Metal package PMT with Cooler Photon Counting Head H7421 Series The H7421 series are photon counting head devices containing a metal package photomultiplier tube having a GaAsP/GaAs photocathode and a thermoelectric cooler. The thermoelectric cooler reduces thermal noise generated from the photocathode which also offers a high quantum efficiency, allowing measurement to be made with a good S/N ratio even at very low light levels. The H7421-40 has high sensitivity on wavelength from 300 nm to 720 nm. The H7421-50 is sensitive over a wide spectral range from 380 nm to 890 nm. The photomultiplier tube is maintained at a constant temperature by monitoring the output from a thermistor installed near the photomultiplier tube and regulating the current to the thermoelectric cooler. Heatsink with fan (A7423) sold separately Product Variations Type No. H7421-40 H7421-50 Spectral Response 300 nm to 720 nm 380 nm to 890 nm Features GaAsP photocathode, QE 40 % at peak wavelength GaAs photocathode, QE 12 % at peak wavelength Specifications Parameter Input Voltage Max. Input Voltage for Main Unit Max. Input Current for Main Unit Max. Input Voltage for Thermoelectric Cooler Max. Input Current for Thermoelectric Cooler Effective Area Peak Sensitivity Wavelength Count Sensitivity Count Linearity *1 Typ. Dark Count *2 *3 Max. Pulse-pair Resolution Output Pulse Width Typ. Output Pulse Height *4 Min. Recommended Load Resistance Signal Output Logic Operating Ambient Temperature Storage Temperature Weight H7421-40 +4.5 to +5.5 +6 50 2.6 2.2 5 580 7.8 × 105 1.5 × 106 100 300 70 30 3.6 3.0 50 Positive logic +5 to +35 -20 to +50 340 H7421-50 800 3.9 × 105 1.5 × 106 125 375 Unit V V mA V A mm nm -1·pW-1 s s-1 s-1 ns ns V Ω — °C °C g *1: Random pulse, at 10 % count loss *2: PMT setting temperature 0 °C, used with C8137, M9011 and A7432 *3: After 30 minute storage in darkness *4: With input voltage +5 V, Load resistance 50 Ω and Coaxial cable RG-174/U (450 mm) Cooling Specifications Parameter Cooling Method Max. Cooling Temperature (∆T) *5 Cooling Time *5 *5: Input current to thermoelectric cooler = 2.0 A H7421-40/H7421-50 Thermoelectric cooling 35 Approx. 5 Unit — °C min 40 Photon Counting Heads Characteristics (Count sensitivity, Dark count) 106 TPHOB0042ED H7421-40 H7421-50 1000 TPHOB0046EB COUNT SENSITIVITY (s-1·pW-1) DARK COUNT (s-1) 105 H7421-50 (when used with C8137, A7423) 100 H7421-40 (when used wih C8137, A7423) 104 103 10 5 10 15 20 25 30 35 40 45 AMBIENT TEMPERATURE (°C) 102 300 400 500 600 700 800 900 1000 WAVELENGTH (nm) Block Diagram POWER INPUT for THERMOELECTRIC COOLER THERMISTOR AMP COMPARATOR + THERMOELECTRIC COOLER PMT HV POWER SUPPLY VOLTAGE DIVIDER CIRCUIT – LLD. PULSE SHAPER OUTPUT TO PULSE COUNTER RL = 50 Ω POSITIVE LOGIC GND +5 V TPHOC0042ED Dimensional Outlines (Unit: mm) 8-M3 DEPTH: 2 POWER INPUT TAJIMI PRC03-23A10-7M ACROSS SECTION WINDOW PHOTOCATHODE 5 26.0 ± 0.2 22.2 ± 0.2 M25.4 P=1/32" C-MOUNT 10.2 7.2 PMT 18.5 ± 0.2 25.4 ± 0.2 53.6 ± 0.2 TOP VIEW SIGNAL OUTPUT BNC-R 15.0 ± 0.2 4.0 ± 0.1 6.4 ± 0.2 7.4 ± 0.2 14.8 ± 0.2 16.3 ± 0.2 O-RING GROOVE (S-28 O-RING INCLUDED) 4-M2 DEPTH: 2 4-M3B 36.0 ± 0.3 20.0 ± 0.2 20.0 ± 0.2 PHOTOCATHODE 8-M3C A 5 BCROSS SECTION CCROSS SECTION 6.5 ± 0.2 M3 DEPTH: 4.0 Max. 2.0 ± 0.2 M3 DEPTH: 4.0 Max. GUIDE MARK 16.3 ± 0.2 19.0 ± 0.2 25.0 ± 0.2 104 ± 1 SIDE VIEW F EG D C A B 9.5 ± 0.1 44.0 ± 0.2 56.0 ± 0.3 FRONT VIEW M25.4 P = 1/32" C-MOUNT A: THERMISTOR 1 B: THERMISTOR 2 C: THERMOELECTRIC COOLER + D: THERMOELECTRIC COOLER – E: Vcc LOW VOLTAGE INPUT(+5 V) F: IC* G: GND *IC: Internal Connection (Do not use) TPHOA0021EC TAJIMI PRC03-23A10-7M 41 Metal package PMT with Cooler Options for H7421 Series 1 4 OPTICAL FIBER OPTICAL FIBER ADAPTER A7412 HEATSINK with FAN A7423 2 SOLD SEPARATELY POWER SUPPLY UNIT with TEMPERATURE CONTROL M9011 6 CN01 CHK1 6 DIRECT INPUT POWER SUPPLY UNIT with TEMPERATURE CONTROL C8137 5 SAMPLE C-MOUNT LENS C-MOUNT ADAPTER A7413 PHOTONCOUNTING HEAD H7421 SERIES 3 ON ON SIGNAL CABLE E1168-05 COUNTER G Heatsink with Fan A7423 The temperature of the H7421 outer case rises due to the thermoelectric cooler housed in the case. The A7423 heatsink efficiently radiates away this heat to prevent a temperature rise in the H7421. The A7423 can be easily installed onto the H7421 with four M3 screws. Apply a coat of heat conductive grease onto the joint surface shared by the H7421 and A7423. Parameter Value Unit Input Voltage V 12 During Lock mA 140 Input Current mA 90 During Operation V 10.2 to 13.8 Operating Voltage g 120 Weight G Signal Cable E1168-05 This signal cable comes attached to a BNC connector for easily connecting the H7421 to external equipment. G Optical Fiber Adapter (FC Type) A7412 The A7412 is an FC type optical fiber connector that attaches to the light input window of the H7421. The A7412 can easily be secured in place with four M2 screws. G C-Mount Adapter A7413 The A7413 mount adapter is used when a C-mount lens protruding 4 mm or more from the flange-back must be installed onto the H7421. G Power Supply Unit with Temperature Control M9011 The M9011 is an on-board type power supply unit. By just connecting to 12 V supply, the M9011 provides power necessary to operate the H7421 series. The M9011 also controls the thermoelectric cooler in the H7421 series so that the output and noise can be maintained at constant levels even when the ambient temperature changes. The thermoelectric cooler and PMT operation can be controlled from an external device by connecting it to the I/O connector on the M9011. Parameter Value 35 Max. Cooling Temperature (∆T) 12 Input Voltage 1.2 Max. Input Current 15.8 Max. Power Consumption 5 Main Circuit Output Voltage 2.2 Max. Output Current for Thermoelectric Cooler 12 Output Voltage for Fan Thermoelectric Cooler Non-insulated TTL level input Control Signal PMT Non-insulated TTL level input Input Voltage Fan Non-insulated TTL level input Error Signal Thermoelectric Non-insulated TTL level output Output Voltage Cooler PMT 5 LED Output Error 5 0 Setting Cooling Temperature 120 Weight (excluding cables) 42 Unit °C V A V·A V A V — — V °C g G Power Supply Unit with Temperature Control C8137 The C8137 is a power supply unit with a temperature control function. Just connecting to an AC source of 100 V to 240 V generates the output voltages for the thermoelectric cooler and the A7423 fan, needed for operating the H7421. The photomultiplier tube temperature can be maintained to 0 °C by monitoring the thermistor and regulating the output current for the thermoelectric cooler. Parameter Max. Cooling Temperature (∆T) Setting Cooling Temperature (preset at factory) Input Voltage Input Voltage Frequency Power Consumption Main Circuit Output Voltage Max. Current for Thermoelectric Cooler Output Voltage for Fan Weight Value 35 0 AC 100 to AC 240 50/60 30 +5 2.2 12 1 Unit °C °C V Hz V·A V A V kg ON SW2 SW1 2345 CABLES (Included with M9011 or C8137) GND Vcont CHK2 CHK3 Don't use CHK4 CN5 Vcont-ADJ CN04 CN03 INPUT 12 V 1 INPUT AC 100 V to AC 240 V TPHOC0040EE Photon Counting Heads H7421 Series Options (Unit: mm) 1 Heatsink with Fan A7423 LEAD LENGTH 50 ± 10 4 Optical Fiber Adapter (FC Type) A7412 4- 2.2 4 TAPERED DEPTH 1.5 2.5 M8 P=0.75 16.5 ± 0.3 22.2 ± 0.2 52.0 ± 0.5 9.5 15 FRONT VIEW 3.0 4-M2 L=3 5.5 3 9 SIDE VIEW LIGHT SHIELD SHEET (THICKNESS: 0.5) TOP VIEW 4-M3 (SUPPLIED) TACCA0190EB 5 C-Mount Adapter A7413 24.5 ± 0.5 40 ± 1 8 30 92.0 ± 0.5 SIDE VIEW TACCA0188ED 4 14 TACCA0191EA 2 Power Supply Unit with Temperature Control M9011 70.0 ± 0.1 4-M2 15.0 ± 0.1 28.0 ± 0.2 6 Power Supply Unit with Temperature Control C8137 POWER PHOTOSENSOR SWITCH 40.0 ± 0.1 SW1 CN01 ON 50.0 ± 0.2 ON ON SW2 1 2345 CN03 160.0 ± 0.5 FRONT VIEW 46.0 ± 0.5 CN5 CHK1 CHK2 CHK3 CHK4 CN04 GND Vcont Don't use Vcont-ADJ 5.0 ± 0.1 100.0 ± 0.3 22 19.2 ± 0.3 53.6 ± 0.3 JST XMR-02V M25.4 P=1/32" C-MOUNT M25.4 P=1/32" C-MOUNT FAN CABLE 1500 ± 50 212 ± 1 SIDE VIEW MODULE OUTPUT AC INPUT 12.5 MODULE CABLE 1500 ± 50 POWER CABLE 1000 ± 50 EXTERNAL I/O CABLE FAN OUTPUT REAR VIEW FUSE POWER CABLE 1000 ± 50 1500 -0 +50 EXTERNAL I/O HOUSING TACCA0252EA FAN CABLE 1500 -0 +50 AC CABLE 3 Signal Cable E1168-05 1800 to 2000 TACCA0237EB BNC-P 1500 - +50 0 BNC-P TACCA0148EA 8 42 6 43
H7421
1. 物料型号: - H7421系列包含两种型号:H7421-40和H7421-50。

2. 器件简介: - H7421系列是包含金属封装光电倍增管(PMT)的光子计数头设备,具有GaAsP/GaAs光阴极和热电制冷器。热电制冷器减少光阴极产生的热噪声,提供高量子效率,即使在极低光照水平下也能获得良好的信噪比。

3. 引脚分配: - 文档中提到了8个主要的接口标记:A(热敏电阻1)、B(热敏电阻2)、C(热电制冷器+)、D(热电制冷器-)、E(Vcc低压)、F(IC)、G(GND)。其中IC表示内部连接,不使用。

4. 参数特性: - H7421-40和H7421-50的主要参数包括输入电压、最大输入电流、最大输入电压和电流针对热电制冷器、有效面积、峰值灵敏度波长、计数灵敏度、计数线性、暗计数、脉冲对分辨率、输出脉冲宽度、输出脉冲高度、推荐负载电阻、信号输出逻辑、操作环境温度、存储温度和重量。

5. 功能详解: - H7421系列通过监测安装在PMT附近的热敏电阻的输出,并调节热电制冷器的电流来维持恒定温度。H7421-40在300nm至720nm波长范围内具有高灵敏度,而H7421-50在380nm至890nm波长范围内具有宽光谱响应。

6. 应用信息: - H7421系列适用于需要在极低光照水平下进行测量的应用,如生物成像、光谱学、化学分析等。

7. 封装信息: - 提供了详细的尺寸图,包括深度、顶部视图、前视图、侧视图和横截面视图。封装类型为C-mount,具有M3螺纹和O型圈槽。
H7421 价格&库存

很抱歉,暂时无法提供与“H7421”相匹配的价格&库存,您可以联系我们找货