GMS-10 RVS Series
Oxygen Sensors
Accurate oxygen measurement
DESCRIPTION
The GMS-10 is a high-accuracy oxygen sensor suitable for
produces the 700 °C required for the ZrO2 to operate as an O2
many applications. Originally developed for boiler
pump. The time taken for the pump to achieve specific
applications, the product range has been extended to offer
minimum and maximum pressures within this chamber is a
versions suitable for oxygen monitoring on board commercial
measure of the partial oxygen pressure of the environment.
and military aircraft, exhaust gas testing for automotive and
medical test equipment. The sensor employs two ZrO2 discs
The GMS-10 has to be operated by an electronic measuring
with a small hermetically sealed chamber in between. One of
circuit that controls the sensor operation and signal
the discs functions as a reversible oxygen pump, which is
processing. This circuit can either be incorporated into the
used to successively fill and empty the chamber. The second
customer’s own electronics, or be purchased as a separate
disc measures the ratio of the partial pressure difference and
box of electronics from Honeywell.
generates a corresponding sense voltage. A heat element
FEATURES
TYPICAL APPLICATIONS
• High-accuracy measurement
• Heating boiler control
• Low power consumption
• Industrial process control
• No reference gas required
• Control of aircraft on board oxygen generation systems
• Linear output signal
• Automotive exhaust gas diagnostics
• No need for temperature stabilization
• Service instruments
• Function testing and calibration in ambient air
• Monitoring environmental oxygen
• Long life
WARNING
MISUSE OF DOCUMENTATION
• The information presented in this product sheet is for
reference only. Do not use this document as a product
installation guide.
• Complete installation, operation, and maintenance
information is provided in the instructions supplied with
each product.
WARNING
PERSONAL INJURY
DO NOT USE these products as safety or emergency stop
devices or in any other application where failure of the
product could result in personal injury.
Failur
e to comply with these instr
uctions could rresult
esult in
Failure
instructions
Failur
e to comply with these instr
uctions could rresult
esult in
Failure
instructions
death or serious injur
y.
injury
death or serious injur
y.
injury
MOUNTING DIMENSIONS IN MM/INCHES
(FOR REFERENCE ONLY)
WARRANTY/REMEDY
Oxygen pressure range
2 mbar – 3 bar
Honeywell warrants goods of its
manufacture as being free of defective
materials and faulty workmanship.
Honeywell’s standard product warranty
applies unless agreed to otherwise by
Honeywell in writing; please refer to your
order acknowledgement or consult your
local sales office for specific warranty
details. If warranted goods are returned
to Honeywell during the period of
coverage, Honeywell will repair or
replace, at its option, without charge
those items it finds defective. The
for
egoing is buyer
emedy and
foregoing
buyer’’s sole rremedy
is in lieu of all war
ranties,
essed
warranties, expr
expressed
or implied, including those of
mer
chantability and fitness for a
merchantability
par
ticular purpose. In no event shall
particular
Honeywell be liable for consequential,
special, or indir
ect damages.
indirect
While we provide application
assistance personally, through our
literature and the Honeywell web site, it
is up to the customer to determine the
suitability of the product in the
application.
Specifications may change without
notice. The information we supply is
believed to be accurate and reliable as
of this printing. However, we assume no
Operational temperature
700 °C
responsibility for its use.
Stand by temperature
500 °C (typ.)
Sensitivity
1.05 ms/mbar
Accuracy
< 5 mbar
Response time
< 15 s
Warm up time
< 100 s
Warm up time (from stand by)
< 20 s
83,0
3.27
150,0
5.91
5,0
0.20
Ø 12,0
0.47
WIRING DIAGRAM
Heater
(3&5)
Yellow
Pump
(1)
Red
Common
(2)
Black
Sense
(4)
Blue
TECHNICAL DATA
Electrical Characteristics
Sensor voltage levels (recommended ), V
45-64-85 mV
Pump current (recommended), A
40 uA
Heater supply, V/A
4.0 V (1.7A)
Heater supply (stand by)
2.0 V
Pump resistance at 700 °C, Ohm
dc 1 kOhm (typ.)
ac 1 kHz 120 Ohm (typ.)
Operational specifications
Permissable gas temperature
-100 /+250 °C
Gas flow rate
0-10 m/s
SALES AND SERVICE
Honeywell serves its customers through
a worldwide network of sales offices,
representatives and distributors. For
application assistance, current
specifications, pricing or name of the
nearest Authorized Distributor, contact
your local sales office or:
E-mail: info.sc@honeywell.com
Mechanical characteristics
Inter
net: www.honeywell.com/sensing
Internet:
Repetitive permissable acceleration
5g
Incidental permissable acceleration
30 g
Phone and Fax:
Europe UK
+44 (0) 1698 481481
+44 (0) 1698 481676 Fax
USA/Canada +1-800-537-6945
+1-815-235-6545 Fax
International
Automation and Contr
ol Solutions
Control
Sensing and Control
Honeywell
Newhouse Industrial Estate
Motherwell, ML1 5SB UK
www.honeywell.com
100458-EN Issue 2 UK07 GLO
September 2004
Copyright 2004 Honeywell International Inc.
+1-815-235-6847