0
登录后你可以
  • 下载海量资料
  • 学习在线课程
  • 观看技术视频
  • 写文章/发帖/加入社区
会员中心
创作中心
发布
  • 发文章

  • 发资料

  • 发帖

  • 提问

  • 发视频

创作活动
EVAL-ADAS1000SDZ

EVAL-ADAS1000SDZ

  • 厂商:

    AD(亚德诺)

  • 封装:

    -

  • 描述:

    BOARD EVAL ECG AFE FOR ADAS1000

  • 数据手册
  • 价格&库存
EVAL-ADAS1000SDZ 数据手册
EVAL-ADAS1000SDZ User Guide UG-426 One Technology Way • P.O. Box 9106 • Norwood, MA 02062-9106, U.S.A. • Tel: 781.329.4700 • Fax: 781.461.3113 • www.analog.com Evaluating the ADAS1000 ECG Front End for Demonstration and Development FEATURES GENERAL DESCRIPTION Biopotential signals in; digitized signals out 6 electrodes (5 acquisition channels and 1 driven lead) 2 × ADAS1000 (master/slave) for 1 to 12 lead electrode measurements AC and dc lead off detection Pace detection Optional thoracic impedance measurement (internal/external) Selectable reference lead Lead or electrode data available Calibration features (DAC and 1 mV square/sinewave) Low or high speed data output rates Serial interface SPI®-/QSPI™-/DSP-compatible The primary function of the EVAL-ADAS1000SDZ evaluation board is to demonstrate the ADAS1000 integrated electrocardiogram (ECG) device for medical instrumentation. A full description of the ADAS1000 is available in the data sheet, and should be consulted when utilizing this user guide. This evaluation board is used to evaluate the ADAS1000 and all variants. CUSTOMER INTERFACING OPTIONS Direct access to ADAS1000 serial interface (J4) USB interface via Analog Devices, Inc., Blackfin®-based SDP-B controller board and GUI APPLICATIONS ECG Patient monitor Holter monitor Cardiac defibrillators PLEASE SEE THE LAST PAGE FOR AN IMPORTANT WARNING AND LEGAL TERMS AND CONDITIONS. The evaluation board is ideal for exploring concepts and adopting the ADAS1000 into advanced medical systems. The board can be operated as part of an end user’s system via the ADAS1000 serial peripheral interface (SPI), or as a standalone evaluation of the ADAS1000 using the support of the Analog Devices system demonstration platform controller board (SDP-B), and a standard PC (running Windows® Vista, Windows 7 32-bit and 64-bit, or Windows 8) to run the Analog Devices evaluation software. The SDP-B board, which is a DSP-based controller board is separate from the evaluation board, is required to run the Analog Devices evaluation software, and is used for data transfer from evaluation board to PC via USB interface. It does not come as part of the evaluation kit and can be ordered separately (EVAL-SDP-CB1Z). The SDP-B board can be reused with many other Analog Devices evaluation modules. For more information, visit the Analog Devices healthcare home page. Rev. B | Page 1 of 44 UG-426 EVAL-ADAS1000SDZ User Guide TABLE OF CONTENTS Features .............................................................................................. 1 ADAS1000 Software Operation.................................................... 10 Customer Interfacing Options ........................................................ 1 Quick Operation of ADAS1000 Software ............................... 10 Applications ....................................................................................... 1 Detailed Description of Main Control Panel.......................... 10 General Description ......................................................................... 1 Control of all Registers .............................................................. 12 Revision History ............................................................................... 2 ECG Capture ............................................................................... 14 Conditions Regarding the Use of This Product in Healthcare Applications................................................................................... 3 ECG Capture with Digital Post Processing............................. 15 Evaluation Kit Contents............................................................... 3 Respiration .................................................................................. 17 Hardware Requirements .............................................................. 3 Pace .............................................................................................. 18 Evalution Board Software ................................................................ 5 Understanding Pace in the ADAS1000 ................................... 19 Installing the Software: Overview .............................................. 5 Write to a File .............................................................................. 20 Installing the Software: Details ................................................... 5 Test Tones .................................................................................... 23 Connecting to the Board ............................................................. 6 Common-Model Level/Wilson Central Terminal ................. 25 Running the Software .................................................................. 6 Detailed Description ...................................................................... 26 Evaluation Board Hardware ............................................................ 7 Overview of Schematics ............................................................ 26 Key Features .................................................................................. 7 Pace Interface (Optional) .......................................................... 26 Connectors .................................................................................... 7 Power............................................................................................ 26 Connecting an ECG Signal ......................................................... 8 Flexible Respiration Feature on Master ADAS1000 .............. 27 RESET Buttons .............................................................................. 8 Troubleshooting.......................................................................... 33 LED................................................................................................. 8 Evaluation Board Schematics........................................................ 35 Leads OFF Control ..................................................................... 16 Jumpers .......................................................................................... 9 REVISION HISTORY 8/2012—Revision 0: Initial Version 11/2018—Rev. A to Rev. B Changes to General Description Section ...................................... 1 Deleted Clamp Section ............................................................................. 34 Deleted Figure 45; Renumbered Sequentially ..................................... 34 3/2014—Rev. 0 to Rev. A Changed ADAS1000SDZ to EVAL-ADAS1000 SDZ Throughout ......................................................................... Universal Change to Title .................................................................................. 1 Changes to Figure 46 ...................................................................... 35 Changes to Figure 52 ...................................................................... 41 Changes to Figure 53 ...................................................................... 42 Rev. B | Page 2 of 44 EVAL-ADAS1000SDZ User Guide UG-426 WALL POWER, BENCH SUPPLY OR BATTERY (J7 OR J9) 4.5V TO 5.5V CIRCUIT/LINK AREA FOR ENHANCED RESPIRATION GND ON BOARD REGULATOR AND LDOS AVDD: 3.3V IOVDD: 3.3V INPUT RC COMPONENT SPACE FOR CABLE MODEL, FILTER MODEL ADAS1000 RESET OPTIONAL HEADER CONNECTOR FOR RESPIRATION (EXT) (J8) EXT_RESP_RA EXT_RESP_LA EXT_RESP_LL ADAS1000 MASTER RA, LA, LL, V1, V2, RLD RLD ECG1 (LA) ECG 2 (LL) ECG 3 (RA) ECG 4 (V1) ECG 5 (V2) ADAS1000 DIRECT MASTER INTERFACE FOR FAST PACE CONNECTOR (OPTIONAL) (J6) CRYSTAL DB15 CONNECTOR FOR ECG (J1) ADJUSTABLE RLD GAIN SETTING COMPONENTS ECG 6 (V3) ECG 7 (V4) ECG 10 (SPARE) ADAS1000 DIRECT SPI CONNECTOR (OPTIONAL) (J4) SDP RESET SDP CONNECTOR 10810-001 ADAS1000 SLAVE V3, V4, V5, V6, SPARE ECG 8 (V5) ECG 9 (V6) Figure 1. Functional Block Diagram CONDITIONS REGARDING THE USE OF THIS PRODUCT IN HEALTHCARE APPLICATIONS In addition to the terms found at the end of this document, the following shall also apply to your use of the board and design: This evaluation board design is being provided “as is” without any expressed or implied representations or warranties of any kind and the use of this board or design shall impose no legal obligation on Analog Devices, Inc., and its subsidiaries, employees, directors, officers, servants and agents. In addition, it is understood and agreed to that the evaluation board or design is not authorized for use in safety critical healthcare applications (such as life support) where malfunction or failure of a product can be expected to result in personal injury or death. This board must not be used for diagnostic purposes and must not be connected to a human being or animal. It must not be used with a defibrillator or other equipment that produces high voltages in excess of the supply rails on the board. This evaluation board is provided for evaluation and development purposes only. It is not intended for use or as part of an end product. Any use of the evaluation board or design in such applications is at your own risk and you shall fully indemnify Analog Devices, Inc., its subsidiaries, employees, directors, officers, servants and agents for all liability and expenses arising from such unauthorized usage. You are solely responsible for compliance with all legal and regulatory requirements connected to such use. EVALUATION KIT CONTENTS • • • EVAL-ADAS1000SDZ board Medical-grade universal ac-to-dc wall adaptor (+5 V) CD that includes • Self-installing graphical user interface (GUI) software that allows users to read/write to ADAS1000 and to stream data • Electronic version of the ADAS1000 data sheet • Electronic version of the EVAL-ADAS1000SDZ documentation HARDWARE REQUIREMENTS • • • Rev. B | Page 3 of 44 Power supply: +5 V (ac-dc adaptor provided) Patient simulator or similar device SDP board, a controller board for data transfer to PC. This can be ordered separately (EVAL-SDP-CB1Z) UG-426 EVAL-ADAS1000SDZ User Guide POWER CONNECTOR, 5V (J9) EXTRA LINKS/CIRCUIT RELATED TO EXTERNAL RESPIRATION MEASUREMENTS OPTIONAL BENCH SUPPLY FOR ADAS CHIPS (J6) ADAS1000 EVALUATION BOARD OPTIONAL BENCH SUPPLY (5V) (J7) 5V WALL ADAPTOR EXTERNAL RESPIRATION CONNECTOR (J8) ON BOARD DC-DC REGULATORS RESET FOR ADAS CIRCUIT ELECTRODE CONNECTOR DB15 (J1) SDP BOARD PATIENT CABLE ADAS1000 MASTER ELECTRODES LA, RA, LL, V1, V2, RLD RLD EXT COMPONENTS ADAS1000 SLAVE ELECTRODES V3, V4, V5, V6 SPARE ADAS1000 SPI INTERFACE(J4) RESET BUTTON FOR SDP BOARD Figure 2. ADAS1000 Evaluation Board/SDP Board Rev. B | Page 4 of 44 USB CABLE TO PC USB CONNECTOR 10810-002 SPACE FOR CABLE/ESIS MODELLING EVAL-ADAS1000SDZ User Guide UG-426 EVALUTION BOARD SOFTWARE INSTALLING THE SOFTWARE: OVERVIEW The EVAL-ADAS1000SDZ kit includes self-installing software on CD. Install the software prior to connecting the SDP board to the USB port of the PC. This ensures that the SDP board is recognized when it connects to the PC. 1. Start the Windows operating system and insert the CD. 2. The installation software should launch automatically. If it does not, run the setup.exe file from the CD. 10810-005 INSTALLING THE SOFTWARE: DETAILS Proceed through the installation steps allowing the software and drivers to be placed in the appropriate locations. Connect the SDP board to the PC only after the software and drivers have been installed. Figure 5. License Agreement 10810-006 There are two parts to the software installation procedure. The first part is installing the software related to the ADAS1000 evaluation board as shown in Figure 3. 10810-003 Figure 6. Monitoring Progress Now, install the software related to the SDP controller board (see Figure 7). 10810-004 10810-007 Figure 3.Getting Started with the Software Installation Figure 4. Destination Directory Rev. B | Page 5 of 44 Figure 7. SDP Software Install UG-426 EVAL-ADAS1000SDZ User Guide If the SDP board is not connected to the USB port when the software is launched, a connectivity error is displayed (see Figure 9). Simply connect the evaluation board to the USB port of the PC, wait a few seconds, click Rescan, and follow the instructions that appear. Plug in the SDP board via the USB cable provided and allow the new Found Hardware Wizard to run; this detects and loads drivers for the SDP board. If necessary, check that the drivers and the board are connected correctly by looking at the Device Manager of the PC. 10810-008 10810-009 The Device Manager can be found by right clicking on My Computer > Manage > Device Manager from the list of System Tools. The SDP board should appear under ADI Development Tools. Figure 8. Device Manager Figure 9. Pop-Up Window Error 2. When the board is found, click Select. CONNECTING TO THE BOARD Follow these steps to power up and start interfacing to the board using the software: After installation is completed, plug the EVALADAS1000SDZ into the SDP controller board using J2 of the ADAS1000 board. Use the plastic screws provided on the ADAS1000 evaluation board to fix the two boards securely in place. 2. Power up the evaluation board as described in the Evaluation Board Hardware section and the Power Connections section. D10 and LED1should appear lit. 3. Plug the SDP board into the PC using the USB cable included in the box. 4. When the software detects the evaluation board, proceed through any dialog boxes that appear in order to finalize the installation. 10810-010 1. Figure 10. Hardware Selection The software connects to the board and displays the following: RUNNING THE SOFTWARE 1. Select Start > All Programs > Analog Devices > ADAS1000 > ADAS1000_Software. To uninstall the program, select Start > Control Panel > Add or Remove Programs > ADAS1000_Software. 10810-011 Follow these steps to run the program: Figure 11. Wait Message Once the board has been correctly detected, the ADAS1000 panel opens (see Figure 13). Rev. B | Page 6 of 44 EVAL-ADAS1000SDZ User Guide UG-426 EVALUATION BOARD HARDWARE Table 1. Power Supply Connection (Choose Only One) KEY FEATURES • • • • • • • 5 or 10 ECG electrode paths capable of demonstrating 1 lead to 12 leads of ECG data AC and DC lead off detection Respiration and pace measurement and display Calibration features and display Test tones Real-time ECG electrode or lead display on PC screen via Analog Devices ADAS1000 GUI (lead calculation available). Recording of ECG data for offline review Power Connections AVDD Supply Range 4.5 V to 5.5 V (may be supplied from wall adaptor, battery or bench supply). If applied, this is the only supply rail required by the board. +3 V to +5.5 V IOVDD ADCVDD DVDD 1.65 V to 3.6 V 1.8 V ±5% 1.8 V ±5% Electrode Connector–J1 This connector provides the primary analog input interface to which customer proprietary lead sets are connected. Choose one of the following: • Alternative Supplies (J5) Parameter +5 V CONNECTORS There are a number of options for supplying power to the board. The simplest is directly from the J9 dc jack connector with the wall adapter which is provided as part of the kit. • Supply Requirement Primary Supply (J7 or J9) J9: dc jack–requires 5 V at 250 mA. Note that this provides power to the SDP board in addition to the ADAS1000 related circuitry. J7: screw terminal (2 inputs)–requires 5 V at 250 mA. Provides power to the on-board dc-dc convertors which supply all circuitry on board and also powers the SDP board. • J5: screw terminal (6 inputs). These inputs are optional supply inputs for the ADAS1000 devices and the remainder of the circuitry on the board • AVDD = 3 V to 5.5 V • IOVDD = 1.65 V to 3.6 V • AGND = DGND = 0 V • ADCVDD and DVDD are optional supplies. They can be supplied from the ADAS1000 on-chip regulators. Alternatively, the regulators can be disabled and the user can drive ADCVDD and DVDD directly via J5. If ADCVDD and DVDD are driven directly, then the following supplies are required. • ADCVDD (optional) = 1.8 V • DVDD (optional) = 1.8 V 10810-012 Alternatively, choose Figure 12. J1, DB15 Connector Connector J1 is a DB15 female connector and mates with a D-SUB plug. All ADAS1000 electrode connections are made available here for both master and slave devices in addition to the Right Leg Drive (RLD_OUT) and Shield Drive (SHIELD) pins. Note that every effort was made to supply input protection to the electrode pins sufficient for the application; however, the intent was not to offer this module as a true medical solution. Therefore, no defibrillation pulses or voltages outside the ADAS1000 operating range should be applied to the input connector/board. Rev. B | Page 7 of 44 UG-426 EVAL-ADAS1000SDZ User Guide CONNECTING AN ECG SIGNAL Table 3. SPI Connector, J4 The user needs to connect a signal source to the evaluation board connector J1 for measurement purposes. Ideally, this would be a patient simulator. In demonstrations at Analog Devices, the PS420 patient simulator (from Fluke Biomedical Division of Fluke Electronics Corporation) is used. Pin No. 1 2 3 4 5 6, 12, 13, 14 7 8 9 10 11 Note that the board is not designed for direct connection to patients or animals for testing. Users should connect the appropriate signal to the ECG electrode inputs and RLD_OUT electrode. Table 2. Electrode Connector, J1 Pin No. 1 2 3 4 5 6 7 8 9 Mnemonic V2 V3 V4 V5 V6 SHIELD CE NC RA 10 11 12 13 LA LL V1 Spare 14 15 RLD NC Description Analog input, Master ECG5_V2 Analog input, Slave ECG1_V3 Analog input, Slave ECG2_V4 Analog input, Slave ECG3_V5 Analog input, Slave ECG4_V6 Output of shield driver Common electrode, Master CM_IN Not connected Analog input, right arm, Master ECG3_RA Analog input, left arm, Master ECG1_LA Analog input, left leg, Master ECG2_LL Analog input, Master ECG4_V1 Analog input, chest electrode or auxiliary bio-potential input, Slave ECG5 Right leg drive, RLD_OUT Not connected SDP Interface Connector, J2 The purpose of this connector is to facilitate interfacing with the Analog Devices SDP1Z control board which is USB controlled. This control board is specific to the operation of this module as a standalone evaluation and learning platform. This connector is not intended for customer-specific interfacing. Main ADAS SPI, J4 This connector provides the ADAS1000 digital interface pins so that the device may be used in standalone mode (without the SDP control board). The user may use this connecter to interface to the device in order to develop their own code and evaluate the ADAS1000 directly. Note that on the board, the /CS, SDI, and SDO paths for each device are separate for ultimate flexibility in control of the devices. When controlled via the SDP board, the /CS line is shared (LK12 inserted). When using multiple devices, the SDI and SDO paths can be shared, and each device can be controlled via its own /CS line, allowing for easy control with minimum wires. Mnemonic PD RESET SDI_1 SDI_0 SDO_1 DGND SDO_0 CS_0 CS_1 SCLK DRDY Device both both slave master slave both master master slave both master Description Power down, active low Device reset, active low Serial data input Serial data input Serial data output Digital ground Serial data output Chip select master Chip select slave Clock input Data ready, active low Timing Characteristics Refer to the ADAS1000 product data sheet for information regarding the required waveforms and behavior of the SPI interface pins when preparing to interface directly to the ADAS1000 SPI interface. Pace Interface/GPIO Connector, J6 This connector provides the optional secondary interface available from the master device for the purposes of the customer-based digital pace detection algorithm. It is a master interface providing MSCLK, MSDO, and MCS outputs to be read by a host controller. It provides ECG data captured at 128 kHz data rate. Pin No. 1 2 Mnemonic GPIO3 GPIO2/MSDO 3 GPIO1/MSCK 4 GPIO0/MCS 5, 6 DGND Description Reconfigurable IO Reconfigurable IO/master interface MSDO Reconfigurable IO/master interface MSCK Reconfigurable IO/master interface MCS Digital ground RESET BUTTONS There are two reset buttons on the board. SDP reset is used for a reset of the SDP board and ADAS reset is used for reset of ADAS1000 devices to default/power-on configuration. LED There is one LED (D10) on the board, which is lit when the board is powered from J7 (+5 V connector). Rev. B | Page 8 of 44 EVAL-ADAS1000SDZ User Guide UG-426 JUMPERS There are a number of jumpers included on this board for flexibility and ease of configuration. These jumpers allow the user to easily drive the main ADAS1000 SPI interface directly without concern about other SPI-controlled components. On receipt of the board, the jumpers are in the default state as described in Table 4. Table 4. Default State of Jumpers Jumper Respiration Jumper LK1 LK2 LK3 LK4 LK19 LK15 LK16 Power Supply LK6 LK7 LK8 VREG EN LK9 LK17 LK18 Shield LK10 Reset Function Interface LK11 LK12 LK13 LK14 Description External Respiration Measurement. LK1 is used to connect the EXT_RESP_XX paths to the ECG channels external to the device (requires that LK4 be connected if using the external paths). External Respiration Measurement. LK2 is used when using external capacitors for the respiration circuit; LK1 needs to be closed also (and, optionally, LK4). External Respiration Measurement. LK3 is used to bring either of the ECG channels (LA or LL) to the input of the AD8226 in-amp (part of an optional respiration circuit). External Respiration Measurement. LK4 connects the external respiration paths to the respiration header or, alternatively, to the ECG paths via LK1. External Respiration Measurement. Connect (A) when using external respiration circuit (AD8226 and AD8606). Can be disconnected for all other respiration options. For External Respiration Using an External Instrumentation Amplifier (Optional Circuit). The evaluation board uses the AD8226 along with a buffer, AD8606. This arrangement allows the output of the amplifier to drive the EXT_RESP_RA input. Use with LK16. For External Respiration Using an External Instrumentation Amplifier (Optional Circuit). The evaluation board uses the AD8226 along with a buffer, AD8606. This arrangement allows the output of the amplifier to drive the EXT_RESP_LA input. Use with LK15. AVDD Path from On-Board LDO (ADP151). Open if powering the board from J5 screw terminals. Closed if supply board from either wall adaptor input or J7 (5 V) supply input. AVDD Path to MASTER ADAS1000. Use to measure supply current in AVDD path to MASTER ADAS1000. IOVDD Path from On-Board LDO (ADP151). Open if powering the board from J5 screw terminals. Closed if supply board from either wall adaptor input or J7 (5 V) supply input. VREG_EN–ADAS1000 On-Chip Regulators (DVDD, ADCVDD). They are enabled when VREG_EN is high, disabled when low, and may be overdriven. If overdriving, close LK9 and use J5 to supply DVDD and ADCVDD supply rails (1.8 V). ADCVDD net is shared to both MASTER and SLAVE. This allows an external ADCVDD to be applied. Each ADAS1000 has its own on-chip regulator for ADCVDD. If both MASTER and SLAVE are inserted on board and LK9 is open, then LK17 should be open. Each ADAS1000 has its Own On-Chip Regulator for DVDD. If both MASTER and SLAVE are inserted on board and LK9 is open, then LK18 should be open (to stop the regulators trying to fight each other). Link in Shield Path. The Shield pin is a shared pin with the external respiration drive for LA. Therefore, when using the SHIELD drive directly, the user can connect LK10 to Link A. Alternatively, if using external respiration feature, the user can connect the shield of the patient cable directly to GND by inserting LK10 in Position B. Link in Reset Path. When using the SDP board, both MASTER and SLAVE are driven with the same CSB (LK12 inserted). If driving the MASTER and SLAVE from J4, open LK12. When using the SDP board, both MASTER and SLAVE drive different SD0 paths (LK13 open). When using the SDP board, both MASTER and SLAVE are driven with different SDI paths (LK14 open). If driving the MASTER and SLAVE from J4 and the same SDI path, close LK14. Rev. B | Page 9 of 44 Condition A, B, C inserted A, B, C inserted Inserted (A) Open Inserted (A) Closed Closed Closed Closed Closed Open Open Open Inserted (B) Closed Closed Open Open UG-426 EVAL-ADAS1000SDZ User Guide ADAS1000 SOFTWARE OPERATION QUICK OPERATION OF ADAS1000 SOFTWARE 1. 2. 3. Launch the ADAS1000 software. The main panel shown in Figure 13 opens. Click Default Settings to power up the ADAS1000, configuring the device into a known condition. To start streaming ECG data, go to Stream ECG and a pop-up window (see Figure 18) opens and the GUI automatically starts reading ECG data from the board. DETAILED DESCRIPTION OF MAIN CONTROL PANEL the ECGCTL, CMREFCTL, and FRMCTL registers as shown in Table 5. All other registers remain at their power-on default settings. When finished using the software, click QUIT to close the window. Table 5. Commands Sent to Master Device by Default Settings Button Register ECGCTL Word 0xF800BE CMREFCTL 0xE0000A FRMCTRL 0x079000 When the software is launched, the main window of the EVALADAS1000SDZ software opens, as shown in Figure 13. The evaluation board automatically detects if LK12 is inserted, so it knows if it needs to read from a single device (master, LK12 = open) or from both devices (master and slave, LK12 = inserted). On the main panel, the user can access pop-up windows which allow access to all register controls, streaming (ECG, respiration, pace, and lead off), and the write to file window. The ADAS1000 powers up with channels disabled and in power-down mode. A number of writes are required to different registers to start the device up and begin streaming data from the device. Within the main window, there is a Default Settings button which allows quick configuration of the device. This configures Rev. B | Page 10 of 44 Conditions All ECG channels enabled. Single-ended input. Gain setting, GAIN 0 = 1.4. VREF buffer enabled. Low noise/high performance mode. Convert enabled. LA, LL, RA are selected to contribute to VCM. Reference drive is enabled and applied to RLD_OUT electrode. Internal common mode is used and driven out on CM_OUT. Shield drive is enabled. Frame includes: all ECG words, pace detect, respiration magnitude, leads off, GPIO, and CRC. Data format is vector mode. Every frame is output at frame rate of 2 kHz. EVAL-ADAS1000SDZ User Guide UG-426 CHECKS IF MASTER AND SLAVE DEVICES ARE PRESENT ACCESS ALL REGISTERS DEFAULT SETTINGS—WHEN LAUNCHING SOFTWARE—HIT THIS BUTTON TO POWER UP THE ADAS1000 INTO KNOWN CONDITION EXIT AND CLOSE GUI Figure 13. Main Evaluation Board Control Window Rev. B | Page 11 of 44 10810-013 THESE BUTTONS OPEN POP-UP WINDOWS FOR EACH FUNCTION UG-426 EVAL-ADAS1000SDZ User Guide CONTROL OF ALL REGISTERS To delve further into the register control, click Program All RW Reg from the main control panel. This produces a pop-up window giving access to different tabs for each control register. When moving back and forth between tabs, note that each time you click on a tab, the device reads the appropriate register and updates the Read register listing accordingly.       A Write Read Reg writes the data in the write panel and reads back all the registers to confirm the write. Caution: the write panel may not match the contents of the ADAS1000 registers, so if using this panel to update particular registers, quickly do a ReSync Write Reg first to ensure that the write and read panels match, and then proceed to make your changes. Update the register contents with any changes. Click Default Settings to return the part to the default state. Save a register setting for reload and reuse later using the Save Reg State and Load Reg State. When finished with this window, click QUIT to close the window. SLAVE WRITE CONTROL UPDATE THE REGISTER WITH THE WRITE CONTENTS 10810-014 MASTER WRITE CONTROL Figure 14. Read/Write Control Register Overview Within the tabs of this window all the individual control registers can be accessed. The ECG control register is shown in Figure 15, giving access to master and slave control. Note that for certain conditions, the slave device must match the master device configuration. As a result, the software locks out some of the slave control to ensure that the settings of the master and slave will always match (for example, gain setting, high performance mode, clock source, and so on). Rev. B | Page 12 of 44 UG-426 10810-015 EVAL-ADAS1000SDZ User Guide Figure 15. ECG Control Register Rev. B | Page 13 of 44 UG-426 EVAL-ADAS1000SDZ User Guide ECG CAPTURE The ECG Capture tab allows users to view the ECG patterns at the different data rates, in vector or electrode mode. The tab panel on the right allows quick, easy configuration of the different parameters—filter, data rate, power mode, vector/electrode in the panel on the right. 1. Once the appropriate registers are configured to start a capture, proceed to the ECG capture window. The ADAS1000 starts streaming ECG data. The signal shown in Figure 16 is a 60 bpm signal coming from the Fluke PS420 simulator. REMOVE ELECTRODES/LEADS IF DESIRED USE LABVIEW FILTER TO INSERT HPF FOR DIAGNOSTIC OR MONITOR BANDWIDTH EASY ACCESS TO DIFFERENT CONTROL PARAMETERS SELECT MASTER OR SLAVE DEVICE TO VIEW ECG LABVIEW CHART TOOLS TO ANALYSE DATA 10810-016 2kHz DATA RATE Figure 16. ECG Capture Window with Extra Filtering • • • • 2. Quick settings allow the user control over the common features, such as data rate, electrode/vector data, gain, and filter frequency. This particular display is set up for lead (vector) display. There is a tab that allows users to apply LabVIEW® filters. There is also a tab that allows user to view the captured heart rate. To zoom in on data to view in more detail, the LabVIEW chart tools are available for use. Simply stop streaming and then use the graph zoom for X or Y to zoom in the chart of interest. Rescaling of the Y-axis can also be done by first right clicking and turning off the autoscale feature. Rev. B | Page 14 of 44 EVAL-ADAS1000SDZ User Guide UG-426 ECG CAPTURE WITH DIGITAL POST PROCESSING The ECG capture window allows the user to use a LabVIEW filter (0.05 Hz) to view diagnostic bandwidth in addition to ac coupling the signal (using LabVIEW VI). Insert a LabVIEW HPF for ac coupling. (ADAS1000 is a dc coupled design). Insert a LabVIEW HPF or 0.5 Hz or 0.05 Hz. 10810-017 1. 2. Figure 17. Different Tabs Providing Differing Controls Within the ECG Capture Window. Controls include: LabVIEW filters, the test tone feature, and graph control, respectively. Note that the software also allows users to detect heart rate. This applies to LEAD II (see Figure 18). When finished with the ECG capture window, click QUIT to close the window. HEART RATE DISPLAY 10810-018 3. 4. Figure 18. ECG Capture Window with Heart Rate Shown Rev. B | Page 15 of 44 UG-426 EVAL-ADAS1000SDZ User Guide LEADS OFF CONTROL The lead off tab allows the user to control which lead off feature is operating (ac or dc) and programs the current and threshold levels. For dc lead off, remember the VCM is 1.3 V; therefore the upper threshold should be in excess of 1.3 V to ensure capture of a lead off event. For ac lead off, the threshold levels are represented in terms of amplitude by multiplying π/2. Note that the levels may need to be adjusted to find the appropriate levels to detect ac leads amplitude. Configure the mode of lead off detection (either ac or dc). Set the current levels. Program the threshold levels of the detection circuitry (applies to ac lead off). Write the changes to the register by clicking Update, which is available within the LOFF Threshold Levels tab. When finished, click QUIT to close the window. PROGRAM THE APPROPRIATE CURRENT LEVELS AC : 0nA TO 100nA DC : 0nA TO 70nA PROGRAM THE THRESHOLDS FOR DETECTION QUICK SETUP—DC LEAD OFF CHOOSE AC OR DC 10810-019 1. 2. 3. 4. 5. Figure 19. Leads Off Control and Display Chart Rev. B | Page 16 of 44 EVAL-ADAS1000SDZ User Guide UG-426 RESPIRATION The Respiration window allows configuration of the respiration feature and display of the respiration rate. 2. 3. Use Quick Respiration Setup to quickly configure the device. Figure 20 shows the respiration signal capturing a respiration signal provided by the Fluke PS420. Configure the settings in the respiration register for many different options. See Figure 20 and see the Flexible Respiration Feature on Master section for a description. When finished, click QUIT to close the window. CONFIGURATION SETTINGS FOR RESPIRATION MEASUREMENT QUICK RESPIRATION SETUP— PRE-CONFIGURES SETTINGS 10810-020 1. Figure 20. Respiration Magnitude Chart Rev. B | Page 17 of 44 UG-426 EVAL-ADAS1000SDZ User Guide PACE The Pace register control offers a number of different features. The pace algorithm is three instances of a digital algorithm, therefore it can run on 3 leads at one time. The algorithm is designed to detect pace widths that range from 100 µs to 2 ms and amplitudes of 400 µV to 1000 mV. For pace capture, the software is streaming data at the chosen data rate. The data rate chosen does not have an effect on the ability of the pace detection algorithm to detect a pulse since the pace algorithm always processes the 128 kHz frame rate. 4. 5. Choose which lead each pace algorithm analyses. Configure the different threshold levels for the desired ranges. Use Quick Pace Setup to program the pace control and thresholds to default levels, and to start to stream pace data. When a pace signal is detected, the frame header flags it. The pace algorithm makes a measure of the height and width of the detected pace and provides that information for readback. There are two ways of returning the width and height information from the register reads, one from within the frame and another from a direct read of the pace height/width register which is a more accurate result. Pace Validation Filter 1 and Filter 2 are for noise and MV pulse filtering. The pace width filter rejects signals 2 ms. If using the 2 kHz data rate, the pace signal may be filtered out by the programmable LPF available in this data rate. Adjust the Low Pass Filter setting to allow the pace signal to be more visible. Note there are three pace algorithms and three pace windows to view. Each algorithm has its own threshold register settings. When finished, click QUIT to close the window. 3 PACE WINDOWS TO VIEW SCREEN CAPTURE OF PACE ON PACE 1 LEAD II PACE LEVELS CONTROL FILTER ENABLES CHOOSE THE LEADS FLAGS PACE DETECTED HEIGHT WIDTH 10810-021 1. 2. 3. Figure 21. Pace Window Rev. B | Page 18 of 44 EVAL-ADAS1000SDZ User Guide UG-426 UNDERSTANDING PACE IN THE ADAS1000 Consult the ADAS1000 data sheet for full details on the pace detection operation. Under Pace Controls, as shown in Figure 21, a user has access to set up three different instances of pace to analyze the particular leads of interest (choices of LEAD I, II, III, and VF). The pace algorithm is a state machine that operates on the 128 kHz 16-bit data. When enabled, it works to identify the leading edge and trailing edge of the pace pulse and can output the width and height of that pulse as detected on the surface of the skin. The configuration registers are shown in Figure 22. SET THE DESIRED MINIMUM PACE AMPLITUDE TO DETECT, THRESHOLD = N × VREF /GAIN/216 SET THE EDGE THRESHOLD, THE EDGE TRIGGER/THRESHOLD IS USED TO FIND A LEADING EDGE THRESHOLD = N × VREF /GAIN/216 10810-022 SET THE LEVEL THRESHOLD WHICH IS USED TO FIND THE LEADING EDGE PEAK THRESHOLD = N × VREF /GAIN/216 Figure 22. Pace Levels Tab Rev. B | Page 19 of 44 UG-426 EVAL-ADAS1000SDZ User Guide WRITE TO A FILE A Write to File tab allows storage of a data capture over a period of time. Select either raw (date read back and provided in decimal format), voltages (software calculates the corresponding voltage of each lead/electrode), or parsed (where parsed breaks out the header word and the remaining words are provided in decimal format) file for offline processing. CHOOSE FORMAT TO SAVE 10810-024 CHOOSE HOW MANY SECONDS OF CAPTURE TO SAVE Figure 23. Write to File Tab When you click Write to File, a Choose file to write window opens (see Figure 24). Select the appropriate location and name the file with the name of your choice. Using an extension like .dat or .xls works well here. The data is formatted tab delimited as shown in Figure 24. This data can be copied into Microsoft® Excel® and processed offline. CHOOSE THE LOCATION AND NAME THE FILE 10810-025 1. 2. Figure 24. Choose the Location and Name the File Rev. B | Page 20 of 44 EVAL-ADAS1000SDZ User Guide UG-426 RAW 10810-026 When you save raw data, this saves data in decimal codes. The information from the header file is shown first (details about lead status, pace, fault, overflow, and so on) followed by the lead words and any other words that are enabled in the Frame Control register (pace, respiration, lead off, and so on). Figure 25. Data Stored in File When Saved as RAW Data Parsed 10810-027 Saving data as parsed saves the ECG data in decimal codes. The information from the header file is shown first (details about lead status, pace, fault, overflow, and so on) and parsed out into each individual bit. This is followed by the lead/electrode words and any other words that are enabled in the Frame Control register (pace, respiration, lead off, and so on). Figure 26. Data Stored in File When Saved as Parsed Data Rev. B | Page 21 of 44 UG-426 EVAL-ADAS1000SDZ User Guide Voltage 10810-028 Voltage processes the electrode data and provides the equivalent voltage level of each electrode. The header and other words within the frame are stripped out of this data capture. Figure 27. Data Stored in File When Saved as Voltage Rev. B | Page 22 of 44 EVAL-ADAS1000SDZ User Guide UG-426 TEST TONES The ADAS1000 has built-in test tones that can put out a 10 Hz or 150 Hz 1 mV sinewave, in addition to a 1 mV calibration pulse. 1. Configure the ADAS1000 for electrode mode in the ECGCTL register (see Figure 15) to be able to see these signals correctly. If the ADAS1000 is configured for vector/lead mode, then the test tone signals are subtracted from each other. Also, in the CMREFCTL register, none of the electrodes should be configured to contribute to the common-mode signal. 2. Use the Test Tone tab to configure this data. 10810-029 A 10 Hz sine wave is shown in Figure 28. Applying an offset of 2 mV helps to enable viewing. Figure 28. Viewing the Internal 10 Hz Sine Wave Test Tone Applied to the Master Device ECG Channels Rev. B | Page 23 of 44 UG-426 EVAL-ADAS1000SDZ User Guide 10810-030 Figure 29 shows the test tone signal programmed with a 1 Hz square wave (with an offset of 3 mV to enable viewing). Again, electrode mode is needed for correct viewing, and none of the electrodes should be configured to contribute to the common-mode signal. Figure 29. Viewing the Internal 1 Hz Square Wave Test Tone Applied to the Master Device ECG Channels Rev. B | Page 24 of 44 EVAL-ADAS1000SDZ User Guide UG-426 COMMON-MODEL LEVEL/WILSON CENTRAL TERMINAL Common-Mode Level/Wilson Central Terminal The ADAS1000 allows flexible configuration of the common-mode signal, in that any of the electrodes can be used to generate the common-mode level VCM. When no electrodes are selected to contribute to the common mode level, then the VCM = VCM_REF which is the internal 1.3 V level. The Wilson central terminal of (RA + LA + LL)/3 can be configured here also as shown in Figure 30. The VCM can be brought out to the CM_OUT pin. The VCM level used internally can come from the internal VCM level (as arranged by the selection of electrodes). Alternatively, it can be sourced externally from the CM_IN pin (for example, if using multiple ADAS1000 devices and wishing to share the VCM across them–the master device could provide the CM_OUT to the slave devices CM_IN pin, so they are all referenced to the same common-mode level). WILSON CENTRAL TERMINAL CONFIGURATION ENABLE THE VCM TO APPEAR ON CM_OUT PIN SELECT WHERE CM IS COMING FROM 10810-031 • • • Figure 30. CMREFCTL 0x05 Table Rev. B | Page 25 of 44 UG-426 EVAL-ADAS1000SDZ User Guide DETAILED DESCRIPTION OVERVIEW OF SCHEMATICS instruments, and absorbs the dc or ac lead-off currents injected on the ECG electrodes. The right leg reference signal may also be redirected onto any of the other electrodes (in absence of right leg electrode connection). Gold pin connectors are provided for replacement purposes. Nominal values used here are RZ = 100 kΩ, CZ = 2 nF, RFB = 3.9 MΩ, and RIN = 39 kΩ. Refer to Figure 46 for schematics related to the master device. The master device provides ECG electrodes RA, LA, LL, V1, V2, and RLD (see Figure 45) while the slave device services ECG electrodes V3, V4, V5, V6, and a spare channel. ECG Protection VREG EN The ADAS1000 device has standard ESD cells on board. In addition, SP724 SCR/diode protection arrays are used on the ECG input paths; however, they are not provided for defib protection purposes. VREG_EN = 1 (LK9 open) allows the internal LDO supplies to power the DVDD and ADCVDD rails. If using the J5 connector to supply DVDD and ADCVDD rails directly to the part, VREG_EN = 0 (LK9 closed) disables the on-board regulators. Optional Component Space on ECG path If using the J5 connector to supply DVDD and ADCVDD rails directly to the part, also remove LK17 and LK18. Optional component space is provided for user-supplied cable and filter modelling in addition to pull-down resistors to RLD. Note that these limit the detection of dc leads of function because, if used, any off electrode would then be sitting at the RLD level and, thus, may no longer be detectable by the dc lead off circuit, particular at low current levels). PACE INTERFACE (OPTIONAL) An optional pace interface is available via Connector J6. This is a master interface and provides fast 128 kHz ECG data for external digital pace algorithm purposes. Right Leg Drive POWER The right leg drive amplifier or reference amplifier is used as part of a feedback loop to force the patient’s common-mode voltage close to the ADAS1000 series internal 1.3 V reference level (VCM_REF). This centers all the electrode inputs relative to the input span, providing maximum input dynamic range. It also helps to reject noise and interference from external sources such as fluorescent lights or other patient-connected The ADP2503 buck-boost regulator is included to allow for battery powered operation from 3 × AA (4.5 V) batteries. ADP151 LDOs provide the AVDD = 3.3 V and IOVDD =3.3 V rails required. IOVDD = 3.3 V as the SDP board interface expects 3.3 V input levels (see Figure 48). EXTERNALLY SUPPLIED COMPONENTS CZ TO SET RLD LOOP GAIN 2nF 40kΩ RIN* RLD_SJ RZ 100kΩ RFB* 4MΩ RLD_OUT CM_OUT/WCT 10kΩ SW2 10kΩ SW3 10kΩ SW4 10kΩ SW5 10kΩ ELECTRODE LL ELECTRODE RA ELECTRODE V1 ELECTRODE V2 SW6 CM_IN OR CM BUFFER OUT VCM_REF (1.3V) – SW1 ELECTRODE LA + 10kΩ RLD_INT_REDIRECT *EXTERNAL RESISTOR RIN IS OPTIONAL. IF DRIVING RLD FROM THE ELECTRODE PATHS, THEN THE SERIES RESISTANCE WILL CONTRIBUTE TO THE RIN IMPEDANCE. WHERE SW1 TO SW5 ARE CLOSED, RIN = 2kΩ. RFB SHOULD BE CHOSEN ACCORDINGLY FOR DESIRED RLD LOOP GAIN. 10810-100 ADAS1000 Figure 31. Right Leg Drive Configuration on Evaluation Board Rev. B | Page 26 of 44 EVAL-ADAS1000SDZ User Guide UG-426 FLEXIBLE RESPIRATION FEATURE ON MASTER ADAS1000 Respiration measurement is made in the master device. The respiration path is equipped with much flexibility for evaluation purpose. Figure 32 and Figure 33 illustrate the different methods of respiration measurement via different paths. The control of the respiration function is described in the ADAS1000 data sheet. Internal Drive/Measure via the ECG Paths 2. Configure the RESPCTL register (0x3) as follows: RESPCAP = 0 (internal). Choose a relevant lead for measurement (RESPSEL), drive frequency (RESPFREQ), measurement gain (RESPGAIN), and so on. Figure 32. RESPCTL Controls–Respiration Cap Configuring the Links Note that the ADAS1000 respiration measurement is only made on one lead at any one time. However, the choice of lead is programmable: choose either LEAD I, LEAD II, or LEAD III. For simplicity, Figure 33 allows the flexibility of programming the ADAS1000 respiration register to measure on all 3 leads (again, only 1 lead at any one time). This may not be the practice in end applications, therefore if a user is only interested in one particular lead during evaluation of the respiration function, adjust the links accordingly. Note that Figure 33 shows RC components which may represent input filtering or cable model. ADAS1000 LK4 EXT_RESP_RA A EXT_RESP_LL B EXT_RESP_LA EXT_RESP_LA EXT_RESP_LL EXT_RESP_RA C SHIELD SHIELD/RESP_DAC_LA A B C C B A LK1 LK2 RESPIRATION PATH FROM ECG ELECTRODES LK1, LK2, LK3, LK4, LK19 OPEN RESP_DAC_LL RESP_DAC_RA LA LA LL LL RA RA CABLE FILTER Figure 33. Respiration Drive Using Internal Capacitor, Respiration Measurement Using ECG Path Rev. B | Page 27 of 44 10810-034 1. 10810-033 The primary respiration method uses internal respiration drive and measure. In some applications, this will require no external components and uses the standard ECG electrodes. In the evaluation board, a number of links are used to allow a user to evaluate different respiration configurations. For this instance, LK1, LK2, LK3, LK4, and LK19 can be open. UG-426 EVAL-ADAS1000SDZ User Guide Internal Drive/Measure via the ECG/EXT_RESP Paths Configuring the Links This configuration allows the user to potentially bypass any ESIS filtering in the ECG path. Note that the input range of the EXT_RESP_XX pins must not be exceeded. Note that the ADAS1000 respiration measurement is only made on one lead at any one time. In the ext_resp path, the EXT_RESP_RA is always enabled internally and the user has a choice of EXT_RESP_LA or EXT_RESP_LL giving options of LEAD 1 or LEAD III. For simplicity, Figure 35 allows the flexibility of programming the ADAS1000 respiration register to measure on these 2 leads (again, only 1 lead at any one time). This may not be the practice in the end application, therefore if you are only interested in one particular lead during evaluation of the respiration function, adjust the links accordingly. In this example, LK4 and LK1 would be inserted and LK2 would be open. Note that Figure 35 shows RC components which may represent input filtering or cable model. Figure 34. RESPCTL Controls–Respiration Lead Select ADAS1000 LK4 EXT_RESP_LA A EXT_RESP_LL B EXT_RESP_RA EXT_RESP_LA EXT_RESP_LL EXT_RESP_RA C SHIELD SHIELD/RESP_DAC_LA C B A A B C LK1 LK2 RESPIRATION PATH FROM ECG ELECTRODES USING EXT_RESP PATH LK1, LK4, CLOSED LK2, LK3, LK19 OPEN DRIVE/ MEASURE (J1) RESP_DAC_LL RESP_DAC_RA LA LA LL LL RA RA CABLE FILTER 10810-036 2. 3. Configure the RESPCTL register (0x3) as follows: RESPCAP = 0 (internal capacitor), RESPSEL = 11 (EXT_RESP path selected). Choose which path to measure on RESPSELEXT. Choose the appropriate drive frequency (RESPFREQ), measurement gain (RESPGAIN), and so on. 10810-035 1. Figure 35. Respiration Drive Using Internal Capacitor, Respiration Measurement Using ECG Paths via the EXT_RESP Path Rev. B | Page 28 of 44 EVAL-ADAS1000SDZ User Guide UG-426 Internal Drive/Measure via the EXT_RESP Path Configuring the Links This configuration allows the user to measure directly from the respiration connector provided on the board (J8). Note that the ADAS1000 respiration measurement is only made on one lead at any one time. In the ext_resp path, the EXT_RESP_RA is always enabled internally and the user has a choice of EXT_RESP_LA or EXT_RESP_LL giving options of LEAD 1 or LEAD III. For simplicity, Figure 37 allows the flexibility of programming the ADAS1000 respiration register to measure on these 2 leads (again, only 1 lead at any one time). This may not be the practice in the end application, therefore if you are only interested in one particular lead during evaluation of the respiration function, adjust the links accordingly. In this example, LK4 would be inserted and LK1 and LK2 would be open. Note that Figure 37 shows RC components which may represent input filtering or cable model. Figure 36. RESPCTL Controls–Respiration Lead Select ADAS1000 LK4 EXT_RESP_LA DRIVE/ MEASURE (J8) A EXT_RESP_LL B EXT_RESP_RA EXT_RESP_LA EXT_RESP_LL EXT_RESP_RA C SHIELD SHIELD/RESP_DAC_LA A B C C B A LK1 LK2 RESPIRATION PATH FROM EXT_RESP CONNECTOR (J8) USING EXT_RESP PATH LK4, CLOSED LK1, LK2, LK3, LK19 OPEN RESP_DAC_LL RESP_DAC_RA LA LA LL LL RA RA CABLE FILTER Figure 37. Respiration Drive Using Internal Capacitor, Respiration Measurement via EXT_RESP Path Rev. B | Page 29 of 44 10810-038 2. 3. Configure the RESPCTL register (0x3) as follows: RESPCAP = 0 (internal capacitor), RESPSEL = 11 (EXT_RESP path selected). Choose which path to measure on RESPSELEXT. Choose the appropriate drive frequency (RESPFREQ), measurement gain (RESPGAIN), and so on. 10810-037 1. UG-426 EVAL-ADAS1000SDZ User Guide External Drive/Measure via the ECG Path This configuration allows the user to drive via external capacitors provided on the board and measure back through the ECG path. This mode requires external capacitors in the RESPDAC_XX paths. Note that when this mode is enabled, RESPDAC_RA is always enabled. 10810-039 2. Configure the RESPCTL register (0x3) as follows: RESPOUT = 0/1 (0 = RESPDAC_LL, 1 = RESPDAC_LA), RESPCAP = 1 (external capacitor), RESPSEL = XX (Lead I, II, III). Choose appropriate drive frequency (RESPFREQ), measurement gain (RESPGAIN), and so on. Figure 38. RESPCTL Controls–Respiration Cap Configuring the Links Note that the ADAS1000 respiration measurement is only made on one lead at any one time. However, the choice of lead is programmable, choose either LEAD I, LEAD II or LEAD III. For simplicity, the drawing that follows allows the flexibility of programming the ADAS1000 respiration register to measure on all 3 leads (again, only 1 lead at any one time). This may not be the practice in the end application, therefore if you are only interested in one particular lead during evaluation of the respiration function, adjust the links accordingly. In this example, LK4 would be open and LK1 and LK2 would be inserted. ADAS1000 LK4 EXT_RESP_LA A EXT_RESP_LL B EXT_RESP_RA EXT_RESP_LA EXT_RESP_LL EXT_RESP_RA C SHIELD SHIELD/RESP_DAC_LA C B A A B C LK1 LK2 DRIVE EXTERNAL RESPIRATION DRIVE USING EXTERNAL CAPACITOR MEASURE ON ECG PATHS. RESP_DAC_LL RESP_DAC_RA LK1, 2 CLOSED LK3, 4, 19 OPEN LA LA MEASURE LL LL RA RA CABLE FILTER Figure 39. Respiration Drive Using External Capacitor, Respiration Measurement Using ECG Paths Rev. B | Page 30 of 44 10810-040 1. EVAL-ADAS1000SDZ User Guide UG-426 External Drive/Measure via the EXT_RESP Path This configuration allows the user to drive via external capacitors provided on the board and measure back through the EXT_RESP path. Note that when this mode is enabled, RESPDAC_RA is always enabled. 10810-041 2. Configure the RESPCTL register (0x3) as follows: RESPOUT = 0/1 (0 = RESPDAC_LL, 1 = RESPDAC_LA), RESPCAP = 1, (external capacitor), RESPSEL = 11 (external respiration path), RESPEXTSEL = 0/1 (0 = EXT_RESP_LL, 1 = EXT_RESP_LA). Choose the appropriate drive frequency (RESPFREQ), measurement gain (RESPGAIN), and so on. Note that the selected RESPDAC_LL or RESPDAC_LA need to match the selected EXT_RESP_LL/LA selected when measuring on the external range. The example for register settings in Figure 41 shows RESPDAC_LA and EXT_RESP_LA selected, therefore LK2AC, LK4AC and LK1AC need to be inserted. Figure 40. RESPCTL Controls–Respiration Cap and Respiration Lead Select Configuring the Links Note that the ADAS1000 respiration measurement is only made on one lead at any one time. In the ext_resp path, the EXT_RESP_RA is always enabled internally and a user has a choice of EXT_RESP_LA or EXT_RESP_LL giving options of LEAD 1 or LEAD III. For simplicity, Figure 41 allows the flexibility of programming the ADAS1000 respiration register to measure on these 2 leads (again, only 1 lead at any one time). This may not be the practice in the end application, therefore if you are only interested in one particular lead during evaluation of the respiration function, adjust the links accordingly. ADAS1000 LK4 EXT_RESP_LA A EXT_RESP_LA MEASURE EXT_RESP_LL B EXT_RESP_RA EXT_RESP_LL EXT_RESP_RA C SHIELD SHIELD/RESP_DAC_LA C B A A B C LK1 LK2 DRIVE EXTERNAL RESPIRATION DRIVE USING EXTERNAL CAPACITOR MEASURE ON EXT RESP PATHS. RESP_DAC_LL RESP_DAC_RA LK1, LK2, LK4 CLOSED LK3, LK19 OPEN LA LA LL LL RA RA CABLE FILTER 10810-042 1. Figure 41. Respiration Drive Using External Capacitor, Respiration Measurement Using ECG Paths via the EXT_RESP Path Rev. B | Page 31 of 44 UG-426 EVAL-ADAS1000SDZ User Guide External Drive/Measure Using External Instrumentation Amplifier Stage Bypassing Internal ADAS1000 Respiration In-Amp This configuration allows the user to drive via external capacitors provided on the board and measure back through the EXT_RESP path. 1. 2. the gain as desired. The range of gain needs to be limited such that the input range
EVAL-ADAS1000SDZ 价格&库存

很抱歉,暂时无法提供与“EVAL-ADAS1000SDZ”相匹配的价格&库存,您可以联系我们找货

免费人工找货