RT 500
Flexible split-core Rogowski coil for the electronic measurement
of AC waveform currents, with galvanic separation between the
primary circuit (power) and the secondary circuit (measurement).
The patented Perfect Loop Technology definitely solves the problem
of the Rogowski coil opening by introducing a specific magnetic
device. This dramatically reduces both the error due to the position of
the measured conductor within the sensor, and the error due to the
proximity of external conductors.
Features
Electrical data
IPN
M
Vout
Primary nominal rms current
500
A
Mutual inductance transfer ratio 1) (± 5 %) dI
0.064 µH
P
Output voltage (instantaneous) 2) Vout = M ×
V
dt
Vout
LS
RS
Output voltage (sinusoidal wave) 2) Vout = 2 × π × M × f × IP AC
V
Example: @ 50 Hz, 500 A
Vout = 2 × π × 64 × 10-9 × 50 × 500 = 10.05 mV
Coil inductance (± 5 %)
125
µH
Coil resistance TA = 20 °C typical
40
Ω
High cut-off frequency 2)
700
kHz
Linearity error
Temperature coefficient of M
Positioning error 3) max
Advantages
●● Thin, lightweight and flexible
●● Easy installation suitable even for
the most confined spaces
●● Very low positioning error.
Applications
Accuracy - Dynamic performance data
εL
TCM
εP
●● Split core type
●● Ø 70 mm sensing aperture for
non-contact measurement
●● 1.5 m output cable
●● Insulating plastic case recognized
according to UL 94-V0.
none
30
0.65
ppm/K
%
General data
TA
Ambient operating temperature (90 % RH)
- 10 .. + 65
°C
TS
Ambient storage temperature
- 25 .. + 70
°C
m
Mass
85
g
IPxx
Protection degree
IP 2X
Standards
Safety IEC 61010-1: 2001 2nd edition
IEC 61010-2-032: 2002
IEC 61010-031: 2002 + Amd 1: 2008
Notes: 1) LEM provides uncalibrated coils with up to 5 % manufacturing tolerance.
Calibration must be performed within the final application.
2)
Open circuit condition (no load).
3)
Considering a primary conductor of at least Ø 15 mm perpendicular to
the RT loop.
●● Power meters
Current measurement for active
Power calculation.
●● Energy sub-meters
For energy efficiency monitoring,
consumption analysis and costs
allocation.
●● Power quality monitoring
●● Condition monitoring
(e.g. Motor loads such as
conveyers, pumps or HVAC).
●● Distributed measurement
systems.
Application domain
●● Energy Solutions.
N° 97.F9.50.000.0
11December2013/version 2
Page 1/4
LEM reserves the right to carry out modifications on its transducers, in order to improve them, without prior notice
www.lem.com
UL 61010-1: Ratings and assumptions of certification
File # E330077-A2-UL
Standards
●●
●●
CAN/CSA-C22.2 No. 61010-1, 2nd Edition, 2004-07 (Electrical Equipment for Measurement, Control, and laboratory
Use; Part 1: General Requirements)
UL 61010-1, 2nd Edition, 2005-07-22 (Electrical Equipment for Measurement, Control, and laboratory
Use; Part 1: General Requirements)
Ratings
Parameter
Symbol
Primary involved potential
Unit
Value
V AC/DC
300
Max surrounding air temperature
TA
°C
50
Primary current
IP
A
According to series primary
current
Output voltage
Vout
V AC
0 to 30
Conditions of acceptability
For use only in or with complete equipment where the acceptability of the combination is determined by Underwriters
Laboratories Inc. When installed in an end-product, consideration must be given to the following:
1 - This component has been judged on the basis of the creepage and clearances required in the indicated Standards, which
would cover the component itself if submitted for Listing: UL 61010-1 2nd Ed., CAN/CSA 22.2 No 61010-1 2nd Ed.
2 - The need for the following shall be considered in the end-product: Markings and documentation that comply with Clause 5.
3 - The following tests shall be performed in the end-product evaluation: Dielectric Strength.
4 - The unit is considered acceptable for use in a max ambient of: 50 °C.
Marking
Only those products bearing the UR Mark should be considered to be Recognized and covered under UL's Follow-Up Service.
Always look for the Mark on the product.
Page 2/4
11December2013/version 2
LEM reserves the right to carry out modifications on its transducers, in order to improve them, without prior notice
www.lem.com
RT 500
Insulation coordination
U b
Rated insulation rms voltage 1)
with IEC 61010-1 standards following conditions:
- Reinforced insulation
- Over voltage category CAT III
- Pollution degree PD2
- Heterogeneous field
U d
Rms voltage for AC insulation test 2), 50 Hz, 1 min
ÛW
Impulse withstand voltage 1.2/50 µs
dCp
Creepage distance
dCI
Clearance
CTI
Comparative Tracking Index (group I) 3)
300
V
3.5
6.5
Min
6
6
600
kV
kV
mm
mm
V
Notes: 1) Between primary conductor and ground. If insulated cable is used for the
primary circuit, the voltage category could be improved according to the
primary cable insulation category
(please refer to the cable manufacturer’s indications).
2)
Between primary (completely filling the hole) and secondary
3)
Plastic parts.
Safety and warning notes
In order to guarantee safe operation of the transducer and to be able to make
proper use of all features and functions, please read these instructions thoroughly!
Safe operation can only be guaranteed if the transducer is used for the purpose
it has been designed for and within the limits of the technical specifications.
Ensure you get up-to-date technical information that can be found in the latest
associated datasheet under www.lem.com.
Caution! Risk of danger
Ignoring the warnings can lead to serious injury and/or cause damage!
The electric measuring transducer may only be installed and put into
operation by qualified personnel that have received an appropriate training.
The corresponding national regulations shall be observed during
installation and operation of the transducer and any electrical conductor.
The transducer shall be used in electric/electronic equipment with respect
to applicable standards and safety requirements and in accordance with all
the related systems and components manufacturers’ operating instructions.
Caution! Risk of electrical shock
When operating the transducer, certain parts of the module may carry
hazardous live voltage (e.g. primary conductor). The user shall ensure
to take all measures necessary to protect against electrical shock. The
transducer is a build-in device containing conducting parts that shall not be
accessible after installation. A protective enclosure or additional insulation
barrier may be necessary. Installation and maintenance shall be done with
the main power supply disconnected except if there are no hazardous live
parts in or in close proximity to the system and if the applicable national
regulations are fully observed.
Safe and trouble-free operation of this transducer can only be guaranteed if transport,
storage and installation are carried out correctly and operation and maintenance are
carried out with care.
11December2013/version 2
LEM reserves the right to carry out modifications on its transducers, in order to improve them, without prior notice
Page 3/4
www.lem.com
Dimensions RT 500 (in mm)
Connection
Ip
IP
VS+
VS-
Fig. 1
Remarks
Mechanical characteristics
●● General tolerance
●● Output cable length
± 1 mm
3000 mm
WARNING!
Do not stress the coil by applying any kind of
mechanical force (ie. twisting, puncturing, excessive
pressure, tight bending, etc.) which will dramatically
degrade the device’s accuracy.
●● Vout= VS+- VS- is positive when dIP/dt flows in the direction
of the arrow (see fig. 1)
●● Due to low positioning error (εP), the device does not
need to be physically fastened around the primary
conductor. Should the device be secured, make sure no
mechanical stress is applied to the coil itself.
Page 4/4
11December2013/version 2
LEM reserves the right to carry out modifications on its transducers, in order to improve them, without prior notice
www.lem.com