TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
030AAC
Highlights & Features
•
•
•
•
•
•
•
•
•
MDS-030AAC05, 12, 15, 24 Meet Efficiency Level VI
Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 62368-1
Compliant with IEC 60601-1-2 4th Ed. Requirements
Meets Limited Power Source (LPS) requirements
Low touch current. Suitable for type BF applications
Detachable AC plug with multiple country options
IP22 ingress protection rating
2 × MOPP isolation
500K hours MTBF
Safety Standards
CB Certified for worldwide use
Model Number:
MDS-030AAC☐
Unit Weight:
150 grams
Dimensions (W × L × H): 53.5 × 88.0 × 27.5 mm
2.11 × 3.46 × 1.08 inch
General Description
The MDS series of external power supply comes with universal AC input at 85 Vac to 264 Vac. Other features include low touch current,
risk management report available and the electric shock protection comply with 2 × MOPP. The MDS series is certified for EMC
standards according to EN/BS EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment and EN/BS EN 55032
for Information Technology Equipment (ITE) radio-frequency equipment.
The MDS series come with both medical and ITE safety approvals including UL/cUL and IEC 60950-1 CB certification as a Limited
Power Source (LPS), and CCC approval. Designs are fully compliant with RoHS Directive for environmental protection.
Model Information
Medical AC-DC Adapter
Model Number
MDS-030AAC05
Input Voltage Range Efficiency Level
85-264Vac
Level VI
Rated Output Voltage
5 Vdc
Rated Output Current
3A
MDS-030AAC12
Level VI
12 Vdc
2A
MDS-030AAC15
Level VI
15 Vdc
2A
MDS-030AAC24
Level VI
24 Vdc
1.25 A
Model Numbering
MDS –
030
AAC
Delta Medical
Power Supply
Max wattage in the
product series
Family Code
Output Voltage (Single
Output)
Revision Code1
Revision Code2
- 05 for 5 V
DC plug type and output cable length
Country duck-head type
- 12 for 12 V
A:2.1 x 5.5 x10mm DC plug
A:China ; B:United States;
B:2.1 x 5.5 x12mm DC plug
C:European; D:United Kingdom;
- 15 for 15 V
- 24 for 24 V
Both A & B are 1200mm cable length
E:Australia; G:Korea; H: India,
J :Argentina ; K:Brazil;
M:South Africa
* MDS-030AAC15 BB is not available.
1
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (August 2022, Rev.14)
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Specifications
Model Number
MDS-030AAC12
MDS-030AAC05
MDS-030AAC15
MDS-030AAC24
Input Ratings / Characteristics
Nominal Input Voltage
100-240 Vac
Input Voltage Range
85-264 Vac
Nominal Input Frequency
50-60 Hz
Input Frequency Range
47-63 Hz
Input Current (max)
Average Efficiency (min)
115 Vac 0.5 A
0.6 A
0.8 A
0.8 A
230 Vac 0.3 A
0.4 A
0.6 A
0.6A
115 Vac 81.4%
86.21%
87.0%
87.0%
0.1 W
0.1 W
0.1 W
230 Vac
Standby Power (max)
115 Vac 0.1W
230 Vac
Inrush Current (typ.)
60 A @ 115 Vac and 100 A @ 230 Vac
Touch Current (max)
0.1 mA @ 264 Vac NC1), 0.3 mA @ 264 Vac SFC2)
Output-PE (protective earth)
leakage current for Type BF
application (max)
0.1 mA @ 264 Vac NC1), 0.5 mA @ 264 Vac SFC2)
1)
2)
NC: Normal condition
SFC: Single fault condition
Model Number
MDS-030AAC12
MDS-030AAC05
MDS-030AAC15
MDS-030AAC24
Output Ratings / Characteristics
Nominal Output Voltage
5 Vdc
12 Vdc
15 Vdc
Output Total Regulation
24 Vdc
±6%
Output Current
3A
2A
2A
1.25 A
Output Power
15 W
24 W
30 W
30 W
Line Regulation
±1%
Load Regulation
±5%
Ripple & Noise (max)
100 mVpk-pk
@ Rated load
150 mVpk-pk
@ Rated load
200 mVpk-pk
@ Rated load
200 mVpk-pk
@ Rated load
Start-up Time (max)
3000 ms @ 115 Vac
Hold-up Time (min)
10 ms @ 115 Vac
Dynamic Response (Overshoot &
Undershoot O/P Voltage)
±10% @ 50-100% load
Capacitive load (max)
4700 uF
*Periodic and Random Deviation.
Ripple & Noise measurement circuit with 20 MHz BW
PSU
5cm
0.47uF Ceramic
10uF Electrolytic
Load
Scope
Test Fixture
2
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (August 2022, Rev.14)
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Model Number
MDS-030AAC05
MDS-030AAC12
MDS-030AAC15
MDS-030AAC24
Mechanical
Case Chassis
PC
Case Cover
PC
Dimensions (W × L × H)
53.5 × 88.0 × 27.5 mm (2.11 × 3.46 × 1.08 in)
Unit Weight
150 grams (5.3 ounces)
Indicator
NA
Cooling System
NA
Terminal
Input Detachable type AC plug
Output Barrel type. Dimensions 2.1 × 5.5 × 10 mm (see page 5)
Environment
Surrounding Air
Temperature
Operating 0°C to +40°C
Storage -40°C to +85°C
Operating Humidity
5 - 95% RH (Non-Condensing)
Operating Altitude
5,000 meters (16400 feet)
Shock Test (Non-Operating)
50 G, 11 ms, 3 shocks for each direction
Vibration (Non-Operating)
5-500 Hz, 2.09 Grms, 20 minute for each three axis
Protections
Overvoltage (max)
150%, Latch Mode
Overload / Overcurrent (max)
250% of rated load current, Hiccup Mode,
(Non-Latching, Auto-Recovery)
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Over Temperature
Short Circuit
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Degree of Protection
IP22
Protection Against Shock
Class II
Reliability Data
MTBF (min)
3
500 K Hours based on Telecordia SR-332 (at 100 Vac, Max. load and 25°C Ambient)
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (August 2022, Rev.14)
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Model Number
MDS-030AAC05
MDS-030AAC12
MDS-030AAC15
MDS-030AAC24
Safety Standards / Directives
Medical Safety
ITE Safety
CE
UKCA
Galvanic Isolation
IEC 60601-1 Edition 3.1 (2012), EN 60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 NO.
60601-1:14, ANSI/AAMI ES 60601-1:2005/(R)2012
UL60601-1 (File No. E356265)
IEC 60950-1 (Ed.2,2005), IEC 62368-1, GB4943.1-2011, GB 9254-2008, GB 17625.1-2003
UL 60950-1 (File No. E131881), LPS (Limited Power Source)
In conformance with EMC Directive 2014/30/EU and Low Voltage Directive 2014/35/EU
EN 60601-1: 2006 + A11: 2011 + A1: 2013 + A12: 2014 & EN 60601-1-2: 2015
In conformance with Electrical Equipment (Safety) Regulations 2016 and Electromagnetic
Compatibility Regulations 2016, Medical Devices Regulations 2002(UK MDR 2002)
Input to 4000 Vac
Output
EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements)
EMC / Emissions
Harmonic Current Emissions
EN/BS EN 55011,EN/BS EN 55032,
FCC Title 47:Class B
IEC61000-3-2
Immunity to
Voltage Flicker
IEC61000-3-3
Electrostatic Discharge
IEC61000-4-2
Level 4 Criteria A1)
Air Discharge: 15 kV
Contact Discharge: 8 kV
Radiated Field
IEC61000-4-3
Criteria A1)
80MHz-2700MHz, 10 V/m AM modulation
385MHz-5785MHz, 28 V/m Pulse mode and other
modulation
Electrical Fast Transient / Burst
IEC61000-4-4
Level 3 Criteria A1): 2 kV
Surge
IEC61000-4-5
Conducted
IEC61000-4-6
Level 3 Criteria A1)
Differential Mode3): 1 kV
Level 2 Criteria A1)
150 kHz-80 MHz, 3 Vrms, 6 Vrms at ISM bands and
Amateur radio bands
Power Frequency Magnetic Fields
IEC61000-4-8
Voltage Dips
IEC61000-4-11
Criteria A1)
Magnetic field strength 30 A/m
Criteria A1)
0% UT, 0.5 cycle (10 ms)
, 0o/45o/90o/135o/180o/225o/270o/315o/360o
Criteria B2)
0% UT, 1 cycle (20 ms), 0o
Criteria B2)
70% UT, 25 cycle (500 ms), 0o
Criteria B2)
0% UT, 250 cycle (5000 ms), 0o
1) Criteria A: Normal performance within the specification limits
2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test.
3) Symmetrical: Differential mode (Line to line)
4
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (August 2022, Rev.14)
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Block Diagram
Dimensions
W × H × H: 53.5 × 88.0 × 27.5 mm
Note:
The plug’s polarity is
5
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (August 2022, Rev.14)
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Duck Head Type:
6
CN type
US/JP type
EU type
UK type
AU
KR type
IN type
AR type
BZ type
SA type
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (August 2022, Rev.14)
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Functions
Start-up Time
The time required for the output voltage (Vo) to reach 90% of its set value, after the input AC voltage is applied.
Rise Time
The time required for the output voltage (Vo) to change from 10% to 90% of its steady state value.
Hold-up Time
Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time
required for the output to reach 95% of its set value, after the input voltage is removed.
Graph illustrating the Start-up Time, Rise Time, and Hold-up Time
Inrush Current
Dynamic Response
Inrush current is the input current that occurs when the input
voltage is first applied. For AC input voltages, the maximum
peak value of inrush current will occur during the first half cycle
of the applied AC voltage. This peak value decreases
exponentially during subsequent cycles of AC voltage.
The power supply output voltage will remain within ±3% of its
steady state value, when subjected to a dynamic load change
from 50 to 100% of its rated current.
7
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (August 2022, Rev.14)
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Overload & Over current Protections
Overvoltage Protection
The power supply’s Overload (OLP) and Over current (OCP)
Protections will be activated when output current is between
110% and 250% of IO (Max load). Upon such an occurrence, VO
will start to drop. Once the power supply has reached its
maximum power limit, the protection will be activated and the
power supply will go into “Hiccup mode” (Auto-Recovery). The
power supply will recover once the fault condition causing the
OLP and OCP is removed and IO is back within the specified limit.
The power supply’s overvoltage circuit will be activated when its
internal feedback circuit fails. The output voltage shall not exceed
its specifications defined on Page 3 under “Protections”. Power
supply will latch off, and require removal/re-application of input
AC voltage in order to restart.
Over Temperature Protection
As mentioned above, the power supply also has Over
Temperature Protection (OTP). This is activated when the
overload condition persists for an extended duration and the
output current is below the overload trigger point but >100% load.
In the event of a higher operating condition at 100% load, the
power supply will run into OTP when the surrounding air
temperature is higher than the operating temperature. When
activated, the output voltage will go into hiccup mode until the
input voltage is removed; then, reapplied, and the surrounding air
temperature drops to its normal operating temperature.
Additionally, if the IO is 110% for a prolong period of
time (depending on the load), the Over Temperature Protection
(OTP) will be activated due to high temperature on critical
components. The power supply will then go into hiccup mode until
the fault is removed; and, the input voltage is removed, then
reapplied.
Short Circuit Protection (Auto-Recovery)
The power supply’s output OLP/OCP function also provides
protection against short circuits. When a short circuit is applied,
the output current will operate in “Hiccup mode”, as shown in the
illustration in the OLP/OCP section on this page. The power
supply will return to normal operation after the short circuit is
removed.
8
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (August 2022, Rev.14)
TECHNICAL DATASHEET
Medical AC-DC Adapter
MDS-030AAC Series / MDS-030AAC□
Certificate
Typically, the input current waveform is not sinusoidal due to the periodical peak charging of the input capacitor.
In industrial environment, complying with EN 61000-3-2 is only necessary under special conditions. Complying
to this standard can have some technical drawbacks, such as lower efficiency as well as some commercial
aspects such as higher purchasing costs. Frequently, the user does not profit from fulfilling this standard,
therefore, it is important to know whether it is mandatory to meet this standard for a specific application.
In addition to a UL Total Certification Program (TCP) approved client laboratory for IEC62368-1. Delta also has
participated UL Client Test Data Program (CDTP) for IEC 60601.
Meet Level V Efficiency Requirement for MDS-030AAC07
Meet Level VI Efficiency Requirement for MDS-030AAC05, 12, 15, 24
Meets Limited Power Source (LPS) requirement
Attention
Delta provides all information in the datasheets on an “AS IS” basis and does not offer any kind of warranty through the information for
using the product. In the event of any discrepancy between the information in the catalog and datasheets, the datasheets shall prevail
(please refer to www.DeltaPSU.com for the latest datasheets information). Delta shall have no liability of indemnification for any claim
or action arising from any error for the provided information in the datasheets. Customer shall take its responsibility for evaluation of
using the product before placing an order with Delta.
Delta reserves the right to make changes to the information described in the datasheets without notice.
Manufacturer and Authorized Representatives Information
Manufacturer
Thailand
Delta Electronics (Thailand) PCL.
909 Pattana 1 Rd., Muang, Samutprakarn, 10280 Thailand
Taiwan
Delta Electronics, Inc.
3 Tungyuan Road, Chungli Industrial Zone, Taoyuan County
32063, Taiwan
Authorized Representatives
The Netherlands
Delta Greentech (Netherlands) B.V.
Zandsteen 15, 2132 MZ Hoofddorp, The Netherlands
9
United Kingdom
Delta Electronics Europe Limited
1 Redwood Court, Peel Park Campus,
East Kilbride, Glasgow, G74 5PF, United Kingdom
All parameters are specified at 25°C ambient unless otherwise indicated.
www.DeltaPSU.com (August 2022, Rev.14)