MT1000
PVDF HEAT SHRINK TUBING
APPLICATIONS
PROFILE
• Abrasion protection for laparoscopic
and in-vivo instruments
• Shrink ratio <
_ 2:1
• High performance insulation for
electrosurgical devices
• Strain relief applications
• Full recovery at 175°C (347°F) minimum
• Supports sterilization environments: gamma, ethylene oxide (ETO), steam, dry heat
and autoclave
• Manufactured to ISO 10993 standards
• Registered with the FDA: MAF-444
• Custom sizing, colors, finishing and value-add options available
• Radiopacity can be customized
ABOUT
• MT1000 is a crosslinked polyvinylidene fluoride (PVDF) heat shrink tubing. PVDF
offers excellent chemical and abrasion resistance, high dielectric strength and
superior tensile strength. Its homogeneous structure (properties evenly distributed)
contributes to its consistency and high performance, making our MT1000 tubing
essentially free from flaws, defects, pinholes, seams, cracks or inclusions
• MT1000 is rigid and highly-lubricious, and works very well at providing abrasion
protection for rigid laparoscopic and in-vivo instruments
MEDICAL / HEAT SHRINK TUBING — MT1000
HEAT SHRINK TUBING — MT1000
TABLE 1: DIMENSIONS
Standard Sizes
As Supplied
Recovered
Inside Diameter Minimum (D)
Inside Diameter Maximum (d)
Wall Thickness (in., mm.) (W)
Size
in.
mm.
in.
mm.
Minimum
3/64
0.046
1.17
0.023
0.58
.008
0.20
0.012
Maximum
0.31
.010
Normal
0.25
1/16
0.063
1.60
0.031
0.79
.008
0.20
0.012
0.31
.010
0.25
3/32
0.093
2.36
0.046
1.17
.009
0.20
0.012
0.31
.010
0.25
1/8
0.125
3.18
0.062
1.58
.009
0.20
0.012
0.31
.010
0.25
3/16
0.187
4.75
0.093
2.36
.009
0.20
0.012
0.31
.010
0.25
1/4
0.250
6.35
0.125
3.18
.011
0.28
0.015
0.38
.013
0.33
3/8
0.375
9.53
0.187
4.75
.011
0.28
0.015
0.38
.013
0.33
1/2
0.500
12.70
0.250
6.35
.011
0.28
0.015
0.38
.013
0.33
TABLE 2: PROPERTIES
Property
Unit
Requirement
Test Method
Physical
Dimensions*
inches (mm)
In accordance with Table 1
Longitudinal change*
percent
+0, -10 maximum
ASTM D 2671
Concentricity as supplied*
percent
70 minimum
ASTM D 2671
psi (MPa)
5000 minimum (34.5)
percent
150 minimum
psi (MPa)
1 x 105 minimum (690)
percent
50 minimum
Tensile strength*
Ultimate elongation*
Secant modulus* (expanded)
Heat resistance
168 hours at 250 ± 5°C (482°F)
Followed by test for: Ultimate elongation
ASTM D 2671, 20”/minute
ASTM D 2671
ASTM D 2671, 20”/minute
Electrical
Dielectric strength: Sizes 3/64 through 1/2
Dielectric strength: Sizes 3/4 through 2
Dielectric withstand 3000V, 60Hz
volts/mil
800 minimum (31.500)
kV/mm
600 minimum (23.600)
sec
60 minimum
ASTM D 2671
ASTM D 2671
Chemical
Fluid resistance
24 hours at 23 ± 3°C (73 ± 5°F) Isopropyl alcohol
5% saline solution Disinfectant
Followed by tests for: Dielectric strength
Sizes 3/64 through 1/2
Sizes 3/4 through 2
Tensile strength
Heavy metals analysis
Cadmium, Mercury, Lead, Bismuth, Antimony
ASTM D 2671
kV/mm
700 minimum (27.600)
500 minimum (19.700)
psi (MPa)
5000 minimum (34.5)
ASTM D 2671, 2”/minute
ppm
1 maximum
(total of all metals)
USP XXII
Physiochemical tests-plastic (Note 1)
*Denotes lot acceptance testNote
1: Sample preparation and extraction is per USP XXII. Metals analysis may be colorimetric as described in USP XXII or by equivalent quantitative analytical method.
TE.com/medical
TE Connectivity, TE connectivity (logo), Raychem and TE are trademarks owned or licensed by the TE Connectivity Ltd. family of companies.
All other logos, products and/or company names referred to herein might be trademarks of their respective owners.
While TE has made every reasonable effort to ensure the accuracy of the information in this brochure, TE does not guarantee that it is error-free, nor does TE make any other
representation, warranty or guarantee that the information is accurate, correct, reliable or current. TE reserves the right to make any adjustments to the information contained
herein at any time without notice. TE expressly disclaims all implied warranties regarding the information contained herein, including, but not limited to, any implied warranties of
merchantability or fitness for a particular purpose. The dimensions in this catalog are for reference purposes only and are subject to change without notice. Specifications are subject
to change without notice. Consult TE for the latest dimensions and design specifications.
© 2024 TE Connectivity. All Rights Reserved.
04/24 med-mt-1000-hst-ss-en
MEDICAL / HEAT SHRINK TUBING — MT1000
很抱歉,暂时无法提供与“895513-000”相匹配的价格&库存,您可以联系我们找货