IN-SITU
ZIRCONIA OXYGEN ANALYZER
ZSB
DATA SHEET
This oxygen analyzer is used to continuously measure oxygen concentration in combustion exhaust gas of industrial boilers or furnaces, and is ideally suited for combustion moni-toring and control. The analyzer consists of a flow guide tube with a detector, which is directly inserted into a flue, etc. to introduce the gas to be analyzed, a transducer, which performs control, signal processing, display, input to and output from outside, and communication, and an auto calibration/blow unit. The detector of a unit sensor structure facilitates on-site maintenance. The transducer provided with an unconventional sensor diagnostic function and sensor renewal function ensures long-term stable detecting operation.
FEATURES
1. No need for gas sampling devices Since the sensor unit is directly inserted into a flue, etc., gas-sampling devices such as gas aspirator and dehumidifier are not required, which ensures high response speed. 2. Easy maintenance The sensor in a unit structure mounted to the detector can be replaced easily. Since the detector and the flow guide tube are installed separately, the filter at the tip of the detector can be replaced easily, and detector and the flow guide tube can be maintained separately depending on the degree of formation of rust, for example. 3. High reliability ensured by the sensor diagnostic and sensor renewal functions Depending on the components in the gas to be measured, the sensor electrode may degrade. The transducer diagnoses the degree of degradation of the sensor and renews it electrically, thus ensuring high reliability and long-term stable measurement. 4. Improved safety Detecting a break of the thermocouple for heater control installed at the sensor, power supply to the detector is stopped. Furthermore, external contact input stops power supply to the detector in an emergency. Those functions along with the key lock function are equipped as standard to ensure improved safety. 5. Simple operation The transducer can be operated and various settings can be made using an interactive system, and display language is available in English, Japanese, and Chinese.
SPECIFICATIONS
General
Measuring object: Oxygen in noncombustible gas Measuring method: Directly insert type zirconia system Measuring range: 0 to 2 ···Setting range at option 2 in 50 vol% O2 (in 1 vol% O2 steps) Within ±0.5%FS Within ±2%FS Within 4 to 7 sec, for 90% (from calibration gas inlet) More than 10 min. 4 to 20mA DC (allowable load resistance 500Ω or less) or 0 to 1V DC (output resistance more 100Ω or more) Rated voltage; 100 to 120V AC (operating voltage 90 to 132V AC) Rated frequency; 50/60Hz
Repeatability: Linearity: Response time: Warmup time: Analog output:
Power supply:
Date
EDS3-124c
Apr 1, 2011
ZSB
Power consumption: Maximum 240VA (Detector: approx. 200VA, Converter: approx. 40VA) Normal 70VA (Detector: approx. 50VA, Converter: approx. 20VA) Steel plates (main frame) Wall-mounting type: 700 X 400 X 180mm Self-standing type: 1500 X 530 X 550mm Wall-mounting type: Approx. 27 kg (excluding cable, detector and standard gas) Self-standing type: Approx. 48 kg (excluding cable, detector and standard gas) Munsell 5Y7/1, semi-gloss Mass: Finish color: Approx. 1.6kg Silver and SUS metallic color 1.5 to 2 L/min Blowdown air inlet pressure: 200 to 300kPa
Calibration gas flow:
Material:
Outer dimensions (H x W x D):
Converter (ZKM3)
Concentration value indication: Digital indication in 4 digits Contact output signal: (1) Contact specification; 6 points, 1a 250V AC/3A or 30V DC/3A (2) Contact function; • Under maintenance • Instrument anomalies Note1) • Alarm Note2) • Zero calibration gas valve • Span calibration gas valve • Under blowdown Note3) Note1) The following Instrument errors (1) Thermocouples break (2) Sensor break (3) Temperature fault (4) Calibration fault (5) Zero/span fault (6) Output error the contact ON Note2) Alarm selects just one as mentioned below (1) High (2) Low (3) H/L (4) HH (5) LL, it turns ON while operating. Note3) Under blow down, it is available in case of option, and it turns ON while operating. Contact input signal: (1) Contact specification; 3 points (the following option) ON; 0V (10mA or less), OFF; 5V (2) Contact function; • External hold • Calculation reset • Heater OFF • Blow down (option) • Inhibition of calibration • Calibration start • Range change Calibration method: (a) Manual calibration with key operation (b) Auto. calibration (option) Calibration cycle; 00 day 00 hour to 99 days 23 hours (c) All calibration • Settable range Zero gas; 0.010 to 25.00% O2 Span gas: 0.010 to 50.00% O2 • Recommended calibration gas concentration Zero gas; 0.25 to 2.0% O2 Span gas; 20.6 to 21.0% O2 (oxygen concentration in the air)
Mass:
Finish color:
Mounting method: Wall-mounting or self-standing
Detector (ZFK8)
Measured gas temperature: Flow guide tube system; −20 to +600ºC (for general-use, corrosive gas) Measured gas pressure: −3 to +3kPa Flow guide tube: • With or without blowdown nozzle Flange; JIS5K 65A FF Insertion length; 0.3, 0.5, 0.75, 1m • For high particulate gas (with blowdown nozzle) With or without cover Flange; JIS5K 80A FF Insertion length; 0.3, 0.5, 0.75, 1m Operating temperature: −10 to +60ºC for Primary detecting element −5 to +100ºC for detector 125ºC or less at detector flange surface with power applied Storage temperature: −20 to +70ºC Structure: Dust/rain-proof structure (IEC IP66 equivalent, except the filter part at the tip) Use a heat insulation cover in cold climates (as specified) Filter: Alumina (filtering accuracy 50µm) and quartz paper Main materials of gas-contacting parts: Zirconia, SCS14 (SUS316 equivalent), platinum, SUS304 Calibration gas inlet: φ6mm tube join Reference air inlet: φ6mm tube join Detector mounting: Horizontal plane ±45º, ambient air should be clean. Outer dimensions (L × max. dia.): 210mm × 100mm
Calibration gas:
2
Blowdown:
A function for blowing out dust with compressed air that has deposited in the flow guide tube. Blowdown can be performed only for a predetermined time and at prede-termined intervals. Blowdown cycle; 00 hour 00 minute to 99 hours 59 minutes Blowdown time; 0 minute 00 second to 0 minutes 999 seconds Output signal hold: Output signal is held during manual/ auto calibration, blowdown, sensor recovery proc-essing, warm-up, PID auto tuning, and while maintenance mode setting is "available". The hold function can also be released. Communication function: RS232C (MODBUS) standard specification RS485 (MODBUS) (option)
Combustion efficiency display: When you select this display, "rich mode display" will be simultaneously displayed. This function calculates and displays combustion efficiency from oxygen concentration and measured gas temperature. Thermocouple (R) is required for temperature measurement. Operating temperature: −20 to +55ºC Operating humidity: 95% RH or less, non condensing Storage temperature: −30 to +70ºC Storage humidity: 95% RH or less, non condensing Outer dimensions (H × W × D): 182 × 163.5 × 70.6mm Approx. 2kg (excluding cable and detector) (IP66) Mounting method: Mounted flush on panel Mass:
3
ZSB
CODE SYMBOLS
12345678
ZSB 2-
9 10 11 12 13
14 15
Description
4th digit 1 2 5th digit A B
Converter mounting (4th digit) Wall-mounting type (with indication window) Self-stand type (with indication window) Sensor, Power supply, Cal. Gas inlet (5th digit) For general use without protection cover, 100/115 V AC, 50/60 Hz and inlet for φ6 mm tube For general use with protection cover, 100/115 V AC, 50/60 Hz and inlet for φ6 mm tube * Protection cover should be selected when ambient temperature is -10°C or lower. Sensor cable + calibration gas tube (6th digit) None 6 m cable (without conduit) + 6 m Teflon tube (φ6/φ4) 10 m cable (without conduit) + 10 m Teflon tube (φ6/φ4) 15 m cable (without conduit) + 15 m Teflon tube (φ6/φ4) 20 m cable (without conduit) + 20 m Teflon tube (φ6/φ4) 6 m cable (with conduit) + 6 m Teflon tube (φ6/φ4) 10 m cable (with conduit) + 10 m Teflon tube (φ6/φ4) 15 m cable (with conduit) + 15 m Teflon tube (φ6/φ4) 20 m cable (with conduit) + 20 m Teflon tube (φ6/φ4) 6 m cable (4-core cable only. Heater cable should be prepared by customer.) + 6 m Teflon tube (φ6/φ4) 10 m cable (4-core cable only. Heater cable should be prepared by customer.) + 10 m Teflon tube (φ6/φ4) 15 m cable (4-core cable only. Heater cable should be prepared by customer.) + 15 m Teflon tube (φ6/φ4) 20 m cable (4-core cable only. Heater cable should be prepared by customer.) + 20 m Teflon tube (φ6/φ4) * When exclusive-special conduit and heater cable are not used, the following conduit should be used. Conduit with outer diameter φ23 and inner diameter φ16 as specified in JIS C8411 For preparing two heater cables, their rating should be 3 A or more. Calibration gas unit (7th digit) Automatic calibration unit + zero gas + filter regulator Automatic calibration unit + zero gas + air standard gas Manual calibration unit + zero gas + filter regulator Manual calibration unit + zero gas + air standard gas Flow guide tube (9th, 10th and 11th digit) (Flange) (Application) None None SUS304 For general use SUS304 For general use SUS304 For general use SUS304 For general use SUS316 For corrosive SUS316 For corrosive SUS316 For corrosive SUS316 For corrosive SUS316 Without high particulate cover SUS316 Without high particulate cover SUS316 Without high particulate cover SUS316 Without high particulate cover Others Others Output signal (12th digit) 4 to 20mA DC 0 to 1V DC Language (13th digit) Japanese English Chinese Communication functions (14th digit) RS232 RS485 Reference gas (15th digit) None φ6mm tube joint (Length) None 300mm 500mm 750mm 1000mm 300mm 500mm 750mm 1000mm 300mm 500mm 750mm 1000mm Others (Note)
6th digit Y A B C D E F G H J K L M
7th digit 1 2 3 4
9th, 10th, 11th digits
0 5 5 5 5 5 5 5 5 6 6 6 6 Z
Y A A A A B B B B D D D D Z
0 3 5 7 1 3 5 7 1 3 5 7 1 Z
With blow-back unit With blow-back unit With blow-back unit With blow-back unit
12th digit B E 13th digit J E C 14th digit 1 2 15th digit Y 1
4
DEVICE CONFIGURATION
The device to be combined differs according to the conditions of the gas to be measured. Please select the flow guide tube, referring to the table shown below.
Measured gas Application Temperature Gas Flow Dust Less than 0.2g/Nm3 5 to 20m/s Less than 1g/Nm3 Less than 25g/Nm3 Note For boiler. Fuel; gas, oil When moisture content in measured gas is low. When moisture content in measured gas is low. With blow-down Device configuration Type at 4th to 7th digits ZSB ZSB ZSB 2– 2– 2– Flow guide tube at 9th to 11th digits 5A 5B 6D
Boiler, refuse incinerator, sludge incinerator, 600ºC or less etc.
Note (1) Dust volume is approximate value. (2) Reference air port specification should be selected when standard gas concentration changes. (3) For optional specifications, contact the manufacturer separately.
SCOPE OF DELIVERY
1) Standard delivery 1 set x zirconia oxygen analyzer 1 set x instruction manual 1 set x standard accessories · 1 x O -ring (for detector) · 1 x ceramic filter (for detector) · 2 each x fuse (3.15 A, 0. 5 A) · 1 set x inlet (for standard gas connection) · 1 x polyethylene tube (for standard gas connection) · 1 x Toaron tube (for standard gas connection) · 1 set x flow guide tube accessories (bolt, nut, spring washer x 4 pieces each) 2) Items to be prepared separately · Replacement detector · Spare (detector) Spares for 1-year operation 1 x O -ring 2 x ceramic filter 1 x Filter frame Spares for 2-year operation 2 x O -ring 4 x ceramic filter 2 x Filter frame
CAUTIONS
· If combustible gas (CO, H2 etc.) exists in the measured gas, error will occur due to burning at the sensor section. The inclusion of corrosive gas (Si vapor, alkaline metal, P Pb etc.) will shorten the , life of the sensor. · When the measured gas temperature is high (+300ºC or higher), the flange should be separated from the furnace wall in order to bring the detector flange surface temperature below the specified value (+125ºC). The flow guide tube should be attached in the direction in which the gas flow to the detector decreases. · If much dust is contained in the gas, the flow guide tube should be connected obliquely from the top. This tube should also be connected in the direction where gas flow into the detector section is reduced. · In application to refuse incineration, automatic blow-down of the flow guide tube should be avoided (in order to protect the tube from corrosion due to drain formation). Blow-down should be carried out manually when indication change becomes small during stop of the incinerator.
5
ZSB
CONFIGURATION
Rc1/4 (upper side blow port should be selected) Blow-down, 200 to 300kPa Solenoid valve (not supplied) Reduction valve (unnecessary when supply air pressure is 200 to 300kPa) Supply air
Heat insulation cover
Flow guide tube Detector (ZFK8)
15ASGP tube equivalent or φ10/φ8mm Teflon pipe or steel pipe (not supplied)
15ASGP tube equivalent or φ10/φ8mm Teflon pipe or steel pipe (not supplied)
Gas temperature. 600ºC max. 6 Ref. air 5 1 2 3 4 White Black White Red Rainproof flexible conduit. (Max. 20m)
Blue Gas outlet
Yellow
Gas temperature. 600ºC max. Flow guide tube for high particulate
Cable gland
Gas outlet T1
R1
1 Gas temperature. 600ºC max. Flow guide tube for high particulate with cover 2 + O2 –
O2 input
Converter (ZKM) 11 12
ZERO GAS
3 4 + TC1 –
Thermocouple input
5 +
or AO
6
–
7
8
9
10
13
14
TM-1
ZV
*1
Analog output 4 to 20mA DC or 0 to 1V DC Thermocouple for combustion control
Fault contact output
Blow HEATER contact Calibrating output gas contact output Alarm Maintenance contact Calibrating gas contact output contact output output
MAINTE COM
SPAN GAS
Contact input
DI1 DI2
ALARM
AC power supply
N AC L
+ TC2 – φ6mm Teflon tube for calibration gas (not supplied) 15 1 TM-2 2 3 Auto breaker 16
DI3 DICOM
SV
17
18
19
20
21
22
23
24
25
26
27
28
Converter
Maintenance signal during calibration/maintenance
R
1
T
2
E
3 Earth 3 TGS When auto calibration is selected Joint Solenoid valve Zero Flowmeter Solenoid valve (0.2L/min)
T1 12
4
5
6
7
8
9
10
11
R1
AC power supply *1 Use a screw for mounting of converter main unit (1.5L/min)
2
Zero
Pressure regulator
O2/N2
Reference air
Span
Standard gas for zero point
Flowmeter
Container box
Span Pressure regulator
Supply air Reduction valve (mounted on the main unit)
Air or O2/N2
Standard gas for span point When air standard gas is selected
6
OUTLINE DIAGRAM
(Unit : mm)
Self-standing type: ZSB2
Instrumentation air inlet (for blow-back) Rc1/2
400
180
100
{490}
External connection terminal (converter) 700
G1/2 Rc1/2
Rc1/4 Instrumentation air inlet (for span calibration gas)
Rc1/4 Instrumentation air outlet (for span calibration gas) Filter regulator (selectable by code symbol)
760
4-φ15 490 530 350 550
30
Front view
Right side view
(Front) Cable gland
exclusive for detector
EXTERNAL CONNECTION DIAGRAM (POWER SUPPLY)
1
R
2
T
3
E
4
5
6
1500
External connection terminal (power supply)
Instrumentation air outlet (for blow-back) Regulator + solenoid valve (selectable by code symbol)
{350}
Anchor plan, anchor bolts (option) M12×160×50
7
8
9
10 11 12
Power Ground supply
Maintenance signal Calibration status/ maintenance status
Inlet (for φ6 mm tube)
(Rear)
Cable gland A15A
(Box bottom view)
7
ZSB
Wall-mounting type: ZSB2
Instrumentation air inlet (for blow-back) Rc1/2
255 180 300 4-φ15
400
External connection terminal (converter) 700
G1/2 Rc1/2
Instrumentation air outlet (for blow-back) Regulator + solenoid valve (selectable by code symbol)
External connection terminal (power supply)
Rc1/4 Instrumentation air inlet (for span calibration gas)
Rc1/4 Instrumentation air outlet (for span calibration gas) Filter regulator (selectable by code symbol)
Front view
Right side view
Rear view
(Front)
Cable gland exclusive for detector
EXTERNAL CONNECTION DIAGRAM (POWER SUPPLY)
1
R
2
T
3
E
4
5
6
7
8
9
10 11 12
Inlet (for φ6 mm tube)
(Rear)
Cable gland A15A
Power Ground supply
Maintenance signal Calibration status/ maintenance status
(Box bottom view)
8
500
Converter: ZKM3
163.5 70.6 10 143.5
Oxygen Analyzer
MODE
ESC
ENTER
POWER
Made in Japan
TYPE OUTPUT DC POWER AC SER. NO. MFD 50/60Hz 60VA
OXYGEN TRANSM ITTER
172
変換器へ供給する電源電圧は、接続される 検出器の電源電圧に合わせてください。
注 意
FUSE T 0.5A
POWER
Power supply must be selected the same voltage of detector.
CAUTION
HEATER
FUSE T 2.5A
POWER
ON OFF
+ O2 − + TC1 − + AO − FAULT BLOW ZV HEATER 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 ALARM MAINAC COM SV + TC2 − DI1 DI2 DI3 DIN L TENANCE COM
4-φ4.5 MTG. holes
External terminal (TM1)
O2 sensor input Exclusive cable Analog output 4-20mA DC or 0-1V DC Fault contact output contact capacity 250V AC/3A 30V DC/3A Blow contact output 250V AC/3A 30V DC/3A O2 sensor heater power output O2 sensor thermocouple input
1
2 O2
3
4 TC1
5
+
−
+
−
+
or
AO
6
7
8
9 BLOW
10
11 ZV
ZERO GAS
12
13
14
−
FAULT ALARM MAINTE
HEATER
SPAN GAS
Ground terminal M4
+
15
TC2
−
16
DI1 17
DI2 18
DI3 DICOM 19 20 21 22 23
COM 24 25
SV 26
N 27
AC
Thermocouple input type R Contact input (combustion control) Short circuit : on (Current 5mA) Open : off (Voltage 5V)
Alarm contact output 250V AC/3A 30V DC/3A
Maintenance contact output 250V AC/3A 30V DC/3A
Calibrating gas contact output contact capacity 250V AC/3A 30V DC/3A
AC Power supply input (100 to 240V AC)
COMMUNICATION TERMINAL (TM2) /INSERTION TERMINAL Terminal number 1 RS232C RS485 TXD TRX+ 2 RXD TRX– 3 GND GND Remarks Standard Option
182
L 28
Fuji Electric Co.,Ltd.
9
ZSB
Detector: ZFK8
Approx.132
φ 80 φ 67
6-φ 6
Approx.62 5
Approx.130 Temp.Proof cover (To order)
Terminal box
EXTERNAL CONNECTION DIAGRAM
φ 79 φ 26 φ 85
Heater Thermocouple Element output
Filter Approx.30
125.5
(?17.5)
20
TK4J4664R0
made in japan
40
5 S-
H
2
4 T-
10
45°
° 60
1 3 2 4 Black / White/ red / White/ 2-core wire
5 6 yellow / blue
4-core wire
Ground-wire Screw: M4
Exclusive-special cable
Calibration gas inlet Ref. Air inlet (to order) (To order) SUS316,for φ 6/φ 4 tube SUS316,for φ 6/φ 4 tube
Exclusive Cable gland
Replacement Detector element
Sensor unit (ZFK8YY)
(132.7)
(88.4)
3.5
4-φ 3.5(Mount for detector)
φ 35
45°
S+ H 6 1
T+
3
Flow guide tube (for general-use)
12 Approx. L φ155 Approx. 20 φ130 φ67 Gas inlet φ34
ZSB
2-5A 3 5 7 1
Gas outlet
6-M5 detector side
Code 11th Oxygen detector L (m)
3 0.3
5
7
1
Z L= (to order) 8-φ15 MTG. holes
0.5 0.75 1.0 3.3 4.1 4.8
MASS 2.7 Approx (kg)
Flow guide tube (for corrosive gas)
φ155 12 3 Approx. L Approx. 40 φ130 φ67 Gas inlet φ67 φ60.5
20
ZSB
2-5B 3 5 7 1
Gas outlet
6-M5 detctor side
Code 11th Oxygen detector L (m)
3 0.3
5
7
1
Z L= (to order) 8-φ15 MTG. holes
0.5 0.75 1.0 4.5 6.1 7.6
MASS 3.3 Approx (kg)
11
ZSB
Flow guide tube (for high particulate)
4-Rc 1/4 with plug
φ180
φ1
45
Blow down air inlet At installation, select upper blow port at one position (as a measure for drain). 8-φ19 MTG. hole
φ67
φ29
Code 11th
3 0.3
5
7
1
Z L= (to order)
.5
L (m)
0.5 0.75 1.0 5.6 7.0 8.3
Flange: JIS 5K80A FF ZSB MTG. position
MASS 4.5 Approx (kg)
Packing 185
ZSB
Oblong hole 50
Gas outlet
Approx. L
2-6D 3 5 7 1
30
Gas flow Applicable companion flange height about 160 mm (Closing gas outlet should be avoided.) Tube (50A SCH40)
Caution on Safety *Before using this product, be sure to read its instruction manual in advance.
International Sales Div Sales Group Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fujielectric.com Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425 http://www.fjielectric.com/products/instruments/ Information in this catalog is subject to change without notice.