EC4-2000-SO2
Industrial Sulfur Dioxide Sensor
(Application : Portable Gas Detectors)
PERFORMANCE
OUTLINE
Range
0 – 2000 ppm
Maximum Overload
All Dimensions are in mm
All tolerances are ±0.2mm
4000 ppm
Sensitivity
12 to 28 nA/ppm SO2
Zero in Air at 20°C
< ± 0.4 µA
(-20°C to +40°C)
0 – 4 ppm SO2
equivalent
Resolution
5 ppm SO2
Zero Deviation
Linearity
Linear across range
Response Time (T 90)
< 60 s
Long-term Output Drift
< 2% signal / month
Warranty
1 year
Bias
No Bias
OPERATING CONDITIONS
Temperature Range
Operating Humidity
Pressure range
Operating Circuit
Recommended Storage
Temperature
Storage life
Expected Operating Life
-20°C to +50°C
15 – 90% RH
(non-condensing)
90 – 110 kPa
(nominal)
See Electrochemical Toxic
Sensor Application Note
0°C to 20°C
6 months in original
packing
2 years in air
PHYSICAL CHARACTERISTICS
Weight
5 g approx.
Orientation Sensitivity
Dimensions
None
See outline
___________________________________________________________________________________________
Whilst SGX has taken care to ensure the accuracy of the information contained herein it accepts no responsibility for the consequences of any use thereof and also
reserves the right to change the specification of goods without notice. SGX accepts no liability beyond the set out in its standard conditions of sale in respect of
infringement of third party patents arising from the use of SGX products in accordance with information contained herein.
In case of modification of the product, SGX disclaims all liability.
SGX Europe sp .z o.o. Poland REGON: 362332227
DS-0243, Issue 2, 29-Jul-2016, Page 1
SGX Europe Sp. z o.o.
Building 11
Ligocka St. 103,
40-568 Katowice,
Poland
T: +48 (0) 32 438 4778
E: sales.is@sgxsensortech.com
www.sgxsensortech.com
CROSS-SENSITIVITY DATA
Gas
Concentration Level (ppm)
Response as ppm SO2
Carbon Monoxide
300
很抱歉,暂时无法提供与“EC4-2000-SO2”相匹配的价格&库存,您可以联系我们找货