User’s Guide
MultiMaster™ Digital MultiMeter Series With PC Interface Models: MM560 MM570
WARRANTY EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for three years from date of shipment (a six month limited warranty applies on sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department at (781) 890-7440 ext. 210 for authorization. A Return Authorization (RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied.
Copyright © 2002 Extech Instruments Corporation. All rights reserved including the right of reproduction in whole or in part in any form. MM Series V2.2 9/03 P/N: 7M1C-0271-A000
1
Introduction Congratulations on your purchase of Extech model MM560 or MM570 digital multimeter. Properly used, this meter will provide many years of reliable service. MultiMaster™ Meters provides high resolution, high accuracy, true rms measurements of AC Voltage and Current, DC Voltage and Current, AC+DC, Resistance, Frequency, Capacitance, Duty Cycle, dBm, Peak, %4-20, Diode, and Continuity. The MultiMaster™ also offers RS-232 PC Interface data acquisition capabilities as an option. Model MM570 offers dual input Temperature measurements and very wide bandwidth. Safety International Safety Symbols
!
Caution ! Refer to the explanation in this Manual Caution ! Risk of electric shock Earth (Ground) Double Insulation or Reinforced insulation Fuse AC, Alternating Current or Voltage DC, Direct Current or Voltage AC+DC, Current or Voltage
Safety Precautions 1. Improper use of this meter can cause damage, shock, injury or death. Read and understand this users manual before operating the meter. 2. Make sure any covers or battery doors are properly closed and secured. 3. Always remove the test leads before replacing the battery or fuses. 4. Inspect the condition of the test leads and the meter itself for any damage before operating the meter. Repair or replace any damage before use. 5. To reduce the risk of fire or electric shock, do not expose this product to rain or moisture. 6. Do not exceed the maximum rated input limits.
2
7. Use great care when making measurements if the voltages are greater than 25VAC rms or 35VDC. These voltages are considered a shock hazard. 8. Always discharge capacitors and remove power from the device under test before performing Capacitance, Diode, Resistance or Continuity tests. 9. Remove the battery from the meter if the meter is to be stored for long periods. 10. To avoid electric shock, do not measure AC current on any circuit whose voltage exceeds 1000V AC. 11. Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the electrical contacts. Other means should be used to ensure that the terminals are not "live". 12. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. 13. The product is intended only for indoor use. PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY OVERVOLTAGE CATEGORY III Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation. OVERVOLTAGE CATEGORY IV Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. Note – Examples include electricity meters and primary over-current protection equipment CENELEC Directives: These instruments conform to CENELEC lowvoltage directive 73/23/EEC and Electromagnetic compatibility directive 89/336/EEC.
3
Specifications
DC VOLTAGE Accuracy Range MM570 MM560 500.00 mV, 0.02% + 2d 0.03% + 2d 5.0000V, 50.000V 500.00V 0.04% + 2d 0.05% + 2d 1000.0V 0.05% + 2d 0.1% + 2d NMRR: >60dB @ 50/60Hz, CMRR: >120dB @ DC, 50/60Hz, Rs=1kΩ Input impedance: 10MΩ, 30pF nominal (80pF nominal for 500mV range) DC CURRENT Range Accuracy Burden Voltage 0.15% + 20d 500.00µA 0.15mV/µA 0.1% + 20d 5000.0µA 0.15mV/µA 50.000mA 0.15% + 10d 3.3mV/mA 500.00mA 0.1% + 20d 3.3mV/mA 5.0000A 0.5% + 10d 0.03V/A 10.000A* 0.5% + 20d 0.03V/A *10A continuous, 20A for 30 seconds max with a 5 minute cool down interval RESISTANCE Accuracy Range MM570 MM560 0.07% + 10d 0.1% + 6d 500.00Ω 0.07% + 2d 5.0000kΩ, 50.000kΩ, 500.00kΩ 0.2% + 6d 0.4% + 6d 5.0000MΩ 2.0% + 6d 2.0% + 6d 50.000MΩ Open circuit voltage: