circuits or for an ADC input. +VS R1 A3 R2 To limit the power consumption of the voltage divider, the use of large resistors is recommended, such as 10 MΩ. The designer must keep in mind that high resistor values make it easier for interfering signals to appear at the input of the reference buffer. To minimize noise pickup, it is recommended to place the resistors close to each other and as near as possible to the REFIN terminal. Furthermore, use a capacitor in parallel with the lower resistor on the divider for additional filtering, as shown in Figure 52. A large capacitor results in better noise filtering but takes longer to settle the reference after power-up. The total time the reference takes to settle within 1% can be estimated with the formula RLDFB tSETTLE_REFERENCE = 5 × 150kΩ REFOUT *LIMIT CURRENT TO LESS THAN 10µA. FAST RESTORE CIRCUIT Figure 51. Typical Configuration of Right Leg Drive Circuit In two electrode configurations, A2 can be shut down by setting RLD SDN low for additional power savings. If left in shutdown, it is recommended to leave both RLD and RLDFB floating. Alternatively, RLD can be used to bias the inputs through 10 MΩ resistors, as described in the Leads On or Off Detection section. When the AD8233 is in shutdown and dc leads off detection mode, RLD pulls down towards ground. This pulldown acts as an LOD wake-up function, pulling the inputs down when the electrodes are reconnected. REFERENCE BUFFER Because of the low cutoff frequency used in high-pass filters in ECG applications, signals may require several seconds to settle. This settling time can result in a delay for the user after a step response, such as when the electrodes are first connected. This fast restore function is implemented internally, as shown in Figure 53. The output of the instrumentation amplifier is connected to a window comparator. The window comparator detects a saturation condition at the output of the instrumentation amplifier when its voltage approaches 0.1 V from either supply rail. FR B3 S1 +VS – 0.1V +IN SWITCH TIMING B5 The AD8233 operates from a single supply. To simplify the design of single-supply applications, the AD8233 includes a reference buffer to create a virtual ground between the supply voltage and the system ground. The signals present at the output of the instrumentation amplifier are referenced around this voltage. For example, if there is zero differential input voltage, the voltage at the output of the instrumentation amplifier is this reference voltage. IA C5 –IN Rev. D | Page 18 of 32 S2 IAOUT 0.1V LOD C1 Figure 53. Fast Restore Circuit 13737-053 A2 VCM 13737-051 18 D5 R1 × R2 × C1 R1 + R2 Disabling the AD8233 with the shutdown terminal does not discharge this capacitor. 1nF TO DRIVEN R* ELECTRODE A3 C1 Figure 52. Setting the Internal Reference C4 RLD REFIN 13737-052 The common-mode signal that is present across the inputs of the instrumentation amplifier is derived from the transconductance amplifier, GM1. It is then connected to the inverting input of A2 through a 150 kΩ resistor. Data Sheet AD8233 SATURATION DETECTED NO SATURATION tS1 S1 tS2 S2 13737-054 tRST LEADS OFF LEADS ON Figure 54. Timing Diagram for Fast Restore Switches (Time Base Not to Scale) If this saturation condition is present when both input electrodes are attached to the subject, the comparator triggers a timing circuit that automatically closes Switch S1 and Switch S2. See Figure 54 for the fast restore switches timing diagram. 10MΩ B5 IA C5 Figure 55. Circuit Configuration for Two Electrode DC Leads Off Detection For three electrode dc mode, each input must have a pull-up resistor connected to the positive supply. During normal operation, the potential of the subject must be inside the common-mode range of the instrumentation amplifier, which is only possible if a third electrode is connected to the output of the right leg drive amplifier. +VS If, by the end of the timing, the saturation condition persists, the cycle repeats. Otherwise, the AD8233 returns to its normal operation. If either of the leads off comparator outputs indicates that an electrode is disconnected, the timing circuit is prevented from triggering because it is assumed that no valid signal is present. To disable fast restore, drive the FR pin low or tie it permanently to GND. LEADS ON OR OFF DETECTION The AD8233 includes leads off detection. The AD8233 features ac and dc detection modes that both work with two and three electrode configurations. Ultralow power comparators allow the leads on or off detection to remain functional in shutdown mode, creating power savings at the system level when the LOD output is used as a wake-up signal for the microcontroller. DC Leads On or Off Detection The dc leads off detection mode can be used in two or three electrode configurations. This mode works by sensing when either instrumentation amplifier input voltage is within 0.27 V from the positive rail. The lowest power use case for the AD8233 is two electrode dc mode. A pull-up resistor on +IN and a pull-down resistor on −IN create a voltage divider when the electrodes are connected, setting the input common mode to midsupply. When the electrodes disconnect, the comparator monitoring +IN sets LOD high when the input pulls to +VS. 13737-055 10MΩ 10MΩ 10MΩ B5 IA C5 TO DRIVEN ELECTRODE D5 RLD 13737-056 These two switches (S1 and S2) enable two different 10 kΩ resistor paths: one between HPSENSE and IAOUT, and another between SW and REFOUT. During the time Switch S1 and Switch S2 are enabled, the internal resistors appear in parallel with their corresponding external resistors, forming high-pass filters. The result is that the equivalent lower resistance shifts the pole to a higher frequency, delivering a quicker settling time. The fast restore settling time depends on how quickly the internal 10 kΩ resistors of the AD8233 can drain the capacitors in the high-pass circuit. Smaller capacitor values result in a shorter settling time. +VS Figure 56. Circuit Configuration for Three Electrode DC Leads Off Detection The AD8233 indicates when any electrode is disconnected by setting the LOD pin high. To use this mode, connect the AC/DC pin to ground. AC Leads On or Off Detection The ac leads off detection mode is useful when using two electrodes. A conduction path must exist between the two electrodes, which is usually formed by two resistors, as shown in Figure 57. These resistors also provide a path for bias return on each input. Connect each resistor to REFOUT or RLD to maintain the inputs within the common-mode range of the instrumentation amplifier. Rev. D | Page 19 of 32 AD8233 Data Sheet STANDBY OPERATION +VS A2 The AD8233 includes a shutdown pin (SDN) that further enhances the flexibility and ease of use in portable applications where low power consumption is critical. A logic level signal can be applied to this pin to switch to shutdown mode. IA C5 10MΩ 13737-057 10MΩ C3 REFOUT Figure 57. Circuit Configuration for Two Electrode AC Leads Off Detection The AD8233 detects when an electrode is disconnected by forcing a 100 kHz current into the input terminals. This current flows through the external resistors from IN+ to IN− and develops a differential voltage across the inputs, which is then synchronously detected and compared to an internal threshold. The recommended value for these external resistors is 10 MΩ. Low resistance values make the differential drop too low to be detected and lower the input impedance of the amplifier. When the electrodes are attached to the subject, the impedance of this path must be less than 3 MΩ to maintain the drop below the threshold of the comparator. To use the ac leads off mode, tie the AC/DC pin to the positive supply rail. Although REFOUT is at a constant voltage value, using the RLD output as the input bias may be more effective in rejecting common-mode interference at the expense of additional power. Driving the SDN pin low places the AD8233 in shutdown mode and draws less than 1 µA of supply current, offering considerable power savings. To enter normal operation, drive SDN high. When not using this feature, permanently tie SDN to +VS. During shutdown operation, the AD8233 cannot maintain the REFOUT voltage, but it does not drain the REFIN voltage, thereby maintaining this additional conduction path from the supply to ground. When emerging from a shutdown condition, the charge stored in the capacitors on the high-pass filters can saturate the instrumentation amplifier and subsequent stages. The use of the fast restore feature helps reduce the recovery time and, therefore, minimize the amount of time powered on in power sensitive applications. Using leads on or off detection in shutdown mode allows system level power saving. The microcontroller enters sleep mode when the electrodes are disconnected, and the LOD signal acts as an interrupt to wake up the microcontroller. An example of this functionality is shown in Figure 59. In three electrode ac leads off detection mode, shown in Figure 58, pull-up resistors are not required, which improves the input impedance of the circuit. This mode is beneficial for dry electrode applications. The ac mode currents contribute flicker noise (1/f noise) to the system. Depending on the application, use ac leads off detection as a spot check and then switching to dc mode for improved ECG acquisition. OUT SDN LOD 3MCU WAKES UP AND SETS SDN HIGH. 1AT LEAST ONE ELECTRODE IS OFF AND THEREFORE LOD OUTPUT IS HIGH. MCU IS OFF AND SDN IS LOW. (AD8233 SHUTDOWN CURRENT < 1µA) 4AD8233 IS ACTIVE (~50µA) AND MONITORING ECG. +VS 2LOD GOES LOW WHEN BOTH ELECTRODES ARE CONNECTED. A FALLING EDGE AT LOD WAKES UP THE MCU. A2 250mV/DIV 400ms/DIV B5 IA C5 13737-059 B5 Figure 59. Electrode Connection and System Wakeup Sequence D5 RLD 13737-058 INPUT PROTECTION TO DRIVEN ELECTRODE Figure 58. Circuit Configuration for Three Electrode AC Leads Off Detection The ac leads off detection mode continues to function in shutdown mode as well. To keep the power under 1 µA, the clock is disabled and the ac currents become dc currents. The current source on +IN is 250 nA, while the current sink on –IN is −300 nA. The stronger pull-down current on −IN acts as a wake-up function, pulling LOD low when the electrodes are reconnected. All terminals of the AD8233 are protected against ESD. In addition, the input structure allows dc overload conditions that are a diode drop above the positive supply and a diode drop below the negative supply. Voltages beyond a diode drop of the supplies cause the ESD diodes to conduct and enable current to flow through the diode. Use an external resistor in series with each of the inputs to limit current for voltages beyond the supplies. In either scenario, the AD8233 safely handles a continuous 5 mA current at room temperature. For applications where the AD8233 encounters extreme overload voltages, such as in cardiac defibrillators, use external series resistors and gas discharge tubes (GDTs). Neon lamps are commonly used as an inexpensive alternative to GDTs. Rev. D | Page 20 of 32 Data Sheet AD8233 These devices can handle the application of large voltages but do not maintain the voltage below the absolute maximum ratings for the AD8233. A complete solution includes further clamping to either supply using additional resistors and low leakage diode clamps, such as BAV199 or FJH1100. As a safety measure, place a resistor between the input pin and the electrode that is connected to the subject to ensure that the current flow never exceeds 10 µA. Calculate the value of this resistor to be equal to the supply voltage across the AD8233 divided by 10 µA. In addition, excessive noise on the supply pins can adversely affect performance. As in all linear circuits, bypass capacitors must be used to decouple the chip power supplies. Place a 0.1 μF capacitor close to the supply pin. A 1 μF capacitor can be used farther away from the device. In most cases, the capacitor can be shared by other integrated circuits. Excessive decoupling capacitance increases power dissipation during power cycling. INPUT REFERRED OFFSETS RADIO FREQUENCY INTERFERENCE Because of its internal architecture, the instrumentation amplifier must always be used with the dc blocking amplifier, labeled HPA in Figure 50. Radio frequency (RF) rectification is often a problem in applications where there are large RF signals. The problem appears as a dc offset voltage at the output. The AD8233 has a 15 pF gate capacitance and 10 kΩ resistors at each input. This forms a low-pass filter on each input that reduces rectification at high frequency (see Figure 60) without the addition of external elements. The dc blocking amplifier attenuates the input referred offsets present at the inputs of the instrumentation amplifier, as described in the Theory of Operation section. However, this attenuation only occurs when the dc blocking amplifier is used as an integrator. In this case, the input offsets from the dc blocking amplifier are dominant and appear directly at the output of the instrumentation amplifier. +IN If the dc blocking amplifier is used as a follower instead of its intended function as an integrator, the input referred offsets of the in-amp are amplified by a factor of 100. 10kΩ CG –IN 10kΩ AD8233 IAOUT LAYOUT RECOMMENDATIONS 13737-060 CG Figure 60. RFI Filter Without External Capacitors For increased filtering, additional resistors can be added in series with each input. They must be placed as close as possible to the instrumentation amplifier inputs. These can be the same resistors used for overload and patient protection. POWER SUPPLY REGULATION AND BYPASSING The AD8233 is designed to be powered directly from a single 3 V battery, such as a CR2032. The AD8233 can also operate from rechargeable Li-Ion batteries, but the designer must take into account that the voltage during a charge cycle may exceed the absolute maximum ratings of the AD8233. To avoid damage to the device, use a power switch or a low power, low dropout regulator, such as the ADP150 or ADP160. It is important to follow good layout practices to optimize system performance. In low power applications, most resistors are of a high value to minimize additional supply current. The challenge of using high value resistors is that high impedance nodes become even more susceptible to noise pickup and board parasitics, such as capacitance and surface leakages. Keep all of the connections between high impedance nodes as short as possible to avoid introducing additional noise and errors from corrupting the signal. To maintain high CMRR over frequency, keep the input traces symmetrical and length matched. Place safety and input bias resistors in the same position relative to each input. In addition, the use of a ground plane significantly improves the noise rejection of the system. For WLCSP layout best practices, refer to the AN-617 Application Note. Rev. D | Page 21 of 32 AD8233 Data Sheet APPLICATION INFORMATION ELIMINATING ELECTRODE OFFSETS 50 To achieve offset rejection, connect a resistor/capacitor (RC) network between the output of the instrumentation amplifier (HPSENSE) and HPDRIVE, as shown in Figure 61. C 30 20dB PER DECADE 20 R A4 10 IAOUT 0 0.01 IN+ HPA B5 GM1 GM2 S1 R ELECTRODE OFFSETS 99R C1 = REFOUT Figure 61. Eliminating Electrode Offsets This RC network forms an integrator that feeds any dc signals that are not filtered back into the instrumentation amplifier, thus eliminating the offsets without saturating any node and maintaining high signal gain. In addition to blocking offsets present across the inputs of the instrumentation amplifier, this integrator also works as a high-pass filter that minimizes the effect of slow moving signals, such as baseline wander. The cutoff frequency of the filter is given by the following equation: fC = 10 100 Figure 62. Frequency Response of a Single-Pole DC Blocking Circuit 13737-061 IN– 1 FREQUENCY (Hz) C5 VCM 0.1 10kΩ 13737-062 B4 HPSENSE 100 2πRC As with any high-pass filter with low frequency cutoff, a fast change in dc offset requires a long time to settle. If such a change saturates the instrumentation amplifier output, the S1 switch briefly enables the 10 kΩ resistor path, thus moving the cutoff frequency to fC = 100(R + 104 ) 2πRC(104 ) For values of R greater than 100 kΩ, this expression can be approximated by fC = 1 200 πC This higher cutoff frequency reduces the settling time and enables faster recovery of the ECG signal. For more information, see the Fast Restore Circuit section. HIGH-PASS FILTERING where: R is in Ω. C is in farads. The filter cutoff is 100 times higher than is typically expected from a single-pole filter. Because of the feedback architecture of the instrumentation amplifier, the typical filter cutoff equation is modified by a gain of 100 from the instrumentation amplifier. The AD8233 can implement higher order high-pass filters. A higher filter order yields better artifact rejection, but increased signal distortion and more passive components on the PCB. Two-Pole High-Pass Filter A two-pole architecture can be implemented by adding a simple ac coupling RC at the output of the instrumentation amplifier, as shown in Figure 63. C1 A5 HPDRIVE D4 B4 A4 IAOUT HPSENSE HPA +IN TO NEXT STAGE C2 R1 S1 10kΩ SW 10kΩ R2 S2 B5 C5 REFOUT C3 –IN = REFOUT Figure 63. Schematic for a Two-Pole High-Pass Filter Rev. D | Page 22 of 32 13737-063 A5 HPDRIVE 40 MAGNITUDE (dB) The instrumentation amplifier in the AD8233 is designed to apply gain and to filter out near dc signals simultaneously. This capability allows the device to amplify a small ECG signal by a factor of 100 while rejecting electrode offsets as large as ±300 mV. Data Sheet AD8233 The right side of C2 connects to the SW terminal. As with S1, S2 reduces the recovery time for the ac coupling network by placing 10 kΩ in parallel with R2. See the Fast Restore Circuit section for additional details on switch timing and trigger conditions. If the passive network is not buffered, the network exhibits higher output impedance at the input of a subsequent low-pass filter, as with Sallen-Key filter topologies. Careful component selection results in reliable performance without a buffer. See the Low-Pass Filtering and Gain section for additional information on component selection. Additional High-Pass Filtering Options In addition to the topologies explained in previous sections, an additional pole may be added to the dc blocking circuit for the rejection of low frequency signals. This configuration is shown in Figure 64. C1 R1 The cutoff frequency is located at fC = 10 2π R1 × C1 × R2 × C2 The selection of RCOMP to be 0.14 times the value of the other two resistors optimizes the filter for a maximally flat pass band. Reduce the value of RCOMP to increase the Q and, consequently, the peaking of the filter. A very low RCOMP value may result in an unstable circuit. The selection of values based on these criteria results in a transfer function similar to what is shown in Figure 66. When additional low frequency rejection is desired, a highorder, high-pass filter can be implemented by adding an ac coupling network at the output of the instrumentation amplifier, as shown in Figure 65. The SW terminal is connected to the ac coupling network to obtain the best settling time response when fast restore engages. C1 R1 TO NEXT STAGE C3 R2 TO NEXT STAGE R2 RCOMP RCOMP A5 HPDRIVE A4 A5 HPDRIVE +IN IAOUT HPA 10kΩ S1 B4 D4 SW 10kΩ IAOUT D4 B4 HPSENSE A4 HPSENSE SW 10kΩ C2 HPA +IN S1 10kΩ C2 R3 S2 B5 S2 B5 C5 REFOUT C3 –IN C5 13737-065 REFOUT C3 = REFOUT Figure 65. Schematic for a Three-Pole, High-Pass Filter Figure 64. Schematic for an Alternative Two-Pole, High-Pass Filter 60 With this circuit topology, the filter attenuation reverts to a single-pole roll-off at very low frequencies. Because the initial roll-off is 40 dB per decade, this reversion to 20 dB per decade has little impact on the ability of the filter to reject out-of-band low frequency signals. The designer may choose different values to achieve the desired filter performance. To simplify the design process, use the following recommendations as a starting point for component value selection. R1 = R2 ≥ 100 kΩ C1 = C2 RCOMP = 0.14 × R1 40dB PER DECADE 40 MAGNITUDE (dB) An extra benefit of this circuit topology is that it allows a lower cutoff frequency with lower R and C values. The resistor, RCOMP, can also be used to control the quality factor (Q) of the filter to achieve narrow band-pass filters (for heart rate detection) or maximum pass-band flatness (for cardiac monitoring). 20 60dB PER DECADE 20dB PER DECADE 0 –20 –40 –60 0.01 40dB PER DECADE 0.1 THREE-POLE FILTER TWO-POLE FILTER 1 FREQUENCY (Hz) 10 100 13737-066 = REFOUT 13737-064 –IN Figure 66. Frequency Response of the Circuits Shown in Figure 64 and Figure 65 Careful analysis and adjustment of all of the component values in practice is recommended to optimize the filter characteristics. To reduce the value of RCOMP, increase the peaking of the active filter to overcome the additional roll-off introduced by the ac coupling network. Proper adjustment yields the best pass-band flatness. Rev. D | Page 23 of 32 AD8233 Data Sheet Table 6. Comparison of High-Pass Filtering Options Filter Order 1 2 2 3 Figure to Reference Figure 61 Figure 63 Figure 64 Figure 65 1 2 Component Count 2 4 5 7 Capacitor Sizes/Values Large Large Smaller Smaller Low Frequency Rejection Good Better Better Best Signal Distortion1 Low Medium Medium Highest Output Impedance2 Low Higher Low Higher The signal distortion is for the equivalent corner frequency location. Output impedance refers to the drive capability of the high-pass filter before the low-pass filter. Low output impedance is desirable to allow flexibility in the selection of values for a low-pass filter, as explained in the Low-Pass Filtering and Gain section. The design of the high-pass filter involves trade-offs between signal distortion, component count, low frequency rejection, and component size. For example, a single-pole, high-pass filter results in the least distortion to the signal, but the associated rejection of low frequency artifacts is the lowest of the available filter options. Table 6 compares the recommended filtering options. narrow-band applications to increase peaking and the selectivity of the band-pass filter. A common design procedure is to set R1 = R2 = R and C1 = C2 = C, simplifying the expressions for the cutoff frequency and Q to fC = 1/(2πRC) Q= LOW-PASS FILTERING AND GAIN The AD8233 includes an uncommitted op amp that can be used for extra gain and filtering. For applications that do not require a high order filter, a simple RC low-pass filter is sufficient, and the op amp can buffer or further amplify the signal. FROM IN-AMP STAGE R FILTERED SIGNAL A1 C 13737-067 REFOUT Figure 67. Schematic for a Single-Pole, Low-Pass Filter and Additional Gain A Sallen-Key filter topology can be implemented for applications that require a steeper roll-off or a sharper cutoff frequency, as shown in Figure 68. R2 FILTERED SIGNAL A1 R3 REFOUT R4 13737-068 C2 Figure 68. Schematic for a Two-Pole, Low-Pass Filter The following equations describe the low-pass cutoff frequency (fC), gain, and Q: fC = These design equations only hold true in a case where the output impedance of the previous stage is much lower than the input impedance of the Sallen-Key filter. The design equations do not hold true when using an ac coupling network between the instrumentation amplifier output and the input of the lowpass filter without a buffer. To connect these two filtering stages properly without a buffer, make the value of R1 at least 10 times larger than the resistor of the ac coupling network (labeled as R2 in Figure 63). The ability of AD8233 to drive capacitive loads makes it ideal for driving an ADC without an additional buffer. However, depending on the input architecture of the ADC, a simple, lowpass RC network may be required to decouple the transients from the switched capacitor input that are typical of modern ADCs. This RC network also acts as an additional filter that can help reduce noise and aliasing. Follow the recommended guidelines from the ADC in use for the selection of proper R and C values. Table 7 lists compatible ADCs by category. Table 7. Compatible ADCs by Category 1 2π R1 × C1 × R2 × C 2 ADCs AD7091 AD7988-1 AD7682 AD7689 Gain = 1 + R3/R4 Q= Q can be controlled by setting the gain with R3 and R4, but this setting limits the gain to be less than 3. The circuit becomes unstable for gain values equal to or greater than 3. A simple modification that allows higher gains is to make the value of C2 at least four times larger than C1. Driving ADCs C1 FROM IN-AMP STAGE R1 1 3 − Gain R1 × C1 × R2 × C2 R1 × C2 + R2 × C2 + R1 × C1(1 − Gain) Changing the gain has an effect on Q and vice versa. Common values for Q are 0.5, to avoid peaking, or 0.7 for maximum flatness and a sharp cutoff frequency. Use a high Q value in Rev. D | Page 24 of 32 Microcontrollers ADuCM350 ADuCM3029 ADuCM4050 Optical/ Bio-Z Sensors ADPD1081 ADPD188GG ADPD1080 ADPD4000 AD5940 Accelerometers ADXL363 Data Sheet AD8233 APPLICATION CIRCUITS AD8233 Heart Rate Measurement (HRM) Next to the Heart C Figure 69. Driving an ADC DRIVEN ELECTRODE A driven electrode (or reference electrode) is often used to minimize the effects of common-mode voltages induced by the power line and other interfering sources. The AD8233 extracts the common-mode voltage from the instrumentation amplifier inputs and makes it available through the RLD amplifier to drive an opposing signal into the patient. This functionality maintains the voltage between the patient and the AD8233 at a near constant, greatly improving the CMRR. For wearable exercise devices, the AD8233 is typically placed in a pod near the heart. The two sense electrodes are placed under the pectoral muscles, and no driven electrode is used. Because the distance from the heart to the AD8233 is small, the heart signal is strong and there is less muscle artifact interference. In this wearable device configuration, space is at a premium. By using as few external components as possible, the circuit shown in Figure 70 is optimized for size. 0.22µF ELECTRODE INTERFACE 10MΩ As a safety measure, place a resistor between the RLD pin and the electrode connected to the subject to ensure that current flow never exceeds 10 µA. Calculate the value of this resistor to be equal to the supply voltage across the AD8233 divided by 10 µA. The AD8233 implements an integrator formed by an internal 150 kΩ resistor and an external capacitor to drive this electrode. The choice of the integrator capacitor is a trade-off between line rejection capability and stability. It is recommended that the capacitor be small to maintain as much loop gain as possible, around 50 Hz and 60 Hz, which is typical for line frequencies. For stability, it is recommended that the gain of the integrator be less than unity gain at the frequency of any other poles in the loop, such as those formed by the capacitance and the safety resistors of the patient. The suggested application circuits use a 1 nF capacitor, which results in a loop gain of about 20 at line frequencies, with a crossover frequency of about 1 kHz. In a 2-lead configuration, the RLD pin amplifier can be shut down or used to drive the bias current resistors on the inputs. Although not as effective as a true driven electrode, this configuration can provide some common-mode rejection improvement if the sense electrode impedance is small and well matched. MEASURING SURFACE ELECTROMYOGRAPHY (EMG) OR ELECTROENCEPHALOGRAPHY (EEG) HPDRIVE 180kΩ 180kΩ 10MΩ HPSENSE +IN IAOUT –IN REFIN +VS RLDFB 10MΩ +VS 10MΩ 0.1µF 1nF GND RLD SW 0.1µF 10MΩ AD8233 FR OPAMP+ AC/DC REFOUT RLD SDN OPAMP– SDN OUT LOD +VS TO DIGITAL INTERFACE 13737-070 ADC SIGNAL OUTPUT Figure 70. Circuit for HRM Next to the Heart A shorter distance from the AD8233 to the heart makes this application less vulnerable to common-mode interference. However, because RLD is not used to drive an electrode, it can be used to improve the common-mode rejection by maintaining the midscale voltage through the 10 MΩ bias resistors. Alternatively, tie RLD SDN low to save power, and tie the bias resistors to REFOUT. A single-pole, high-pass filter is set at 7 Hz, and there is no lowpass filter. No gain is used on the output op amp, which reduces the number of resistors for a total system gain of 100, as shown in Figure 71. 70 60 50 MAGNITUDE (dB) Due to its flexible architecture, the AD8233 filters can be configured to measure other biopotential signals, such as surface EMG or nondiagnostic EEG (alpha or beta waves). The frequency range of signals for surface EMG is typically 2 Hz to 500 Hz for skeletal muscles and 0.01 Hz to 1 Hz for smooth muscles. When measuring wider bandwidth signals inclusive of 50 Hz or 60 Hz, consider lower gain settings. EEG signals have a shared frequency range with ECG signals. However, the amplitude for EEG signals is about 10 times smaller than those for ECG, and as such, require a lower noise solution. Alpha waves (8 Hz to 13 Hz) and beta waves (14 Hz to 40 Hz) can be measured with the AD8233 by setting the high-pass filter at 7 Hz, similar to ECG measurement at the hands (see Figure 72). The 7 Hz cutoff frequency helps remove additional 1/f noise, which lowers the noise floor for the EEG measurement. Rev. D | Page 25 of 32 40 30 20 10 0 0.1 1 10 100 1k 10k FREQUENCY (Hz) Figure 71. Frequency Response for HRM Next to the Heart Circuit 13737-071 R D1 13737-069 A1 AD8233 Data Sheet The input terminals in this configuration use two 180 kΩ resistors to protect the user from fault conditions. Two 10 MΩ resistors provide input bias. Use higher values for electrodes with high output impedance, such as cloth electrodes. The overall narrow-band nature of the two-pole, low-pass filter combination distorts the ECG waveform significantly. Therefore, it is only suitable to determine the heart rate, and not to analyze the ECG signal characteristics. Figure 70 also shows two 10 MΩ resistors to set the midscale reference voltage. If there is already a reference voltage available, it can be driven into the REFIN input to eliminate these two 10 MΩ resistors. The low-pass filter stage also includes a gain of 11, bringing the total system gain close to 1100. Because the ECG signal is measured at the hands, it is weaker than when measured closer to the heart. Exercise Application, HRM at the Hands The RLD circuit drives to the third electrode, which can also be located at the hands, to cancel common-mode interference. In this application, the heart rate signal is measured at the hands with stainless steel electrodes. The arm and upper body movement of the user create large motion artifacts, and the long lead length makes the system susceptible to commonmode interference. A very narrow band-pass characteristic is required to separate the heart signal from interference. The circuit in Figure 75 is designed for monitoring the shape of the ECG waveform. To obtain an ECG waveform with minimal distortion, the AD8233 is configured with a 0.5 Hz, single-pole, high-pass filter, followed by a two-pole, 40 Hz, low-pass filter. A third electrode is driven for optimum common-mode rejection. 0.22µF 10MΩ HPDRIVE 10MΩ 180kΩ 180kΩ 0.22µF 360kΩ RL IAOUT –IN REFIN 60 10MΩ 0.1µF 50 AD8233 FR OPAMP+ AC/DC REFOUT RLD SDN OPAMP– SDN 22nF 100kΩ 3.3nF 10MΩ 0.1µF 1MΩ 100kΩ +VS GND RLD SW 1MΩ +VS RLDFB 1nF 70 10MΩ MAGNITUDE (dB) RA HPSENSE +IN 1MΩ OUT +VS LOD SIGNAL OUTPUT 10 0 0.01 The circuit shown in Figure 72 uses a two-pole, high-pass filter set at 7 Hz. A two-pole, low-pass filter at 24 Hz follows the high-pass filters to eliminate any other artifacts and line noise. 70 60 MAGNITUDE (dB) 50 1k 13737-073 10 100 10 100 1k In addition to 40 Hz filtering, the op amp stage is configured for a gain of 2, resulting in a total system gain of 200. Keeping the gain lower helps with any motion artifacts picked up in band. To optimize the dynamic range of the system, the gain level is adjustable, depending on the input signal amplitude (which may vary with electrode placement) and ADC input range. 20 10 1 Figure 74. Frequency Response of Holter Monitor Circuit 30 FREQUENCY (Hz) 0.1 FREQUENCY (Hz) 40 1 30 20 Figure 72. Circuit for HRM at Hands 0 0.1 40 TO DIGITAL INTERFACE 13737-072 LA 13737-075 +VS Holter Monitor Configuration Figure 73. Frequency Response for HRM Circuit Taken at the Hands Rev. D | Page 26 of 32 Data Sheet AD8233 3.3µF +VS ELECTRODE INTERFACE 10MΩ 10MΩ HPDRIVE 150kΩ LA 300kΩ RL +IN IAOUT –IN +VS RLDFB 10MΩ 10MΩ REFIN 0.1µF GND AD8233 SW 1MΩ 0.1µF 10MΩ 1nF RLD 499kΩ 10MΩ +VS (= +2.5V) 150kΩ RA HPSENSE +VS FR OPAMP+ AC/DC REFOUT RLD SDN OPAMP– SDN ...TO MCU OUT LOD ...TO MCU 5.6nF 1MΩ 5.6nF 1MΩ +VS 13737-074 ...TO 11-13 ENOB ADC 0.5Hz TO 40Hz GAIN = ×200 Figure 75. Holter Monitor Circuit 0.22µF ADP150 +VS ELECTRODE INTERFACE VOUT 1.8V TO 3.5V HPDRIVE 150kΩ LA 150kΩ RA 10MΩ 10MΩ 10MΩ 1nF 0.22µF +IN IAOUT –IN +VS 100kΩ 1MΩ RLDFB REFIN 11nF 1MΩ 1µF VBATT 10MΩ 10MΩ 0.1µF 0.1µF GND RLD AD8233 FR OPAMP+ AC/DC +VS REFOUT RLD SDN +VS OPAMP– SDN SPI0_RDY/GPIO30 OUT LOD XINT0_WAKE1/GPIO16 22nF 100kΩ 1µF 10MΩ +VS SW 150kΩ HPSENSE VIN GND ADuCM3029 +VS 0.1µF VBAT_ADC 7Hz TO 26Hz GAIN = ×1100 0.1µF VBAT_ANA1 VBAT_ANA2 VBAT_DIG1 ADC0_VIN0 SELECT VREF_ADC = 2.5V 0.1µF 0.1µF 0.1µF VBAT_DIG2 4.7µF ADXL362 1µF 0.1µF VS VDDI/O GND XINT0_WAKE0/GPIO15 INT2 CS SPI1_CS0_GPIO25 MISO SPI1_MISO_GPIO24 MOSI SPI1_MOSI_GPIO23 SCLK SPI1_CLK_GPIO22 TO HOST, MEMORY OR DISPLAY GND_VREFADC GND_ANA GND_DIG 0.47µF CS SPI2_CS0_GPIO21 TX SPI2_MISO_GPIO20 RX SPI2_MOSI_GPIO19 CLK SPI2_CLK_GPIO18 Figure 76. Portable Heart Rate and Activity Monitor Circuit Rev. D | Page 27 of 32 VDCDC_OUT 0.47µF VLDO_OUT 13737-076 +VS VREF_ADC AD8233 Data Sheet Portable Heart Rate and Activity Monitor System to one of the current inputs of the ADPD1080 through a 200 kΩ resistor to convert the voltage output of the AD8233 into a current. The ADPD1080 is configured to alternately measure the photodiode signal and the ECG signal from the AD8233 on consecutive timeslots to provide fully synchronized PPG and ECG measurements. Data can be read out of the on-chip FIFO or straight from data registers. The ADPD1080 channel used to process the ECG signal is set up in TIA ADC mode, and the input bias voltage must be set to the 0.90 V setting using Bits[5:4] of Register 0x42 if the ECG signal is on Time Slot A, or Register 0x44 on Time Slot B. The TIA gain setting can be set to optimize the dynamic range of the signal path. The channel used to process the PPG signal is configured in its normal operating mode. Figure 77 shows a plot of a synchronized ECG and PPG measurement using the AD8233 with the ADPD1080. The circuit in Figure 76 shows the AD8233 configured as an HRM circuit for two electrode applications. The AD8233 implements a two-pole, low-pass filter with a cutoff frequency of 7 Hz, and a two-pole, low-pass filter with a cutoff frequency of 26 Hz. The total signal gain in the pass band is 1,100. The output of this HRM circuit is sampled by the 12-bit ADC embedded in the ADuCM3029, an ultralow power (
AD8233ACBZ-R7 价格&库存

很抱歉,暂时无法提供与“AD8233ACBZ-R7”相匹配的价格&库存,您可以联系我们找货

免费人工找货