ULTRASONIC FLOWMETER (2-Path System type)
DATA SHEET
This flowmeter is a fixed type ultrasonic flowmeter based on transit-time measuring method. The flowmeter is a compact and light-weight instrument incorporating the latest electronics and high speed digital signal processing technologies, realizing high performance and easy operation. This high performance flowmeter is used for 2-path system only.
FSH, FSG, FSD
FEATURES
1. Compact and light-weight Thanks to the adoption of the latest electronics the flow transmitter size and mass are 1/2 of our traditional instrument. 2. Full variety of sensors The flowmeter can be used with various types of sensors applicable for wide range of pipe size (ø13 to ø6000mm) and fluid temperature (-40 to +200°C). 3. High accuracy The flowmeter is designed for high accuracy (± 1.0%). The adoption of new sound velocity measurement system permits measurements of fluids of unknown sound velocity, and also slightly affection from fluid temperature and pressure. 4. Excellent resistance against aerated flow Fuji's unique ABM feature improves measurement reliability for different flow like slurries, sludge, raw sewage and bubble-contained flow (acceptable up to air bubble of 12% volume at 1m/s velocity). 5. Quick response With the use of high-speed micro-processor suited for digital signal processing, the fast response time is realized. 6. Multi-lingual The following languages are supported for display: Japanese (Katakana), English, German French, and Spanish. 7. Excellent performance and easy operation LCD and function keys are allowing easy configuration and trouble shooting. − LCD with back light − Easy mounting of sensor − Trouble shooting Flow transmitter (FSH)
Detector (FSD)
Detector (FSG)
SPECIFICATIONS
Operational specifications
System configuration: The system is composed of two detectors (Model: FSG, FSD) and one Flow transmitter (Model: FSH), realizing two-path measurement.
Applicable fluid:Homogenous liquid where the ultrasonic signal can be transmitted Bubble quantity: 0 to 12vol% (for pipe size 50A, water, velocity 1m/s) Fluid turbidity: 10000mg/L max. Type of flow: Fully-developed turbulent or laminar flow in a full-filled pipe Applicable flow pipe: Material: P lastics (PVC, FRP, etc.) or Metals (carbon steel, SS, copper, aluminum, etc.) 50 to 6000 mm Pipe size: (inside diameter) Liner: Tar epoxy, mortar, etc. Straight pipe length: Typically 10D for upstream and 5D for downstream. Refer to JEMIS-032 in detail. Velocity: 0 to ±0.3 ········±32 m/s Power supply: 100 to 240 Vac+10%/-15%, 50/60Hz or 20 to 30 Vdc
Date
EDS6-141d
Apr. 1, 2011
FSH, FSG, FSD
Signal cable: High-frequency coaxial cable and 3-wire shielded cable. Maximum cable length: 150m Allowable temperature limit: 80ºC Coaxial cable (4 pcs) Environment: N on-explosive environment without direct sunlight, corrosive gas and heat radiation Ambient temperature: -10 to +50ºC for flow transmitter, -20 to +80ºC for detector Ambient humidity: 95%RH or less for flow transmitter, 100%RH or less for detector Grounding: Class D (less than 100 ohm) Arrester: Surge absorbers for outputs and power supply incorporated as standard Display device: Graphic LCD (number of pixels: 240x64) with back light Display language: Japanese, English, French, German or Spanish selectable Velocity/Flow rate display: Display of velocity and/or flow rate with flow direction Data: up to 10 digits (decimal point to be counted as 1 digit) Unit: Metric/English system selectable
Metric system Velocity Flow rate m/s L/s, L/min, L/h, L/d, kL/h, ML/d, m3/s, m3/min, m3/h, m3/d, km3/d, Mm3/d, BBL/s, BBL/min, BBL/h, kBBL/d, MBBL/d English system ft/s ft3/s, ft3/min, ft3/h, Mft3/d, gal/s, gal/min, gal/h, Mgal/d, BBL/s, BBL/min, BBL/h, BBL/d, kBBL/d, MBBL/d
Performance specifications
Accuracy :
Detector Common type Pipe size (diameter) Applicable pipe material Flow velocity Accuracy ±1.0% of rate ±0.02m/s ±1.0% of rate ±0.0075m/s ±1.5% of rate ±0.03m/s ±1.5% of rate ±0.0113m/s
Note: ˝gal˝ means US gal. Total display: Display of forward or reverse total Data: u p to 10 digits (decimal point to be counted as 1 digit) Unit: Metric/English system selectable
Metric system Total mL, L, m3, km3, Mm3, mBBL, BBL, kBBL English system ft3, kft3, Mft3, gal, kgal, mBBL, BBL, kBBL, ACRf
2 to 32m/s 0 to 2m/s 0.75 to 32m/s 0 to 0.75m/s Plastic, metal pipe 2 to 32m/s ø50 to below ø300 0 to 2m/s FSGS31 FSGS50 0.75 to 32m/s ø300 to ø6000 mm 0 to 0.75m/s FSGS32 ø50 to below ø300 FSD32 FSGS51 ø300 to ø6000 mm
Response time: Transit time method: Power consumption: 20W or less 0.5sec
Configuration: Fully configurable on keyboard by menudriven software Zero adjustment: Set zero/Clear available. Damping: 0 to 100s (every 0.1s) configurable for analog output and display Low flow cut off: 0 to 5 m/s configurable Alarm: Hardware fault/process fault can be tied to digital output Burnout: Analog output: Hold/Upper limit/Lower limit/ Zero/Not-used selectable Total: Hold/Count selectable Timer: 0 to 900s (every 1s) configurable Bi-directional range: Forward and reverse ranges configurable independently Hysteresis: 0 to 20% of working range configurable Working range applicable to digital output Auto-2 ranges: Forward 2 ranges configurable independently Hysteresis: 0 to 20% of working range configurable Working range applicable to digital output Flow switch: Lower and upper switching points configurable independently Acting point applicable to digital output Total switch: +total switching point configurable Acting point applicable to digital output
Functional specifications
Analog output: 4 to 20 mAdc (1 point) Max. load resistance: 1k ohm Digital output: + total, -total, alarm, acting range, flow switch or total switch -- arbitrarily selectable Mechanical relay contact: 1 point with socket (replaceable) Normally closed/open selectable Capacity:240 Vac/30 Vdc, 1 A Total pulse: less than 1 p/s (Pulse width: 50, 100 or 200 ms selectable) Transistor open collector: 2 points Capacity: 30 Vdc, 0.1 A Normally off/on selectable Total pulse: less than 1000 p/s (Pulse width: 0.5, 1, 2, 5, 10, 20, 50, 100 or 200 ms selectable) Communication interface: RS-232C equivalent / RS-485 (selectable) Number of connectable units: one (RS-232C)/ up to 31 (RS-485) Baud rate: 9600/19200/38400 bps selectable Parity: none/odd/even selectable Stop bit: 1 or 2 bits selectable Distance: up to 15 m (RS-232C)/up to 1k m (RS-485) Data: velocity, flow rate, +total, -total, status (standard), velocity profile (option)
2
Physical specifications
Enclosure protection: Flow Transmitter: IP67, Detector: FSG (common type): IP67 (Silicon compound is filled on the terminal part when wiring) FSG (submersible type): IP68 (submersible in water for 5 days) FSD (high temperature type): IP52 Flow Transmitter: wall mount Mounting: Detector: clamped on pipe surface Acoustic coupler: Silicon compound (RTV) Flow Transmitter: aluminum alloy Material: Detector:
Detector Sensor housing Sensor cover Guide rail FSGS3 FSGS5 FSD32 PBT PBT SUS304 SUS304 SUS304 – SUS304 + plastic – SUS304 + aluminum alloy
Signal cable:
Dimensions:
FLY8, FLY9 (applicable detector: FSG, FSD) • Structure: High frequency coaxial cable (double shield) • Sheath: Black flame-resisting PVC • Outer diameter: ø7.5mm • Termination: M3.5 amp terminal (flow transmitter side) and M4 amp terminal (FLY8).Note, however, that the detecter side of FSD32 is provided with BNC connector (FLY9). • Mass: Approx. 90g/m Flow transmitter FSH: H240×W247×D134mm Detector: H46×W410×D50mm (FSGS3) H67×W78×D84mm (FSGS5) H205×W530×D52mm (FSD32) Flow transmitter Detector: 0.6kg 1.2kg 1.6kg : 5kg (FSGS3) (FSGS5) (FSD32)
Mass:
Conditions on straight pipe
Classification
10D or more
( D : Inside diameter of pipe) Downstream side L L 10D 5D
Upstream side
90 bend
Detector L L 50D 10D
Tee 10D or more L Diffuser
0.5D or more D
10D or more
30D
L
5D
1.5D L Reducer 10D L 5D
L Various Valve
30D
L
10D
In case that flow control valve exists on upstream side. Stop valve Check valve Pump
P
In case that flow control valve exists on downstream side.
L
50D
(Note) The source : JEMIS-032
3
FSH, FSG, FSD
MEASUREMENT PRINCIPLE
Ultrasonic pulses are propagated aslant between the upstream and downstream sensors, detecting the time difference due to flow for measurement.
Upstream sensor
MOUNTING OF DETECTOR
Ultrasonic oscillator Plastic wedge Pipe
Lining Output signal Flow transmitter
t1 Velocity V
t2
D
Downstream sensor
CONFIGURATION DIAGRAM
(1) 2-path system (V method)
Cable Detector Flow transmitter 4-20mA DC Contact (up to 3 points) Detector
(2) 2-path system (Z method)
Cable Flow transmitter 4-20mA DC Contact (up to 3 points)
DETECTOR SELECTION GUIDE
Type FSGS32 Note) FSGS31 –40 to 80 FSGS51 Note) FSGS50 Fluid Mounting temperature [°C] method V V Z –40 to 200 V Z Classification of piping materials
50
Inner diameter of piping ø (mm)
13 25 50 100 200 250 300 300 400 1000 3000 6000
50
Px, P M ,
200 200
Px, P M , Px, P M ,
250
3000 6000
Px, P M ,
150
FSD32
Px, P M ,
400
Px : PP PVDF , P : Plastic (PVC, etc.) M : Msetallic piping (steel pipe, copper pipe, aluminum, etc.)
Note: The ultrasonic signal cannot be transmitted easily when the classification of piping material is Px or the turbidity is high. In such a case, a preliminary check by a portable ultrasonic flowmeter is recommended.
4
CODE SYMBOLS
12345678 9 10 11 12
12345678 9 10 11 12 13
F SGS
A1
3 3 5 5 2 1 1 0 B C D E F G H J K L M N P Q R Z
1
Description
Type (5th and 6th digits) Small sensor 2MHz (ø50 to ø300) V method Small sensor 1MHz (ø50 to ø300)*2 Large sensor 1MHz (ø200 to ø6000) V or Z Large sensor 0.5KHz (ø200 to ø6000)*2 method Dedicated signal cable (9th digit) 10m 20m 30m 40m 50m 60m 70m 80m 90m 100m 110m 120m 130m 140m 150m Specified length (Contact us if length is more than 150m. Max. length is 300m.) Acoustic coupler (10th digit) Silicon rubber (KE348) Silicone grease (G40M) *5 Additional specification (11th digit) None Tag plate Wire rope for mounting (12th digit) Specify it in the case of FSGS41 or FSGS5. None Nominal diameter: up to ø500mm Nominal diameter: up to ø1000mm Nominal diameter: up to ø1500mm Nominal diameter: up to ø3000mm Can be specified Nominal diameter: up to ø6000mm only for FSGS5
F SH
E
YA
1
SYY
Description
Type (4th digit) Standard
A 1 4 Y
Use (6th digit) Transit time measurement (two path) Power supply (7th digit) 100 to 240 Vac, 50/60 Hz 20 to 30 Vdc Conduit connection (10th digit) G1/2 and G3/8 (female screw) with water-proof connection Y A B C Parameter setting, Tag Plate (12th digit) None With setting With setting and Tag Plate With Tag Plate
12345678
FSD
320
1
Description
Type (4th, 5th and 6th digits) High-temperature sensor *1 (ø50 to ø400), V or Z method
A C Y A
Y
Belt, acoustic coupler (7th digit) *3 For fixed type
12345678 9 10 11 12 13
Y A B C D E
F SGS
3 3 5 5 2 1 1 0
Y1
Y
1
Description
Type (5th and 6th digits) Small sensor 2MHz (ø50 to ø300) V method Small sensor 1MHz (ø50 to ø300)*2 V or Z Large sensor 1MHz (ø200 to ø6000) Large sensor 0.5KHz (ø200 to ø6000)*2 method
Y A B C Y A
Acoustic coupler (10th digit) None*4 Silicon rubber (KE348W) Silicone-free grease (HIGH-Z) *5 Silicone grease (G40M)*5 Additional specification (11th digit) None Tag plate Wire rope for mounting (12th digit) Specify it in the case of FSGS41 or FSGS5. None Nominal diameter: up to ø500mm Nominal diameter: up to ø1000mm Nominal diameter: up to ø1500mm Nominal diameter: up to ø3000mm Can be specified Nominal diameter: up to ø6000mm only for FSGS5
*1: For turbid fluid or old pipe, cast iron pipe, mortar lining pipe or others through which the ultrasonic signal could not be transmitted easily, use an optional guide rail (TK4C6164C1), and carry out mounting by Z method. Applicable diameter range V method: ø50 to ø250 Z method: ø150 to ø400 *2: For aging pipes, cast iron pipes or mortar-lined pipes that interrupts the propagation of ultrasonic signals, select FSGS31 or FSGS50. *3: As for the acoustic coupler, grease (KS62M) for high temperatures is attached. *4: Silicone rubber (KE-348W) is provided as a standard accessory to fill the wiring mold. (It can also be used as an acoustic coupler.) If an additional acoustic coupler is required, select one among A, B and C. *5: Normally select silicone rubber as acoustic coupler. Siliconerubber in tube (100g) is furnished. If you place an order for several units, 1 tube may suffice for every 2 units. Select silicone-free grease for semiconductor anufacturing equipment or the like that is vulnerable to silicone. The silicone-free grease is water-soluble and, therefore, cannot be used in environment exposed to water or on piping subjected to a condensation. Since the grease does not set, a periodic maintenance (cleaning, refilling every about 6 months at normal temperature) is necessary.
Y A B C D E
5
FSH, FSG, FSD
Note: Must be procured unless the sensor is a submersible type. • For detector FSG and FSD
12345678
FLY
8 9 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 1 1 1 1 1 1 Z 0 1 1 2 2 3 3 4 4 5 5 6 6 7 7 8 8 9 9 0 1 2 3 4 5 Z 5 0 5 0 5 0 5 0 5 0 5 0 5 0 5 0 5 0 5 0 0 0 0 0 0 Z
1
Description
Type of sensor (4th digit) Small and large sensor (for FSG) Hight temp sensor (for FSD) Cable length (5,6 and 7th digit) 5m 10 m 15 m 20 m 25 m 30 m 35 m 40 m 45 m 50 m 55 m 60 m 65 m 70 m 75 m 80 m 85 m 90 m 95 m 100 m 110 m 120 m 130 m 140 m 150 m Others (contact us)
Note: Preapare a detector and dedicated signal cable (2 set). Loader software for PCs Equipped as standard • Works on PC/AT compatible machines. • O peration on PC98-series machines (NEC) cannot be guaranteed. • Operation on self-made PCs or shop-brand PCs cannot be guaranteed. • Major functions: Setting/changing of various parameters for the main unit • O/S: Windows2000/XP • Memory requirement: 128MB or more • Disk unit: Windows2000/XP-compatible CD-ROM drive • Hard disk drive capacity: Free space of 52MB or more Note: PC loader communication cable (type ZZP*TK4H6253, Specifications: D-sub 9 pin receptacle, cable length 3m) is separately required.
SCOPE OF DELIVERY
•Flow transmitter (Type: FSH): Flow transmitter CD-ROM (Instruction manual, Loader software) •Detector (Type: FSG, FSD): S ensor unit, Mounting belt, Silicon compound •Signal cable (Type: FLY): Cable (2 wires)
ITEMS DESIGNATED ORDERING
1. Detector code symbols 2. Flow transmitter code symbols 3. Signal cable code symbols 4. For large sensor: Mounting pipe size 5. Tag No. as necessary 6. If parameter setting is specified, send back the attached parameter specification table duly filled.
6
OUTLINE DIAGRAM (Unit:mm)
4-Ø 9
Mtg. plate
240
Duosonics
3
134
72 14 233
5-G3/8 Female screw
33 33
30 30 30 48 48
24 48
4-G1/2 Female screw
CONNECTION DIAGRAM
Power supply terminals (AC) 1 2 3 4 L N 100 to 240V AC 50/60Hz Power supply terminals (DC) 4 1 2 3 1 B 2 B 3 A 4 5 SERIAL SERIAL No. 4 5 6 RS485 TRX2 (–) TRX1 (+) SHILD RS232C TXD RXD COM Communication cable (option) Cap (red) Cap (yellow) Cap (black) 6 7 8 9
(273)
352
332
30
+
Iout
–
+
–
20 to 30V DC
1 HF
2 GND
3 HF
4 GND
5 HF
6 GND
7 HF
8 GND
+
DO1
1
–
2
+
DO2
3
–
4
+
DO3
5
–
6
7
UP STR
DOWN STR
UP STR
DOWN STR
LINE1
LINE2
Open collector
Relay contact
7
FSH, FSG, FSD
OUTLINE DIAGRAM (Unit:mm)
Scale(inch)
Connector
Saddle
Scale(mm)
Name plate Distance between two sensors 0 to 330
Lock nut Element holder Cursor
φ26 90max. 205max.
33±0.5
44±1
Detector FSD32 (High-temperature sensor)
Frame 5 P
Name plate 47 Spacing : 0 to 270mm
410±2 400
Sensor 5
5 P
Frame
530±2
410±2 400
Sensor 5
Name plate 50±1 33
47
Spacing : 0 to 270mm
Chain Mounting spring
50±1 33
Chain Mounting spring
46±1
Frame guide (10)
(10)
VIEW FROM P
46±1
VIEW FROM P
Frame guide
(Common type)
(Submersible type) Detector FSGS3 (Small sensor)
Name plate
Name plate
84±1
84±1
114
114
78±1 Sensor
Wire rope Mounting spring 67±1
78±1 Sensor
L(cable length (m) : specified by order) Wire rope Mounting spring 67±1
(Common type)
8
(Submersible type) Detector FSGS5 (Large sensor)
Ø19
ø19
L(cable length (m) : specified by order)
Signal cable : FLY8 (For FSG)
8
3-ø4.3
To Detector ø11 L(cable length (m) : specified by order)
30
ø7.3
To Flow transmitter
3-ø3.7 L : By specification when ordered. 5.5
Signal cable : FLY9 (For FSD)
ø14.5 BNC connector 46
Loader cable : ZZP*TK4H6253
34 Connector
ø11 L(cable length (m) : specified by order) ø7.3
30
67
To Detector
To Personal computer
3.2±0.2 (m)
ø6.9
To Flow transmitter
To Flow transmitter Terminal cap (Red) 3ø3 .7 Terminal cap (Yellow) Terminal cap (Black) Red TXD 6.6 Yellow RXD RS-232C 9 Black COM
3-ø3.7 5.5
L : By specification when ordered.
FSH, FSG, FSD
No. 1 2 Setting item Outer diameter Pipe material Settable range 10.00 to 6200.00mm (0.393 to 244.100 inch) 12 menus Pipe S.V. : 1000 to 3700m/s (3280 to 12140 ft/s) Initial value 60.00mm (2.362 inch) PVC Settable value [mm, inch] Carbon steel, Stainless steel, PVC, Copper, Castiron, Aluminum, FRP Ductileiron, PEEK, , PVDF, Acrylic Others (Sound velocity : [m/s, ft/s]) [mm, inch] No lining, Tar epoxy, Mortar, Rubber, Teflon, Pyrex glass, PVC, Others (Sound velocity : [m/s, ft/s]) [mm, inch] Water Water, Seawater, DIST. water, Ammonia, Alcohol, Benzene, Bromide, Ethanol, Glycol, Kerosene, Milk, Methanol, Toluol, Lube oil, Fuel oil, Petrol, Others (Sound velocity : [m/s, ft/s]) (Kinematic viscosity [x10-6m2/s, ft2/s]) m/s, L/s, L/min, L/h, L/d, kL/d, ML/d, m3/s, m3/min, m3/h, m3/d, km3/d, Mm3/d, BBL/s, BBL/min, BBL/h, BBL/d, KBBL/d, MBBL/d, (ft/s, ft3/s, ft3/min, ft3/h, ft3/d, kft3/d, Mft3/d, gal/s, gal/min, gal/h, gal/d, kgal/d, Mgal/d, BBL/s, BBL/min, BBL/h, BBL/d, kBBL/d, MBBL/d) Single, Auto 2, Bi-dir, Bi-dir Auto 2 [ ]
PiPing specification
3 4
Wall thickness Lining material
0.10 to 100.00mm (0.003 to 3.940 inch) 8 menus Lining S.V. : 1000 to 3700m/s (3280 to 12140 ft/s)
4.00mm (0.157 inch) No lining
5 6
Lining thickness Kind of Fluid
0.01 to 100.00mm (0.000 to 3.940 inch) 17 menus
Fluid S.V. : 500 to 2500m/s (1641 to 8203 ft/s) Kinematic viscosity : 0.001 to 999.9999 x 10-6m2/s (0.0107 to 10763.9088 x 10-6ft2/s) 7 Range unit 19 menus m/s (ft/s)
8 9
Range type Full scale or Full scale 1
4 menus In terms of flow velocity 0.00 ··· 0.30 to 32.00m/s (0.98 to 104.98 ft/s) In terms of flow velocity 0.00 ··· 0.30 to 32.00m/s (0.98 to 104.98 ft/s) 0.00 to 20.0% -20 to 0% 100 to 120% 5 menus 0 to 900sec 0.00 to 5.00m/s (0.00 to 16.40 ft/s) in terms of flow velocity 0 to 900sec 0.0 to 100.0sec 7 menus
Single 2.00m/s (6.56 ft/s) 4.00m/s (13.12 ft/s) 10.00% -20% 120% Hold 10sec 0.00m/s (0.00 ft/s) 0sec 5.0sec Flowrate (m3/s)
Output setting
10
Full scale 2
[
]
11 12 13 14 15 16
Range HYS. Output limit LO. Output limit HI. Output burnout Burnout timer Rate limit
% % % Not use, Hold, Upper, Lower, Zero sec [ ]
17 18 19
Rate limit timer Damping
sec sec Flow velocity, Flowrate, Total forward, Total reverse, F : Total pulse, R : Total pulse, Flow rate (%)
20
Display setting
1 : Display kind
2 : Display kind
7 menus
Flow velocity (m/s) Flow velocity, Flowrate, Total forward, Total reverse, F : Total pulse, R : Total pulse, Flow rate (%) 0.01m/s (0.03 ft/s) [ ]
21
Low flow cut
0.00 to 5.00m/s (0.00 to 16.40 ft/s) in terms of flow velocity
10
No. 22 23
Setting item Total mode Total unit 3 menus 8 menus
Settable range
Initial value Total stop mL
Settable value Total stop, Total run, Total reset mL, L, m3, km3, Mm3, mBBL, BBL, kBBL, ft3, kft3, Mft3, kgal, gal, mBBL, BBL, kBBL, ACRF [ [ [ [ [ Not use, Hold sec 50, 100, 200 0.5, 1.0, 2.0, 5.0, 10.0, 20.0, 50.0, 100.0, 200.0 [ ] ] ] ] ] ]
24 25
Total rate F : Total preset
0.000 to 999999.999 0.000 to 9999999999.999 0.000 to 9999999999.999 0.000 to 9999999999.999 0.000 to 9999999999.999 2 menus 0 to 900sec 3 menus 9 menus In terms of flow velocity 0.00 to 32.00m/s (0.00 to 104.98 ft/s) In terms of flow velocity 0.00 to 32.00m/s (0.00 to 104.98 ft/s) 0 to 20% 15 menus
0.000 0.000 0.000 0.000 0.000 Hold 10sec 50ms 50ms 0.00m/s
Total Flow switch
26 27 28 29 30 31 32 33
F : Total SW R : Total preset R : Total SW Output burnout Burnout timer Pulse width 1 Pulse width 2 Flow sw high
34
Flow sw low
4.00m/s
[
]
35 36
Flow sw HYS. Output DO1
10% Not use Not use, Signal error, F : Total pulse, R : Total pulse, F : Total alarm, R : Total alarm, F : Total overflow, R : Total overflow, Flow SW high, Flow SW Low, Full scale2, AO range over, Pulse range over, R : Flow direction, Device error Normal, Reverse Not use, Signal error, F : Total pulse, R : Total pulse, F : Total alarm, R : Total alarm, F : Total overflow, R : Total overflow, Flow SW high, Flow SW Low, Full scale2, AO range over, Pulse range over, R : Flow direction, Device error Normal, Reverse Not use, Signal error, F : Total pulse, R : Total pulse, F : Total alarm, R : Total alarm, F : Total overflow, R : Total overflow, Flow SW high, Flow SW Low, Full scale2, AO range over, Pulse range over, R : Flow direction, Device error Normal, Reverse
%
37 38
Mode DO1 Output DO2
2 menus 15 menus
Normal Not use
39 40
Status output
Mode DO2 Output DO3
2 menus 15 menus
Normal Not use
41
Mode DO3
2 menus
Normal
11
FSH, FSG, FSD
No. 42 43 44
Serial com.
Setting item System unit Language COM. speed COM. parity COM. stop bit Serial method Station No.
Measuremant mode
Settable range 2 menus 5 menus 3 menus 3 menus 2 menus 2 menus 31 menus 2 menus 3 menus 4 menus
Initial value Metric English 38400BPS None 1 bit RS232C 1 1 Path (2 path) Metric, English
Settable value
Japanese, English, German, French, spanish 9600BPS, 19200BPS, 38400BPS None, Even, Odd 1 bit, 2bits RS232C, RS485 1 to 31 2 Path Note2 Average, Line 1, Line 2 FSW12, FSW21, FSW40, FSW50, FLW11/FSG31, FLW41/FSG41, FLW12/FSG32, FLD12/FSD12, FLD22/FSD22, FLW32/FSD32, FLS50/FSG50, FLW51/FSG51
45
System
46 47 48 49 50 51
Measurement mode AO Definition
Line 1 (Average) Note2 FSW12 (FLW51/FSG51) Note2
Sensor Type
Note1: When total pulse output has been selected for DO1, DO2 or DO3 specify total pulse value and total pulse width so that conditions 1 and 2 shown below are satisfies. Condition 1 : Flow span-1*[m3/s] total pulse value*[m3] Flow span-1*[m3/s] total pulse value*[m3] 1000 [In the case of DO1 and DO2] 1 [In the case of DO3] 1000 2 × total pulse width [ms]
Condition 2 :
* In the case of 2 ranges, perform calculations using either flow span-1 or flow span-2, whichever is greater. Note2: Settable value at the time of shipment.
Caution on Safety *Before using this product, be sure to read its instruction manual in advance.
International Sales Div Sales Group Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fujielectric.com Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425 http://www.fjielectric.com/products/instruments/ Information in this catalog is subject to change without notice.