PID-10.6eV 40
Datasheet
The SGX PID Sensor
The SGX PID (Photo-Ionisa�on Detec�on) sensors are designed to offer the most accurate method to detect VOCs in
the PPM (parts per million) and PPB (parts per billion) range. Our PIDs are not cross sensi�ve to common
atmospheric gasses (CO2, CO, Sox, NOx, water vapour). Furthermore, it offers high humidity resistance and long-term
(temperature) stability. Our sensors are used across a wide number of industries and applica�ons, wherever you
have a need to detect gases and vapours.
Quality, Safety, Responsibility
SGX Europe Sp. z o.o.
T: +48 (0) 32 438 4778
Building 11
Ligocka St. 103,
40-568 Katowice,
Poland
E: sales.is@sgxsensortech.com
www.sgxsensortech.com
Technical Specifications
Performance
Sensitivity
>30 mV/ppm
Range
>40 ppm
Minimum Detection Limit
1 ppb
Response Time
T90 < 8s
Supply Voltage (using internal regulator)
3.2 - 10 V
(non intrinsically safe 10 - 18 V)
PID-10.6eV 40
3 - 3.6 V
Supply Voltage (using a regulated power supply)
20 - 32 mA
Current
(130 for 100 ms at start up)
Power Consumption
100 mW at 3.3 V
Output Signal (using internal regulator)
Output Signal (using a regulated power supply)
0 - 3.2 V
0 to rail voltage - 0.1 V
Environmental Details
Temperature Range
-40°C to +65°C
Operating Humidity Range
0-99% RH
Features
• Wide dynamic range
• Anti-contamination design
• Humidity resistant
• 10,000 hours lamp life
• Lamp out diagnostic
• Serviceable electrode stack and lamp
• Intrinsically safe
• Series 4/A package
Lifetime Details
10 000 hours
Lamp Life
>5 years
Expected Lifetime
10.6 eV lamp and PID Body
= 12 months
Warranty
Order Code
Supply Voltage
PID-10.6eV-40A
3.2V to 3.6V
PID-10.6eV-40B
3.6V to 10V
Typical Applications
• Industrial health and safety
• Confined space entry
• Air quality monitoring
• Arson investigation
• Indoor air quality monitoring
• HVAC
• Leak detection
• Soil contamination and remediation
Patents & Certification
Patents
US 7.046.012 EC 1474681
Europe
ATEX II 1G Ex ia IIC Ga (-40 °C � Ta � +55 °C) @ 1.1W
Worldwide
IECEx II 1G Ex ia IIC Ga (-40 °C � Ta � +55 °C) @ 1.1W
North America
UL913 (8th) Edition Class I, II, III, Division 1 CSA-C22.2 No157-92
Eurasian Customs Union
№ EAЭC RU C-GB.HA65.B.00635/20
© SGX Sensortech 2012-2022
DS-0415-PID-10.6eV 40, Issue 1, 17-Feb-2022, Page 1/2
SGX Europe Sp. z o.o.
T: +48 (0) 32 438 4778
Building 11
Ligocka St. 103,
40-568 Katowice,
Poland
E: sales.is@sgxsensortech.com
www.sgxsensortech.com
Dimensions
Temperature Curve
Sensor Development Kit (SDK)
Responsivity vs Responsivity at 20 °C
Sensor Development Kit is used as an efficient
way to test and integrate the SGX Sensortech
photoionisation detection sensors into any
application.
SDK includes everything needed to run PID
including an integration board, sensor board,
gas delivery hood and a PC application.
Temperature °C
Accessories and Replacement Parts
Category
Type
PID Spares
PID Spares
PID Spares
Accessories
Accessories
Accessories
Accessories
Lamp
PID-Lamp-10.6eV
Long-life Lamp spare for 10.6eV and 10.0eV PID sensors
Electrode Stack
PID-Stack-40
Electrode Stack spare for PID-10.6eV-40A/B
Lamp spring
PID-spring
Lamp spring
Stack removal tool
PID-Stack-tool
Stack removal tool
Lamp cleaning kit
PID-Lamp-kit
Lamp cleaning kit
Gas chamber
PID-chamber
Gas Delivery Chamber
PCB board
PID-PCB
PID connection PCB (not includes PID sensor)
Part number
Description
DISCLAIMER:
SGX Europe Sp. z o.o. reserves the right to change design features and specifica�ons without prior no�fica�on. We do not accept any legal responsibility for customer
applica�ons of our sensors. SGX Europe Sp. z o.o. accepts no liability for any consequen�al losses, injury or damage resul�ng from the use of this document, the
informa�on contained within or from any omissions or errors herein. This document does not cons�tute an offer for sale and the data contained is for guidance only
and may not be taken as warranty. Any use of the given data must be assessed and determined by the user thereof to be in accordance with federal, state and local laws
and regula�ons. All specifica�ons outlined are subject to change without no�ce.
SGX Europe Sp. z o.o. sensors are designed to operate in a wide range of harsh environments and condi�ons. However, it is important that exposure to high
concentra�ons of solvent vapours is to be avoided, both during storage, fi�ng into instruments and opera�on. When using sensors on printed circuit boards (PCBs),
degreasing agents should be used prior to the sensor being fi�ed. SGX Europe Sp. z o.o. makes every effort to ensure the reliability of its products. Where life safety is
a performance requirement of the product, we recommend that all sensors and instruments using these sensors are checked for response to gas before use.
Copyright© 2012-2022 SGX Sensortech All rights reserved.
Trademarks and registered trademarks are the property of their respective owners.
No part of this publication may be reproduced, distributed, or transmitted in any form or by any means, including photocopying, recording, or other electronic or mechanical methods,
without the prior written permission of the publisher, except in the case of brief quotations embodied in critical reviews and certain other non-commercial uses permitted by copyright law.
For permission requests or technical support please contact or write to the publisher, addressed “Attention: Permissions Coordinator,”.
© SGX Sensortech 2012-2022
DS-0415-PID-10.6eV 40, Issue 1, 17-Feb-2022, Page 2/2
很抱歉,暂时无法提供与“PID-10.6EV-40A”相匹配的价格&库存,您可以联系我们找货
免费人工找货- 国内价格 香港价格
- 1+6295.280951+786.65646