Additional Resources:
Product Page
date
08/25/2020
page
1 of 4
SERIES: SWM6-E-USB │ DESCRIPTION: AC-DC POWER SUPPLY
FEATURES
•
•
•
•
•
•
•
•
up to 6 W power
EN 60601-1 4th edition
compact vertical or horizontal mount
universal input (90~264 Vac)
single regulated output from 5~12 Vdc
over voltage and short circuit protections
DOE Level VI / CoC Tier 2 efficiency
2 MOPP compliant
MODEL
output
voltage
output
current
output
power
ripple
and noise1
(Vdc)
max
(A)
max
(W)
max
(mVp-p)
SWM6-5-E-I38
5
1.2
6
100
VI
SWM6-6-E-I38
6
1
6
120
VI
SWM6-9-E-I38
9
0.66
6
180
VI
12
0.5
6
240
VI
SWM6-12-E-I38
Notes:
efficiency
level2
1. At full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, output terminated with a low ESR 47 μF electrolytic and 0.1 μF multilayer capacitor.
2. CoC Tier 2 compliant.
PART NUMBER KEY
SWM6 - XX - E X - I38 - CXX
Base Number
Reserved for Custom
Configurations
Output Voltage
AC Input Type:
Europe
Orientation:
H = horizontal
V = vertical
cui.com
USB Type A Inlet
Additional Resources:
Product Page
CUI Inc │ SERIES: SWM6-E-USB │ DESCRIPTION: AC-DC POWER SUPPLY
date 08/25/2020│ page 2 of 4
INPUT
parameter
conditions/description
min
voltage
typ
90
frequency
47
max
units
264
Vac
63
Hz
current
at 115 Vac, full load
at 230 Vac, full load
0.6
0.3
A
A
inrush current
at 115 Vac, full load, cold start
at 230 Vac, full load, cold start
30
60
A
A
leakage current
at 240 Vac, 50 Hz
0.1
mA
no load power consumption
at 115/230 Vac, 60/50 Hz
0.075
W
max
units
OUTPUT
parameter
conditions/description
min
line regulation
typ
±1
load regulation
%
±5
start-up time
at 100 Vac, to 90% of rated output voltage
rise time
at nominal input voltage, full load, from 10% to
90% of output voltage
hold-up time
at nominal input voltage, full load
8.3
parameter
conditions/description
min
over voltage protection
output shutdown, auto recovery
over current protection
output shutdown, auto recovery
short circuit protection
output shut down, auto recovery
%
3
s
50
ms
ms
PROTECTIONS
typ
max
units
200
%
max
units
SAFETY & COMPLIANCE
parameter
conditions/description
isolation voltage
input to output
min
isolation resistance
input to output at 500 Vdc for 3 seconds
safety approvals
medical TUV EN 60601-1
EMI/EMC
CE
MTBF
as per MIL-HDBK-217F
RoHS
2011/65/EU
typ
4,000
Vac
50
MΩ
100,000
hours
ENVIRONMENTAL
parameter
conditions/description
min
operating temperature
storage temperature
typ
max
units
0
40
°C
-20
85
°C
operating humidity
non-condensing
10
90
%
storage humidity
non-condensing
5
90
%
altitude
5,000
cui.com
m
Additional Resources:
Product Page
CUI Inc │ SERIES: SWM6-E-USB │ DESCRIPTION: AC-DC POWER SUPPLY
date 08/25/2020│ page 3 of 4
MECHANICAL
parameter
conditions/description
dimensions
horizontal models: 55 x 27 x 55
vertical models: 53 x 41 x 46
min
inlet plug
Europe
weight
horizontal models
vertical models
color
black
typ
units
mm
mm
60
65
MECHANICAL DRAWINGS
units: mm
tolerance: ±1 mm
Vertical Models
46
Horizontal Models
max
cui.com
g
g
Additional Resources:
Product Page
CUI Inc │ SERIES: SWM6-E-USB │ DESCRIPTION: AC-DC POWER SUPPLY
date 08/25/2020│ page 4 of 4
REVISION HISTORY
rev.
description
date
1.0
initial release
06/27/2017
1.01
company logo updated
08/25/2020
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
techsupport@cui.com
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
CUI offers a one (1) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical component is any component of a
life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to affect its safety or effectiveness.
很抱歉,暂时无法提供与“SWM6-12-EV-I38”相匹配的价格&库存,您可以联系我们找货
免费人工找货