SERIES: PSK-150 │ DESCRIPTION: INTERNAL AC-DC POWER SUPPLY
FEATURES
•
•
•
•
•
•
EN 60601 3.1 certified
4,000 Vac isolation
2 MOPP
adjustable output voltage
power factor > 90%
Class I or Class II
MODEL
output
voltage
output
current
output
power
ripple
and noise1
efficiency2
(Vdc)
max
(A)
max
(W)
max
(mVp-p)
typ
(%)
PSK-150-12
12
12.50
150
120
93.0
PSK-150-24
24
6.25
150
240
93.5
PSK-150-48
48
3.12
150
480
93.5
Note:
1. Ripple and noise are measured at 20 MHz BW with 47 uF aluminum electrolytic capacitor and 0.1 uF ceramic capacitor on the output.
2. At 230 Vac after 30 minutes of operation.
PART NUMBER KEY
PSK-150 - XX
Output Voltage
Base Number
cui.com
Additional Resources:
Product Page
|
3D Model
|
PCB Footprint
CUI Inc │ SERIES: PSK-150 │ DESCRIPTION: AC-DC POWER SUPPLY
date 07/18/2022 │ page 2 of 6
INPUT
parameter
conditions/description
voltage
(see derating curve)
min
typ
90
frequency
47
max
units
264
Vac
63
Hz
current
at 115 Vac
at 230 Vac
2.5
1.25
A
A
inrush current
at 115 Vac
at 230 Vac
45
90
A
A
leakage current
at 264 Vac
0.1
mA
power factor
at 230 Vac, full load
0.5
W
max
units
6,000
2,000
330
μF
μF
μF
0.9
no load power consumption
OUTPUT
parameter
conditions/description
capacitive load
12 Vdc output
24 Vdc output
48 Vdc output
initial set point accuracy
at full load
min
line regulation
load regulation
0%~100% load
hold-up time
at 115 Vac, 90% output
temperature coefficient
adjustability
remote ON/OFF
typ
±2
%
±1
%
±1
%
10
ms
±0.05
%/°C
±5
%
ON: tie pin 4 (REM) to pin 6 (-Vo) or leave floating
OFF: apply +5 Vdc to pin 4 (REM) with respect to pin 6 (-Vo)
PROTECTIONS
parameter
conditions/description
over voltage protection
auto recovery
over current protection
auto recovery, hiccup
short circuit protection
level 1 (nominal): continuous, auto recovery
level 2 (instantaneous high current): latch
over temperature protection
auto recovery
min
typ
max
units
min
typ
max
units
SAFETY & COMPLIANCE
parameter
isolation voltage
conditions/description
input to output
4,000
5,656
Vac
Vdc
input to PE ground
2,000
2,828
Vac
Vdc
output to PE ground
1,500
2,121
Vac
Vdc
safety approvals
certified to 60601: IEC/EN/UL
safety class
Class I (with PE) and Class II (without PE)
conducted emission
EN 55011 Class B
radiated emissions
EN 55011 Class I Class B / Class II Class A
ESD
IEC/EN61000-4-2
radiated immunity
IEC/EN61000-4-3
EFT/burst
IEC/EN61000-4-4
surge
IEC/EN61000-4-5
cui.com
Additional Resources:
Product Page
|
3D Model
|
PCB Footprint
CUI Inc │ SERIES: PSK-150 │ DESCRIPTION: AC-DC POWER SUPPLY
date 07/18/2022 │ page 3 of 6
SAFETY & COMPLIANCE (CONTINUED)
conducted immunity
IEC/EN61000-4-6
voltage dips and interruption
IEC/EN61000-4-11
MTBF
per MIL-HDBK-217F at 25 °C
RoHS
yes
250,000
hours
ENVIRONMENTAL
parameter
conditions/description
min
operating temperature
(see derating curve)
storage temperature
storage humidity
DERATING CURVES
cui.com
typ
max
units
-30
70
°C
-30
85
°C
0
95
%
Additional Resources:
Product Page
|
3D Model
|
PCB Footprint
CUI Inc │ SERIES: PSK-150 │ DESCRIPTION: AC-DC POWER SUPPLY
date 07/18/2022 │ page 4 of 6
MECHANICAL
parameter
conditions/description
dimensions
109.0 x 58.5 x 35.0
min
365
cooling
natural convection
MECHANICAL DRAWING
units: mm
general tolerance: ±0.5
PIN CONNECTIONS
PIN
Function
1
AC (N)
2
AC (L)
3
4
ON/OFF
5
+Vo
6
-Vo
7
Trim
max
units
mm
weight
Note:
typ
If the Remote ON/OFF function (pin 4) is not
required, tie directly to -Vo (pin 6) or leave
floating/open.
cui.com
g
Additional Resources:
Product Page
|
3D Model
|
PCB Footprint
CUI Inc │ SERIES: PSK-150 │ DESCRIPTION: AC-DC POWER SUPPLY
date 07/18/2022 │ page 5 of 6
BLOCK DIAGRAM
Figure 1
APPLICATION NOTES
Table 1
Part no.
TrimUp
-Vo
TrimDown
+Vo
PSK-150-12
5% / 0%
34KΩ ~ 10MΩ
0% / -5%
10MΩ ~ 106KΩ
PSK-150-24
5% / 0%
37.4KΩ ~ 10MΩ
0% / -5%
10MΩ ~ 270KΩ
PSK-150-48
5% / 0%
38KΩ ~ 10MΩ
0% / -5%
10MΩ ~ 640KΩ
cui.com
Additional Resources:
Product Page
|
3D Model
|
PCB Footprint
CUI Inc │ SERIES: PSK-150 │ DESCRIPTION: AC-DC POWER SUPPLY
date 07/18/2022 │ page 6 of 6
REVISION HISTORY
rev.
description
date
1.0
initial release
12/17/2021
1.01
adjustability updated, medical icon added
04/07/2022
1.02
mechanical drawing updated
07/18/2022
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
techsupport@cui.com
CUI offers a two (2) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical component is any component of a
life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to affect its safety or effectiveness.
很抱歉,暂时无法提供与“PSK-150-24”相匹配的价格&库存,您可以联系我们找货
免费人工找货