MT-FEP
FEP Heat Shrink Tubing
Applications
• Process aid for reflowing catheter shafts
• Process aid for joint bonding
PROFILE
• Shrink ratio <
_ 1.6:1
• Full recovery at 210°C (410°F) minimum
• Tight longitudinal change control as low as ± 2%
• Manufactured to ISO 10993 standards
• Custom sizing, finishing options available
• Translucent for high optical clarity
• Color blending option available
ABOUT
• MT-FEP is a fluorinated ethylene propylene (FEP) heat shrink tubing. FEP offers excellent consistency, high
dielectric strength and is chemically inert. MT-FEP is lubricious and semi-rigid with shrink ratios up to 2:1*.
FEP is the industry gold standard for reflowing catheter shafts and bonding joints.
• For our MT-FEP, we can control longitudinal growth ± 2% ensuring consistency on lot-to-lot, reducing waste
and cost.
*Upper limit on select applications. Optimal shrink ratio is 1.6:1
TABLE 1: DIMENSIONS
As Supplied
Size
Recovered
Inside Diameter
Minimum (D)
Inside Diameter
Maximum (d)
in.
mm.
in.
mm.
Minimum
in.
mm.
Wall Thickness (W)
Maximum
in.
mm.
Nominal
in.
mm.
1/32
0.035
0.9
0.025
0.6
0.006
0.15
0.010
0.25
0.008
0.20
3/64
0.045
1.1
0.032
0.8
0.006
0.15
0.010
0.25
0.008
0.20
1/16
0.063
1.6
0.040
1.0
0.006
0.15
0.010
0.25
0.008
0.20
3/32
0.093
2.4
0.056
1.4
0.006
0.15
0.010
0.25
0.008
0.20
1/8
0.125
3.2
0.075
1.9
0.007
0.18
0.013
0.33
0.010
0.25
3/16
0.188
4.8
0.115
2.9
0.007
0.18
0.013
0.33
0.010
0.25
1/4
0.250
6.4
0.150
3.8
0.007
0.18
0.013
0.33
0.010
0.25
3/8
0.375
9.5
0.225
5.7
0.009
0.23
0.015
0.38
0.012
0.30
1/2
0.500
12.7
0.300
7.6
0.011
0.28
0.019
0.48
0.015
0.38
MEDICAL /// Heat Shrink Tubing — MT-FEP
Heat Shrink Tubing — MT-FEP
TABLE 3: PROPERTIES
Property
Unit
Requirement
Test Method
Physical
Dimensions*
inches (mm)
In accordance with Table 1
Longitudinal change*
percent
Concentricity as supplied*
percent
Tensile strength*
Ultimate elongation*
Secant modulus* (expanded)
Heat resistance
168 hours at 175°C (347°F)
Followed by test for:
Ultimate elongation
Electrical
Dielectric strength
psi (MPa)
percent
psi (MPa)
+0, -10 maximum
70 minimum (2:1 Exp. ratio)
60 minimum (3:1 Exp. ratio)
2000 minimum (10:3)
200 minimum
2.5 x 104 maximum (172)
percent
100 minimum
ASTM D 2671,
20"/minute
volts/mil
(volts/mm)
500 minimum (19.7)
ASTM D 2671
sec
60 minimum
ASTM D 2671
Dielectric withstand
3000V, 60Hz
Chemical
Fluid resistance
24 hours at 23 ± 3°C (77 ± 5°F)
Isopropyl alcohol
5% saline solution
Disinfectant
Followed by tests for:
Dielectric strength
ASTM D 2671
ASTM D 2671
ASTM D 2671,
20"/minute
ASTM D 2671
ASTM D 2671
volts/mil
(volts/mm)
400 minimum (15.7)
Tensile strength
psi (MPa)
1000 minimum (6.9)
Heavy metals analysis
Cadmium
Mercury
Lead
Bismuth
Antimony
ppm
1 maximum
(total of all metals)
ASTM D 2671
USP XXII
Physiochemical
tests-plastic
(Note 1)
*Denotes lot acceptance test
Note 1: Sample preparation and extraction is per USP XXII. Metals analysis may be colorimetric as described in USP XXII or by
equivalent quantitative analytical method.
te.com/medical
AMERICAS
EMEA
ASIA
+1 866 251 3352
+49 6151 607 1999
China +86 400 820 6015
Japan +81 44 844 8041
Korea +82 2 3415 4506
Singapore +86 512 6255 4384
MEDICAL /// Heat Shrink Tubing — MT-FEP
TE Connectivity, TE, TE connectivity
(logo) and Raychem are trademarks.
© 2020 TE Connectivity Ltd. family
of companies. All Rights Reserved.
很抱歉,暂时无法提供与“ER0322-000”相匹配的价格&库存,您可以联系我们找货