Corcom Product Guide
Catalog: 1654001 Issue Date: 06.2011
Multi-purpose Medical Filter for Power Line Noise Protection
MV Series
UL Recognized CSA Certified VDE Approved
MV Series
• Multi-purpose medical filter • Improved Line to Ground performance • A good solution to emission or immunity problems • Meets leakage current requirements of UL2601 for health care equipment
Specifications
Maximum leakage current each Line to Ground: @ 120 VAC 60 Hz: .07 mA @250 VAC 50 Hz: .13 mA Hipot rating (one minute): Line to Ground: Line to Line: Rated Voltage (max): Operating Frequency: Rated Current: 2250 VDC 1450 VDC 250 VAC 50/60 Hz 3 to 20A
Operating Ambient Temperature Range (at rated current Ir): -10°C to +40°C In an ambient temperature (Ta) higher than +40°C the maximum operating current (Io) is calculated as follows: Io = Ir (85-Ta)/45
Ordering Information 6 MV 1
Input / Output Style 1 – .250 [6.3] spade terminals MV Series Current Rating 3, 6, 10 or 20A
Electrical Schematic
LINE LOAD
Available Part Numbers
3MV1 10MV1 6MV1 20MV1
54
Dimensions are in inches and millimeters unless otherwise specified. Values in italics are metric equivalents. Dimensions are shown for reference purposes only. Specifications subject to change.
For email, phone or live chat, please go to te.com/help corcom.com
Corcom Product Guide
Catalog: 1654001 Issue Date: 06.2011
Multi-purpose Medical Filter for Power Line Noise Protection (continued)
MV Series
Case Styles MV1 (3, 6, 10A) Performance Data Typical Insertion Loss
Measured in closed 50 Ohm system
3MV
db 100
6MV
db 100 90 80
1
RFI Power Line Filters
1 10 30 Frequency in MHz
B
90 80 70 60 50
70 60 50 40 30 20 10 0.1 1 10 30 Frequency in MHz 0 0.1
C
D E A
40 30 20 10 0
10MV
db 100 90 80 70 60
20MV
db 100 90 80 70 60 50 40 30 20 10
20MV1
50 40 30 20 10 0 0.1 1 10 30 Frequency in MHz
0
0.1
1
10 30 Frequency in MHz
E
DB
Common Mode / Asymmetrical (L-G) Differential Mode / Symmetrical (L-L)
Minimum Insertion Loss
Measured in closed 50 Ohm system
2.00 50.8
Common Mode / Asymmetrical (Line to Ground)
C
A
Case Dimensions
Part No. 3MV1 6MV1 10MV1 20MV1 A
B
C
D
E
Differential Mode / Symmetrical (Line to Line)
55
Dimensions are in inches and millimeters unless otherwise specified. Values in italics are metric equivalents. Dimensions are shown for reference purposes only. Specifications subject to change.
For email, phone or live chat, please go to te.com/help corcom.com