ARTESYN
LPS200-M SERIES
250 Watts
Advanced Energy’s Artesyn LPS200-M series medicallyapproved AC-DC power supplies are available with a nominal
main output of 5 V, 12 V, 15 V, 24 V or 48V. Auxiliary 12 V fan
output is also provided. LPS200-M power supplies provide
125 Watts of output power with free air convection cooling or
up to 250 Watts with 30 CFM of forced air. All models have a
power factor of 0.99 typical and a 3 x 5 x 1.32 inch form factor.
AT A GLANCE
Total Power
125 to 250 Watts
Input Voltage
90 to 264 VAC
# of Outputs
Single
SPECIAL FEATURES
SAFETY
Medical and ITE safeties
TUV
62368, 60601-1
Active power factor correction
UL
62368, 60601-1
3” x 5” footprint
cULus
62368, 60601-1
Less than 1U high
CB
Certificate & report
EN61000-3-2 compliant
CE
Mark (LVD)
Remote sense
UKCA Mark
Power fail
Adjustable main output
Built-in Class B EMI filter
Overvoltage protection
Overload protection
Thermal overload protection
Isolated 12 V Fan output
LPX200 enclosure kit available
RoHS compliant
©2022 Advanced Energy Industries, Inc.
LPS200-M
ELECTRICAL SPECIFICATIONS
Input
Input range
90 to 264 VAC; 120 to 300 VDC
Frequency
47 to 63 Hz
Inrush current
50 A max., cold start @ 25°C
Efficiency
86% typical at full load
EMI/RFI
FCC Class B conducted; CISPR22 Class B conducted;
EN55022 Class B conducted; VDE0878PT3 Class B conducted
Power factor
0.99 typical
Safety ground Leakage current
275 µA @ 50/60 Hz, 264 Vac input
Output
Maximum power
125 W for convection (100 W for LPS202-M); 250 W (200 W for LPS202-M) with 30CFM forced
air
Adjustment range
± 10% minimum on the main outputs
Fan output
12 V @ 0.5 A convection, @ 1 A forced air isolated, ± 10%
Hold-up time
16 ms @ 250 W load, 120 Vac input
Overload protection
Short circuit protection on all outputs. Case overload protected @ 110% to 160% above rating
Overvoltage protection
15% to 50% above nominal output
Logical Control
Power failure
Open collector logic signal goes high 100 to 500 msec after main output; it goes low at least 6
msec before loss of regulation
Remote sense
Compensates for 0.5 V lead drop min. Will operate without remote sense connected. Reverse
connection protected.
ENVIRONMENTAL SPECIFICATIONS
Operating temperature
0°C to 50°C ambient derate each output as 2.5% per degree from 50° to 70°C. -20°C start up
Storage temperature
-40°C to +85°C
Electromagnetic susceptibility
Designed to meet EN61000-4; -2, -3, -4, -5, -6, -8, -11 Level 3
Humidity
Operating; non-condensing 10% to 95% RH
Vibration
IEC68-2-6 to the levels of IEC721-3-2
MTBF calculated
1 million hours at full load and 25°C ambient conditions. 230 Vac input, Bellcore
2 advancedenergy.com
LPS200-M
ORDERING INFORMATION
Model
Number
Output
Voltage
Minimum
Load
Maximum Load with
Convection Cooling
Maximum Load with
30CFM Forced Air
LPS202-M
5V
0A
20 A
LPS203-M
12 V
0A
10.3 A
LPS204-M
15 V
0A
8.3 A
LPS205-M
24 V
0A
5.2 A
LPS208-M
48 V
0A
2.6 A
5.2 A
Ripple
P/P(PARD)3
Peak Load1
Regulation2
40 A
44 A
± 2%
50 mV
20.8 A
22 A
± 2%
120 mV
16.6 A
18 A
± 2%
150 mV
10.4 A
11.5 A
± 2%
240 mV
5.8 A
± 2%
480 mV
1 Peak current lasting