Embedded Power for Business-Critical Continuity
NLP65 Medical Series
Single, dual and triple output
Total Power: 65W Input Voltage: 85 - 264 VAC # of Outputs: Single, dual, triple
Rev.06.18.09_104 NLP65 Medical 1 of 4
Special Features
• 85 VAC to 264 VAC universal input range • Harmonic current correction as standard • Maximum component height 1.26 inches • UL, CSA and VDE safety approvals • Overvoltage and short circuit protection • 5 x 3 x 1.26 inch (127.0 x 76.2 x 32mm) footprint • Available RoHS compliant • 2 year warranty
Electrical Specifications
Input
Input voltage range: Input frequency range: Input current: (cold start) Safety ground leakage current: Input current: Input fuse: 120 Vac 230 Vac 264 Vac, 60 Hz 120 Vac 230 Vac Universal input (see Note 2) 85 - 264 Vac 47 - 63 Hz 17 A max. 32 A max 95 µA 1.05 A rms 0.51 A rms 250 Vac F 3.15 A Natural convection 65 W max. See table At turn-on Main output 25% step at 0.1 A/µs Main outputs Cyclic operation 1.0 s, max 5.0% max. dev., 1ms recovery to 1.0% ± 0.02%/°C 125%, ± 10% Yes
Output
Output power: Total regulation: (line and load) Rise time:
Safety
• VDE0705/EN60601-1/IEC1010 File No. 10401-3336-0156/32480 Licence No. 121949 • UL1950 File No. E147937 • CSA C22.2 No. 950 File No. LR41062C
Transient response: Temperature coefficient: Overvoltage protection: Short circuit protection:
Embedded Power for Business-Critical Continuity
All specifications are typical at nominal input, full load at 25° C unless otherwise stated.
EMC Charateristics
Conducted emissions: Radiated emissions: ESD air: ESD contact: Surge: Fast transients: Radiated immunity: Conducted immunity: EN55022, FCC part 15 EN55022, FCC part 15 EN61000-4-2, level 3 EN61000-4-2, level 4 EN61000-4-2, level 3 EN61000-4-4, level 3 EN61000-4-3, level 3 EN61000-4-6, level 3 120 Vac, 60 Hz 120 Vac, 65 W Input/output Input/chassis Fixed 283 g (10 oz) MIL-HDBK-217F 150,000 hours Level A Level A Perf. criteria 1 Perf. criteria 1 Perf. criteria 1 Perf. criteria 1 Perf. criteria 2 Perf. criteria 2 16 ms @ 65 W 72% typical 4000 Vac 1500 Vac 100 kHz, ± 5 kHz
Rev.06.18.09_104 NLP65 Medical of 4
General Specifications
Hold-up time: Efficiency: Isolation voltage: Switching frequency: Approvals and standards: Weight: MTBF demonstrated:
EN60601, UL2601, CSA 22.2 No. 125
Environmental Specifications
Thermal performance: Operating ambient (See derating curve) Non-operating 0 °C to 50 °C, ambient, convection cooled 50 °C - 70 °C ambient, convection cooled Peak (0 °C to 50 °C, 60 s) Relative humidity: Altitude: Vibration (See Note 5): Shock Non-condensing Operating Non-operating 5 - 500 Hz per MIL-STD-810E 0° C to +70 °C -40 °C to +85 °C 65 W Derate to 50% load See table 5 to 95% RH 10,000 feet max. 30,000 feet max. 2.4 G rms approx. 516.4 Part IV
0°C 10°C 20°C 30°C 40°C 50°C 60°C 70°C 75W 20 CFM FORCED AIR COOLING 65W DERATING CURVE Output Power (Watts)
NATURAL CONVECTION COOLING
38W 32W
Embedded Power for Business-Critical Continuity
Ordering Information
Output Voltage +5 V +12 V –12 V +5 V +15 V –15 V +5 V +24 V +5 V +12 V +5 V +12 V +15 V +24 V Output Current Max (1) 7A 2.5 A 0.5 A 7A 2.2 A 0.65 A 7A 2A 7A 2.5 A 10 A 5.4 A 4.4 A 2.7 A Peak 9.1 A 3.3 A 0.81 A 9.1 A 2.9 A 0.85 A 9.1 A 2.6 A 9.1 A 3.3 A 13 A 7A 5.7 A 3.5 A Fan (10) 8A 3A 1A 8A 2.5 A 0.8 A 8A 2A 8A 3A 12 A 6.5 A 5.3 A 3.5 A Ripple (4) 50 mV 150 mV 120 mV 50 mV 150 mV 150 mV 50 mV 240 mV 50 mV 150 mV 50 mV 120 mV 150 mV 240 mV Total Regulation (6) ±2.0% ±5.0% ±5.0% ±2.0% ±5.0% ±5.0% ±2.0% ±5.0% ±2.0% ±5.0% ±2.0% ±2.0% ±2.0% ±2.0% NLP65-9905J NLP65-9912J NLP65-9915J NLP65-9924J NLP65-9929J NLP65-9920J NLP65-9910J Model Number (11, 12) NLP65-9908J
Rev.06.18.09_104 NLP65 Medical of 4
Notes
Natural convection cooling. Models NLP65-9929J, NLP65-9908J, NLP65-9910J must not exceed 62.5 Watts continuous output power with natural convection. Model NLP65-9920J not to exceed 65 Watts continuous output power with natural convection. 2 When the input voltage is less than 90 Vac the operating temperature range is 0 °C to +40 °C. The ripple and regulation specifications may not be met. 3 Peak output current lasting less than 60 seconds with duty cycle less than 5%. During peak loading, output voltage may exceed total regulation limits. 4 Figure is peak-to-peak for convection power rating. Output noise measurements are made across a 20 MHz bandwidth using a 6 inch twisted pair, terminated with a 10 µF electrolytic capacitor and a 0.1 µF ceramic capacitor. 5 Three orthogonal axes, random vibration 10 minutes for each axes, 2.4 G rms 5 Hz to 500 Hz. 6 To maintain stated regulation then: for single output units I ≥ 0.2 A I max. for multiple output units 0.25 ≤ I(A)/I(B) ≤ 5, for I(A) ≥ 0.2 A I(A) max. 7 For optimum reliability, no part of the heatsink should exceed 120 oC, and no semiconductor case temperature should exceed 130 oC. 8 CAUTION: Allow a minimum of 1 second after disconnecting line power when making thermal measurements. 9 This product is only for inclusion by professional installers within other equipment and must not be operated as a stand alone product. 10 Maximum continuous output power for all multiple output models must not exceed 75 Watts with 20 CFM forced air cooling at 50 oC. 11 The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant. 12 NOTICE: Some models do not support all options. Please contact your local Emerson Network Power representative or use the on-line model number search tool at http://www.PowerConversion.com to find a suitable alternative. 1
Embedded Power for Business-Critical Continuity
Americas
5810 Van Allen Way Carlsbad, CA 92008 USA Telephone: +1 760 930 4600 Facsimile: +1 760 930 0698
Rev.06.18.09_104 NLP65 Medical 4 of 4
5.00 (127.00) 4.550 (115.57) J1 3N 1L 3.00 (76.20) J2 C19 6 T1 2.550 (64.770) 0.225 (5.715)
Europe (UK)
J3 1
Waterfront Business Park Merry Hill, Dudley West Midlands, DY5 1LX United Kingdom Telephone: +44 (0) 1384 842 211 Facsimile: +44 (0) 1384 843 355
Asia (HK)
14/F, Lu Plaza 2 Wing Yip Street Kwun Tong, Kowloon Hong Kong Telephone: +852 2176 3333 Facsimile: +852 2176 3888
For global contact, visit:
Mounting Hole Diameter 0.156 (4.00) ALL DIMENSIONS IN INCHES (mm) MAXIMUM COMPONENT HEIGHT 1.26 (32.00) TOLERANCE: .xxx 0.005" .xx 0.02"
www.PowerConversion.com techsupport.embeddedpower@ emerson.com
While every precaution has been taken to ensure accuracy and completeness in this literature, Emerson Network Power assumes no responsibility, and disclaims all liability for damages resulting from use of this information or for any errors or omissions.
Pin Connections
J1 Pin 1 Pin 2 Pin 3 J2 Pin 1 Safety Ground AC Line No Pin AC Neutral AC (J1) DC (J3)
Input and Output Connectors
Molex 26-60-4030 type Molex 26-60-4060
Mating Connectors
Molex 09-50-3031 or equivalent with Molex 08-50-0105 or equivalent crimp terminals Molex 09-50-3061 with Molex 2478 phosphor bronze crimp terminals or equivalent.
Emerson Network Power. The global leader in enabling business-critical continuity. AC Power Connectivity DC Power Embedded Computing Embedded Power
Output Pin Connections
J3 Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 SINGLE No Connection V (A) V (A) Return Return No Connection DUAL V (B) V (A) V (A) Return Return No Pin TRIPLE V (B) V (A) V (A) Return Return V (C)