Model A-205
Flush Diaphragm Pressure
Transducer
DESCRIPTION
Models A-205 gage pressure transducers are manufactured with
environments. Model A-205 has welded electrical connectors
a unitized stainless steel diaphragm. The advantage of this type
as an integral part of the transducer body and recommended
of design is that a thin diaphragm and heavy sidewalls are made
for applications involving rough handling or where an all-welded
from one piece of stainless steel. This unitized diaphragm is rug-
stainless steel transducer is required.
ged, but at the same time can be made thin enough to measure
low pressures. Available pressure ranges span from 100 psig
to 10000 psig. These models can be used in corrosive fluid
FEATURES
•
100 psig to 10000 psig range
•
Flush mount design
•
1/4-18 NPT mounting thread
•
Miniature footprint
•
Stainless steel
•
CE approved
All units have four (4) active bonded strain gages arranged in a
Wheatstone-bridge configuration.
Model A-205
PERFORMANCE SPECIFICATIONS
RANGE CODES
Characteristic
Measure
Range Code
Available ranges
Pressure ranges
100, 150, 200, 300, 500, 750, 1000,
1500, 2000, 3000, 5000, 7500, 10000
psig
BR
100 psi
CJ
150 psi
CL
200 psi
3
Accuracy2
0.5 % full scale
Non-linearity and hysteresis
±0.5 % full scale3
CP
300 psi
Non-repeatability
±0.1 % full scale3
CR
500 psi
Output
2 mV/V
CT
750 psi
Infinite
CV
1000 psi
DJ
1500 psi
DL
2000 psi
DN
3000 psi
Resolution
ENVIRONMENTAL SPECIFICATIONS
Characteristic
Measure
DR
5000 psi
Temperature, operating
-54 °C to 149 °C [-65 °F to 300 °F]
DT
7500 psi
Temperature, compensated
16 °C to 71 °C [60 °F to 160 °F]
DV
10000 psi
Temperature effect, zero
0.01 % full scale/°F3
Temperature effect, span
0.02 % reading/°F3
OPTION CODES
ELECTRICAL SPECIFICATIONS
Characteristic
Measure
Strained gage type
Bonded foil
Excitation (calibration)
5 Vdc
Excitation (acceptable)
Up to 5 Vdc or (Vac)
Range Code
Many range/option combinations are available in
our quick-ship and fast-track manufacture programs. Please see http://sensing.honeywell.com/
TMsensor-ship for updated listings.
Pressure ranges
100, 150, 200, 300, 500, 750, 1000, 1500, 2000,
3000, 5000, 7500, 10000 psi
Temperature
compensation
1a. 60 °F to 160 °F
1b. 30 °F to 130 °F
1c. 0 °F to 185 °F
1d. -20 °F to 130 °F
1e. -20 °F to 200 °F
1f. 70 °F to 250 °F
1i. -65 °F to 250 °F
1g. 70 °F to 325 °F
1h. 70 °F to 400 °F
Insulation resistance
5000 mOhm @ 50 Vdc
Bridge resistance
350 ohm
Shunt calibration data
Included
Electrical termination (std)
PTIH-10-6P or equivalent (hermetic
stainless)
Internal
amplifiers
2u. Unamp., mV/V output
Mating connector (not incl.)
PT06A-10-6S or equiv. (AA111)
Pressure ports
5b. 1/4-18 NPT male
Electrical
termination
6a. Bendix PTIH-10-6P
(or equivalent) 6 pin
(max 400 °F)
6e. Integral cable: Teflon (-65 °F to 475 °F)
Special
calibration4
9a. 10 point (5 up/5 down)
20 % increments @ 70 °F 4
9b. 20 point (10 up/10 down)
10 % increments @ 70 °F 4
Wetted
diaphragm
10e. Iconel X-750
Shock and
vibration
44a. Shock and vibration resistance
MECHANICAL SPECIFICATIONS
Characteristic
Measure
Media
Gases, liquids compatible with wetted
parts
Overload, safe
50 % over capacity
Proof pressure
400 % full scale (max. 20000 psi)
Dead volume
Flush diaphragm (consult factory for
charged volume)
Wetted parts material
17-4 PH stainless steel
Weight
2 oz
Case material
Stainless steel
2 Honeywell • Sensing and Control
6h. Integral cable:
Silicone (-65 °F to
300 °F)
6i. Integral underwater
cable: (max 180 °F)1
6v. Phoenix connector
on end of cable
Flush Diaphragm Pressure Transducer
Mounting dimensions and characteristics
TYPICAL SYSTEM DIAGRAM
For reference only
Wiring codes
Connector
Connector, unamplified
A
(+) excitation
B
(+) excitation
C
(-) excitation
D
(-) excitation
E
(-) output
F
(+) output
Special requirements (consult factory)
Have a special requirement? New case pressure, different cable lengths,
electrical connectors, or materials? Consult our factory by calling +1
614-850-5000 (800-848-6564). Customization is key to our test and
measurement business. Special outputs, wiring codes, and calibrations
are all standard to us.
Honeywell • Sensing and Control 3
Model A-205
NOTES
Flush Diaphragm Pressure Transducer
1. Not available with option 1e, 1f, 1g, 1h or 1i.
2. Accuracies stated are expected for best-fit straight line for all errors, including linearity, hysteresis & non-repeatability thru zero.
3. For ranges less than 100 psi, consult factory for specifications.
4. Not available for all ranges.
Warranty. Honeywell warrants goods of its manufacture as
being free of defective materials and faulty workmanship.
Honeywell’s standard product warranty applies unless agreed
to otherwise by Honeywell in writing; please refer to your
order acknowledgement or consult your local sales office for
specific warranty details. If warranted goods are returned to
Honeywell during the period of coverage, Honeywell will repair
or replace, at its option, without charge those items it finds
defective. The foregoing is buyer’s sole remedy and is in lieu
of all warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
While we provide application assistance personally, through our
literature and the Honeywell web site, it is up to the customer to
determine the suitability of the product in the application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this printing.
However, we assume no responsibility for its use.
For more information about Sensing and Control products, visit
www.honeywell.com/sensing or call +1-815-235-6847
Email inquiries to info.sc@honeywell.com
WARNING
PERSONAL INJURY
• DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARNING
MISUSE OF DOCUMENTATION
• The information presented in this catalogue is for
reference only. DO NOT USE this document as product
installation information.
• Complete installation, operation and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
Sensing and Control
Automation and Control Solutions
Honeywell
1985 Douglas Drive North
Golden Valley, MN 55422 USA
+1-815-235-6847
www.honeywell.com/sensing
008616-1-EN IL50 GLO
May 2008
Copyright © 2008 Honeywell International Inc. All rights reserved.