ULTRASONIC LIQUID LEVEL METER
DATA SHEET
An ultrasonic liquid level meter measures the liquid level by the propagation time of the ultrasonic pulses emitted out of the ultrasonic detector are reflected on the surface of the measurement object and are received as echoes by the detector again. It provides high reliability and it permits easy installation, running and maintenance, as it does not make contact with the measurement object and it is free of mechanical moving parts. It deals with uses of a broad range including water level measurement of water, wastewater, tide level, rivers and reservoirs and level measurement of highly viscous liquid in chemical tanks, slurry agitation tanks, roll mills and food tanks.
FQH-2,FQG-2,FLY
FEATURES
1. No liquid contact Permits easy installation and maintenance. 2. Detector of waterproof structure and of excellent corrosion resistance It is of no problem against adhesion of water drops and splashes, and it is optimum for wastewater, alkali, acid and high viscosity liquid. 3. Not affected by fixed reflection noise Not affected by obstacles such as projections of detector mounting area, walls and ladders because of the DGL (dynamic gain limit) function that automatically learns the noise level at all times and masks noise so that no noise is detected. 4. Minimum influence of electrical noise Not affected by electrical noise or noise from agitating blades, as such noise is largely reduced by statistical processing. 5. Use of high gain amplifier Permits use at a pump well involving scum that could be hardly measured with conventional meters and also at a place of large attenuation such as sludge storage tank.
Detector (FQG)
Converter (FQH)
Display:
SPECIFICATIONS
Converter (Type: FQH)
Structure: Material grade: Coating color: Accuracy: Output: IP65 Aluminum alloy Blue (cover), silver (case) ± 1%/FS or better 4 to 20mA DC (insulated, load resistance 0 to 1kΩ)
LCD display (16 characters, 2 lines, with back light) Level measured value (m, %), detector peripheral temperature, ultrasonic wave received signal intensity, etc. are displayed. Dumping: 2 to 600 seconds (once every second) Power failure recovery processing: Memory backup with lithium cell (valid for about 10 years) Power supply: 100 to 240V AC, 50/60Hz or 20 to 30A DC Power consumption: 35VA or less (AC power supply) 12W or less (DC power supply) Contact output: Upper limit and lower limit alarms, error alarms, up to 8 points (optional), relay contact output, contact capacity AC 250V, 5A, normally open contact Working ambient temperature: -10 to50˚C Working ambient humidity: 90% RH or less External dimensions: H320 X W240 X D137mm Mounting method: Wall surface mounting Mass: About 7 kg Lightning measure: Power supply, output signals and alarm outputs are provided with arresters Others: Provided with self-diagnostic function
Detector (Type: FQG)
Structure: Composition: IP67 With built-in preamplifier, transmitter and
EDS7-64
Date Mar. 10, 2003
FQH-2,FQG-2,FLY
resistance thermometer bulb Measuring distance: Up to 12 m (blind distance: 0.3 m) Material grade: Heat resistant PVC (color: gray) Matching flange: JIS 5K 150A or larger External dimensions: ø140 X 180mm Mass: About 1.2 kg Lightening measure: With arrester Working ambient temperature: -20 to 80˚C Working ambient humidity: 95% RH or less
Dedicated cable (between converter and detector) (Type: FLY5)
Twisted paired shield cable 2 pairs x 0.5 mm2 shield (length up to 300 m) Outside diameter: ø10.5mm (Max ø11mm) Cable end treatment: With M3.5 crimp-style terminal (at both ends)
CONFIGURATION
Converter
Dedicated cable (up to 300m) Detector Blind distance (0.3m) Zero-point distance (up to 12m) Full level output 20mA DC Span (up to 11.7m) Zero level output 4mA DC
Output signal (DC4 to 20mA) Error alarm Upper limit, lower limit alarms Optional
Power supply (100 to 240V AC or 24V DC) *1 Error alarm An alarm is output on occurrence of set numeric value data error, cable open circuit or reception error. (Also indicated in LCD display screen.)
USES
Level measurement of waterworks, wastewater, rivers, dams and reservoirs and tide level Level measurement of food, high viscosity fluid and chemical storage tanks Level measurement of dehydrated slurry and sludge tanks Level measurement of agitation tanks such as slurry agitation tanks Level measurement of high viscosity fluid in tanks such as roll mills
Typical detector installation
* Detectors are of watertight structure. Avoid exposure to direct sunlight and rain by means of a roof or the like, when they are used outdoors. [Mounting with flange] [Mounting with square flange]
400 mm min.
[Mounting on wave guide]
200 mm max.
ø200
200 mm max.
150A flange or larger
ø200 or more
Sound absorbing cartridge Wave guide pipe of ø200 or larger
* Side wall must be free from standout and sticking material
* Inquire us separately regarding wave-guide pipe and sound absorbing cartridge.
CODE SYMBOLS
12345678 9 10 11 12
Specification Application Level meter for liquid/slurry
FQH 1 F
0
2-
1 2 3 4 5 6 7 8 9 10 F QG 02 A 1 1 Y T 0 1 2
Specification Structure Heat-resisting PVC immersionproof Measurable distance (frequency) 12 m (30 kHz) max. Resistance bulb Incorporated Tag indication No Yes Detector fixture None L-shaped fixture, mounted on 1 inch pipe L-shaped fixture, suspended from flat plate
Structure Immersion-proof 3 7 1 0 5 8 Y A Y T Power source 100 to 240 V AC, 50/60 Hz 20 to 30 V DC Current output (4 to 20 mA DC) 1 point Contact closure command None 5 points 8 points Wiring port With waterproof gland With union (for PRICA) gland Tag indication No Yes
12345678 FLY5 1 5 005 Specification Applicable detector Ultrasonic level meter detector Detector-converter cable length 5m 5 m steps up to 100 m. 10 m steps for 100 to 300 m. (Ex.: 150 for 150 m.) 300 m
300
OUTLINE DIAGRAM (Unit : mm)
Converter (Type: FQH)
74 2-ø12 74 2-ø12
Detector (Type: FQG)
4-ø7 mounting holes TAG nameplate Main nameplate
420* 320
240
450
Ball chain (SUS305)
120 ø90
ø1 24
Main nameplate (SUS304)
70
95
3
134
12
12 12 12 . If provided with gland for PRICA tube (JIS C8309, No. 17)
150 50
25
ø140
28
(Unit: mm)
TAG nameplate (SUS304 except where noted)
25
100
(Unit: mm)
180
. If provided with waterproof gland . * Adjustable to 450 mm
Cable lead-in port (nylon 66) G1/2
FQH-2,FQG-2,FLY
CONNECTION DIAGRAM
Converter (FQH)
L AC LINE N
Precautions against obstacles
DC power connection
1 DC 2 20 to 30 V DC
AC power connection
Source 100 to 240 V AC, 50/60 Hz
(1)
(2)
Mount the detector 400mm or more away from the side wall Get at least θ of 8° away from shelf, step or standout, if any
Dedicated cable (FLY 5) (Twisted pair shield cable)
θ
A R B Blue White/blue Yellow White/yellow Blue White/blue Yellow White/yellow A B C D E Detector (FQG)
SENSOR
C T D SH E
400mm or more
Shelf
OUT1
CURRENT
OUT2
OUT3
+ + + A1
Current output (4 to 20 mA DC)
(Terminal screw M3.5)
AL1 B1 A2 AL2 B2 A3 AL3 B3 A4 AL4
FAIL alarm Relay output 1 (device error alarm) Relay output 2 High (low) limit alarm
Detector fixture (option)
L-shaped fixture made of SUS304 On 1 inch pipe or suspended from flat plate e 90 Cable lead-in Fixture 90 170 f [Mounting size] 4-ø7, or M6 Tap
øg
Relay output 3 High (low) limit alarm
Relay output 4 High (low) limit alarm B4 A5
ALARM (option)
AL5 B5 A6 AL6 B6 A7 AL7 B7 A8 AL8 B8
Relay output 5 High (low) limit alarm
100 42
4-ø7
Relay output 6 High (low) limit alarm
Relay output 7 High (low) limit alarm
Relay output 8 High (low) limit alarm
(Terminal screw M3.5)
Scope of delivery
Converter main unit, detector main unit, dedicated cable (of specified length)
What are specified at the time of placement of an order
1. Detector type FQG_____________ 2. Converter type FQH_____________ 3. Dedicated cable FLY 4. Span distance __________m 5. Zero-point distance __________m 6. Tag No. designation ______________
øh
(Boring of flanges, etc.)
Caution on Safety *Before using this product, be sure to read its instruction manual in advance.