0
登录后你可以
  • 下载海量资料
  • 学习在线课程
  • 观看技术视频
  • 写文章/发帖/加入社区
会员中心
创作中心
发布
  • 发文章

  • 发资料

  • 发帖

  • 提问

  • 发视频

创作活动
ADS1298RECGFE-PDK

ADS1298RECGFE-PDK

  • 厂商:

    BURR-BROWN(德州仪器)

  • 封装:

    -

  • 描述:

    KIT PERFORMANCE DEMO FOR ADS1298

  • 数据手册
  • 价格&库存
ADS1298RECGFE-PDK 数据手册
ADS1298RECG-FE ECG Front-End Performance Demonstration Kit User's Guide Literature Number: SBAU181B March 2011 – Revised Janurary 2016 Contents 1 2 3 4 ADS1298R Overview ............................................................................................................. 7 1.1 Important Disclaimer Information....................................................................................... 7 1.2 Introduction................................................................................................................ 8 1.3 Supported Features ...................................................................................................... 8 1.4 Features Not Supported in Current Version .......................................................................... 9 1.5 ADS1298R Hardware .................................................................................................... 9 Quick Start ........................................................................................................................ 10 2.1 Default Jumper/Switch Configuration ................................................................................ 10 2.2 ADS1298RECG-FE Operation ........................................................................................ 12 Using the ADS1298RECG-FE Software .................................................................................. 13 3.1 Application User Menu ................................................................................................. 14 3.2 Top-Level Application Controls 3.3 About Tab ................................................................................................................ 15 3.4 ADC Register Tab ...................................................................................................... 16 3.5 Analysis Tab............................................................................................................. 24 3.6 Save Tab................................................................................................................. 33 ....................................................................................... 14 ADS1298RECG-FE Input Signals .......................................................................................... 35 4.1 Input Short Testing ..................................................................................................... 35 ............................................................................................ 35 ................................................................................................... 36 4.4 Normal Electrode Input ................................................................................................ 37 4.5 MVDD Input, RLD Measurement, RLD Positive Electrode Driver, and RLD Negative Electrode Driver...... 37 4.6 Lead Derivation ......................................................................................................... 38 4.7 Wilson Center Terminal (WCT) ....................................................................................... 38 4.8 Right Leg Drive ......................................................................................................... 38 4.9 PACE Detection ......................................................................................................... 39 5 Evaluation of the ADS1298R Respiration Function ................................................................. 41 5.1 Introduction .............................................................................................................. 41 5.2 Testing with Onboard Circuitry ........................................................................................ 41 5.3 Testing with Patient Simulator ........................................................................................ 44 6 ADS1298R Daughter Card Hardware Details .......................................................................... 46 6.1 Jumper Description ..................................................................................................... 46 6.2 Power Supply ........................................................................................................... 48 6.3 Clock ..................................................................................................................... 49 6.4 Reference ................................................................................................................ 49 6.5 Analog Output Signals ................................................................................................. 49 6.6 Digital Signals ........................................................................................................... 50 6.7 Analog Input Signals ................................................................................................... 50 Appendix A BOM, Layout, and Schematics .................................................................................... 52 A.1 ADS1298R Front-End Board Schematics .......................................................................... 52 A.2 Printed Circuit Board Layout ......................................................................................... 53 2 4.2 Internal Test Signals Input 4.3 Temperature Sensor Table of Contents SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated www.ti.com A.3 ECG Cable Details .................................................................................................... 53 A.4 Bill of Materials ........................................................................................................ 55 Appendix B External Optional Hardware ....................................................................................... 57 B.1 Optional External Hardware (Not Included) ........................................................................ 57 .................................................................... 59 Appendix C Software Installation ................................................................................................. 60 C.1 Minimum Requirements .............................................................................................. 60 C.2 Installing the Software ................................................................................................ 60 Revision History .......................................................................................................................... 61 B.2 ADS1298R Power-Supply Recommendations SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Contents 3 www.ti.com List of Figures 1 ADS1298R Kit ................................................................................................................ 9 2 ADS1298RECG-FE Default Jumper Locations......................................................................... 10 3 Software Start Screen/About Tab ........................................................................................ 13 4 User Menu - File Item ...................................................................................................... 14 5 User Menu - Help Item 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 4 .................................................................................................... Top Level Controls ......................................................................................................... Lead-Off Status Registers Display Window............................................................................. Channel Registers Tab .................................................................................................... Internal Reference and Buffer Connections ............................................................................ Lead-Off Excitation Options ............................................................................................... Input Multiplexer for a Single Channel .................................................................................. LOFF and RLD Tab ........................................................................................................ LOFF_STATP and LOFF_STATN Comparators ....................................................................... GPIO and OTHER Register Tab ......................................................................................... Wilson Central and Augmented Lead Routing Diagrams ............................................................. Device Registers Settings ................................................................................................. Scope Tool Features ....................................................................................................... Scope Analysis Tab (Noise Levels for Each Channel Shown) ....................................................... Zoom Tool Options ......................................................................................................... Histogram Bins for 12-Lead ECG Signal ................................................................................ Statistics for the Signal Amplitude of Eight ECG Channels .......................................................... FFT Graph of Normal Electrode Configuration ......................................................................... AC Analysis Parameters: Windowing Options.......................................................................... FFT Analysis: Input Short Condition ..................................................................................... Changing the User-Defined Dynamic Range for Channel 1 .......................................................... FFT Plot Using Zoom Tool ................................................................................................ ECG Display Tab Showing LEAD I-III and Augmented Leads ....................................................... ECG Signal Zoom Feature for Six Leads ............................................................................... ECG Signal Zoom Feature for Lead 1 ................................................................................... Save Tab .................................................................................................................... Example of Internal Test Signals Viewed on the ECG Display Tab ................................................. Internal Temperature Sensor ............................................................................................. Eight-Channel Read of Internal Temperature .......................................................................... Normal Electrode ECG Connection in ECG Display Tab ............................................................. Digitization of PACE Signal Using ADS1298R ......................................................................... Respiration Evaluation with Onboard Circuitry ......................................................................... Channel 1 Δ Impedance Measurement ................................................................................. Results After Low-Pass Filtering ......................................................................................... Channel 1 Result for RB= 500Ω and ΔR = 1Ω .......................................................................... Channel 1 Result for RB= 500Ω and ΔR = 0.1Ω........................................................................ ADS1298R Front-End Block Diagram ................................................................................... Fluke Simulator Configuration ............................................................................................ Top Component Placement ............................................................................................... Bottom Component Placement and Routing ........................................................................... Internal Ground Plane (Layer 2) .......................................................................................... Internal Power Plane (Layer 3) ........................................................................................... ECG Cable Schematic ..................................................................................................... List of Figures 14 14 15 16 17 17 18 19 20 21 22 23 24 24 25 26 26 27 27 28 28 29 31 32 32 34 35 36 36 37 40 41 42 42 45 45 46 51 53 53 53 53 54 SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated www.ti.com 48 15-Pin, Shielded Connector from Biometric Cables ................................................................... 57 49 15-Pin, Twisted Wire Cable to Banana Jacks .......................................................................... 57 50 15-Pin, Twisted Wire Cable ............................................................................................... 57 51 Cardiosim ECG Simulator Tool ........................................................................................... 58 52 Recommended Power Supply for ADS1298R .......................................................................... 59 53 Initialization of ADS1298R ................................................................................................ 60 54 License Agreement 55 Installation Process 56 ........................................................................................................ ........................................................................................................ Completion of ADS1298R Software Installation ........................................................................ SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated List of Figures 60 61 61 5 www.ti.com List of Tables ADS1298RECG-FE Default Jumper/Switch Configuration 2 ADS1298R Lead Measurements ......................................................................................... 38 3 Derived Lead Calculations ................................................................................................ 38 4 RLD Jumper Options 39 5 ADS1298RECG-FE Default Jumper/Switch Configuration 46 6 7 8 9 10 11 12 13 6 ........................................................... 1 ...................................................................................................... ........................................................... Power-Supply Test Points ................................................................................................. Analog Supply Configurations (AVDD/AVSS) .......................................................................... Digital Supply Configurations (DVDD/DGND) .......................................................................... CLK Jumper Options ....................................................................................................... External Reference Jumper Options ..................................................................................... Test Signals ................................................................................................................. Serial Interface Pinout ..................................................................................................... Bill of Materials: ADS1298R ............................................................................................. List of Tables 11 48 48 49 49 49 50 50 55 SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated User's Guide SBAU181B – March 2011 – Revised Janurary 2016 ADS1298R This user's guide describes the characteristics, operation, and use of the ADS1298R. This EVM is an evaluation module for the ADS1298R device, an eight-channel, 24-bit, low-power, integrated analog frontend (AFE) designed for patient monitoring and portable and high-end electrocardiogram (ECG) and electroencephalogram (EEG) applications. The ADS1298R is intended for prototyping and evaluation. This user's guide includes a complete circuit description, schematic diagram, and bill of materials. The following related documents are available through the Texas Instruments web site at www.ti.com. Device Literature Number ADS1298R, ADS1296R, ADS1294R SBAS495 1 ADS1298R Overview 1.1 Important Disclaimer Information CAUTION NOTICE: The ADS1298R is intended for feasibility and evaluation testing only in laboratory and development environments. This product is not for diagnostic use. This product is not for use with a defibrillator. The ADS1298R is to be used only under these conditions: • The ADS1298R is intended only for electrical evaluation of the features of the ADS1298R device in a laboratory, simulation, or development environment. • The ADS1298R is not intended for direct interface with a patient, patient diagnostics, or with a defibrillator. • The ADS1298R is intended for development purposes ONLY. It is not intended to be used as all or part of an end equipment application. • The ADS1298R should be used only by qualified engineers and technicians who are familiar with the risks associated with handling electrical and mechanical components, systems, and subsystems. • You are responsible for the safety of yourself, your fellow employees and contractors, and your coworkers when using or handling the ADS1298R. Furthermore, you are fully responsible for the contact interface between the human body and electronics; consequently, you are responsible for preventing electrical hazards such as shock, electrostatic discharge, and electrical overstress of electric circuit components. Pentium III, Celeron are registered trademarks of Intel Corporation. Microsoft, Windows are registered trademarks of Microsoft Corporation. SPI is a trademark of Motorola Inc. All other trademarks are the property of their respective owners. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 7 ADS1298R Overview 1.2 www.ti.com Introduction The ADS1298R is intended for evaluating the ADS1298R low-power, 24-bit, simultaneously sampling, eight-channel front-end for ECG and EEG applications. The digital SPI™ control interface is provided by the MMB0 Modular EVM motherboard (Rev. C or higher) that connects to the ADS1298R ECG FE evaluation board (Rev. A). The ADS1298R is NOT a reference design for ECG and EEG applications; rather, its purpose is to expedite evaluation and system development. The output of the ADS1298R yields a raw, unfiltered ECG signal. The MMB0 motherboard allows the ADS1298R to be connected to the computer via an available USB port. This manual shows how to use the MMB0 as part of the ADS1298R, but does not provide technical details about the MMB0 itself. Throughout this document, the abbreviation EVM and the term evaluation module are synonymous with the ADS1298RECG-FE. 1.3 Supported Features Hardware Features: • Configurable for bipolar or unipolar supply operation • Configurable for internal and external clock and reference via jumper settings • Configurable for AC- or DC-coupled inputs • Configurable for up to 12 standard ECG leads • External Right Leg Drive (RLD) Reference (VCC – VEE)/2 • External shield drive amplifier • External Wilson central voltage • Easy connectivity to popular ECG simulators • On-board respiration circuitry for Impedance Pneumography evaluation Software Features: • Designed to display 12 lead ECG data • Analysis tools including a virtual oscilloscope, histogram, FFT, and ECG display • File printing for post-processing of raw ECG data • Sets the ADS1298R register settings via easy-to-use graphic user interface (GUI) software 8 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R Overview www.ti.com 1.4 Features Not Supported in Current Version • • • • 1.5 Real-time data processing AC lead-off detection filters QRS detection algorithms Software PACE detection algorithms ADS1298R Hardware Figure 1 shows the hardware included in the ADS1298R kit. Contact the factory if any component is missing. The latest software is available on the TI website at http://www.ti.com; you should verify that you have the latest software before using the device. Figure 1. ADS1298R Kit The complete kit includes the following items: • ADS1298R ECG FE printed circuit board (PCB), Rev A • MMB0 (Modular EVM motherboard, Rev C or higher) SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 9 Quick Start 2 www.ti.com Quick Start This section provides a QuickStart guide to quickly begin evaluating the EVM using the ADS1298RECGFE software. 2.1 Default Jumper/Switch Configuration Figure 2 shows the jumpers found on the ADS1298RECG-FE EVM and the respective factory default conditions for each. Figure 2. ADS1298RECG-FE Default Jumper Locations Table 1 lists the jumpers and switches and the factory default conditions. 10 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Quick Start www.ti.com Table 1. ADS1298RECG-FE Default Jumper/Switch Configuration Jumper Default Position Description JP1 Installed RLD feedback JP2 Installed 1-2 AVDD selected for bipolar supply operation selected (AVDD = +2.5V) JP3 Header Not Installed External Vref buffer not connected JP4 Installed EVM +5V provided from J4 (power header) JP5 Open PWDN pin controlled from J5 header (pulled up to DVDD) Header Not Installed (Pins 1-2 shorted on PCB) DC-coupled input signals JP15 Installed 2-3 Shield drive is open JP16 Installed Wilson Central Terminal (WCT) connected to INM for CH1 and CH4-8 JP17 Header Not Installed ECG shield drive is connected to AGND JP18 Installed 2-3 CLK connected to OSC1 JP19 Installed 1-2 OSC1 enabled JP20 Installed 2-3 AVSS selected for bipolar supply operation (AVSS = -2.5V) JP21 Installed 1-2 CS connected to DSP via J3.1 JP22 Installed 2-3 START comes from J3.14 JP23 Installed 1-2 CLKSEL set to 0 JP24 Installed 2-3 DVDD supply = 3.3V JP25 Header Not Installed No external reference selected Installed 1-2 (top) WCT connected to CH8- input Installed 3-4 (bottom) ECG_V1 connected to CH8+ input Installed 1-2 (top) WCT connected to CH7- input Installed 3-4 (bottom) ECG_V5 connected to CH7+ input Installed 1-2 (top) WCT connected to CH6- input Installed 3-4 (bottom) ECG_V4 connected to CH6+ input Installed 1-2 (top) WCT connected to CH5- input Installed 3-4 (bottom) ECG_V3 connected to CH5+ input Installed 1-2 (top) WCT connected to CH4- input Installed 3-4 (bottom) ECG_V2 connected to CH4+ input Installed 1-2 (top) ECG_RA connected to CH3- input Installed 3-4 (bottom) ECG_LL connected to CH3+ input Installed 1-2 (top) ECG_RA connected to CH2- input Installed 3-4 (bottom) ECG_LA connected to CH2+ input Open 1-2 (top) Not connected Open 3-4 Not connected Installed 5-6 CH1- input connected to source of JP35 JP6 to JP14 JP26 JP27 JP28 JP29 JP30 JP31 JP32 JP33 Installed 7-8 (bottom) CH1+ input connected to source of JP36 JP34 Installed 1-2 Connected onboard RESP circuit to CH1- signal input mux (RESPMOD-) JP35 Installed 1-2 Connected onboard RESP circuit to CH1+ signal input mux (RESPMOD+) JP36 Installed 1-2 RESP control signal for U11/U12 controlled by MSP430 (U14) SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 11 Quick Start 2.2 www.ti.com ADS1298RECG-FE Operation To prepare to evaluate the ADS1298R with the ADS1298RECG-FE, complete the following steps: 1. Verify the jumpers on the ADS1298RECG-FE are as shown in Figure 2 (note that these settings are the factory-configured settings for the board). 2. Verify that the jumpers on the MMB0 motherboard are configured as shown below: • MMB0 J13A → Open • MMB0 J13B → Open • Refer to Appendix B2 for details about the MMB0 power supply 3. Install the ADS1298RECG-FE software using the latest software version. The latest software can be downloaded from the ADS1298RECG-FE product page. Double click the installer and follow the instruction to complete the software installation. For detailed installation information and screenshots, see Appendix C. 12 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com 3 Using the ADS1298RECG-FE Software The ADS1298RECG-FE software provides complete control over all the settings of the ADS1298R. By using the user interface (UI), the ADS1298R control registers can be manipulated to evaluate the various options available on the device. Figure 3 shows the starting UI screen of the software. The UI consists of a user menu (Section 3.1), a few top-level controls (Section 3.2), and a tabbed interface, with different functions available on the different tabs. The tabs are: • About (Section 3.3) • ADC Register (Section 3.4.3) • Analysis (Section 3.5) • Save (Section 3.6) Figure 3. Software Start Screen/About Tab The user can adjust the settings when the software is not acquiring data. During acquisition, all controls are disabled and settings may not be changed. When a setting is changed via a control, the settings are immediately updated on the device and EVM. Settings in the software correspond to settings described in the ADS1298R product data sheet. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 13 Using the ADS1298RECG-FE Software 3.1 www.ti.com Application User Menu The application user menu is located along the top of the application menu. It consists of two items: File and Help. File Menu (see Figure 4) The File menu provides several options: • Capture Screen takes a screen capture of the current view of the application and saves it as to a file specified by the user. • Save Configuration Settings saves the current states of the ADS1298R control registers for reloading at a later time. This file is different from the save register on the Save tab (see Section 3.6), which saves the current register map to a tab-delimited text file. • Load Configuration Settings loads a previously save configuration setting file and initializes the hardware and software to the settings within the configuration file. The configuration file must be a file saved previously from this application using the Save Configuration Settings command, not a file from the Save tab. • Exit closes the application. Help Menu (see Figure 5) The Help menu provides the About option, which display the software and firmware version that is currently being used. Please have this information if you need to request assistance or have a question regarding the software or hardware. Figure 4. User Menu - File Item 3.2 Figure 5. User Menu - Help Item Top-Level Application Controls Several controls/indicators are located along the top of the UI screen (see Figure 6). The controls and indicators are described below. Figure 6. Top Level Controls The Data Rate indicator displays the current data rate of the ADS1298R. The data rate can be configured in CONFIG1 control register (see Section 3.4.2.1). The Progress indicator will display the current progress of data transfer to the PC during acquisition cycles. The Samples/CH control allows for the selection of the number of points, per channel, to collect during an acquisition cycle. Keep in mind the value entered into this control in relation to the current data rate. Large numbers of samples, coupled with slower data rates, can take time to acquire. The ACQUIRE control starts the acquisition process. When pressed, the software will collect the requested number of samples from the ADS1298R. All points collected during an acquisition process will be contiguous points. 14 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com The CONTINUOUS control starts a repeated acquisition process. This function acquires the requested samples and repeats the data acquisition until the button is turned off. Within a single acquisition cycle, the points will be contiguous, but from acquisition to acquisition, there may be points missing. The Analysis Data input referred checkbox changes the displayed data that is read from the ADC. Checking the box displays the data input referred, while not checking displays the data as converted. The Show/Poll Lead Off Status displays a window (see Figure 7) that shows the status of the Lead-Off status registers, LOFF STATP and LOFF STATN, of the ADS1298R. When the lead for the channel is disconnected, the corresponding channel LED changes from green to red. Figure 7. Lead-Off Status Registers Display Window 3.3 About Tab The About tab provides software information to the user. Important safety warning, restrictions, and disclaimers for the software and hardware are shown and should be followed during the evaluation of this product. Additional indicators are present to provide device information (Device ID and Rev) and software information (Firmware Version). The About tab should be the first screen displayed at startup (see Figure 3). SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 15 Using the ADS1298RECG-FE Software 3.4 www.ti.com ADC Register Tab The ADC Register tab provides controls to manipulate ADC control registers of the ADS1298R. Details of the control registers are provided in the product datasheet. The ADC Register tab consists of a few controls and several sub-tabs that further divide the control registers into different functions groups. The sub-tabs are: • Channel Registers tab (Section 3.4.2) • LOFF and RLD tab (Section 3.3) • GPIO and Other Registers tab (Section 3.4.4) • Register Map tab (Section 3.4.5) Figure 8. Channel Registers Tab 3.4.1 Standby and Reset Controls The Standby control allows the user to place the ADS1298R in standby. The Reset control allows the user to reset the ADS1298R. The Reset Mode determines which mode is executed when the Reset control is pressed. Device Defaults resets the device to the device defaults; Programmed Defaults resets the device, then writes the default values for using this software application. 16 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com 3.4.2 Channel Registers Tab (ADC Register) The Channel Register tab provides access to control registers that control different properties/settings for the ADC channels. The control register are grouped into two groups: Global Channel Registers and Channel Control Registers. 3.4.2.1 Global Channel Registers The Global Channel Registers box includes Configuration Register 1 (CONFIG1), Configuration Register 2 (CONFIG2), Configuration Register 3 (CONFIG3), and Lead Off Control Register (LOFF). The upper half of Figure 8 shows the section of the UI panel that allows manipulation and control of these registers. Configuration Register 1 enables the user to control the resolution mode, enable the daisy-chain configuration options, and program the data rate. Configuration Register 2 enables the user to select an internal square wave test source amplitude of ±1mV or ±2mV and a frequency of DC, 2Hz (fCLK/221), or 4Hz (fCLK/220). Configuration Register 3 controls the bandgap reference (illustrated in Figure 9) and right leg drive (RLD) options. This register enables the user to select between an external or internal reference voltage, enable/disable the internal reference buffer, toggle between a 2.4V or a 4.0V output voltage, and to enable/disable the RLD as well as choose whether the RLD voltage is provided internally or externally. 22mF VCAP1 R1 (1) Bandgap 2.4V or 4V R3 VREFP (1) 10mF R2 (1) VREFN AVSS To ADC Reference Inputs Figure 9. Internal Reference and Buffer Connections The Lead-Off Control Register allows the user to configure the threshold for the lead-off comparator, resistive pull-up or current-source excitation, the lead-off current magnitude, and DC or AC detection. Figure 10 illustrates a simplified diagram of the resistive pull-up and excitation options for the lead-off detect feature. AVDD AVDD ADS129x ADS129x 10MW INP INP PGA INN PGA INN 10MW a) Pull-Up/Pull-Down Resistors b) Current Source Figure 10. Lead-Off Excitation Options SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 17 Using the ADS1298RECG-FE Software 3.4.2.2 www.ti.com Channel Control Registers The Channel Control Registers box allows the user to uniquely configure the front-end MUX for each ADC channel. Additionally, at the top of the Channel Control Registers box (see Figure 8) is the option to globally set all channels to the same setting. The channel-specific MUX is illustrated in Figure 11. ADS129x MUX INT_TEST TESTP_PACE_OUT1 INT_TEST MUX[2:0] = 101 TestP TempP MvddP (1) MUX[2:0] = 100 MUX[2:0] = 011 From LoffP MUX[2:0] = 000 VINP MUX[2:0] = 110 EMI Filter To PgaP MUX[2:0] = 010 AND RLD_MEAS MUX[2:0] = 001 (AVDD + AVSS) 2 MUX[2:0] = 111 MUX[2:0] = 000 VINN RLDIN From LoffN MUX[2:0] = 001 To PgaN MUX[2:0] = 010 AND RLD_MEAS RLD_REF MvddN (1) TempN MUX[2:0] = 011 MUX[2:0] = 100 MUX[2:0] = 101 TestN INT_TEST TESTN_PACE_OUT2 INT_TEST Figure 11. Input Multiplexer for a Single Channel 18 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com 3.4.3 LOFF and RLD Tab (ADC Register) The LOFF and RLD tab provides control over the Lead-Off Detection and Current Control Registers and the Right Leg Derivation Control Registers. The tab and controls are shown in Figure 12. Figure 12. LOFF and RLD Tab 3.4.3.1 Lead-Off Detection and Current Direction Control Registers The first two arrays of controls (Lead Off Sense) enable lead-off detection for both the positive and negative channels, LOFF_SENSP and LOFF_SENSN. By pressing the buttons, lead-off detection is enabled for each channel individually and for each input (positive and negative). Set All LOFFP Bits and Set All LOFFN Bits allow the user to turn on or off all the enable bits at once instead of clicking each individual channel control. The third array of controls (Lead Off Current Direction) determines the current direction used for lead-off detection when an excitation signal is selected as a pull-up/pull-down resistor. Each channel is controlled individually by selecting the button that corresponds to the desired channel to manipulate. When the button is not illuminated, LOFF_FLIP = 0 (INP is pulled-up to AVDD and INN is pulled-down to ground). When the button is pressed/illuminated, LOFF_FLIP = 1 (INP is pulled-down to ground and INN is pulledup to AVDD). Further details of these registers and lead-off function are located in the Applications Section of the device data sheet. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 19 Using the ADS1298RECG-FE Software www.ti.com Figure 10 describes the mode for Lead-Off Detection (that is, resistive or current source) and the 4-bit DAC settings to configure the lead-off threshold. Figure 13 illustrates the connections from the positive and negative inputs to the lead-off comparators. The output of the comparators is viewed by using Show/Poll Lead Off Status control as described in Section 3.2 LOFF_STATP VINP VINN PGA To ADC LOFF_STATN 4-Bit DAC COMP_TH[2:0] Figure 13. LOFF_STATP and LOFF_STATN Comparators 3.4.3.2 Right Leg Drive Derivation Control Registers The Right Leg Drive Derivation Control Registers enable the user to set any combination of positive and/or negative electrodes to derive the RLD voltage that is fed to the internal right leg drive amplifier. 20 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com 3.4.4 GPIO and OTHER Registers Tab (ADC Register) The GPIO and Other Registers tab, located under the Analysis tab, includes controls for GPIO1 through GPIO4, respiration phase and frequency, routing of the Wilson amplifiers, and derivation of the Goldberger terminals. Figure 14 shows the GPIO and OTHER Registers tab and all controls contained on the tab. Figure 14. GPIO and OTHER Register Tab The General-Purpose I/O Register (GPIO) controls the four general-purpose I/O pins. Each GPIO can be set as an input or an output via GPIOCx controls. If the output is selected, the GPIODx control is enabled allowing the user to set the value to output. If the GPIO is selected as an input, the GPIODx control is disabled and shows the value of the GPIO. If any of the GPIOs are selected as inputs, the Read GPIO control is enabled which allows the GPIODx values to be updated to the current GPIO value. The PACE Detect Register does not enable a special PACE measurement mode. The register allows for enabling and configuration of the PACE amplifiers. PACE Amplifier 1 can connect to input channels 1-4 and Pace Amplifier 2 can connect to input channels 5-8. The Configuration 4 Register allows control over the Respiration Frequency, WCT connection to the RLD and lead-off comparator enable status. The Respiration Control Register allows the user to configure the respiration modulation and demodulation on Channel 1, Respiration Phase, and set the Respiration Control Signal. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 21 Using the ADS1298RECG-FE Software 3.4.4.1 www.ti.com Wilson Central and Augmented Lead Registers The Wilson Central Voltage (an average voltage between the right arm [RA], left arm [LA], and left leg [LL] connections) can be derived from any combination of positive and negative terminals from channels 1-4 and routed to the WCT pin. Likewise, the Augmented Leads (AVF, AVL, AVR) may be derived from channels 1-4 and routed to the negative terminal of channels 5, 6, and 7. Figure 15 shows these configurations; Figure 15a illustrates the central lead routing, and Figure 15b shows the augmented lead routing. (b) Wilson Augmented Lead Routing (a) Wilson Central Lead Routing IN1P IN1N IN2P IN2N IN3P IN3N IN4P IN4N To Channel PGAs 8:1 MUX 8:1 MUX Wcta Wctb 8:1 MUX WCT2[2:0] Wctc 8:1 MUX WCT2[5:3] 30kW 8:1 MUX WCT1[2:0] 30kW Wctb To Channel PGAs 8:1 MUX WCT2[2:0] WCT2[5:3] WCT1[2:0] Wcta IN1P IN1N IN2P IN2N IN3P IN3N IN4P IN4N Wctc avF_ch4 30kW WCT ADS1298 80pF AVSS ADS1294/6/8 avF_ch6 avF_ch5 avF_ch7 IN5P IN5N IN6P IN6N IN7P IN7N To Channel PGAs Figure 15. Wilson Central and Augmented Lead Routing Diagrams 22 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com 3.4.5 Register Map (ADC Register) The Register Map tab is a helpful debug feature that allows the user to view the state of all the internal registers. This tab is illustrated in Figure 16. Refresh Registers control updates the register map values with the current register settings of the ADS1298R. Figure 16. Device Registers Settings SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 23 Using the ADS1298RECG-FE Software 3.5 www.ti.com Analysis Tab The Analysis tab provides access to the different analysis options that are available using the software. The different analyses are grouped by sub-tabs: • Scope tab (Section 3.5.1) • Histogram tab (Section 3.5.2) • RESP tab (Section 3.5.4) • FFT tab (Section 3.5.3) • ECG tab (Section 3.5.5) 3.5.1 Scope Tab (Analysis) The Scope tool is useful for examining the exact amplitude of the measured input signals from each channel. Additionally, users can determine the noise contribution from each channel at a given resolution, and review the sampling rate, the PGA gain, and the input signal amplitude. Figure 17 illustrates the Scope tool features. Figure 17. Scope Tool Features In the Scope Analysis window, as Figure 18 illustrates, the different noise levels are displayed when the MUX is selected as Input Short, PGA gain is set to 6 (default), and the sample rate is set to 500 samples per second (SPS). Figure 18. Scope Analysis Tab (Noise Levels for Each Channel Shown) 24 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com 3.5.1.1 Zoom Tool The zoom tool allows the user to zoom in either on all channels simultaneously or on a single channel. Figure 19 shows an example of the waveform examination tool with the magnifying glass zoomed in on Channel 2. In this case, the tool makes it much easier to determine that the noise seen on the ECG waveform is a result of 50Hz/60Hz line cycle noise. Figure 19. Zoom Tool Options SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 25 Using the ADS1298RECG-FE Software 3.5.2 www.ti.com Histogram Tab (Analysis) The Histogram tool is used primarily to see the bin separation of the different amplitudes of the ECG waveform harmonics. Figure 20 illustrates the histogram output for a 12-lead signal. The same ECG Signal Zoom analysis may be used on the histogram plots for a more detailed examination of the amplitude bins. Figure 20. Histogram Bins for 12-Lead ECG Signal Figure 21 shows the Histogram Analysis window that is displayed when the Histogram Analysis button (at the bottom of the screen in Figure 20) is clicked. The analysis window shows the mean, VRMS, and VPP channel amplitude bins. Figure 21. Statistics for the Signal Amplitude of Eight ECG Channels 26 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com 3.5.3 FFT Tab The FFT tool allows the user to examine the channel-specific spectrum as well as typical figures of merit such as SNR, THD, ENOB, and CMRR. Each feature is numbered below and described in detail in the following subsections. Figure 22 illustrates an FFT plot for a normal electrode configuration. Figure 22. FFT Graph of Normal Electrode Configuration 1 - Coherent Frequency Calculator Coherent sampling in an FFT is defined as FAIN/FSAMPLE = NWINDOW/NTOTAL, where: • FAIN is the input frequency • FSAMPLE is the sampling frequency of the ADS1298R • NWINDOW is the number of odd integer cycles during a given sampling period • NTOTAL is the number of data points (in powers of 2) that is used to create the FFT If the conditions for coherent sampling can be met, the FFT results for a periodic signal will be optimized. The Ideal AIN Frequency is a value that is calculated based on the sampling rate, such that the coherent sampling criteria can be met. 2 - AC Analysis Parameters This section of the tool allows the user to dictate the number of harmonics, DC leakage bins, harmonic leakage bins, and fundamental leakage bins that are used in the creation of various histograms. Pressing the Windowing button, illustrated in Figure 23, allows the user to evaluate the FFT graph under a variety of different windows. Note that pressing the Reference button toggles between dBFS (decibels, full-scale) and dBc (decibels to carrier). Figure 23. AC Analysis Parameters: Windowing Options SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 27 Using the ADS1298RECG-FE Software www.ti.com 3 - FFT Analysis Pressing the FFT Analysis button pulls up the FFT Analysis window shown in Figure 24. This window provides calculated parameters obtained from the collected data that may be useful during evaluation. One of the values included in this analysis is the channel-to-channel noise. Figure 24. FFT Analysis: Input Short Condition 4 - User-Defined Dynamic Range This section enables the user to examine the SNR of a specific channel within a given frequency band defined by Low Frequency and High Frequency. The SNR displayed in this window shows under the Dynamic Range heading as Figure 25 illustrates. Figure 25. Changing the User-Defined Dynamic Range for Channel 1 5 - Input Amplitude This field is a user input that is important for accurately calculating the CMRR of each channel. 6 - Zoom Tool As with the Analysis, Histogram, and Scope tool, this zoom function allows a closer examination of the FFT at frequencies of interest, as shown in Figure 26. 28 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com Figure 26. FFT Plot Using Zoom Tool SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 29 Using the ADS1298RECG-FE Software 3.5.4 www.ti.com RESP Tab The Resp tool allows the user to make use of the on-board respiration circuitry to show a simulated respiration signal. Respiration Tab shows the RESP tab tool features. Respiration Tab The bottom left portion of the screen provides controls that may be utilized to add filtering to the displayed respiration signal. NOTE: The filtering performed using this analysis tool is performed in the UI software and is not available in the ADS1298R device. The Low Pass Filter group of controls implement a lowpass digital filter on the collected data. The filter is enabled using the Enable control, and Filter Order and Cutoff Frequency allow for the customization of the filter. The High Pass Filter group of controls implement a digital high-pass filter on the collected data. The high-pass controls function the same as the low-pass controls except that they apply to the high-pass filter. The Unsettled Points to Remove control allows the user to remove a specific number of points from the beginning of the filtered data set. This allows the user to remove the data the isn't filtered and look only at the filtered data. The Simulated Respiration Frequency set the respiration frequency that the on-board respiration circuitry will simulate when collecting data. The Statistical Data group provides two calculated parameters from the displayed data, Mean and Noise. 30 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com 3.5.5 ECG Tab (Analysis) This tool allows the user to examine the input signal according to the different lead configurations. For a detailed description of the lead configurations, see Table 2 in Section 4.6. Figure 27 shows Leads I-III and the Augmented Lead outputs with the input MUX configured in Normal Electrode mode. Figure 27 also shows numerical annotations 1 to 4, which highlight the different features of this tool. These features are described in detail in the following subsections. Figure 27. ECG Display Tab Showing LEAD I-III and Augmented Leads 1 - Plot Set Selection Feature The Plot Set Selection control allows the user to change the visual selection between: • Limbs and Augmented Leads displays LEAD I, LEAD II, LEAD III, aVR, aVL, and AVF signals, • Chest Leads displays V1 - V6 signals. NOTE: For display that shows 6 leads combined, the ECG signals have any DC offset removed and a different offset added to the signal to display the signals as shown. To see the raw ECG data, you can select the individual signals as described below in the Zoom feature (box 4). 2 - ECG Separation Feature The ECG separation control toggles the vertical distance between the input plots. This capability is useful when ECG signals are large and require more separation to avoid overlap, or to collapse the range between signals when the ECG signals are small. 3 - Post Processing Filters Feature The Post Processing Filters Features controls provides a low-pass, a 50Hz/60Hz notch, and a highpass digital filters for post-processing the data from the ADS1298R. To activate each filter, the Enable checkbox should be checked. To disable the filter, the Enable checkbox should be unchecked. Any combination of the three digital filters can be used by enabling the respective filter. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 31 Using the ADS1298RECG-FE Software www.ti.com The low-pass filter controls a digital low-pass filter, whose order and cutoff frequency are controlled using the Filter Order and Cutoff Freq controls in the low-pass filter part of the Post Processing Filters group (left side of the box). The notch filter provides a 50Hz/60Hz notch filter, whose order and 50Hz/60Hz notch selection are controlled using the Filter Order and Notch Freq controls in the low-pass filter part of the Post Processing Filters group (center of the box). The high-pass filter controls a digital high-pass filter, whose order and cutoff frequency are controlled using the Filter Order and Cutoff Freq controls in the high-pass filter part of the Post Processing Filters group (right side of the box). NOTE: The digital filters are not part of the ADS1298R. These are digital filters implemented in the UI to aid in the evaluation of the ADS1298RECG-FE. 4 - Zoom Feature The zoom feature is available to allow the user to navigate and view all signals at the same time, as shown in Figure 28. This tool allows the user to zoom in/out on the horizontal or vertical axis and pan left or right while viewing all ECG signals simultaneously. Figure 28. ECG Signal Zoom Feature for Six Leads Additionally, each ECG signal can be zoomed individually by moving the mouse (which appears as a plus icon) over the lead of interest and clicking on it. A new window opens showing the raw ECG data as read from the ADS1298R. This window provides controls in the lower right corner to zoom in/out or pan right/left to provide a more detailed inspection of the individual ECG signal. Figure 29. ECG Signal Zoom Feature for Lead 1 32 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Using the ADS1298RECG-FE Software www.ti.com 3.6 Save Tab The Save tab provides the user the ability to save the collected data for a record of the evaluation or further analysis. Reference the previous sections for the list of the analysis data available for each analysis. The Analysis to Save group allows the user to save the different analysis calculations that were performed on the data. • Scope Analysis saves the scope analysis data available from the scope analysis pop-up window • FFT Analysis saves the FFT analysis data available from the FFT analysis pop-up window. • Histogram Analysis saves the histogram analysis data available from the histogram analysis pop-up window. • Register Settings saves the current settings from the register map and can be useful to obtain the register values for your specific device configuration. Saving the register map in this format is not be confused with saving your register settings for reloading into the software at another time (see Section 3.1). Each item will be saved if the corresponding checkbox is checked. The Data to Save group allows the user to save the various data sets collected from the ADS1298R. The CH x controls allow the user to specify which channels will be saved for each of the data selections made. • Data - Codes selects the raw data (in codes format) to be saved to a file. • Data - Volts selects the raw data (converted to voltage) to be saved to a file. • FFT Data selects the calculated FFT data to be saved to a file. Note: This is not raw data; it is the frequency bin and magnitude data that was calculated by the UI. • Histogram Data selected the calculated Histogram data to be saved to a file. Note: This is not raw data; it is the code bin and number of occurrences data that was calculated by the UI. The User Comments/Notes group allows the user to indicate a Record Number and User Comments that are saved with each file. This data permits the user to distinguish different data sets from one another. The Directory to Save Files is the directory where all the saved files will be placed. The user can select a directory by pressing the folder button (located to the right of the control). Each data file that is saved is automatically named to prevent overwriting of files. The Save To File button saves all the data files that were selected using the check boxes to the selected directory. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 33 Using the ADS1298RECG-FE Software www.ti.com Figure 30. Save Tab 34 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298RECG-FE Input Signals www.ti.com 4 ADS1298RECG-FE Input Signals NOTE: Before evaluating specific ECG functions, it is recommended that the user acquire data with inputs shorted internally. This configuration ensures that the board is operating properly. 4.1 Input Short Testing By default, the EVM powers up with the individual channels to an internal short with a data rate of 500SPS and a PGA gain of 6. Once the Acquire button is pressed, the Scope Analysis should reflect inputreferred VPP values less than 5µVPP 4.2 Internal Test Signals Input Configuration Register 2 controls the signal amplitude and frequency of an internally-generated square wave test signal. The primary purpose of this test signal is to verify the functionality of the front-end MUX, the PGA, and the ADC. The test signals may be viewed on the ECG Display tab, as Figure 31 shows. Detailed instructions for using the ECG Display tab are provided in Section 3.5.5. Figure 31. Example of Internal Test Signals Viewed on the ECG Display Tab SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 35 ADS1298RECG-FE Input Signals 4.3 www.ti.com Temperature Sensor The internal temperature sensor on the ADS1298R is shown in Figure 32. When the internal MUX is routed to the temperature sensor input, the ADC internal temperature is calculated from the ADC output voltage using Equation 1. AVDD 1x 2x 8x 1x AVSS Figure 32. Internal Temperature Sensor Temperature (°C) = Temperature Reading (mV) - 145,300mV 490mV/°C + 25°C (1) The ADC can be configured to give a temperature reading by selecting the Temperature Sensor option on the Channel Control Registers GUI (see Section 3.4.2.2) and verified using the Scope tab as shown in Figure 33. The number 0.1447V (on the y-axis) can be calculated as a temperature using Equation 1: Temperature = (0.1447 – 0.145300) / 0.00049 + 25 = 23.78°C A more detailed description of the Scope tab is provided in Section 3.5.1. Figure 33. Eight-Channel Read of Internal Temperature 36 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298RECG-FE Input Signals www.ti.com 4.4 Normal Electrode Input The Normal Electrode input on the MUX routes the inputs (VINP and VINN) differentially to the internal PGA, as Figure 11 illustrates. In this mode, an ECG, sine wave, or pulse generator may be connected to test the ADS1298R. Figure 34 shows a typical six-lead output when connected to a 5mVPEAK, 80BPM ECG signal. Figure 34. Normal Electrode ECG Connection in ECG Display Tab 4.4.1 Capturing 12-Lead ECG Signals To capture signals from external inputs: 1. Configure the Channel Input control in Globally Set Channels to Normal Electrode. 2. Connect the 10 ECG electrodes from the Fluke simulator to the EVM through the DB15 connector (J1). Refer to Section A.3 for the ECG cable details. The ECG electrode signals are passed through a single pole RC filter followed by the lead configuration. For ECG signal processing, the electrode signals are routed through J5 to the ADS1298R input. The signal path in the board can be chosen by jumper settings, depending on the application. 4.5 MVDD Input, RLD Measurement, RLD Positive Electrode Driver, and RLD Negative Electrode Driver The MVDD input option allows the measurement of the supply voltage VS = (AVDD + AVSS)/2 for channels 1, 2, 5, 6, 7, and 8; however, the supply voltage for channel 3 will be DVDD/2. As an example, in bipolar supply mode, AVDD = 3.0V and AVSS = –2.5V. Therefore, with the PGA gain = 1, the output voltage measured by the ADC will be approximately 0.25V. The RLD measurement takes the voltage at the RLDIN pin and measures it on the PGA with respect to (AVDD + AVSS)/2. This feature is beneficial if the user would like to optimize the gain of the RLD loop. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 37 ADS1298RECG-FE Input Signals www.ti.com The voltage used to derive the right leg drive for both the positive and negative electrodes may also be measured with respect to (AVDD + AVSS)/2. 4.6 Lead Derivation The EVM is configured to generate the 12 ECG signals using 10 electrodes connected to the eight ADC channels. Lead I, Lead II, and V1-V6 are computed in the analog domain, while the augmented leads and Lead III are computed digitally. The channel assignments are described in Table 2. • LA = Left Arm • LL = Left Leg • RA = Right Arm Table 2. ADS1298R Lead Measurements Lead (1) ADS1298R Input Channels (1) 1 V6 = V6 – WCT 2 LEAD I = LA – RA 3 LEAD II = LL – RA 4 V2 = V2 – WCT 5 V3 = V3 – WCT 6 V4 = V4 – WCT 7 V5 = V5 – WCT 8 V1 = V1 – WCT WCT = (LA + RA + LL)/3 Table 3. Derived Lead Calculations Derived Lead Formula Used to Calculate LEAD III 4.7 LL - RA - LA = LEAD II - LEAD I aVR RA - (LA + LL) / 2 = - (LEAD I + LEAD II) / 2 aVL LA - (RA +LL) / 2 = LEAD I - LEAD II / 2 aVF LL - (RA + LA) / 2 = LEAD II - LEAD I /2 Wilson Center Terminal (WCT) The Wilson Center Terminal voltage is internally generated by the ADS1298R device. The WCT1 and WCT2 registers provide controls to select any of the eight inputs (CH1P to CH4P, CH1M to CH4M) for routing to the three integrated WCT amplifiers. The ADS1298RECG-FE is configured for 12-lead ECG inputs, with the limb electrodes connected as shown in Table 2. During EVM power-up, the firmware configures the device to route CH2P, CH2M, and CH3P (RA, LA, LL) to the internal buffers. This configuration generates a signal at the WCT pin equal to (RA + LA + LL)/3. By installing JP16, the WCT is routed to the single-ended channels to achieve the desired signals. 4.8 Right Leg Drive The RL electrode is driven directly by the RLD signal generated on-chip by the ADS1298R. The bandwidth of the RLD loop is determined by R8 (392kΩ) and C20 (10nF). Users can change these values to set the bandwidth based on their specific application. The loop stability is determined by the user’s specific system. Therefore, adjustment of the feedback component values may be required to ensure stability if additional filtering components and long cables are added before the ADS1298RECG-FE. In a typical application, the RLD signal is implemented as the average of RA, LA, and LL. For system flexibility, the ADS1298R allows the user to select any combination of the electrodes to generate the RLD (see ADS1298R data sheet for more details). 38 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298RECG-FE Input Signals www.ti.com 4.8.1 RLD Common Mode Voltage The RLD common mode voltage can be set to (AVDD+AVSS)/2 or to an externally provided source. If the application requires the common mode to be set to any voltage other than mid-supply, this can be accomplished by setting the appropriate bit in the Configuration 3 Register. On the ADS1298ECG-FE, the external RLDREF voltage is set using resistor R1 and adjustable resistor R2 (R1 and R2 are not installed by default). During power-up, the firmware configures the device for internal RLDREF operation. To configure the RLD circuitry manually, use the following steps and the controls found on the ADC Register tab. 1. Verify that the Channel Input is set to the Normal Electrode mode for all channels (Channel Registers tab). 2. In CONFIG3 control register (Channel Registers tab): • Enable the RLD Buffer (RLD Buffer Power = Enable) • Set the internal RLD reference (RLDREF Signal Source) 3. Select the electrodes for the RLD loop from the Right Leg Drive Derivation Control Registers controls (LOFF and RLD tab) Once these steps are completed, measure and verify that the voltage on either side of R38 is close to mid-supply. This measurement confirms whether the RLD loop is functional. The on-chip RLD signal can be fed back into the ADS1298R by shorting JP1. This RLD signal can then be sent to the ADC (to measure for debug purposes) or to other electrodes for driving (to change the reference drive in case the RL electrode falls off). Refer to the ADS1298R product data sheet for additional details. 4.8.2 Driving the RLD Cable Shield Apart from the RLD signal, the ADS1298RECG-FE also offers three options to drive the cable shield: • In-phase RLD signal • Out-of-phase RLD signal • Board AGND Table 4 summarizes the configuration of JP15 and JP17 for each of the options. Table 4. RLD Jumper Options 4.9 ECG Cable ELEC_SHD signal JP15 JP17 AGND 1-2 Don't Care RLD (0: In phase) 2-3 2-3 RLD (180: Out of phase) 2-3 1-2 PACE Detection The ADS1298R supports data rates up to 32kSPS for software PACE detection, which typically requires a data rate of at least 8kSPS. NOTE: The ADS1298RECG-FE does not include software PACE detection algorithms. The ADS1298R provides the user the flexibility of doing hardware PACE detection with external circuitry. PACE detection can be done simultaneously on two channels: one from the odd channels and one from the even channels. Refer to the ADS1298R product data sheet for additional details. To turn on the PACE buffer and select the channels, set the PACE Register from the GPIO and OTHER Registers tab. The PGA outputs of the selected channels are available at connector J5, pins 1 and 2. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 39 ADS1298RECG-FE Input Signals www.ti.com Figure 35 shows an example waveform created by a Fluke Medsim 300B processed by the ADS1298R at a data rate of 8kSPS. Using higher data rates increases power consumption because all channels must sample at this data rate simultaneously; thus, the PACE buffers offer the flexibility to process PACE signals separately from the ADS1298R. The signal must be AC coupled to obtain the waveform\ shown below. Figure 35. Digitization of PACE Signal Using ADS1298R 40 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Evaluation of the ADS1298R Respiration Function www.ti.com 5 Evaluation of the ADS1298R Respiration Function 5.1 Introduction The ADS1298R allows for the measurement of respiration rate based on the principle of impedance pneumography. The key concept of this approach is to measure the change in impedance of the thoracic cavity during breathing (respiration). The ADS1298REVM provides two ways to test the respiration circuitry: first, using onboard circuitry; and second, using a patient simulator with respiration outputs. This section describe both approaches. 5.2 Testing with Onboard Circuitry For additional information on Impedance Pneumography, please refer to the following application note. 5.2.1 Hardware Configuration The ADS1298REVM has onboard circuitry to test a respiration circuit. A simplified diagram illustrating only the respiration portion of the circuit is shown in Figure 36; the onboard circuitry to generate the respiration signal is within the red dotted box. By default, jumpers JP34 and JP35 connect the onboard circuitry for respiration evaluation. R77 models the baseline impedance of the thoracic cavity. This component is a surface-mount resistor located on the underside of the board; it can be changed if desired. R78 and R79 model the cable resistance (typically 1kΩ for patient monitoring). Capacitors C97 and C98 are used to model the parasitic capacitance that may be present when real cables are used. Capacitors C113 and C114 serve as a secondary means to prevent a single fault (such as a shorted C112 capacitor from a carrier generator) from causing excessive DC currents through the electrodes. The parameter of interest is the change in impedance during breathing. This change in impedance is accomplished by switching 1MΩ impedance in parallel to R77 using an analog switch, U11. Two 1MΩ resistors (R82 and R83) have been added to provide a DC bias to this switch. Switch U11 must have a control signal that toggles between AVSS and AVDD at the desired respiration frequency. This control signal can be fed from either an onboard signal source or an external function generator via JP36. Figure 36. Respiration Evaluation with Onboard Circuitry SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 41 Evaluation of the ADS1298R Respiration Function www.ti.com The remainder of the circuit is necessary for respiration measurements with both the onboard circuitry and patient simulator. Resistors R96 and R97 limit the amount of AC current that flows into the body. Capacitors C108 and C109 block any DC current from flowing into the body from the transmission side. Capacitors C99 and C100 serve the same purpose on the receiver side. The respiration capability is available on Channel 1. The respiration signals are routed to Channel 1 when the two shunts on JP33 are in the respective default locations, shorting pins 5-6 and 7-8 as shown in Figure 36. 5.2.2 Software Configuration The Respiration Control Register works in conjunction with Configuration Register 4 (CONFIG4). Certain bit changes must be done in order to activate the respiration circuitry of the ADS1298R. For internal respiration with an internal clock, set the following: • Configuration Register 4 – Respiration Frequency to 32kHz • CHxSET registers – Normal Electrode – Gain to 6 • RESP register set to: – Respiration Demodulation to Enabled – Respiration Modulation to Enabled – VREF to VREFP – Respiration Phase to 112.5 deg – Respiration Control to Internal Respiration with Internal Clock Next, switch U11 must be toggled at the desired respiration frequency by applying a square wave via JP36. An onboard MSP430G2121 is provided to give approximately 0.1Hz through 0.5Hz. The MSP430 circuitry is selected when JP36 is loaded with a shunt jumper that shorts pins 1-2 (default). An external function generator can also be used for this purpose by applying a signal to SMA connector J6 and moving the shunt jumper on JP36 to cover pins 2-3. Data can be acquired by clicking on the Acquire tab. Figure 37 shows the results for a 0.5Hz toggling of the switch. The expected DC output can be calculated using Equation 2. R77 + R78 +R79 · (VREFP - VREFM) DC_V = R96 + R97 + R77 + R78 +R79 DC_V = 0.33k + 1k +1k · 2.4 = 67.9mV 40k + 40k + 0.33k + 1k +1k (2) The expected peak-to-peak output as a result of the impedance change can be calculated with Equation 3; this value is the current flowing through the body: 2.4 VREFP - VREFM IB = = 29.15mA = R77 + R78 + R79 + R96 + R97 82.33k (3) The Δ impedance and peak-to-peak output can be calculated with Equation 4 and Equation 5. DR = R77 - R77 || R76 = 330 - 329.89 = 0.11W (4) DV = DR · IB = 0.11 · 29.15 = 3.2mV (5) The results shown in Figure 37 are taken using a data rate of 500SPS. To obtain better resolution, these results must be low-pass filtered. Figure 38 shows the result after the signal has been processed through a 2Hz low-pass filter. 42 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Evaluation of the ADS1298R Respiration Function www.ti.com Figure 38. Results After Low-Pass Filtering Figure 37. Channel 1 Δ Impedance Measurement SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 43 Evaluation of the ADS1298R Respiration Function 5.3 5.3.1 www.ti.com Testing with Patient Simulator Hardware Configuration The output from any typical patient simulator can be directly fed into the DB15 connector. The shunts on jumpers JP34 and JP35 should be moved from the respective default positions (shorting pins 1-2) so that these components now short pins 2-3. 5.3.2 Software Configuration We then must write to the following registers before we take any data. Under the ADC Register, Channel Registers tab, the Global Channel Registers should be configured to the software default configuration (see Section 3.4.1 for further details on software default configuration). For internal respiration with an internal clock, set the following: • Channel Control Registers control – Channel Input to Normal Electrode – Gain to 6 • Configuration Register 4 – Respiration Frequency to 32kHz – Pulse Mode Enabled to Disable – Lead-off Comparator Power-down to Enable • RESP register set to: – Respiration Demodulation to Enable – Respiration Modulation to Enable – VREF to VREFP – Respiration Phase to 112.5 deg – Respiration Control to Internal Respiration with Internal Clock There are several options that permit users to choose different base impedance (RB) and delta impedance (ΔR) on the simulator. For illustration purposes, a base impedance RB of 500Ω is chosen. Figure 39 shows the results with RB= 500Ω and ΔR = 1Ω. The expected DC output can be calculated with Equation 6: RB · (VREFP - VREFM) DC_V = RB + R96 +R97 DC_V = 44 0.5k · 2.4 = 14.9mV 0.5k + 40k + 40k (6) Current flowing through the body is calculated by Equation 7: 2.4 VREFP - VREFM IB = = 29.81mA = 80.5k R96 + R97 + RB (7) The peak-to-peak output can then be calculated as shown in Equation 8: DV = DR · IB = 1 · 29.15 = 29.1mV (8) ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Evaluation of the ADS1298R Respiration Function www.ti.com Figure 39 and Figure 40 show the results with RB = 500Ω and ΔR = 1Ω and with RB= 500Ω and ΔR = 0.1Ω, respectively. Figure 39. Channel 1 Result for RB= 500Ω and ΔR = 1Ω Figure 40. Channel 1 Result for RB= 500Ω and ΔR = 0.1Ω These two approaches to performing respiration measurements show that the internal respiration circuitry of the ADS1298R can resolve down to 100mΩ. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 45 ADS1298R Daughter Card Hardware Details 6 www.ti.com ADS1298R Daughter Card Hardware Details The ADS1298RECG front-end evaluation board is configured to be used with the TI MMB0 data converter evaluation platform. The ADS1298RECG-FE board is a four-layer circuit board. The board schematic and layout are provided in Appendix A. The ADS1298R can be used as a demonstration board for standard, 12-lead ECG applications with an input configuration of 10 electrodes. Users can also bypass the 12-lead configuration and provide any type of signal directly to the ADS1298R through a variety of hardware jumper settings (JP26-JP33; see Section A.2). External support circuits are provided for testing purposes such as external references, clocks, lead-off resistors, and shield drive amplifiers. Figure 41 shows the functional block diagram with important jumper names for the EVM. ADS1298REVM Power Connector J4 Shield Drive ADS1298R Signals Connectors Power Mgmt High-pass Filter Right Leg Electrodes MMB0/C5505 Interface Connectors ADS1298R ECG SOC JP6 to J14 Input Signal Connector 3 Limb and 6 Chest J1 Electrodes Diode Clamp Low-pass Filter DB Connector for 10 ECG Electrodes Shield JP26JP33 JP3 External Reference (optional) JP18 External Clock (optional) Figure 41. ADS1298R Front-End Block Diagram 6.1 Jumper Description Table 5 shows the jumpers on the ADS1298RECG-FE and options available for each jumper. Table 5. ADS1298RECG-FE Default Jumper/Switch Configuration (1) 46 Jumper Function Settings JP1 Installed RLD feedback JP2 AVDD supply source 1-2: AVDD selected for bipolar supply operation (AVDD = +2.5V) 2-3: AVDD selected for single supply operation (AVDD = +3.0V) JP3 External Reference Connection (Header Not Installed) (1) Open: External reference not connected Installed: External reference connected JP4 Connect EVM +5V rail to J4(power header) Open: EVM +5V must be supplied externally Installed: EVM +5V supplied from J4 (power header) JP5 PWDN source Open: PWDN pin controlled from J5 header (pulled up to DVDD) Installed: Device is powered down (PWDN pin = AGND) JP6 to JP114 Input signal DC/AC couples (Header Not Installed) 1-2: DC-coupled input signals (Pins 1-2 shorted on PCB) 2-3: AC-coupled input signals (Requires installation of header and cutting PCB short of pin 1-2) JP15 ECG shield drive connected 1-2: ECG shield is grounded (AGND) 2-3: ECG shield is connected to buffer (required U2 installation, otherwise shield connection is open) JP16 Wilson Central Terminal (WCT) connection Open: WCT NOT connected to JP26-30 and JP33 Installed: WCT connected to JP26-30 and JP33 for connection to CH1 and CH4-8 IN- Requires installation of JP25 and references U3/U4 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R Daughter Card Hardware Details www.ti.com Table 5. ADS1298RECG-FE Default Jumper/Switch Configuration (continued) Jumper Function Settings JP17 ECG shield drive buffer input(Header Not Installed) (2) 1-2: ECG shield drive connected to RLDOUT 2-3: ECG shield drive connected to RLDINV JP18 CLK connection 1-2: CLK connected to DSP (J3.17) 2-3: CLK connected to OSC1 JP19 OSC1 Enable 1-2: OSC1 enabled 2-3: OSC1 disabled JP20 AVSS supply source 1-2: AVSS selected for single supply operation (AVSS = 0V (AGND)) 2-3: AVSS selected for bipolar supply operation (AVSS = -2.5V) JP21 CS source 1-2: CS connected to DSP via J3.1 2-3: CS connected to DSP via J3.7 JP22 START source 1-2: START comes from J3.1 2-3: START comes from J3.14 JP23 CLKSEL source 1-2: External Master Clock 2-3: Internal Master Clock (CLKSEL controlled by J3.2 (pulled up to DVDD)) JP24 DVDD Supply Select 1-2: DVDD supply = 1.8V 2-3: DVDD supply = 3.3V JP25 External Reference Selection (Header Not Installed) (3) 1-2: Select U4 as reference source 2-3: Selected U3 as reference source 1-2: CH8- connection Open: Channel input not connected Installed: Channel input connected to WCT (requires JP16 to be installed) 3-4: CH8+ connection Open: Channel input not connected Installed: Channel input connected to ECG_V1 1-2: CH7- connection Open: Channel input not connected Installed: Channel input connected to WCT (requires JP16 to be installed) 3-4: CH7+ connection Open: Channel input not connected Installed: Channel input connected to ECG_V5 1-2: CH6- connection Open: Channel input not connected Installed: Channel input connected to WCT (requires JP16 to be installed) 3-4: CH6+ connection Open: Channel input not connected Installed: Channel input connected to ECG_V4 1-2: CH5- connection Open: Channel input not connected Installed: Channel input connected to WCT (requires JP16 to be installed) 3-4: CH5+ connection Open: Channel input not connected Installed: Channel input connected to ECG_V3 1-2: CH4- connection Open: Channel input not connected Installed: Channel input connected to WCT (requires JP16 to be installed) 3-4: CH4+ connection Open: Channel input not connected Installed: Channel input connected to ECG_V2 1-2: CH3- connection Open: Channel input not connected Installed: Channel input is connected to ECG_RA 3-4: CH3+ connection Open: Channel input not connected Installed: Channel input is connected to ECG_LL 1-2: CH2- connection Open: Channel input not connected Installed: Channel input is connected to ECG_RA 3-4: CH2+ connection Open: Channel input not connected Installed: Channel input is connected to ECG_ LA 1-2: CH1- connection (Chest Lead 6 connected) Open: Channel input not connected Installed: Channel input connected to WCT (requires JP16 to be installed) 3-4: CH1+ connection (Chest Lead 6 connected) Open: Channel input not connected Installed: Channel input connected to ECG_V6 5-6: CH1- connection (Respiration connected) Open: Channel input not connected Installed: Channel input connected to CH1- input signal mux (see JP34) 7-8: CH1+ connection (Respiration connected) Open: Channel input not connected Installed: Channel input connected to CH1- input signal mux (see JP35) JP26 JP27 JP28 JP29 JP30 JP31 JP32 JP33 (2) (3) Requires installation of U2 Requires installation of JP3 and references U3/U4 SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 47 ADS1298R Daughter Card Hardware Details www.ti.com Table 5. ADS1298RECG-FE Default Jumper/Switch Configuration (continued) Jumper 6.2 Function Settings JP34 CH1- Input Signal Mux 1-2: Connect RESP MOD- to JP33 2-3: Connect ECG_LA to JP33 JP35 CH1+ Input Signal Mux 1-2: Connect RESP MOD+ to JP33 2-3: Connect ECG_RA to JP33 JP36 Control signal for U11/U12 for on-board RESP circuitry 1-2: Use MSP430 as control signal source 2-3: Use External clock as control signal source Power Supply The ADS1298RECG front-end EVM mounts on the MMB0 EVM with connectors J2, J3 and J4. The main power supplies (+5V, +3V and +1.8V) for the front-end board are supplied by the host board (MMB0) through connector J4. All other power supplies needed for the front-end board are generated on board by power management devices. The EVM is shipped in +3V unipolar supply configuration. The ADS1298R can operate in a single supply with +3.0V to +5.0V analog supply (AVDD/AVSS) or bipolar mode supply (±1.5V to ±2.5V). An additional digital supply of and +1.8V to +3.0V digital supply (DVDD) is required for operation. The ADS1298REVM power consumption can be measured by removing the JP4 jumper and JP24 jumper to connect an ammeter. By shorting JP5, the ADS1298R can be placed in powerdown mode for low power consumption. Test points TP5, TP6, TP7, TP8, TP9, TP10, and TP14 are provided to verify that the host power supplies are correct. The corresponding voltages are shown in Table 6. Table 6. Power-Supply Test Points Test Point Voltage TP7 +5.0V TP9 +1.8V TP10 +3.3V TP5 +3.0V TP13 +2.5V TP6 –2.5V TP8 GND The front-end board must be properly configured in order to achieve the various power-supply schemes. The default power-supply setting for the ADS1298R is a unipolar analog supply of 3V or a bipolar analog supply of ±2.5V and DVDD of either +3V or +1.8V. Table 7 shows the board and component configurations for each analog power-supply scheme; Table 8 shows the board configurations for the digital supply. Table 7. Analog Supply Configurations (AVDD/AVSS) Unipolar Analog Supply 48 ADS1298R Bipolar Analog Supply AVDD/AVSS 3V 5V ±1.5V ±2.5V JP20 1-2 1-2 2-3 2-3 JP2 2-3 2-3 1-2 1-2 U7 TPS73230 TPS73250 Don't Care Don't Care U9 Don't Care Don't Care TPS73201 TPS73201 U8 Don't Care Don't Care TPS72301 TPS72301 R52 Don't Care Don't Care 21kΩ 47.5kΩ R53 Don't Care Don't Care 78.7kΩ 43kΩ R56 Don't Care Don't Care 23.3kΩ 49.9kΩ R57 Don't Care Don't Care 95.3kΩ 46.4kΩ C87, C66, C62 Optional Optional Optional Optional SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R Daughter Card Hardware Details www.ti.com Table 8. Digital Supply Configurations (DVDD/DGND) 6.3 DVDD +3.0V +1.8V JP24 1-2 2-3 Clock The ADS1298R has an on-chip oscillator circuit that generates a 2.048MHz clock (nominal). This clock can vary by ±2% over temperature. For applications that require higher accuracy, the ADS1298R can also accept an external clock signal. The ADS1298R provides an option to test both internal and external clock configurations. It also provides an option to generate the external clock from either the onboard oscillator or from an external clock source. The onboard oscillator is powered by the DVDD supply of the ADS1298R. Care must be taken to ensure that the external oscillator can operate either with +1.8V or +3.0V, depending on the DVDD supply configuration. Table 9 shows the jumper settings for the three options for the ADS1298R clocks. Table 9. CLK Jumper Options ADS1298R Clock Internal Clock External OSC Clock External Clock JP18 Not Installed 2-3 1-2 JP23 Open or 2-3 1-2 (Disable) Don't Care J3 – pin 17 Don't Care Don't Care Clock Source A 2.048MHz oscillator installed on the EVM for +3V DVDD operation is FXO-HC735-2.048MHz. If operation at +1.8V DVDD is desired, the oscillator will need to be replaced. SiT8002AC-34-18E-2.048 is a possible oscillator for +1.8V DVDD operation. The EVM is shipped with the external oscillator enabled. 6.4 Reference The ADS1298R has an on-chip internal reference circuit that provides reference voltages to the device. Alternatively, the internal reference can be powered down and VREFP can be applied externally. This configuration is achieved with the external reference generators (U3 and U4) and driver buffer. The external reference voltage can be set to either 4.096V or 2.5V, depending on the analog supply voltage. Measure TP3 to make sure the external reference is correct. The setting for the external reference is described in Table 10. Table 10. External Reference Jumper Options Internal Reference External Reference ADS1298R Reference VREF = 2.5V VREFP = 4.096V JP25 Don't Care 2-3 VREFP = 2.5V 1-2 JP3 Not Installed Installed Installed The software uses the VREF value from the Reference Voltage control (CONFIG3 register) in Section 3.4.2.1) to calculate the input-referred voltage value for all the tests. The default value is 2.5V. If the user is using an alternative value, the control must be updated to display the collected data to the proper scale. 6.5 Analog Output Signals Several output signals from the ADS1298R are provided on the J5 header. Table 11 lists the various test signals and their location on the header. The PACEOUT pins can also be used as an auxiliary differential input channel. Alternatively with appropriate user configuration, these pins may provide PACE detection for use with external PACE detection circuitry (see PACE Detect Register in Section 3.4.4). SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 49 ADS1298R Daughter Card Hardware Details www.ti.com Table 11. Test Signals Signal 6.6 J5 Pin Number Signal PACEOUT1 1 2 PACEOUT2 RESERVE 3 4 RESERVE GPIO4 5 6 PWDNB GPIO3 7 8 DAISY_IN GND 9 10 VCC_5V Digital Signals The ADS1298R digital signals (including SPI interface signals, some GPIO signals, and some of the control signals) are available at connector J3. These signals are used to interface to the MMB0 board DSP. The pinout for this connector is given in Table 12. Table 12. Serial Interface Pinout Signal 6.7 J3 Pin Number Signal START/CS 1 2 CLKSEL CLK 3 4 GND NC 5 6 GPIO1 CS 7 8 RESETB NC 9 10 GND DIN 11 12 GPIO2 DOUT 13 14 NC/START DRDYB 15 16 SCL NC 17 18 GND NC 19 20 SDA Analog Input Signals The ADS1298R provides users the option to feed in standard ECG signals from a patient simulator to the DB15 connector, or to feed inputs from any arbitrary signal source directly to the ADS1298R. 50 ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R Daughter Card Hardware Details www.ti.com 6.7.1 Patient Simulator Input The output from any typical patient simulator can be directly fed in to the DB15 connector. For all measurements in this user guide, a Fluke medSim 300B simulator was used, as Figure 42 shows . The simulator is capable of generating ECG signals down to 50µV of amplitude. Particular attention must be given to the common-mode value of the input signal for proper data capture. Refer to the ADS1298R product data sheet for the common-mode range for various programmable gain amplifier (PGA) gain settings. Section 4.4.1 explains the process used to capture 12-lead ECG data. Figure 42. Fluke Simulator Configuration 6.7.2 Arbitrary Input Signals Arbitrary input signals can be connected to the ADS1298R by bypassing the DB15 connector and feeding the signal directly at jumpers JP26-JP33. This requires the removal of the 16 jumpers at JP26-JP33. The input signal must be connected differentially since each ADC channel input is differential. If it is desired to connect single-ended signals, bias the negative input of the channels to a mid-supply voltage. NOTE: Ensure that the single-ended signal has an offset equal to the voltage supplied at the negative input of the channel. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R 51 Appendix A SBAU181B – March 2011 – Revised Janurary 2016 BOM, Layout, and Schematics This section contains the complete bill of materials, printed circuit board (PCB) layouts, and schematic diagrams for the ADS1298R. NOTE: Board layouts are not to scale. These are intended to show how the board is laid out; they are not intended to be used for manufacturing ADS1298R PCBs. A.1 ADS1298R Front-End Board Schematics The ADS1298R schematic is appended to this document. 52 BOM, Layout, and Schematics SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Printed Circuit Board Layout www.ti.com A.2 Printed Circuit Board Layout Figure 43 through Figure 46 show the ADS1298R PCB layouts. A.3 Figure 43. Top Component Placement Figure 44. Bottom Component Placement and Routing Figure 45. Internal Ground Plane (Layer 2) Figure 46. Internal Power Plane (Layer 3) ECG Cable Details Figure 47 shows the details of the recommended ECG cable. Cable details: • 10-lead ECG cable for Philips/HP-snap, button (Part No: 010302013); http://www.biometriccables.com/index.php?productID=692 • 10-lead ECG cable for Philips/HP-Clip-on type (Part No: 010303013A); http://www.biometriccables.com/index.php?productID=693 SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated BOM, Layout, and Schematics 53 ECG Cable Details www.ti.com Another compatible cable for the ADS1298R: HP/Philips/Agilent-compatible 10-lead ECG cable. Figure 47. ECG Cable Schematic 54 BOM, Layout, and Schematics SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Bill of Materials www.ti.com A.4 Bill of Materials Table 13 lists the bill of materials for the ADS1298R. Table 13. Bill of Materials: ADS1298R Item Quantity Reference Designator 1 1 NA 16 C1, C2, C3, C4, C5, C6, C11, C17, C47, C48, C49, C52, C58, C76, C77, C96 Capacitor, ceramic 1µF 25V 10% X5R 0603 3 0 C7, C8, C15, C19, C21, C22, C25, C27, C34, C36, C38, C40, C41, C42, C43, C44, C56, C62, C67, C78, C79, C81, C83, C85, C87, C89, C92, C101 - C107, C110, C111 Not installed 4 1 C9 5 11 C10, C45, C46, C50, C51, C54, C55, C60, C61, C65, C66 6 13 7 Manufacturer Part Number TI 6522881 Murata GRM188R61E105KA12D Capacitor, ceramic 22µF 6.3V 10% X5R 0805 Taiyo Yuden JMK212BJ226KG-T Capacitor, ceramic 10µF 10V 10% X5R 0805 Murata GRM219R61A106KE44D C12, C13. C14, C16, C18, C57, C69, C70, C94, C95, C112, C113, C114 Capacitor, ceramic 0.1µF 50V 10% X7R 0603 Murata GRM188R71H104KA93D 1 C20 Capacitor, ceramic 1500pF 50V 5% C0G 0603 Murata GRM1885C1H152JA01D 8 19 C23, C24, C26, C28, C29, C30, C31, C32, C72, C73, C74, C75, C80, C82, C84, C86, C88, C91, C93 Capacitor, ceramic 47pF 50V 5% C0G 0603 Murata GRM1885C1H470JA01D GRM185R60J225KE26D 2 Description Printed Wiring Board 9 0 C33, C35, C37 Not installed 10 0 C39 Not installed 11 4 C53, C59, C63, C64 Capacitor, ceramic 2.2µF 6.3V 10% X5R 0603 Murata 12 2 C68, C71 Capacitor, ceramic 100µF 10V 20% X5R 1210 Taiyo Yuden LMK325BJ107MM-T 13 1 C90 Capacitor, ceramic 1000pF 50V 5% C0G 0603 Murata GRM1885C1H102JA01D 14 2 C97, C98 Capacitor, ceramic 560pF 50V 5% C0G 0603 Murata GRM1885C1H561JA01D 15 4 C99, C100, C108, C109 Capacitor, ceramic 2200pF 50V 5% C0G 0603 Murata GRM1885C1H222JA01D 16 0 D1 - D10 17 1 D11 Lumex SML-LX0603GW-TR 18 1 J1 FCI D15S13A4GV00LF Tyco 5747845-3 19 2 J2, J3 (Bottom) 10-Pin, Dual Row, SM Header (20 Pos.) Samtec SSW-110-22-F-D-VS-K 20 1 J3 (Top) 10-Pin, Dual Row, SM Header (20 Pos.) Samtec TSM-110-01-T-DV-P 21 1 J4 (Bottom) 5-Pin, Dual Row, SM Header (10 Pos.) Samtec SSW-105-22-F-D-VS-K 22 0 J5 Not Installed 23 1 J6 Connector, SMA Jack Straight PCB Not installed LED 565nm Green DIFF 0603 SMD CONN D-SUB RCPT 15-Position R/A PCB SLD CONN D-SUB RCPT R/A 15-Position Gold Amphenol 132134 Emerson 142-0701-201 Connector, Socket RT ANG 50-Pos .050 (Cut to 4 pins) Mill-Max 851-43-050-20-001000 2-Position Jumper _ 0.1" spacing Samtec TSW-102-07-T-S 3-Position Jumper _ 0.1" spacing Samtec TSW-103-07-T-S 2x2x.1, 2-Pin Dual Row Header Samtec TSW-102-07-T-D 24 1 J7(TOP) 25 4 JP1, JP4, JP5, JP16 26 0 JP3 Not installed 27 0 JP6, JP7, JP8, JP9, JP10, JP11, JP12, JP13, JP14, JP17, JP25 Not installed 28 12 JP2, JP15, JP18, JP19, JP20, JP21, JP22, JP23, JP24, JP34, JP35, JP36 29 7 JP26 - JP32 SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated BOM, Layout, and Schematics 55 Bill of Materials www.ti.com Table 13. Bill of Materials: ADS1298R (continued) Item Quantity Reference Designator 30 1 JP34 31 5 Ferrite bead 470 Ω 0805 Manufacturer Part Number Samtec TSW-104-07-T-D Taiyo Yuden BK2125HM471-T R1, R4, R5, R11, R12, R15, R16, R19, R20, R23, R24, R27, R28, R31, R32, R35, R36, R41, R42, R45, Not installed R46, R47 - R51, R54, R55, R58 - R66, R68, R69, R70, R89, R90, R91, R92 32 0 33 0 R2 9 R3, R71, R72, R73, R74, R88, R93, R94, R95 Resistor, 0.0Ω 1/10W 5% 0603 SMD Yageo RC0603JR-070RL 35 17 R6, R7, R10, R14, R18, R22, R26, R30, R34, R38, R40, R44, R67, R75, R100, R101, R102 Resistor, 10.0kΩ 1/10W 1% 0603 SMD Yageo RC0603FR-0710KL 36 10 R9, R13, R17, R21, R25, R29, R33, R37, R39, R43 Resistor, 22.1kΩ 1/10W 1% 0603 SMD Yageo RC0603FR-0722K1L 37 1 R52 Resistor, 47.5kΩ 1/10W 1% 0603 SMD Yageo RC0603FR-0747K5L 38 1 R53 Resistor, 43.2kΩ 1/10W 1% 0603 SMD Yageo RC0603FR-0743K2L 39 1 R56 Resistor, 49.9kΩ 1/10W 1% 0603 SMD Yageo RC0603FR-0749K9L 40 1 R57 Resistor, 46.4kΩ 1/10W 1% 0603 SMD Yageo RC0603FR-0746K4L 41 1 R76 Resistor, 1.00MΩ 1/8W 1% 0805 SMD Yageo RC0805FR-071ML 42 1 R77 Resistor, 330Ω 1/8W 1% 0805 SMD Yageo RC0805FR-07330RL 43 2 R78, R79 Resistor, 1.00kΩ 1/10W 1% 0603 SMD Yageo RC0603FR-071KL 44 5 R8, R80, R81, R82, R83 Resistor, 1.00MΩ 1/10W 1% 0603 SMD Yageo RC0603FR-071ML 45 4 R84, R85, R86, R87 Resistor, 10.0MΩ 1/10W 1% 0603 SMD Yageo RC0603FR-0710ML 46 2 R96, R97 Resistor, 40.2kΩ 1/10W 1% 0603 SMD Yageo RC0603FR-0740K2L 47 1 R98 Resistor, 330Ω 1/10W 1% 0603 SMD Yageo RC0603FR-07330RL 48 1 R99 Resistor, 47.0kΩ 1/10W 1% 0603 SMD Yageo RC0603FR-0747KL Keystone 5001 Keystone 5000 TI ADS1298RIZXG 34 56 L1 - L5 Description 4x2x.1, 4-Pin Dual Row Header Not installed 49 5 TP1, TP2, TP8, TP11, Test Point PC Mini .040"D Black TP12 50 8 TP3, TP4, TP5, TP6, TP7, TP9, TP10, TP13 51 1 U1 8-Channel, 24-Bit Analog-To-Digital Converter with Integrated ECG Front End 52 0 U2 Not installed 53 0 U3, U4, U5, U13 Not installed 54 1 U6 IC, Unreg Chrg Pump V Inv SOT23-5 TI TPS60403DBVR 55 1 U7 IC LDO Reg 250mA 3.0V SOT23-5 TI TPS73230DBVR 56 1 U8 IC LDO Reg Neg 200mA ADJ SOT23-5 TI TPS72301DBVT 57 1 U9 IC LDO Reg 250mA ADJ-V SOT23-5 TI TPS73201DBV 58 1 U10 IC EEPROM 256kBit 400kHZ 8TSSOP 59 1 U11 60 1 61 1 62 1 OSC1 Test Point PC Mini .040"D Red Microchip 24AA256-I/ST IC Switch SPST SOT23-5 TI TS5A3166DBVR U12 IC Comparator P-P Nanopwr 8-SOIC TI TLV3491AID U14 IC MCU 16-Bit 14TSSOP TI MSP430G2121IPW14R Fox FXO-HC735-2.048MHz BOM, Layout, and Schematics OSC 2.0480 MHz 3.3V HCMOS SMT SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Appendix B SBAU181B – March 2011 – Revised Janurary 2016 External Optional Hardware B.1 Optional External Hardware (Not Included) The input of the ADS1298R requires a DB15 connector. Figure 48 illustrates the most optimal cable connection to the ADS1298R. Figure 49 and Figure 50 show two alternate ways that cables can be constructed to interface with the ADS1298R. Figure 51 shows an alternate testing tool to the instrument used in the tests for this user guide (refer to Section 6.7.1). Figure 48. 15-Pin, Shielded Connector from Biometric Cables SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated External Optional Hardware 57 Optional External Hardware (Not Included) Figure 49. 15-Pin, Twisted Wire Cable to Banana Jacks www.ti.com Figure 50. 15-Pin, Twisted Wire Cable Figure 51. Cardiosim ECG Simulator Tool 58 External Optional Hardware SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated ADS1298R Power-Supply Recommendations www.ti.com B.2 ADS1298R Power-Supply Recommendations If you chose to power the MMB0 board through the wall adapter jack, it must comply with the following requirements: • Output voltage: 5.5 VDC to 15 VDC. • Maximum output current: ≥ 500 mA. • Output connector: barrel plug (positive center), 2.5-mm I.D. x 5.5-mm O.D. (9-mm insertion depth). • Complies with applicable regional safety standards. Figure 52 shows a +6V power-supply cable (not provided in the EVM kit) connected to a battery pack with four 1.5V batteries connected in series. Connecting to a wall-powered source makes the ADS1298R more susceptible to 50Hz/60Hz noise pickup; therefore, for best performance, it is recommended to power the ADS1298R with a battery source. This configuration minimizes the amount of noise pickup seen at the digitized output of the ADS1298R. Figure 52. Recommended Power Supply for ADS1298R SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated External Optional Hardware 59 Appendix C SBAU181B – March 2011 – Revised Janurary 2016 Software Installation C.1 Minimum Requirements Before installing the software, verify that your PC meets the minimum requirements outlined below. • Pentium III®/Celeron® processor, 866MHz or equivalent • Minimum 256MB of RAM (512MB or greater recommended) • USB 1.1-compatible input • Hard disk drive with at least 200MB free space • Microsoft® Windows® XP operating system with SP2 (Windows Vista and Windows 7 are NOT supported) • Mouse or other pointing device • 1280 x 960 minimum display resolution C.2 Installing the Software CAUTION Do not connect the ADS1298R before installing the software on a suitable PC. Failure to observe this caution may cause Microsoft Windows to not recognize the ADS1298R. The latest software is available from the ADS1298RECGFE-PDK product folder on the TI web site. Check the TI web site regularly for updated versions. To install the ADS1298R software: • Download the software from http://www.ti.com/tool/ads1298recgfe-pdk. • Click on the executable file ads129xecg-fe-y.y.y.exe, where y.y.y represents the version number of the software installer Then follow the prompts illustrated in Figure 53 through Figure 56. You must accept the license agreement (shown in Figure 54) before you can proceed with the installation. 60 Software Installation SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Revision History www.ti.com Figure 53. Initialization of ADS1298R Figure 54. License Agreement Figure 55. Installation Process Figure 56. Completion of ADS1298R Software Installation Revision History Changes from A Revision (September 2012) to B Revision .......................................................................................... Page • • • • • Deleted power adapter from Figure 1 .................................................................................................. 9 Deleted bullet in Section 1.5 ..."Universal AC to DC wall adapter"... ............................................................... 9 Changed "No additional power connections are required" in Section 2.2 to "Refer to Appendix B2 for details about the MMB0 power supply". ................................................................................................................... 12 Added new paragraph to Section B.2... "If you chose to power the MMB0 board through the wall adapter jack"... ....... 59 Deleted "provided in the EVM kit" from Section B.2. ................................................................................ 59 NOTE: Page numbers for previous revisions may differ from page numbers in the current version. SBAU181B – March 2011 – Revised Janurary 2016 Submit Documentation Feedback Copyright © 2011–2016, Texas Instruments Incorporated Revision History 61 1 2 3 4 AVDD AVDD R9 D 22.1k AVDD R11 NI R10 47pF C72 47pF AGND AGND JP11 ECG_V2 R12 NI R17 R18 22.1k 10k ECG_V2 NI AVSS AVDD AVSS AVDD C26 47pF AGND AGND R20 NI R27 NI D8 NI C27 NI C28 47pF AVSS R19 NI D5 NI C78 NI 10k C91 6 AVDD AVSS D1 NI 5 JP9 ECG_V4 R25 R26 22.1k 10k ECG_V4 NI AVSS C79 C93 47pF C32 47pF AGND AGND D NI R28 NI JP6 ECG_V6 ECG_V6 NI AVSS AVDD AVDD AVSS AVDD AVDD D2 NI R13 22.1k C81 10k C80 NI C73 47pF 47pF JP10 ECG_V3 R16 NI R21 R22 22.1k 10k ECG_V3 NI C AGND J1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 AVDD R23 NI D6 NI R14 AGND AVDD AVSS R15 NI C83 C82 47pF C29 47pF AGND AGND R31 NI D9 NI NI R24 NI AVSS JP8 ECG_V5 R29 R30 22.1k 10k C85 JP7 C84 47pF C75 47pF AGND AGND ECG_V5 NI NI R32 NI ECG_V1 ECG_V1 NI C AVSS ELEC_V2 ELEC_V3 ELEC_V4 ELEC_V5 ELEC_V6 ELEC_SHD AVSS AVSS ELEC_RA ELEC_LA ELEC_LL ELEC_V1 ELEC_RL AVDD AVDD AVDD R41 NI D3 NI ECG_SHD_DRV JP15 B R39 R40 22.1k 10k AGND AVDD AVSS JP14 C74 47pF AGND AGND R42 NI R35 NI D7 NI C87 NI C86 47pF AVSS ECG_RA NI ECG_RA AVSS R33 R34 22.1k 10k C89 NI B JP12 C88 47pF C30 47pF AGND AGND R36 NI ECG_LL ECG_LL NI AVSS ELEC_RA ELEC_LA AVDD AVDD R45 NI D4 NI A R43 R44 22.1k 10k AVDD AVSS D10 NI C25 NI JP13 C24 47pF C23 47pF AGND AGND AVSS R46 NI ECG_LA NI R38 R37 10k 22.1k C31 47pF C92 NI ECG_LA AGND ECG_RL ECG_RL ti AGND AVSS A 12500 TI Boulevard. Dallas, Texas 75243 Title: Engineer: Tom Hendrick Drawn By: Tom Hendrick FILE: 1 2 3 4 5 ADS1298R ECG FE DOCUMENTCONTROL # REV: 6522881 DATE: 19-Feb-2011 SIZE: 6 SHEET: B 1 OF: 5 1 2 3 4 5 6 AVDD AVDD C21 NI AVDD AVDD C6 C16 C7 C19 C8 C15 1uF 0.1uF NI NI NI NI AVDD 5 AGND U2 NI R4 1 4 ECG_SHD_DRV ECG_SHD_DRV D R1 NI JP1 R3 ECG_RL R64 NI R65 NI R66 NI AVSS C90 IN8N IN8P ECG_V6 U1 ADS1298R ECG_V1 1 3 ECG_LL ECG_RA ECG_LA ECG_LL ECG_RA ECG_LA R88 8 NI AGND 2 2200pF C97 AVDD R83 1M 5 R79 C98 4 C114 7 C96 1uF 6 5 V+ DVDD -IN2 OUT +IN3 N/C AGND V- 2200pF B5 B6 B7 B8 C1 C2 C3 C4 C5 C6 C7 C8 D1 D2 D3 D4 D5 D6 D7 D8 E1 E2 E3 E4 E5 E6 E7 E8 F1 F2 F3 F4 F5 F6 F7 F8 G1 G2 G3 G4 G5 G6 G7 G8 H2 H3 H4 H5 IN1P MSP R80 1M AVSS 3 4 AVSS SPI_DRDY GPIO4 SPI_OUT GPIO2 GPIO3 SPI_CLK SPI_CS SPI_START H6 H7 SPI_IN H8 CLKSEL SPI_DRDY GPIO4 SPI_OUT GPIO2 GPIO3 SPI_CLK SPI_CS DVDD SPI_START SPI_IN 1uF 0 3 R6 10K R7 10K CLK VREFP C10 10uF JP19 1 E/D JP18 RESETB GPIO1 C95 HC735-2.048MHZ EXT_CLK B 2 Output GND AGND EXT_CLK PWDNB DAISY_IN 0.1uF AVSS 0 JP5 C1 NI 1uF AVSS AVSS C33 NI C9 C2 22uF 1uF AGND J5 AVSS 0 C108 2200pF R96 40.2k C109 2200pF R97 40.2k C110 C111 R81 1M NI AVSS PACEOUT1 GPIO4 GPIO3 VCC_5v 2 4 6 8 10 1 3 5 7 9 PACEOUT2 PWDNB DAISY_IN AGND Title: Engineer: Tom Hendrick Drawn By: Tom Hendrick FILE: 4 5 A 12500 TI Boulevard. Dallas, Texas 75243 NI AGND 3 ti NI AVSS AVSS 2 VDD AGND TLV3491 1 OSC1 4 AGND DVDD NI R93 TP12 C PACEOUT2 PACEOUT1 1 2 CLKSEL C11 U13B R87 10M TP11 AGND IN1N 2 4 6 8 7 5 0.1uF AVDD R95 TP2 A8 B4 H1 TP3 NI C100 U12 A7 IN2N IN2P C107 6 ELEC_RA A 2 4 C106 R92 JP35 N/C 1 3 IN3N IN3P NI C103 NI R86 R91 10M NI 560pF A6 AGND NI AVSS N/C R94 C104 NI 1.0k 8 2 4 NI AVDD 2 AGND R89 R90 NI A5 B3 AVSS AVSS NI 560pF R77 330 AVSS A4 B2 4 1.0k A3 IN4N IN4P C105 R85 10M A2 B1 NI C102 R78 GND V+ 1 3 1 3 5 7 U13A 3 TP1 C77 1uF JP33 C99 R82 1M 1 2 4 IN5N IN5P JP32 1 U11 TS5A3166 1 3 0 AVDD C101 R84 10M AVDD 3 2 4 JP31 AVSS C113 C76 1uF PACEOUT2 VCAP4 VREFP PACEOUT1 RESP_MOD_P R58 NI AVSS AGND JP30 ECG_V2 ECG_V2 R76 1M 1 3 ECG_V3 ECG_V3 AGND JP29 ECG_V4 ECG_V4 AGND AVSS IN6N IN6P 2 4 ECG_V5 ECG_V5 B 1uF DVDD A1 JP28 ECG_V1 JP34 C5 0.1uF 0.1uF 1000pF IN7N IN7P 2 4 ECG_V6 0.1uF C12 0.1uF AVDD JP26 2 4 1 3 AVSS C18 1uF VCAP3 JP27 ELEC_LA C17 0.1uF AVSS RESP_MOD_N PWDNB VBG GPIO1 DAISY_IN VCAP2 RESETB R63 NI 1 3 C C14 1uF C3 1uF C13 JP16 DVDD C4 RLDINV RLDOUT RLDIN RLDREF WCT R62 NI 0 R8 1M 1.5nF AVDD R61 NI R2 NI AVDD AVSS C20 R60 NI AVDD AGND AVSS R59 NI D R5 NI C22 AVSS AGND NI NI 2 JP17 ECG_RL AVSS NI 3 ADS1298R ECG FE DOCUMENTCONTROL # REV: 6522881 DATE: 19-Feb-2011 SIZE: 6 SHEET: 2 OF: B 5 1 2 3 4 5 6 D D External Reference 2 3 C34 4 NI N/C VIN R50 C40 7 C35 TEMP GND NI 6 OUT NI NI AVDD 5 TRIM C43 AVSS C41 NI 7 1 U3 NI N/C AVDD N/C 8 Optional External Reference AGND 8 NI 1 2 3 C36 4 NI N/C VIN U4 NI JP25 N/C AVDD N/C C R49 NI NI OUT 6 C37 TEMP GND R47 3 NI JP3 NI VREFP NI NI C38 NI C39 NI C42 NI AGND NI TRIM R51 6 C R48 NI U5 4 7 2 8 AVSS 5 C44 AVSS AVSS NI AVSS NOT INSTALLED B B ti A A 12500 TI Boulevard. Dallas, Texas 75243 Title: Engineer: Tom Hendrick Drawn By: Tom Hendrick FILE: 1 2 3 4 5 ADS1298R ECG FE DOCUMENTCONTROL # REV: 6522881 DATE: 19-Feb-2011 SIZE: 6 SHEET: B 3 OF: 5 1 2 3 4 5 6 C48 D D 3 5 1uF 2 10uF AGND C46 C47 10uF 1uF IN OUT L2 1 GND 3.3uH CFLY- L1 AGND 4 C45 TP4 U6 CFLY+ VCC_5v C50 1uF 10uF TPS60403 VCC_-5v 3.3uH C49 C51 10uF AGND AGND AGND TP5 U7 1 IN OUT L3 5 C52 C 1uF 3 R54 NI EN C53 C54 2.2uF 10uF 3.3uH +3.0V JP2 C55 AVDD C 10uF AGND AGND 2 AGND GND NR/FB AGND 4 C57 TPS73230 R55 NI C56 0.1uF NI AGND AGND TP13 U9 1 IN OUT L5 5 C58 1uF 3 EN C60 2.2uF 10uF 3.3uH 10uF 49.9K 2 AGND GND +2.5V C61 R56 AGND B C59 NR/FB AGND AGND 4 C62 TPS73201 R57 B NI 46.4K AGND AGND TP6 U8 VCC_-5v 2 IN OUT L4 5 C63 2.2uF 3 EN C64 C65 2.2uF 10uF AGND C66 JP20 AVSS 10uF AGND 47.5K GND -2.5V R52 AGND 1 3.3uH NR/FB AGND 4 AGND ECG Power Supplies TPS72301 R53 C67 NI 43.2K ti AGND A AGND A 12500 TI Boulevard. Dallas, Texas 75243 Title: Engineer: Tom Hendrick Drawn By: Tom Hendrick FILE: 1 2 3 4 5 ADS1298R ECG FE DOCUMENTCONTROL # REV: 6522881 DATE: 19-Feb-2011 SIZE: 6 SHEET: B 4 OF: 5 1 2 3 4 5 6 D D DVDD DVDD For MMB0 - JP21 and 23 short 1-2, JP22 short 2-3 C71 100uF R67 DVDD 10K C70 R75 10k J2 1 3 5 7 9 11 13 15 17 19 C 2 4 6 8 10 12 14 16 18 20 JP23 J3 1 3 5 7 9 11 13 15 17 19 SPI_CLK SPI_CS JP21 SPI_IN SPI_OUT SPI_DRDY EXT_CLK DVDD R102 10k CLKSEL TP7 2 4 6 8 10 12 14 16 18 20 VCC_5v JP4 GPIO1 RESETB GPIO2 VCC_3.3V 1 3 5 7 9 TP9 VCC_1.8V JP24 VCC_3.3V TP10 TP8 C68 C69 100uF 0.1uF C VCC_3.3V R100 R101 Dummy Connector SPI_START 10k C94 10k 0.1uF VCC_3.3V U10 AGND JP22 8 6 5 7 SCL SDA VCC SCL SDA WP 1 2 3 4 A0 A1 A2 GND 24AA256-I/ST JP36 R98 J6 330 1 R74 0 AGND 0.1uF U14 1 2 3 4 5 6 7 NI NI NI VCC_3.3V C112 B R68 R69 R70 R71 R72 R73 0 0 0 VCC_3.3V MSP 2 4 6 8 10 AGND 0.1uF J4 VCC GND P1.0 P2.6 1.1 P2.7 1.2 TCK 1.3 TDIO 1.4 1.7 1.5 1.6 R99 14 13 12 11 10 9 SDA 8 SCL 47k B J7 4 3 2 1 4X1X.1 MSP430G2121 5 4 3 2 D11 ti A 12500 TI Boulevard. Dallas, Texas 75243 Title: Engineer: Tom Hendrick Drawn By: Tom Hendrick FILE: 1 2 3 4 5 ADS1298R ECG FE REV: 6522881 B DOCUMENTCONTROL # DATE: 19-Feb-2011 SIZE: 6 SHEET: 5 OF: 5 A STANDARD TERMS AND CONDITIONS FOR EVALUATION MODULES 1. Delivery: TI delivers TI evaluation boards, kits, or modules, including any accompanying demonstration software, components, or documentation (collectively, an “EVM” or “EVMs”) to the User (“User”) in accordance with the terms and conditions set forth herein. Acceptance of the EVM is expressly subject to the following terms and conditions. 1.1 EVMs are intended solely for product or software developers for use in a research and development setting to facilitate feasibility evaluation, experimentation, or scientific analysis of TI semiconductors products. EVMs have no direct function and are not finished products. EVMs shall not be directly or indirectly assembled as a part or subassembly in any finished product. For clarification, any software or software tools provided with the EVM (“Software”) shall not be subject to the terms and conditions set forth herein but rather shall be subject to the applicable terms and conditions that accompany such Software 1.2 EVMs are not intended for consumer or household use. EVMs may not be sold, sublicensed, leased, rented, loaned, assigned, or otherwise distributed for commercial purposes by Users, in whole or in part, or used in any finished product or production system. 2 Limited Warranty and Related Remedies/Disclaimers: 2.1 These terms and conditions do not apply to Software. The warranty, if any, for Software is covered in the applicable Software License Agreement. 2.2 TI warrants that the TI EVM will conform to TI's published specifications for ninety (90) days after the date TI delivers such EVM to User. Notwithstanding the foregoing, TI shall not be liable for any defects that are caused by neglect, misuse or mistreatment by an entity other than TI, including improper installation or testing, or for any EVMs that have been altered or modified in any way by an entity other than TI. Moreover, TI shall not be liable for any defects that result from User's design, specifications or instructions for such EVMs. Testing and other quality control techniques are used to the extent TI deems necessary or as mandated by government requirements. TI does not test all parameters of each EVM. 2.3 If any EVM fails to conform to the warranty set forth above, TI's sole liability shall be at its option to repair or replace such EVM, or credit User's account for such EVM. TI's liability under this warranty shall be limited to EVMs that are returned during the warranty period to the address designated by TI and that are determined by TI not to conform to such warranty. If TI elects to repair or replace such EVM, TI shall have a reasonable time to repair such EVM or provide replacements. Repaired EVMs shall be warranted for the remainder of the original warranty period. Replaced EVMs shall be warranted for a new full ninety (90) day warranty period. 3 Regulatory Notices: 3.1 United States 3.1.1 Notice applicable to EVMs not FCC-Approved: This kit is designed to allow product developers to evaluate electronic components, circuitry, or software associated with the kit to determine whether to incorporate such items in a finished product and software developers to write software applications for use with the end product. This kit is not a finished product and when assembled may not be resold or otherwise marketed unless all required FCC equipment authorizations are first obtained. Operation is subject to the condition that this product not cause harmful interference to licensed radio stations and that this product accept harmful interference. Unless the assembled kit is designed to operate under part 15, part 18 or part 95 of this chapter, the operator of the kit must operate under the authority of an FCC license holder or must secure an experimental authorization under part 5 of this chapter. 3.1.2 For EVMs annotated as FCC – FEDERAL COMMUNICATIONS COMMISSION Part 15 Compliant: CAUTION This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. FCC Interference Statement for Class A EVM devices NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. SPACER SPACER SPACER SPACER SPACER SPACER SPACER SPACER FCC Interference Statement for Class B EVM devices NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: • • • • Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. 3.2 Canada 3.2.1 For EVMs issued with an Industry Canada Certificate of Conformance to RSS-210 Concerning EVMs Including Radio Transmitters: This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Concernant les EVMs avec appareils radio: Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes: (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le fonctionnement. Concerning EVMs Including Detachable Antennas: Under Industry Canada regulations, this radio transmitter may only operate using an antenna of a type and maximum (or lesser) gain approved for the transmitter by Industry Canada. To reduce potential radio interference to other users, the antenna type and its gain should be so chosen that the equivalent isotropically radiated power (e.i.r.p.) is not more than that necessary for successful communication. This radio transmitter has been approved by Industry Canada to operate with the antenna types listed in the user guide with the maximum permissible gain and required antenna impedance for each antenna type indicated. Antenna types not included in this list, having a gain greater than the maximum gain indicated for that type, are strictly prohibited for use with this device. Concernant les EVMs avec antennes détachables Conformément à la réglementation d'Industrie Canada, le présent émetteur radio peut fonctionner avec une antenne d'un type et d'un gain maximal (ou inférieur) approuvé pour l'émetteur par Industrie Canada. Dans le but de réduire les risques de brouillage radioélectrique à l'intention des autres utilisateurs, il faut choisir le type d'antenne et son gain de sorte que la puissance isotrope rayonnée équivalente (p.i.r.e.) ne dépasse pas l'intensité nécessaire à l'établissement d'une communication satisfaisante. Le présent émetteur radio a été approuvé par Industrie Canada pour fonctionner avec les types d'antenne énumérés dans le manuel d’usage et ayant un gain admissible maximal et l'impédance requise pour chaque type d'antenne. Les types d'antenne non inclus dans cette liste, ou dont le gain est supérieur au gain maximal indiqué, sont strictement interdits pour l'exploitation de l'émetteur 3.3 Japan 3.3.1 Notice for EVMs delivered in Japan: Please see http://www.tij.co.jp/lsds/ti_ja/general/eStore/notice_01.page 日本国内に 輸入される評価用キット、ボードについては、次のところをご覧ください。 http://www.tij.co.jp/lsds/ti_ja/general/eStore/notice_01.page 3.3.2 Notice for Users of EVMs Considered “Radio Frequency Products” in Japan: EVMs entering Japan may not be certified by TI as conforming to Technical Regulations of Radio Law of Japan. If User uses EVMs in Japan, not certified to Technical Regulations of Radio Law of Japan, User is required by Radio Law of Japan to follow the instructions below with respect to EVMs: 1. 2. 3. Use EVMs in a shielded room or any other test facility as defined in the notification #173 issued by Ministry of Internal Affairs and Communications on March 28, 2006, based on Sub-section 1.1 of Article 6 of the Ministry’s Rule for Enforcement of Radio Law of Japan, Use EVMs only after User obtains the license of Test Radio Station as provided in Radio Law of Japan with respect to EVMs, or Use of EVMs only after User obtains the Technical Regulations Conformity Certification as provided in Radio Law of Japan with respect to EVMs. Also, do not transfer EVMs, unless User gives the same notice above to the transferee. Please note that if User does not follow the instructions above, User will be subject to penalties of Radio Law of Japan. SPACER SPACER SPACER SPACER SPACER 【無線電波を送信する製品の開発キットをお使いになる際の注意事項】 開発キットの中には技術基準適合証明を受けて いないものがあります。 技術適合証明を受けていないもののご使用に際しては、電波法遵守のため、以下のいずれかの 措置を取っていただく必要がありますのでご注意ください。 1. 2. 3. 電波法施行規則第6条第1項第1号に基づく平成18年3月28日総務省告示第173号で定められた電波暗室等の試験設備でご使用 いただく。 実験局の免許を取得後ご使用いただく。 技術基準適合証明を取得後ご使用いただく。 なお、本製品は、上記の「ご使用にあたっての注意」を譲渡先、移転先に通知しない限り、譲渡、移転できないものとします。 上記を遵守頂けない場合は、電波法の罰則が適用される可能性があることをご留意ください。 日本テキサス・イ ンスツルメンツ株式会社 東京都新宿区西新宿6丁目24番1号 西新宿三井ビル 3.3.3 Notice for EVMs for Power Line Communication: Please see http://www.tij.co.jp/lsds/ti_ja/general/eStore/notice_02.page 電力線搬送波通信についての開発キットをお使いになる際の注意事項については、次のところをご覧くださ い。http://www.tij.co.jp/lsds/ti_ja/general/eStore/notice_02.page SPACER 4 EVM Use Restrictions and Warnings: 4.1 EVMS ARE NOT FOR USE IN FUNCTIONAL SAFETY AND/OR SAFETY CRITICAL EVALUATIONS, INCLUDING BUT NOT LIMITED TO EVALUATIONS OF LIFE SUPPORT APPLICATIONS. 4.2 User must read and apply the user guide and other available documentation provided by TI regarding the EVM prior to handling or using the EVM, including without limitation any warning or restriction notices. The notices contain important safety information related to, for example, temperatures and voltages. 4.3 Safety-Related Warnings and Restrictions: 4.3.1 User shall operate the EVM within TI’s recommended specifications and environmental considerations stated in the user guide, other available documentation provided by TI, and any other applicable requirements and employ reasonable and customary safeguards. Exceeding the specified performance ratings and specifications (including but not limited to input and output voltage, current, power, and environmental ranges) for the EVM may cause personal injury or death, or property damage. If there are questions concerning performance ratings and specifications, User should contact a TI field representative prior to connecting interface electronics including input power and intended loads. Any loads applied outside of the specified output range may also result in unintended and/or inaccurate operation and/or possible permanent damage to the EVM and/or interface electronics. Please consult the EVM user guide prior to connecting any load to the EVM output. If there is uncertainty as to the load specification, please contact a TI field representative. During normal operation, even with the inputs and outputs kept within the specified allowable ranges, some circuit components may have elevated case temperatures. These components include but are not limited to linear regulators, switching transistors, pass transistors, current sense resistors, and heat sinks, which can be identified using the information in the associated documentation. When working with the EVM, please be aware that the EVM may become very warm. 4.3.2 EVMs are intended solely for use by technically qualified, professional electronics experts who are familiar with the dangers and application risks associated with handling electrical mechanical components, systems, and subsystems. User assumes all responsibility and liability for proper and safe handling and use of the EVM by User or its employees, affiliates, contractors or designees. User assumes all responsibility and liability to ensure that any interfaces (electronic and/or mechanical) between the EVM and any human body are designed with suitable isolation and means to safely limit accessible leakage currents to minimize the risk of electrical shock hazard. User assumes all responsibility and liability for any improper or unsafe handling or use of the EVM by User or its employees, affiliates, contractors or designees. 4.4 User assumes all responsibility and liability to determine whether the EVM is subject to any applicable international, federal, state, or local laws and regulations related to User’s handling and use of the EVM and, if applicable, User assumes all responsibility and liability for compliance in all respects with such laws and regulations. User assumes all responsibility and liability for proper disposal and recycling of the EVM consistent with all applicable international, federal, state, and local requirements. 5. Accuracy of Information: To the extent TI provides information on the availability and function of EVMs, TI attempts to be as accurate as possible. However, TI does not warrant the accuracy of EVM descriptions, EVM availability or other information on its websites as accurate, complete, reliable, current, or error-free. SPACER SPACER SPACER SPACER SPACER SPACER SPACER 6. Disclaimers: 6.1 EXCEPT AS SET FORTH ABOVE, EVMS AND ANY WRITTEN DESIGN MATERIALS PROVIDED WITH THE EVM (AND THE DESIGN OF THE EVM ITSELF) ARE PROVIDED "AS IS" AND "WITH ALL FAULTS." TI DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, REGARDING SUCH ITEMS, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OR NON-INFRINGEMENT OF ANY THIRD PARTY PATENTS, COPYRIGHTS, TRADE SECRETS OR OTHER INTELLECTUAL PROPERTY RIGHTS. 6.2 EXCEPT FOR THE LIMITED RIGHT TO USE THE EVM SET FORTH HEREIN, NOTHING IN THESE TERMS AND CONDITIONS SHALL BE CONSTRUED AS GRANTING OR CONFERRING ANY RIGHTS BY LICENSE, PATENT, OR ANY OTHER INDUSTRIAL OR INTELLECTUAL PROPERTY RIGHT OF TI, ITS SUPPLIERS/LICENSORS OR ANY OTHER THIRD PARTY, TO USE THE EVM IN ANY FINISHED END-USER OR READY-TO-USE FINAL PRODUCT, OR FOR ANY INVENTION, DISCOVERY OR IMPROVEMENT MADE, CONCEIVED OR ACQUIRED PRIOR TO OR AFTER DELIVERY OF THE EVM. 7. USER'S INDEMNITY OBLIGATIONS AND REPRESENTATIONS. USER WILL DEFEND, INDEMNIFY AND HOLD TI, ITS LICENSORS AND THEIR REPRESENTATIVES HARMLESS FROM AND AGAINST ANY AND ALL CLAIMS, DAMAGES, LOSSES, EXPENSES, COSTS AND LIABILITIES (COLLECTIVELY, "CLAIMS") ARISING OUT OF OR IN CONNECTION WITH ANY HANDLING OR USE OF THE EVM THAT IS NOT IN ACCORDANCE WITH THESE TERMS AND CONDITIONS. THIS OBLIGATION SHALL APPLY WHETHER CLAIMS ARISE UNDER STATUTE, REGULATION, OR THE LAW OF TORT, CONTRACT OR ANY OTHER LEGAL THEORY, AND EVEN IF THE EVM FAILS TO PERFORM AS DESCRIBED OR EXPECTED. 8. Limitations on Damages and Liability: 8.1 General Limitations. IN NO EVENT SHALL TI BE LIABLE FOR ANY SPECIAL, COLLATERAL, INDIRECT, PUNITIVE, INCIDENTAL, CONSEQUENTIAL, OR EXEMPLARY DAMAGES IN CONNECTION WITH OR ARISING OUT OF THESE TERMS ANDCONDITIONS OR THE USE OF THE EVMS PROVIDED HEREUNDER, REGARDLESS OF WHETHER TI HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES. EXCLUDED DAMAGES INCLUDE, BUT ARE NOT LIMITED TO, COST OF REMOVAL OR REINSTALLATION, ANCILLARY COSTS TO THE PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES, RETESTING, OUTSIDE COMPUTER TIME, LABOR COSTS, LOSS OF GOODWILL, LOSS OF PROFITS, LOSS OF SAVINGS, LOSS OF USE, LOSS OF DATA, OR BUSINESS INTERRUPTION. NO CLAIM, SUIT OR ACTION SHALL BE BROUGHT AGAINST TI MORE THAN ONE YEAR AFTER THE RELATED CAUSE OF ACTION HAS OCCURRED. 8.2 Specific Limitations. IN NO EVENT SHALL TI'S AGGREGATE LIABILITY FROM ANY WARRANTY OR OTHER OBLIGATION ARISING OUT OF OR IN CONNECTION WITH THESE TERMS AND CONDITIONS, OR ANY USE OF ANY TI EVM PROVIDED HEREUNDER, EXCEED THE TOTAL AMOUNT PAID TO TI FOR THE PARTICULAR UNITS SOLD UNDER THESE TERMS AND CONDITIONS WITH RESPECT TO WHICH LOSSES OR DAMAGES ARE CLAIMED. THE EXISTENCE OF MORE THAN ONE CLAIM AGAINST THE PARTICULAR UNITS SOLD TO USER UNDER THESE TERMS AND CONDITIONS SHALL NOT ENLARGE OR EXTEND THIS LIMIT. 9. Return Policy. Except as otherwise provided, TI does not offer any refunds, returns, or exchanges. Furthermore, no return of EVM(s) will be accepted if the package has been opened and no return of the EVM(s) will be accepted if they are damaged or otherwise not in a resalable condition. If User feels it has been incorrectly charged for the EVM(s) it ordered or that delivery violates the applicable order, User should contact TI. All refunds will be made in full within thirty (30) working days from the return of the components(s), excluding any postage or packaging costs. 10. Governing Law: These terms and conditions shall be governed by and interpreted in accordance with the laws of the State of Texas, without reference to conflict-of-laws principles. User agrees that non-exclusive jurisdiction for any dispute arising out of or relating to these terms and conditions lies within courts located in the State of Texas and consents to venue in Dallas County, Texas. Notwithstanding the foregoing, any judgment may be enforced in any United States or foreign court, and TI may seek injunctive relief in any United States or foreign court. Mailing Address: Texas Instruments, Post Office Box 655303, Dallas, Texas 75265 Copyright © 2015, Texas Instruments Incorporated spacer IMPORTANT NOTICE Texas Instruments Incorporated and its subsidiaries (TI) reserve the right to make corrections, enhancements, improvements and other changes to its semiconductor products and services per JESD46, latest issue, and to discontinue any product or service per JESD48, latest issue. Buyers should obtain the latest relevant information before placing orders and should verify that such information is current and complete. All semiconductor products (also referred to herein as “components”) are sold subject to TI’s terms and conditions of sale supplied at the time of order acknowledgment. TI warrants performance of its components to the specifications applicable at the time of sale, in accordance with the warranty in TI’s terms and conditions of sale of semiconductor products. Testing and other quality control techniques are used to the extent TI deems necessary to support this warranty. Except where mandated by applicable law, testing of all parameters of each component is not necessarily performed. TI assumes no liability for applications assistance or the design of Buyers’ products. Buyers are responsible for their products and applications using TI components. To minimize the risks associated with Buyers’ products and applications, Buyers should provide adequate design and operating safeguards. TI does not warrant or represent that any license, either express or implied, is granted under any patent right, copyright, mask work right, or other intellectual property right relating to any combination, machine, or process in which TI components or services are used. Information published by TI regarding third-party products or services does not constitute a license to use such products or services or a warranty or endorsement thereof. Use of such information may require a license from a third party under the patents or other intellectual property of the third party, or a license from TI under the patents or other intellectual property of TI. Reproduction of significant portions of TI information in TI data books or data sheets is permissible only if reproduction is without alteration and is accompanied by all associated warranties, conditions, limitations, and notices. TI is not responsible or liable for such altered documentation. Information of third parties may be subject to additional restrictions. Resale of TI components or services with statements different from or beyond the parameters stated by TI for that component or service voids all express and any implied warranties for the associated TI component or service and is an unfair and deceptive business practice. TI is not responsible or liable for any such statements. Buyer acknowledges and agrees that it is solely responsible for compliance with all legal, regulatory and safety-related requirements concerning its products, and any use of TI components in its applications, notwithstanding any applications-related information or support that may be provided by TI. Buyer represents and agrees that it has all the necessary expertise to create and implement safeguards which anticipate dangerous consequences of failures, monitor failures and their consequences, lessen the likelihood of failures that might cause harm and take appropriate remedial actions. Buyer will fully indemnify TI and its representatives against any damages arising out of the use of any TI components in safety-critical applications. In some cases, TI components may be promoted specifically to facilitate safety-related applications. With such components, TI’s goal is to help enable customers to design and create their own end-product solutions that meet applicable functional safety standards and requirements. Nonetheless, such components are subject to these terms. No TI components are authorized for use in FDA Class III (or similar life-critical medical equipment) unless authorized officers of the parties have executed a special agreement specifically governing such use. Only those TI components which TI has specifically designated as military grade or “enhanced plastic” are designed and intended for use in military/aerospace applications or environments. Buyer acknowledges and agrees that any military or aerospace use of TI components which have not been so designated is solely at the Buyer's risk, and that Buyer is solely responsible for compliance with all legal and regulatory requirements in connection with such use. TI has specifically designated certain components as meeting ISO/TS16949 requirements, mainly for automotive use. In any case of use of non-designated products, TI will not be responsible for any failure to meet ISO/TS16949. Products Applications Audio www.ti.com/audio Automotive and Transportation www.ti.com/automotive Amplifiers amplifier.ti.com Communications and Telecom www.ti.com/communications Data Converters dataconverter.ti.com Computers and Peripherals www.ti.com/computers DLP® Products www.dlp.com Consumer Electronics www.ti.com/consumer-apps DSP dsp.ti.com Energy and Lighting www.ti.com/energy Clocks and Timers www.ti.com/clocks Industrial www.ti.com/industrial Interface interface.ti.com Medical www.ti.com/medical Logic logic.ti.com Security www.ti.com/security Power Mgmt power.ti.com Space, Avionics and Defense www.ti.com/space-avionics-defense Microcontrollers microcontroller.ti.com Video and Imaging www.ti.com/video RFID www.ti-rfid.com OMAP Applications Processors www.ti.com/omap TI E2E Community e2e.ti.com Wireless Connectivity www.ti.com/wirelessconnectivity Mailing Address: Texas Instruments, Post Office Box 655303, Dallas, Texas 75265 Copyright © 2016, Texas Instruments Incorporated
ADS1298RECGFE-PDK 价格&库存

很抱歉,暂时无法提供与“ADS1298RECGFE-PDK”相匹配的价格&库存,您可以联系我们找货

免费人工找货