SGX Europe Sp. Z o.o.
Building 11
Ligocka St. 103,
40-568 Katowice,
Poland
T: +48 (0) 32 438 4778
E: sales.is@sgxsensortech.com
www.sgxsensortech.com
EC4-500-ETO Datasheet
Industrial Ethylene Oxide Sensor
OUTLINE
All dimensions are in mm
All tolerances are ±0.2mm
PERFORMANCE
Sensitivity
Typical Baseline Range
(pure air)
Baseline Drift
T90 Response Time
Measurement Range
58nA ±33nA/ppm
-0.2 to 2.5 µA
0 to 25 ppm equivalent
≤ 120 seconds
0-500 ppm
Linearity
Linear
Resolution (Electronics
dependent)
10 ppm
Bias Voltage
+300mV
OPERATING CONDITIONS
Temperature Range
-20°C to +50°C
Continuous
Pressure Range
Operating Humidity
Range
LIFETIME
Long Term Sensitivity
Drift
Recommended Storage
Temp
Expected Operating Life
© SGX Sensortech 2020
900 to 1100 mbar
15% to 90% RH
< 2% signal per month
0°C to 20°C
24 months in air
DS-0326 EC4-500-ETO, Issue 1, 30-April-2020, Page 1/3
SGX Europe Sp. Z o.o.
Building 11
Ligocka St. 103,
40-568 Katowice,
Poland
T: +48 (0) 32 438 4778
E: sales.is@sgxsensortech.com
www.sgxsensortech.com
CROSS – SENSITIVITY DATA
GAS
Correction factor to ETO
(Sensitivity of ETO/Sensitivity of test gas)
Ethylene Oxide
1.0
Carbon Monoxide
2.5
Ethanol
2.5
Methanol
0.5
Isopropanol
5.0
i-Butylene
2.5
Butadiene
0.9
Ethylene
0.8
Propane
1.7
Vinyl Chloride
1.4
Vinyl Acetate
2.5
Formic Acid
5.0
* See Important Notes
© SGX Sensortech 2020
DS-0326 EC4-500-ETO, Issue 1, 30-April-2020, Page 2/3
SGX Europe Sp. Z o.o.
Building 11
Ligocka St. 103,
40-568 Katowice,
Poland
T: +48 (0) 32 438 4778
E: sales.is@sgxsensortech.com
www.sgxsensortech.com
IMPORTANT NOTES
1. Calibration with cross-sensitivity gas is not recommended. The cross-sensitivity may fluctuate
between ±30% and may differ from batch to batch and within the sensor lifetime. The crosssensitivities are including but not limited to the above gases. The sensor may also respond to other
gases.
2. PCB sockets are recommended for the sensor pin connection. Soldering to the sensor should be
avoided and will invalidate warranty.
3. All performance specification are based upon the following environment conditions +20°C, 50%
RH and 1atm (1013mBar or ambient pressure).
4. Sensors are designed to operate in a wide range of harsh environments and conditions. However,
it is important to avoid exposure to high concentrations of solvent during storage, fitting into
instrumentation and operation. When using sensors on PCBs, degreasing agents should be used
prior to the sensor being fitted.
Warning:
By the nature of the technology used, any electrochemical gas sensor offered by SGX Europe Sp. z o.o. can potentially fail to meet specification
without warning. SGX Europe Sp. z o.o. makes every effort to ensure the reliability of our products of this type, where life safety is a performance
requirement of the product, we recommend that all sensors and instruments using these sensors are checked for response to gas before use.
SGX Europe Sp. z o.o reserves the right to make product changes without notice. No liability is accepted for any consequential losses, injury or
damage resulting from the use of this document or from any omissions or errors herein. The data is given for guidance only. It does not constitute
a specification or an offer for sale. The products are always subject to a program of improvement and testing which may result in some changes
in the characteristics quoted. As the products may be used by the client in circumstances beyond the knowledge and control of SGX Europe Sp.
z o.o, we cannot give any warranty as to the relevance of these particulars to an application. It is the clients’ responsibility to carry out the necessary
tests to determine the usefulness of the products and to ensure their safety of operation in a particular application. Performance characteristics
on this data sheet outline the performance of newly supplied sensors. Output signal can drift below the lower limit over.
© SGX Sensortech 2020
DS-0326 EC4-500-ETO, Issue 1, 30-April-2020, Page 3/3