ARTESYN
NLP250 MEDICAL
SERIES
Single Output
Advanced Energy's Artesyn NLP250-M series of enclosed
medical AC-DC power supplies comprises two single
output models, providing standard voltages of 12 V or
24 V. Both models also provide 5 V standby and 12 V fan
outputs. They accept a universal input of 85–264 Vac and
have a low safety ground leakage current. The main output
is compatible with systems that implement distributed
power and point-of-load (POL) architectures. It can also be
used to power electromechanical devices such as relays,
motors and pumps. Featuring a thermally efficient U-channel
construction that is less than 1U high, the enclosed versions
of NLP250-M power supplies can deliver up to 250 watts
when cooled by 250 LFM of forced air.
SPECIAL FEATURES
Medical safeties
Active PFC and EN61000-3-2
compliant
250 W on main channel with
forced air
Low profile fits 1U applications
U-Channel for maximum
thermal performance
Optional cover (CJ suffix)
5 V standby output
12 V fan output
Integrated control and
monitoring features
DATA SHEET
Total Power:
250 W
Input Voltage:
85 - 264 Vac
# of Outputs:
Single
Compliance to EN55022-B
conducted noise standard
Dual AC fuses
RoHS compliant
Two year warranty
SAFETY
VDE0750/EN60950
IEC950/IEC60601-1
File No. 1177400-3336-0759
UL60601-1
File No. E186249
Certificate No. 40014041
CB Ref DE1-36628
Overcurrent, overvoltage and
overtemperature protection
©2021 Advanced Energy Industries, Inc.
NLP250 MEDICAL
ELECTRICAL SPECIFICATIONS
Input
Input voltage range
Universal input
Input frequency range
85 - 264 Vac
47 - 63 Hz
Input surge current
264 Vac (cold start)
40 A max.
Safety ground leakage current
264 Vac, 50 Hz
150 µA
Input current
120 Vac @ 250 W
230 Vac @ 250 W
2.78 A rms
1.36 A rms
Input fuse
UL/IEC127
T6.4 AH, 250 Vac In live and neutral
Maximum power
200 LFM forced air
250 LFM with cover
250 watts
Adjustment range
Main output
± 5%
Total regulation
(line and load)
Main output
Auxiliary outputs
± 2.0%
± 5.0%
Turn-on delay
@ 120 Vac Input
2.0 s max.
Transient response
Main output
50 - 100%
Step at 0.5 A/µs
5.0% or 250 mV
max. dev., 1 ms max
recovery to 1%
Output
Temperature coefficient
±0.02%/°C
Overvoltage protection
Main output
115%, ± 5%
Short circuit protection
Cyclic operation
Continuous
Minimum output current
Singles
0A
Auxiliary outputs
(See Note 8, page 3)
5 Vsb
12 V (fan)
5 V @ 1.0 A
12 V @ 0.3A
All specifications are typical at nominal input, full load at 25 °C unless otherwise stated
EMC Characteristics (5)
Conducted emissions
EN55022, FCC part 15
Level B
Harmonic current correction
EN61000-3-2
Compliant
Voltage flicker
EN61000-3-3
Compliant
ESD air
EN61000-4-2
Level 3
ESD contact
EN61000-4-2
Level 3
Radiated immunity
EN61000-4-3
Level 3
Fast transients
EN61000-4-4
Level 3
Surge
EN61000-4-5
Level 3
Conducted immunity
EN61000-4-6
Level 3
Voltage dips
EN61000-4-11
Compliant
Hold-up time
85 Vac @ 50 Hz
20 ms @ 250 W
Efficiency
115 Vac @ 250 W
230 Vac @ 250 W
84% typ.
86% typ.
Isolation voltage
Input/output
Input/chassis
4000 Vac
2000 Vac
Safety approvals (see note 6, page 3)
UL/cUL UL60601-1, VDE EN60601-1, CAN/CSA22.2 No. 601-1
General Specifications
Weight
MTBF (@25 ºC)
2 advancedenergy.com
650g (22 oz)
Telcordia SR-332
MIL-HDBK-217F
317,000 hours min.
158,000 hours min.
NLP250 MEDICAL
ENVIRONMENTAL SPECIFICATIONS
Thermal performance
Operating ambient,
0 °C to +70 °C
(See derating curve)
Non-operating
-40 °C to +85 °C
0 °C to 50 °C ambient,
250 W
200 LFM forced air
250 LFM with cover
0 °C to 50 °C ambient,
0 °C to 40 °C with cover
175 W
Convection cooled
50 °C to 70 °C ambient,
Derate linearly
Convection cooled
to 50% load
Relative humidity
Non-condensing
5 - 95% RH
Altitude
Operating
10,000 feet max.
Non-operating
30,000 feet max.
Vibration (See Note 7, page 3)
5-500 Hz
2.5 G rms peak
Shock
Per MIL-STD-810E
516.4 Part IV
ORDERING INFORMATION
Output Voltage
Output Current
Min
Max (free air) (1,4)
12 V
0A
14.6 A
24 V
0A
7.3 A
Ripple (3)
Total Regulation
Model Numbers (9, 10)
21 A
120 mV
± 2.0%
NLP250N-99S12J
10.5 A
240 mV
± 2.0%
NLP250N-99S24J
Max (forced air) (2,4)
Notes
1. Free air convection. Maximum continuous output power not to exceed 175W. Refer to Figure 1 for the derating curve.
2. 200 LFM (250 LFM with cover) forced air cooling from the longer side. Maximum continuous output power not to exceed 250 W.
3. F igure is peak-to-peak for room temperature rating. Output noise measurements are made across a 20 MHz bandwidth using a 6 inch twisted pair, terminated with a 10 µF tantalum
capacitor and a 0.1 µF ceramic capacitor.
4. CAUTION: Allow a minimum of 1 second after disconnecting line power when making thermal measurements. For optimum reliability no part of the heatsink should exceed 115 °C
and no semi-conductor case temperature should exceed 120 °C.
5. N
o external filtering required during conducted emissions testing but some applications may require additional filtering to achieve system compliance. Compliance with radiated
EMI specifications may require mounting in a suitable enclosure.
6. This product is only for inclusion by professional installers within other equipment and must not be operated as a stand alone product.
7. Three orthogonal axes, random vibration 10 minutes for each axes, 2.4 G
8. 5 V sb (standby) output is available whenever AC is present, regardless of remote ON/OFF signal status. 12 V (fan) present when main output is present.
9. The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant. “CJ” suffix indicates covered RoHS version.
10. N
OTICE: Some models do not support all options. Please contact your local Artesyn Embedded Power representative or use the on-line model number search tool at
http://www.artesyn.com to find a suitable alternative.
11. T
his product is a Component Power Supply and is only for inclusion by professional installers within other equipment and must not be operated as a standalone product. EMC
compliance to appropriate standards must be verified at the system level. This product is for sale to OEMs and System Integrators, including through Distribution Channels. It is not
intended for sale to End Users.
advancedenergy.com
3
NLP250 MEDICAL
MECHANICAL DRAWING
CUSTOMER MOUNTING HOLES
MAX. SCREW PENETRATION 0.08in
MAX. TIGHTENING TORQUE
3.1 in-1b(0.36N.m)
6-32
1.140
(28.96)
1
VR7
4.000 [101.60]
J3
JP1
1
J4
0.750 (19.05)
J1
E
L
N
M3.0
5.850 (148.59)
J2 1
0.25 (6.35)
6.250 (158.75)
2.500 (63.50)
M3.0
4 PLACES
7.000 [177.80]
6-32
4 PLACES
4.700 (119.38)
5.500 (139.70)
BOTTOM VIEW
1.500 [38.10]
Figure 1: Derating Curve
Output Power (Watts)
Figure 1b: Derating Curve With Cover
Output Power (Watts)
DERATING CURVE Output Power (Watts)
Output Power (Watts)
Output Power (Watts)
All Models
Forced Air
250 W
24 V Free Air
200 LFM (1.0 m/s)
FORCED AIR COOLING
24/48 V Free Air
250 W
250 LFM (1.25 m/s)
FORCED AIR COOLING
12 V Free Air
175 W
All Models
Forced Air
12 V Free Air
175 W
125 W
0W
0C
125 W
87 W
75 W
NATURAL FREE AIR
CONVECTION COOLING
NATURAL FREE AIR
CONVECTION COOLING
75 W
0W
10 C
20 C
4 advancedenergy.com
30 C
40 C
50 C
60 C
70 C
0C
10 C
20 C
30 C
40 C
50 C
60 C
70 C
NLP250 MEDICAL
CONNECTOR AND MATING CONNECTOR TYPES
Connector
Type
Mating Connector Type
J1
Molex 09-65-2058 (5273 series)void pins 2 and 4 or
equivalent
Molex 09-50-8051 or equivalent with Molex 08-52-0113 or equivalent
crimp terminals
J2
Molex 22-23-2021 (6373 series) or equivalent
Molex 22-01-3027 (2695 series) or equivalent with Molex 08-50-01113
(2759 series) or equivalent crimp terminals
J3
Molex terminal block 387007504 or equivalent
Terminal block contains #6-32 screw with clamp washer suitable for wire
2
size 12-22 awg (0.5-2.5 mm ). Max Torque tp 1.36 Nm (12 in.lb)
J4
Molex 22-23-2091 (6373 series) or equivalent
Molex 22-01-3097 (2695 series) or equivalent with Molex 08-50-0113
(2759 series) or equivalent crimp terminals
PIN CONNECTIONS
J1
Pin 1
Ground/Earth
Pin 2
Live
Pin 3
Neutral
J2
Pin 1
+12 V
Fan Voltage
Pin 2
SGND
Return
Pin 1
Vo
+ Main Output
Pin 2
Vo
+ Main Output
Pin 3
RTN
Main Return
Pin 4
RTN
Main Return
Pin 1
+S
+Vo Remote Sense
Pin 2
-S
Vo Remote Sense
Pin 3
LS
Load Share Signal
Pin 4
PS OFF
Remote ON/OFF signal NO
Pin 5
PS ON
Remote ON/OFF signal NC
Pin 6
SGND
Signal Common
Pin 7
PW OK
Power Good
Pin 8
5 Vsb
Stand-by Voltage
Pin 9
DC OK
DC Power Good Signal
J3
J4
advancedenergy.com
5
ABOUT ADVANCED ENERGY
Advanced Energy (AE) has devoted more than three
decades to perfecting power for its global customers. AE
designs and manufactures highly engineered, precision
power conversion, measurement and control solutions for
mission-critical applications and processes.
Our products enable customer innovation in complex
applications for a wide range of industries including
semiconductor equipment, industrial, manufacturing,
telecommunications, data center computing, and medical.
With deep applications know-how and responsive service
and support across the globe, we build collaborative
partnerships to meet rapid technological developments,
propel growth for our customers, and innovate the future
of power.
For international contact information,
visit advancedenergy.com.
powersales@aei.com (Sales Support)
productsupport.ep@aei.com (Technical Support)
+1 888 412 7832
Specifications are subject to change without notice. Not responsible
for errors or omissions. ©2021 Advanced Energy Industries, Inc. All rights
reserved. Advanced Energy®, AE® and Artesyn™ are U.S. trademarks
of Advanced Energy Industries, Inc.
ENG-NLP250 MEDICAL-235-01 1.07