FC SERIES PT TYPE SIGNAL CONVERTER
DATA SHEET
The FC series PT type signal converter converts inputs from a thermocouple, resistance bulb, DC voltage and current into a signal of isolated 1 to 5V DC or 4 to 20mA DC. In addition, burnout protection in the event of thermocouple breakage, transmission function are avaiable. Its structure is plug-in style.
PTA
SPECIFICATIONS
INPUT SPECIFICATION
Input signal:
Input signal Thermocouple
1 point
Type B(*1) E J K N(*2) R S T Pt 100(*3) JPt 100(*3) mV mA POT Measuring range 0 to 1700°C –200 to 800°C 0 to 750°C –200 to 1200°C 0 to 1300°C 0 to 1600°C 0 to 1600°C –200 to 350°C –200 to 650°C –200 to 500°C 3 to 100mV 0.1 to 10mA 50 to 300Ω Min. span width 900°C 100°C 100°C 100°C 100°C 400°C 400°C 100°C 50°C 50°C 3mA 0.1mA 20Ω
Resistance bulb Voltage Current Potentiometer
Note: (*1) : With the B thermocouple, accuracy guarantee is in the range of 600 to 1700°C. (*2) : N : NICROSIL-NISIL (IEC584) (*3) : Pt 100 : JIS C 1604, 1606, IEC751 JPt 100 : JIS C 1604, 1606 (OLD JIS Pt100) Input type Voltage Thermocouple Current Input resistance Allowable wiring resistance 1MΩ or more 1kΩ or less 10 to 250Ω 10Ω or less Resistance Potentiometer bulb
Connectable units: 31 units (Max.) Code format: Data length……8 bits (binary) Parity bit……even, odd, none Stop bit……one or two bits Transmission distance: 1 km (Max.) 2. Loader interface RS-232C equivalent
POWER SUPPLY
Power supply: 24V DC (20 to 30V DC) 24V AC, +13%, –10%, (47 to 63Hz) 100V AC (85 to 132V AC /47 to 63Hz) 200V AC (187 to 264V AC /47 to 63Hz) as specified. Power consumption: Approx. 3W (at DC power) Approx. 6VA (at AC power)
* : Zero elevation is ± 25mV or less.
OPERATION CONDITION
Ambient temperature: 0 to 50°C Ambient humidity: Less than 90%RH (no condensation) Outline dimension (HxWxD): 96 x 52 x 130mm Mass: Appox. 300g Housing: Plastic housing (color: black) Mounting method: Panel mounting or DIN rail mounting
OUTPUT SPECIFICATION
Output signal: 2 points
Output type Output signal Voltage 1 to 5V DC Current 4 to 20mA DC 600Ω or less
Allowable load 15kΩ or over resistance
TRANSMISSION SPECIFICATION
1. Data transmission Interface: RS-485 Transmission system: Start-stop synchronous system Transmission speed: 2400, 4800, 9600, 19200bps
CHARACTERISTICS
Accuracy: Less than ± 0.1% of full span (in case of 10mV span or more)
EDS9-21c
Date Mar. 25, 1999
PTA
Less than ± 0.2% of full span (in case of less than 10mV span ) Less than ± 0.1% of full span or ± 0.1°C, whichever is larger. (in case of resistance bulb) Linearizing error: Less than ± 0.05% of full span Referece junction compensation accuracy: E, J, K, T ................. ± 1°C R, S, N ..................... ± 2°C Response time: 0.3 sec. Burnout time: 10 sec. or less Ambient temperature effect: Less than ± 0.1% of full span/10°C Power regulation effect: Less than ± 0.1% of full span Dielectric strength: 1500V AC, 1 min. (Input-Output-Transmission-PowerGround) Insulation resistance: 500V DC, 100MΩ or over (Input-Output-Transmission-PowerGround)
STANDARD INPUT RANGE
The input signal and measuring range below can be selected with a front switch. For others, specify separately.
Input signal B R S Measuring range 0 to 1700, 600 to 1700 0 to 1000, 0 to 1200, 0 to 1400, 0 to 1600, 400 to 1400, 600 to 1600, 800 to 1600 0 to 1000, 0 to 1200, 0 to 1400, 0 to 1600, 400 to 1400, 600 to 1600, 800 to 1600 –200 to 100, 0 to 300, 0 to 400, 0 to 500, 0 to 600, 0 to 800, 0 to 1000, 0 to 1200, 300 to 600, 400 to 800, 500 to 1000, 600 to 1200 –200 to 100, 0 to 200, 0 to 300, 0 to 400, 0 to 500, 0 to 600, 0 to 800, 200 to 400, 300 to 600 0 to 200, 0 to 300, 0 to 400, 0 to 500, 0 to 600, 200 tp 400, 300 to 600 –200 to 100, 0 to 300 0 to 300, 0 to 400, 0 to 500, 0 to 600, 0 to 800, 0 to 1000, 0 to 1200, 300 to 600, 400 to 800, 500 to 1000, 600 to 1200 0 to 50, 0 to 100, 0 to 150, 0 to 200, 0 to 300, 0 to 400, 0 to 500, 100 to 300, 200 to 400, 300 to 500, –20 to 30, –50 to 50, –50 to 100, –200 to 50, –200 to 100
K Thermocouple [°C] E J T N
CODE SYMBOLS
12345678 9 10 11
Pt 100 Resistance bulb [°C] Description Input signal Voltage (mV) Current (mA) Thermocouple Resistance bulb Potentiometer (3-wire system)
PTA 1 2 3 4 5
W
1-
R
0 to 50, 0 to 100, 0 to 150, 0 to 200, 0 to 300, 0 to 400, 0 to 500, 100 to 300, 200 to 400, 300 to 500, JPt 100 –20 to 30, –50 to 50, –50 to 100, –200 to 50, –200 to 100 [Ω] [mV] [mA] 10–100–10, 0–100–0, 10–135–10, 0–135–0 0 to 5, 0 to 10, 0 to 20, 0 to 30, 0 to 50, 0 to 100 4 to 20, 0 to 20, 0 to 10
Potentionmeter Voltage Current
W Y L H 1 2 3 Y A R Y
Output signal 1 to 5V DC/4 to 20mA DC Burnout Without Down-scale Up-scale Power supply 24V DC/24V AC (50/60Hz) 100V AC (50/60Hz) 200V AC (50/60Hz) Application For general For zener barrier connection Transmission function With RS-485 Loader interface Without
FUNCTIONAL DIAGRAM
Input Thermocouple Resistance bulb mV, mA Potentinmeter
I+ I– I0 I–
11 12 21 22 13 AC/DC conversion
Transmission circuit
5 6
+
RS-485
–
31 Calculation circuit DC/AC conversion 32 41 42
+
Output (DC 4~ 20mA) Output (DC 1~ 5v)
– + – +
Rcj
Reference junction compensator
14 23 24 Switch panel Loader interface Power supply circuit
43 44 34
Power supply
–
OUTLINE DIAGRAM (Unit: mm)
Terminal screw M4 Main unit Terminal screw M2.5 (for transmission)
74
94
35
Terminal
48 52 100 130
2
Stopper
Panel mounting
DIN rail mounting
2-M4
DIN rail
40 52
Stopper
PTA
EXTERNAL CONNECTION DIAGRAM
Thermocouple input
Rcj
Resistance bulb input
+ 11 12
13 14
11 12
13 14
21 22 + 5
RS-485
23 24
21 22 + 5
RS-485
23 24
+ 33 34
+ 33 34
6
6
31 32 + OUTPUT(mA)
31 32 + OUTPUT(mA)
41 42 + OUTPUT(V)
43 44
POWER
41 42 + OUTPUT(V)
43 44
POWER
Potentiometer Input
Voltage•Current Input
INPUT
11 12 21 22 + 5
RS-485
13 14
+ 11 12
13 14
23 24
21 22 + 5
RS-485
23 24
+ 33 34
+ 33 34
6
6
31 32 + OUTPUT(mA)
31 32 + OUTPUT(mA)
41 42 + OUTPUT(V)
43 44
POWER
41 42 + OUTPUT(V)
43 44
POWER
RANGE OF DELIVERY
Main unit and socket
ORDERING INFORMATION
1. Input signal specification 2. Measuring range
Caution on Safety *Before using this product, be sure to read its instruction manual in advance.
Head Office
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fesys.co.jp/eng
Instrumentation Div. International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187 http://www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
Printed in Japan