EVAL-CN0409-ARDZ Shield Overview
o Turbidity Measurement
Nonratio Measurement Method (Low Turbidity)
Ratio Measurement Method (Medium to High Turbidity)
o Vial Holder
o Test Vials
Cleaning Procedure
Indexing Vials
Measurement Procedure
o Calibration Solutions
o Software
Schematic, PCB Layout, Bill of Materials
EVAL-CN0409-ARDZ Shield Overview
CN0409 uses the ADPD105 photometric front end and a network of 860nm Infrared emitters and
silicon PIN photodiodes to achieve a water turbidity measurement system. The system can measure
low to high water turbidity levels ranging from 0 FTU to 1000 FTU (limited by available solution). The
IR LED and photodiode network is arranged in such a way that it can support two of the most
recognized turbidity measurement standards – ISO7027 (both ratio and non-ratio) and GLI Method.
With three-point calibration, the typical accuracy that the system can achieve is ± 0.50 FTU or ±5%
of the reading – whichever is greater. This accuracy combined with the 0.05 FTU noise level makes
the measurements obtained using this system very reliable.
The ADPD105’s ambient light rejection feature makes this circuit ideal for applications where
accurate, robust, and non-contact turbidity measurements are critical. Applications include chemical
analysis and monitoring natural bodies of water, wastewater and drinking water.
Turbidity Measurement
The International Organization for Standardization (ISO) developed a design standard known as
ISO7027 Water Quality—Determination of Turbidity, which is best known for its requirement of a
monochromatic light source. Most instruments that comply with this standard use an 860 nm LED
light source and a primary detector at an angle of 90°. Additional detection angles are allowed, such
as a detector at an angle of 180°, to increase the range of measurable turbidity levels.
The CN0409 hardware allows for the measurement of both 90 degrees and 180 degrees scattering
measurements.
Nonratio Measurement Method (Low Turbidity)
For turbidities between 0 FTU and 40 FTU, the 90° detector provides the most linear response to
scattering. At low turbidity levels, the particles are much smaller than the wavelength of incident
light; therefore, they exhibit a symmetrical scattering distribution. As the number and size of
suspended solids increase within this range, the 90° detector receives a linearly proportional amount
of scattered light. This method is also known as the nonratio ISO7027 because of the use of just one
detector.
Ratio Measurement Method (Medium to High Turbidity)
Higher turbidity levels in the range of 40 FTU to 4,000 FTU require an additional detector to obtain
the same linear response as the nonratio method. Larger particles present in these types of
solutions exhibit an asymmetrical light scattering distribution that results in a higher intensity forward
scattered light. The ISO7027 ratio method of the CN-0409 uses the ratio of the 90° detector and the
180° detector to measure turbidity levels of theoretically up to 4,000 FTU. (Note: The CN0409
solution has only been validated up to 1000 FTU)
Vial Holder
The turbidity calibration and sample vials are held in a mechanical holder that is specifically milled
for the LaMotte style test vial.
The photodiodes(DS1 and DS2) and LEDs(D1 and D2) should be bend slightly backwards to allow
placement of the vial holder. Once in place carefully bend the photodiodes and LEDs into the holes
of vial holder, until they are snug and secure.
The test vial should be screwed into the EVAL-CN0409-ARDZ board, by lining up the holes in the
vial holder with the holes in the board, using the screws provided.
The mechanical outline and dimensions for the vial holder used with the EVAL-CN0409-ARDZ can
be found here: Vial Holder Measurements
Test Vials
The test vials are one of the biggest sources of measurement error within a turbidity system, and
therefore one of the most important factors to consider in obtaining accurate turbidity measurements.
It is recommended to use LaMotte Test Vial as the size fits into the CN0409 board vial holder
The following items are critical to the success of proper turbidity vial preparation and measurement
techniques:
Cleaning Procedure
Indexing the Vials
Measurement Procedure
Cleaning Procedure
To obtain the most accurate results when taking measurements, process below should take into
considerations:
Test vials MUST be meticulously cleaned.
1. Washing the vials with soap and deionized water
2. Soaking the sample vial in Hydrochloric Acid solution
3. Rinsing with ultra-filtered deionized water
4. Polishing with silicone oil.
Indexing Vials
Test vials must also be indexed. After the cleaning process, the vial is used to measure a very low
turbidity solution.
Use a calibrated FTU solution to ensure that the index is properly assessed and assigned.
The position with the lowest measured turbidity should be indexed and this position should be used
for succeeding measurements.
If possible, use a properly indexed test vial solution to ensure the system is working and
calibrated properly.
Measurement Procedure
1. Fill a clean test vial(see cleaning procedure above) up to 10mL of the solution under test.
2. Allow sufficient time for bubbles to escape before placing the cap.
1. This can be done by letting the solution stand for several minutes to allow the
bubbles to vacate.
3. Wipe the test vial with a lint free cloth before inserting into the on board test vial holder to
make sure it is free from fingerprints.
1. Make sure you are holding on the test vial cap when placing on the holder
Calibration Solutions
Calibration solutions are an important baseline to ensure that your measurement system is working
approriately. It's needed in order to perform that calibration calculations and routines within the
software, providing accurate turbidity measurements.
The turbidity calibration solutions used in the evaluation are the:
HI88703-11
Oakton T100
Cole Parmer kit
The calibration solution sets allowed us to do testing and characterization at many different data
points and compare results against many different available products on the market. There are many
different available calibration solutions values within these kits.
0.10, 15.0, 100, 750, and 2000 NTU
0.02, 20.0, 100 and 800 NTU.
0.5, 10, and 40 NTU.
Software
ADICUP360 + CN0409 Demo
Schematic, PCB Layout, Bill of Materials
EVAL-CN0409-ARDZ Design & Integration Files
Schematics
PCB Layout
Bill of Materials
Allegro Project
End of Document