HM1520LF – Relative Humidity Module
•
Tubular form for through wall mounting
•
Product free from Lead, Cr(6+), Cd and Hg
•
Calibrated, linear voltage for easy electronic interface
•
Typical 1 to 1.6 Volt DC output for 0 to 20% RH at
5Vdc supply
•
Ratiometric to voltage supply
•
Patented solid polymer structure
DESCRIPTION
Based on the rugged HS1101LF capacitive humidity sensor, HM1520LF is a dedicated humidity transducer
designed for measurements at low humidity. Direct measurement of dew point or water concentration can be
easily obtained in really cost effective conditions. Direct interface with a micro-controller is made possible with
the module’s linear voltage output.
FEATURES
APPLICATIONS
•
Full interchangeability
•
Low Humidity
•
Not affected by water immersion
•
Meteorology
•
Controlled temperature dependency
…
Humidity Sensor Specific features
Not affected by long period at low humidity values
High resistance to chemical
Fast response time
PERFORMANCE SPECS
MAXIMUM RATINGS
Ratings
Storage Temperature
Storage Humidity
Supply Voltage (Peak)
Humidity Operating Range
Temperature Operating Range
Symbol
Tstg
RHstg
Vs
Value
-30 to 70
0 to 100
10
Unit
°C
% RH
Vdc
RH
Ta
0 to 100
-40 to 60
% RH
°C
Peak conditions: less than 10% of the operating time.
HM1520LF HPC091_C
www.meas-spec.com
1/6
October 2012
HM1520LF – Relative Humidity Module
ELECTRICAL CHARACTERISTICS
(Ta=23°C, Vs=5Vdc +/-5%, R L>1MΩ unless otherwise stated)
Humidity Characteristics
Humidity Measuring Range / Ta -40 to 60°C
Relative Humidity Accuracy (1 to 20% RH) at 23°C
Relative Humidity Accuracy @55% RH at 23°C
Supply Voltage (regulated at 5Vdc*)
Nominal Output @10%RH (Ta = 25°C)
Current Consumption
Temperature Coefficient (10 to 50°C and 1 to 20% RH )
Humidity Average Sensitivity from 5% to 10%RH
Sink Current Capability (RL=33kΩ)
Humidity Hysteresis
Time Constant (at 63% of signal, static) 5% to 10%RH
Warm Up Time (electronic)
Humidity Resolution
Long Term Stability
Output Impedance
Symbol
RH
RH
RH
Vs
Vout
Ic
Tcc
∆Vout/∆RH
Is
τ
tw
Z
Min
0
4.75
1.17
Typ
+/-2
+/-5
5
1.24
1.4
- 0.05
+26
Max
100
+/-3
5.25
1. 31
2
-0.1
150
+/-1
10
150
0.4
0.5
70
Unit
%RH
%RH
%RH
Vdc
V
mA
%RH/°C
mV/%RH
µA
%RH
s
ms
%RH
%RH/yr
Ω
*Maximum power supply ramp up time to Vcc should be less than 4ms.
TYPICAL PERFORMANCE CURVES
HUMIDITY SENSOR
•
HM1520LF Preliminary Specification when used from 1 to 30% RH
•
Those equations can be used above 30% RH and allow to obtain an over all accuracy as described in page
4 of this document in all humidity measuring range.
Temperature (in the range 0 to 50°C) does not affec t HM1520LF measurement when used from 1% to 30%
RH. No temperature compensation is required.
•
HM1520LF HPC091_C
www.meas-spec.com
2/6
October 2012
HM1520LF – Relative Humidity Module
•
Signal output from 1% to 20% RH at 23°C
RH (%)
1
2
3
4
5
6
7
8
9
10
Vout (mV)
1013
1038
1064
1089
1115
1141
1166
1192
1217
1243
RH (%)
11
12
13
14
15
16
17
18
19
20
Vout (mV)
1269
1294
1320
1346
1371
1397
1422
1448
1474
1499
With Vs = 5.0 Volts DC
Calibration data are traceable to NIST standards through CETIAT laboratory.
IMPORTANT NOTICE : HM1520LF is based on HS1101LF capacitive sensor and thus is fully useable on a
large range of relative humidity (1% to 99% RH). In that range, HM1520LF presents a typical accuracy of
+/-5% RH at 55% RH.)
However HM1520LF is a dedicated module for humidity measurements at low Relative Humidity levels.
Thus, HM1520LF is also well adapted to measure water concentrations (ppm) or low dew points when
associated with an ambient temperature probe.
HM1520LF features an optimized accuracy for water concentration below 6000 ppm water or 0°C dew point
at 23°C (equivalent to 20% RH).
•
Typical Output of HM1520LF when measuring water concentration at ambient temperature of 23°C
ppm
275
550
825
1100
1375
1650
1925
2200
2470
2750
Vout (mV)
1013
1038
1064
1089
1115
1141
1166
1192
1217
1243
ppm
3025
3300
3570
3850
4120
4395
4670
4945
5220
5495
Vout (mV)
1269
1294
1320
1346
1371
1397
1422
1448
1474
1499
ppmwater = 10.75 Vout - 10615 with Vout in mV
•
Typical Output of HM1520LF when measuring dew point at ambient temperature of 23°C
DP (°C)
-36
-29
-24
-21
-18.5
-16
-14.5
-13
-11.5
- 10
Vout (mV)
1013
1038
1064
1089
1115
1141
1166
1192
1217
1243
DP (°C)
-9
-7.8
-6.8
-5.8
-4.9
-4.1
-3.2
-2.5
-1.6
-1.1
Vout (mV)
1269
1294
1320
1346
1371
1397
1422
1448
1474
1499
HM1520LF HPC091_C
www.meas-spec.com
3/6
October 2012
HM1520LF – Relative Humidity Module
•
Accuracy of HM1520LF when used from 1% to 95% RH
QUALIFICATION PROCESS
RESISTANCE TO PHYSICAL AND CHEMICAL STRESSES
•
•
•
HM1520LF has passed through qualification processes of MEAS-FRANCE including vibration, shock,
storage, high temperature and humidity, ESD.
Additional tests under harsh chemical conditions demonstrate good operation in presence of salt
atmosphere, SO2 (0.5%, H2S (0.5%), 03, NOx, NO, CO, CO2, Softener, Soap, Toluene, acids (H2SO4,
HNO3, HCI), HMDS, Insecticide, Cigarette smoke, this is not an exhaustive list.
HM1520LF is not light sensitive.
SPECIFIC PRECAUTIONS
•
•
HM1520LF is protected against reversed polarity.
If you wish to use HM1520LF in a chemical atmosphere not listed above, consult us.
HM1520LF HPC091_C
www.meas-spec.com
4/6
October 2012
HM1520LF – Relative Humidity Module
PACKAGE OUTLINE
Dim
Min (mm)
Max (mm)
A
9.75
10.25
B
4.00
4.50
C
53
55
D
10.9
11.4
E*
200
250
* Specific lenght available on request
Wire
Color
Function
W1
White
Ground
W2
Blue
Supply Voltage
W3
Yellow
Humidity Output Voltage
ORDERING INFORMATION
HPP805C031 (MULTIPLE PACKAGE QUANTITY OF 10 PIECES)
HM1520LF – HUMIDITY ANALOG VOLTAGE OUTPUT MODULE
Sample kit of HM1520LF is available through MEASUREMENT SPECIALTIES web site:
http://www.meas-spec.com/humidity-sensors.aspx
Customer Service contact details
Measurement Specialties, Inc - MEAS France
Impasse Jeanne Benozzi
CS 83 163
31027 Toulouse Cedex 3
FRANCE
Tél: +33 (0)5 820 822 02
Fax: +33(0)5 820 821 51
Sales: humidity.sales@meas-spec.com
HM1520LF HPC091_C
www.meas-spec.com
5/6
October 2012
HM1520LF – Relative Humidity Module
Revision
A
B
C
Comments
Characteristics updated
Standardized datasheet format and Current consumption
lowering updated
New MEAS template, MEAS-France contact details
updated
Who
D. LE GALL
Date
January 08
D. LE GALL
April 08
D. LE GALL-ZIRILLI
October 12
The information in this sheet has been carefully reviewed and is believed to be accurate; however, no responsibility is assumed for
inaccuracies. Furthermore, this information does not convey to the purchaser of such devices any license under the patent rights to the
manufacturer. Measurement Specialties, Inc. reserves the right to make changes without further notice to any product herein. Measurement
Specialties, Inc. makes no warranty, representation or guarantee regarding the suitability of its product for any particular purpose, nor does
Measurement Specialties, Inc. assume any liability arising out of the application or use of any product or circuit and specifically disclaims
any and all liability, including without limitation consequential or incidental damages. Typical parameters can and do vary in different
applications. All operating parameters must be validated for each customer application by customer’s technical experts. Measurement
Specialties, Inc. does not convey any license under its patent rights nor the rights of others.
HM1520LF HPC091_C
www.meas-spec.com
6/6
October 2012
很抱歉,暂时无法提供与“HPP805C031”相匹配的价格&库存,您可以联系我们找货