date page
01/19/2012 1 of 4
SERIES: VSKM-S10 Series │ DESCRIPTION: MEDICAL SWITCHING POWER SUPPLY
FEATURES
• • • • • • • Up to 10 W continuous power Compact board mount design Universal input (85~264 Vac / 120~370 Vdc) Single output from 3.3 to 24 V Over current and short circuit protections Full medical approvals Efficiency up to 80%
RoHS
MODEL
VSKM-S10-3R3U VSKM-S10-5U VSKM-S10-9U VSKM-S10-12U VSKM-S10-15U VSKM-S10-24U
Notes:
output voltage
(Vdc)
output current
max (A)
output power
max (W)
ripple and noise1
typ (mVp-p)
efficiency
typ (%)
3.3 5 9 12 15 24
2 2 1.1 0.9 0.7 0.45
6.6 10 10 10 10 10
50 50 50 50 50 50
70 74 76 76 78 80
1. Ripple and noise measured at 20 MHz bandwidth
PART NUMBER KEY
VSKM-S10 - XXU
Base Number Output Voltage
cui.com
CUI Inc │ SERIES: VSKM-S10 Series │ DESCRIPTION: MEDICAL SWITCHING POWER SUPPLY
date 01/19/2012 │ page 2 of 4
INPUT
parameter voltage frequency input current inrush current external input fuse (recommended) leakage current at 110 Vac at 230 Vac at 110 Vac at 230 Vac slow blow, 250 V 230 V ac, 50 Hz conditions/description min 85 120 47 230 150 10 20 2 0.1 typ max 264 370 63 units Vac Vdc Hz mA mA A A A mA
OUTPUT
parameter line regulation load regulation temperature coefficient hold-up time voltage set point accuracy switching frequency at 230 Vac 3.3 V output all other models 10 ~ 100% conditions/description min typ ±0.5 ±1 0.02 50 ±3 ±2 60 max units % % %/°C ms % % kHz
PROTECTIONS
parameter over current protection short circuit protection auto recovery with no damage from a short on any output conditions/description min typ max 110 units %
SAFETY & COMPLIANCE
parameter isolation voltage safety approvals safety class EMI/EMC1 RoHS compliant MTBF conditions/description primary to secondary (for 1 minute) IEC 60601-1, EN 60601-1, UL 60601-1 class II EN 55011 (level B), IEC/EN 61000-4-2 level 4 (8kV/15kV), IEC/EN 61000-4-3, IEC/EN 61000-4-4 level 4 (4kV), IEC/EN 61000-4-5 level 4 (2kV/4kV) yes 25°C 300,000 hrs min 4,000 typ max units Vac
Notes: 1. external EMC application circuit is required
ENVIRONMENTAL
parameter operating temperature storage temperature case temperature operating humidity non-condensing conditions/description min -25 -40 typ max 70 105 95 95 units °C °C °C %
cui.com
CUI Inc │ SERIES: VSKM-S10 Series │ DESCRIPTION: MEDICAL SWITCHING POWER SUPPLY
date 01/19/2012 │ page 3 of 4
DERATING CURVES
1. output power vs. ambient temperature 2. output power vs. input voltage
≈
100 80 60 50 40 20
100 80 60 50 40 20
Load (%)
≈
Load (%)
-40
-30 -25 -20
-10
0
10
50
60
70
70
85
110
125
≈
210
≈
225
240
264
Ambient Temperature (°C)
Input Voltage (Vac)
MECHANICAL
parameter dimensions case material weight conditions/description 2.1 x 1.12 x 0.75 inch (53.5 x 28.5 x 19.0 mm) UL94V-0 47 50 53 g min typ max units
MECHANICAL DRAWING
units: mm [inches] tolerance: ±0.5 [±0.02] pin section tolerance: ±0.10 mm [±0.004]
53.50 2.106
1.20[0.047]
1
19.00 0.748
3 4
6.00 0.236 (TYP)
1.00 0.039 (TYP)
2
53.50 2.106 2 4 1 45.72 1.800 3
recommended PCB footprint (top view) grid: 2.54 mm (0.10 inch)
10.16 0.400
10.16 0.400
10.16 0.400
28.5 1.122
14.25 0.561
PIN CONNECTIONS PIN FUNCTION 1 AC(N) AC(L) 2 3 +Vo 4 -Vo
cui.com
CUI Inc │ SERIES: VSKM-S10 Series │ DESCRIPTION: MEDICAL SWITCHING POWER SUPPLY
date 01/19/2012 │ page 4 of 4
TYPICAL APPLICATION CIRCUIT
*Fuse *NTC
L
NF *MOV
C6
L
C7
+Vo
C1
C2 TVS
+
C8
RL
-Vo
N
N
EXTERNAL CAPACITORS TYPICAL VALUE (Unit: μF) MODEL VSKM-S10-3R3U VSKM-S10-5U VSKM-S10-9U VSKM-S10-12U VSKM-S10-15U VSKM-S10-24U
Notes:
C1 220 μF/10 V 220 μF/10 V 120 μF/25 V 120 μF/25 V 120 μF/25 V 68 μF/35 V
C2 0.1 μF/50 V 0.1 μF/50 V 0.1 μF/50 V 0.1 μF/50 V 0.1 μF/50 V 0.1 μF/50 V
TVS P6KE6.8A P6KE6.8A P6KE12A P6KE20A P6KE20A P6KE30A
1. Output filtering capacitor C1 is an electrolytic capacitor. It is recommended to use high frequency and low impedance electrolytic capacitors. For capacitance and current of capacitor please refer to manufacture’s datasheet. Voltage derating of capacitor should be 80% or above. C2 is ceramic capacitor, it is used to filter high frequency noise. TVS is a recommended component to protect post-circuits (if converter fails). 2. MOV is required to protect the device under surge. 2. It is recommended to use a 2A/250V slow blow FUSE. External input NTC is recommended to use 5D-9. 3. If EMC performance is required, it is recommended to add “EMC filter” at the input end (see EMC Application Figure). C6: X capacitor, recommended parameter 0.1μF/275V; C7,C8: Y capacitor, recommended parameter 2200pF/400V; NF: common model choke, recommended inductance is about 10mH-30mH.
REVISION HISTORY
rev. 1.0 description initial release date 01/16/2012
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters 20050 SW 112th Ave. Tualatin, OR 97062 800.275.4899
CUI offers a two (2) year limited warranty. Complete warranty information is listed on our website.
Fax 503.612.2383 cui.com techsupport@cui.com
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use. CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to affect its safety or effectiveness.
很抱歉,暂时无法提供与“VSKM-S10-15U”相匹配的价格&库存,您可以联系我们找货