Additional Resources:
Product Page
|
3D Model
date
03/01/2023
page
1 of 10
SERIES: VMS-550C │ DESCRIPTION: AC-DC POWER SUPPLY
FEATURES
•
•
•
•
•
•
•
•
•
90~264 Vac and 127~370 Vdc input voltage range
3” x 5” open-frame footprint
320 W under natural convection, 550 W with airflow
-40 ~ 70 °C temperature range
active PFC
5 Vsby, 12 Vfan, power good, remote sense
suitable for BF applications
certified to 60601 safety standard
designed to meet 60335 & 61558 safety requirements
MODEL
output
voltage
output
current
output
power2
ripple
and noise3
efficiency4
(Vdc)
range¹
(Vdc)
max
(A)
max
(W)
max
(mVp-p)
typ
(%)
VMS-550C-12
12
11.4~12.6
41.60
499.2
200
91.0
VMS-550C-15
15
14.25~15.75
33.30
499.5
200
92.0
VMS-550C-24
24
22.8~25.2
22.90
549.6
200
93.0
VMS-550C-27
27
25.65~28.35
20.40
550.8
200
93.5
VMS-550C-36
36
34.2~37.8
15.30
550.8
200
94.0
VMS-550C-48
48
45.6~50.4
11.46
550.0
200
94.0
Notes:
1.
2.
3.
4.
When adjusting the output voltage care should be taken never to exceed the stated output power or output current of the unit.
With 25 CFM of forced air cooling.
At full load, nominal input, 20 MHz bandwidth oscilloscope, tip & barrel method, output terminated with 47 µF electrolytic and 0.1 µF ceramic capacitors.
At 230 Vac.
PART NUMBER KEY
VMS-550C - XX - XXX
Base Number
Output Voltage
cui.com
Chassis:
“blank” = open-frame
CNF = covered
CFS = enclosed with rear fan
Additional Resources:
Product Page
|
3D Model
CUI Inc │ SERIES: VMS-550C │ DESCRIPTION: AC-DC POWER SUPPLY
date 03/01/2023 │ page 2 of 10
INPUT
parameter
conditions/description
min
max
units
voltage
ac input
dc input
90
127
264
370
Vac
Vdc
47
63
Hz
6.5
3.0
A
A
frequency
current
at 90/115 Vac
at 230 Vac
inrush current
at 115 Vac, cold start
at 230 Vac, cold start
leakage current
at 264 Vac
contact leakage current
earth leakage current
power factor correction
at 115 Vac, full load
at 230 Vac, full load
no load power consumption
at 230 Vac, PS-ON signal held low (output disabled)
typ
50
80
A
A
0.1
0.5
mA
mA
0.5
W
max
units
6,000
4,000
3,000
2,000
μF
μF
μF
μF
0.98
0.95
OUTPUT
parameter
conditions/description
min
output capacitance
12, 15, 24 Vdc output models
27 Vdc output model
36 Vdc output model
48 Vdc output model
initial set point accuracy
full load
12,15, 24, 27 Vdc output models
all other output models
line regulation
rated load
load regulation
0 ~ 100% load
hold-up time
at 115 Vac
at 230 Vac
±2
±1
%
%
±0.5
%
±1
%
10
10
temperature coefficient
fan power
typ
ms
ms
±0.03
output power of 12 Vdc/0.5A
%/°C
6
W
max
units
15.6
19.5
31.2
35.1
46.8
60.0
Vdc
Vdc
Vdc
Vdc
Vdc
Vdc
PROTECTIONS
parameter
conditions/description
min
over voltage protection
output shutdown, latching
12 Vdc output model
15 Vdc output model
24 Vdc output model
27 Vdc output model
36 Vdc output model
48 Vdc output model
over current protection
auto recovery, hiccup
short circuit protection
continuous, auto recovery, hiccup, recovery time