Honeywell’s Next Generation Precision Pressure Transducer (PPT2)
combines proven silicon sensor technology with microprocessor-based
signal conditioning to provide an extremely smart pressure transducer.
Available in a compact, rugged design, the PPT2 has many software
features that support a wide range of digital and analog applications.
FEATURES & BENEFITS
HIGHLY ACCURATE
• Accuracy is guaranteed over the
whole operating temperature range
APPLICATIONS
• Simplifies System Design
No additional signal compensation
needed to gain the benefits
of a very accurate sensor
SMART, DIGITAL SENSING
AND CONTROL
• Efficient Data Acquisition
Network up to 89 units
VERSATILE AND CONFIGURABLE
• Works with existing and new systems
0-5V analog and either RS-232
or RS-485 digital output
• Optimizes Output
User-configurable pressure
units, sampling, update rate
• Flags Problems
Internal diagnostics set
flags, indicates errors
USER SELECTABLE
SOFTWARE FEATURES
• Baud Rate, Parity Setting, Continuous
Broadcast, ASCII or Binary Output,
Sensor Temperature Output (°C),
Deadband, Sensitivity, Tare Value,
Configurable Analog Output
CE QUALIFIED. ISO-9001, ISO-14001
• Secondary Air Data
• Flow and Pressure Calibrators
• Altimeters
• Engine Testing
• Instrumentation and
Analytical Equipment
• Flight Testing
• Process Control
• Meteorology
• Research and Development
SPECIFICATIONS
PERFORMANCE
Total Error Band(1)
See Ordering Information
Temperature Range
Operating: -40 to 85ºC Standard (S), -55 to 110°C Extended (E)
Storage: -50 to 100°C Standard (S), -60 to 125°C Extended (E)
Sample Rate(3)
1.0 ms to 42.67min; minimum response delay 2 ms
Resolution
Digital: Up to 0.001% FS, Analog: 0.1 mV typical (15+ bits)
Long Term Stability
0.025% FS per year typical
MECHANICAL
Pressure Units(3)
atm, bar, cmwc, ftwc, hPa, inHg, inwc, kg/cm2, Kpa, mBar, mmHg, MPa,
mwc, psi, user, pfs
Media Compatibility
Suitable for non-condensing, non-corrosive, and non-combustible gases
Weight
4.4 oz. (125 gm) without fittings
ELECTRICAL
Output(3) (4)
RS-232 Digital with 0-5V Analog, RS-485 Digital with 0-5V Analog
Power Requirements
Supply Voltage: 6.0 to 34 VDC, Operating Current: 50 mA maximum
Baud Rate
User configurable between 1200 and 115200 bits/sec
(3)
Bus Addressing
Address up to 89 units
Connector
MIL-C-26482, Shell Size #10, 6-pin, #20 size
(3)
ENVIRONMENTAL
Mechanical Shock
RTCA/DO-160E Sec. 7, Cat. B: up to 20 G 3 sec/direction
Thermal Shock
RTCA/DO-160E Sec. 5, Cat. A, -55 to 110°C
Vibration
RTCA/DO-160E Sec. 8, Cat. S, Curve W: 20G, 10-2000 Hz
Overpressure
(2)
3X FS
Burst Pressure
3X FS
EMC Directive
Compliant, Directive 2004/108/EC, Standards BS EN 61326-1:2006
RoHS
Compliant
(2)
Total Error Band is the sum of worst case linearity, repeatability, hysteresis, thermal effects and calibration errors over the operating temperature range.
Full scale for differential ranges is the sum of + and – ranges. Calibration is traceable to NIST.
(2)
Exposure to overpressure will not permanently affect calibration or accuracy of unit. Burst pressure is the sum of the measured pressure plus
the static pressure and exceeding it may result in media escape.
(3)
User configurable.
(4)
Recommended load impedance of 100 k-ohm or greater.
(1)
PPT DIMENSIONS
0.69
(17.45)
SIGNAL NAME
0.12
(3.17)
2.12
(53.72)
0.17
(4.19)
0.49
(12.45)
2.2
(55.77)
0.47 (11.92)
1.56
(39.50)
0.53
(13.34)
1.06
(26.93)
0.5
(12.70)
1.97
(50.16)
2.2
(55.88)
2.47
(62.86)
0.09
(2.29)
A
B
C
D
E
F
RS-232(TD)/RS-485 (B)
RS-232(RD)/RS-485 (A)
Case Ground
Common Ground (GD)
DC Power In
Analog Output
ORDERING INFORMATION
PPT2 PRECISION PRESSURE TRANSDUCER
FULL SCALE PRESSURE RANGE
Absolute Gauge
Differential Digital Total Error Band(1)
Analog Total Error Band(1)
0002 N/A
2 PSI (1)
±2 PSI
±(0.075%FS + 0.20% Abs. Reading) ±(0.090%FS + 0.20% Abs. Reading)
0005 N/A
5 PSI
±5 PSI
±(0.075%FS + 0.20% Abs. Reading) ±(0.090%FS + 0.20% Abs. Reading)
0010 N/A
10 PSI
±10 PSI
±0.075%FS
±0.090%FS
0015 15 PSI
N/A
N/A
±0.075%FS
±0.090%FS
0020 20 PSI
N/A
N/A
±0.075%FS
±0.090%FS
P1 PRESSURE
P2 PRESSURE
TYPE
A
Absolute
0 (vacuum) to FS
N/A
G
Gauge
Reference to FS
Reference
D
Differential
+FS to -FS rel. to P2
+FS to -FS rel. to P1
P1
PRESSURE CONNECTION (ABSOLUTE, GAUGE, DIFFERENTIAL)
F
Filter (blocks debris)
G
Stainless Swagelok (1/8 inch female)
K
Stainless Swagelok-compatible (1/8 inch male)
R
Brass barbed, right angle (1/8 inch ID tubing)
W
Brass barbed (1/8 inch ID tubing)
X
Brass Swagelok (1/8 inch female)
P2
PRESSURE CONNECTION (GAUGE, DIFFERENTIAL)
F
Filter (blocks debris)
G
Stainless Swagelok (1/8 inch female)
K
Stainless Swagelok-compatible (1/8 inch male)
R
Brass barbed, right angle (1/8 inch ID tubing)
W
Brass barbed (1/8 inch ID tubing)
X
Brass Swagelok (1/8 inch female)
N
Not Applicable (Absolute)
OUTPUTS
2V
RS-232 digital, 0-5V analog
5V
RS-485 digital, 0-5V analog
OPERATING TEMPERATURE RANGE
S
Standard: -40 to 85°C
E
Extended: -55 to 110°C
- OPTIONS
PPT2 0020 A
(1)
W
N
2V
S
C
Power Supply/Data Cable (RS-232 only, See Below)
E
Certificate of Conformance
F
Calibration Certificate
-EF
See application note AN106 “Mechanically Mounting the PPT2 in Legacy PPT Applications,” at
aerospace.honeywell.com/en/learn/products/sensors/precision-pressure.
PRESSURE CONNECTION
OPTIONS
F
G
K
Option C
R
W
X
ESD (electrostatic discharge) sensitive device
Damage may occur when subjected to
high energy ESD. Proper ESD precautions
should be taken to avoid performance
degradation or loss of functionality.
EOS (electrical overstress) sensitive device
Damage may occur when subjected to EOS.
Do not exceed specified ratings to avoid
performance degradation or loss of functionality.
Honeywell reserves the right to make
changes to improve reliability, function or
design. Honeywell does not assume any
liability arising out of the application or use
of any product or circuit described herein;
neither does it convey any license under its
patent rights nor the rights of others.
For more information
aerospace.honeywell.com/en/learn/products/sensors/precision-pressure
Honeywell Aerospace
1944 East Sky Harbor Circle
Phoenix, AZ 85034
aerospace.honeywell.com
ADS-14211
N61-1598-000-002 | 02/20
© 2020 Honeywell International Inc.