Kotron® RF Sensing Probes
DESCRIPTION
Kotron RF Capacitance sensing probes are available in many different configurations to handle a wide variety of application conditions. The sensing probes in this brochure can be combined with Kotron level switches and transmitters.
FEATURES
RIGID PROBES • Available as bare probe or insulated probe • Maximum process temperature: +1000° F (+540° C) – bare probe +400° F (+200° C) – insulated probe • Maximum process pressure 5,000 PSIG (345 bar) • Lengths up to 234 inches (6 meters) • Bare probes can be cut to length in the field • distance from the probe to its ground; • Wetted materials include 316/316L SS, Hastelloy® C, Teflon® (TFE), Halar® (ECTFE) and Kynar® (PVDF) • Ceramic seals for High Temperature/High Pressure designs FLEXIBLE PROBES • Available as bare probe or insulated probe • Maximum temperature: +650° F (+345° C) – bare probe +285° F (+140° C) – insulated probe • Lengths up to 150 feet (45 meters) • Insulated probe length can be adjusted in the field • Wetted materials include 316 SS and Halar® (ECTFE) • Teflon seals for standard probes
®
TECHNOLOGY
The amount of capacitance developed in any application is affected by three variables: • size (surface area) of the probe;
• dielectric of the medium it is measuring. Considering that the probe’s mounting position is fixed, and the dielectric constant of the process media is stable, then the amount of capacitance developed is directly proportional to the level of the process media on the probe. Increasing the surface area (diameter) of the probe and/or decreasing the distance between the probe and its ground reference will increase the capacitance gain.
APPLICATIONS
• Clean or dirty liquids • Viscous liquids • Light slurries • High temperature/pressure liquids • Foods and beverages • Hydrocarbons and solvents • Corrosives, acids, and caustics
• Ceramic seals for High Temperature/High Pressure designs
M A G N E S E A L®
PROBES
3. The Teflon® (TFE) probe insulation is heat-treated which forms the material tight to the probe rod. This process yields better linearity and eliminates “stress-relieving” (elongation) of the material at elevated temperatures. 4. Stability of the outer jacket is maintained by the end of the probe barb which further secures the insulating jacket. The Magneseal design ensures reliability long after initial installation.
Probes are an essential component of the RF Capacitance measurement system. They are critical in development of the proper “capacitor” for reliable level measurement. As importantly, the probe becomes part of the process seal of the vessel; its reliability is crucial. With the development of the Magneseal probe, Magnetrol has taken strides to further ensure this reliability. The Magneseal probe offers the following advantages: 1. A sophisticated compression seal exerts radial pressure between the center rod and the mounting nut yielding bulletproof sealing up to 3000 psig (205 bar). 2. Spring washers maintain the seal particularly during varying temperature and pressure that can degrade other process seals.
PROCESS
MEDIA
GUIDELINES
Non-conductive (dielectric less than 10 or conductivity less than 10 µsiemens/cm) Hydrocarbons, solvents, and bulk solids typically fall into the category of non-conductive media. Initially, when the vessel is empty, the dielectric constant is 1 (air). As the media level rises, the dielectric of the media replaces the air, thus causing the capacitance to increase. This increase is linear with the level increase. A bare probe is usually the best choice for this application. Conductive (dielectric greater than 10 or conductivity greater than 10 µsiemens/cm) Conductive media in conjunction with a bare probe will result in an electrical “short”, causing a transmitter to indicate a high level or a switch to change state. The solution is to use an insulated probe constructed of Teflon®, Kynar®, or Halar®. The conductive media creates an electrical connection between a metal vessel wall and the probe insulation. Like the non-conductive application, the distance between the probe and ground, and the probe diameter is fixed. Instead of measuring the dielectric of the media, we are measuring the dielectric of the probe insulation which is covered by the media.
Air
Bare metal probe
Non-conductive (insulating medium)
Metal vessel wall (ground reference)
Conductive medium (ground reference) Insulated probe
Metal rod probe Conductive media
Probe insulation
Metal vessel wall
2
PROBE
SELECTION
where the probe will be mounted more than 12" (30 cm) from the vessel wall, or when measuring any liquid in non-metallic vessels. The stillwell probe is the most common. If the application requires “no metal” in the process; or, if the liquid is too viscous for a stillwell, use the reference rod probe. The reference wire probe should only be used in clean, conductive, non-coating applications. 4. Use a flexible (cable) probe when the measurement range is greater than 10 feet (3 meters). Rigid (rod) probes are available up to 234 inches (6 meters) in length, but they are sometimes difficult to handle and can be damaged during installation. 5. Use an inactive sheath probe when mounting horizontally through a nozzle.
Probe selection is the most critical part of applying an RF Capacitance device for a given application. The goal is to select the probe that will give the optimum capacitance change per unit level change (pF/inch). The first step in selecting an RF probe is determining the correct configuration for your application. The following guidelines will assist you in this selection. 1. Use bare probes for non-conductive liquids. 2. Use insulated probes for conductive liquids. If you are uncertain about the conductivity value, use an insulated probe. Teflon® has the widest temperature range and material compatibility. Kynar® will maximize the capacitance change. 3. Use a probe with an integral ground reference (Reference Probe) when measuring non-conductive fluids in horizontal vessels, non-conductive liquids
RIGID
PROBES
Inactive sheath probes An inactive sheath is a metallic tube that is tightly coupled to the insulation on the probe rod and attached to the process connection. The sheath “deadens” the portion of the probe covered. It is used when a false capacitance could be developed by interference, such as: • Collection of debris in a nozzle when the probe is horizontally mounted • Falling process media entering the vessel • Use with Retractable Probe Assembly (Hot Tap)
Rigid probes consist of a process connection, probe rod, and the seal. The rod may take many forms, depending on the application. The following is a description of some of the more common rigid probe styles: Bare probes Bare rod probes are typically used in non-conductive process media with a dielectric value less than 10 or a conductivity value less than 10 µsiemens/in. Capacitance is measured from the probe through the process media to the vessel wall. Insulated probes Insulated rod probes are used in conductive process media with a dielectric value greater than 10 or a conductivity value greater than 10 µsiemens/in. Capacitance is typically measured from the probe rod through the insulation to the process media, which is at the same potential as the vessel wall for conductive media. Probes are insulated with Teflon®, Halar®, or Kynar®. When you are uncertain about the dielectric constant of your process media, insulated probes are a wise choice.
Probe rod Insulation Probe rod Insulation
Sheath
Bare Probe
Insulated Probe
Inactive Sheath Probe
3
RIGID
PROBES
CONTINUED
Bent probes Bent rod probes have a variety of uses. They can provide vertical configurations when only side mounting is available. They can also provide horizontal configurations when only top mounting is available. When top mounted, the horizontal section of the probe can be used to create an extremely stable setpoint by developing a very large capacitance change with a small level change.
L1
Reference wire A reference wire is spirally wrapped around an insulated probe to provide a “ground” reference where none exists. It must be used selectively: clean, conductive, and low viscosity processes only. Reference rod probes An insulated reference rod is mounted parallel with the sensing probe. It is typically used with corrosive process media in nonmetallic vessels where no metal can be introduced to the process.
L2
Sanitary fitting
Sanitary probes These probes consist of Teflon® (TFE) insulated rods with a Tri-clover style process connection and are certified by 3A. GUARDED PROBES Guarded probes are only used for switch function in applications where buildup can occur. The guarded probes require additional coating rejection circuitry and can only be used with the following Kotron switches: Model 810, Model 811, and Sentinel Model 822. The guarded probe can be cut to length but requires a minimum of 4 inches (100 mm) below the lower probe insulator.
REFERENCE PROBES This classification covers probes that supply the “second plate of the capacitor” in non-metallic vessels, or linearize an existing reference (i.e. horizontal cylindrical vessels). There are three types of referenced probes offered: Stillwell probes A stillwell is a metallic tube into which a probe is inserted. It can be used to minimize the effect of turbulence in a vessel and increases the capacitance gain by bringing the ground reference closer to the probe. Stillwell probes are recommended for vessels with agitators.
4
FLEXIBLE
PROBES
Insulated flexible and bare flexible probes Rigid probes longer than 10 feet (3 m) are difficult to install and to physically move in the field. Flexible probes are the ideal solution for measuring ranges from 10 feet (3 m) up to 150 feet (45 m). Insulated flexible probes can be used with either conductive or non-conductive media. Bare flexible probes can only be used with non-conductive media. Always verify that the selected electronics is powerful enough to operate at these longer distances.
HEAT
EXTENSION
The heat dissipation graph, at left, depicts the maximum temperatures at which the extension can be used effectively. 1. Determine the maximum process temperature in the application and locate on X axis. 2. Determine the maximum ambient temperature surrounding the heat extension and locate on the Y axis. 3. If the intersecting point on the graph is within the shaded area the heat extension will dissipate enough heat to keep the electronics temperature below +160° F (+70° C). NOTE: The heat extension may be used with all rigid probe configurations and flexible probe configuration 8A2-AA1A-0XX. The heat extension can not be used with guarded probes.
10" (254 mm)
HEAT DISSIPATION GRAPH
For use with Heat extension (89-6593-001)
160
Ambient air temperature at heat extension ° F
150 140 130 120 110 100 90 80 70 60 50 0 200 400 600 800 1000
Process temperature ° F
5
CAPACITANCE
GAIN GRAPHS
PICOFARAD
(PF)
The following pages contain capacitance gain graphs which can be used to determine the proper probe/electronics choice for any given application. To use the graphs, follow the steps below. 1. Determine the dielectric of the process medium being measured. If the dielectric is unknown, use a dielectric of 2 for non-conductive media such as hydrocarbons or dry media, and a dielectric value of 80 for water based, conductive liquids (dielectric values are along the X axis). 2. Choose a probe. Because more than one probe will usually work, consider the other application parameters such as temperature, pressure, material compatibility, etc. 3. Find the graph which covers the chosen probe. Choose the curve on the graph which most closely relates to your particular application. 4. Using the chosen curve, determine the amount of pF/inch or pF/centimeter your application will develop (values are on Y axis). 5. Multiply the pF/inch value by the transmitter span needed in the application. 6. Compare total capacitance generated by the probe against the needed zero and span of the Kotron electronics to be used.
pF per inch cm 20 8.0 18 7.2 Stillwell
16 6.4 14 5.6 12 4.8 10 8 6 4 2 0 4.0 3.2 2.4 1.6 0.8 0 1 2 3 4 5 6 7 Dielectric value of media 8 9 10 Tank diameter inches (cm) 4 (10) 12 24 48 96 (30) (60) (120) (240)
Example: Parameters: a. b. c. d. e. f.
Dielectric = 2.0 Probe = Part No. 8AB-AA2B-072 (with stillwell) pF/inch = 2.00 Electronics = Kotron Two-Wire Transmitter Required application span = 72 in. Electronics span = 50 pF min. to 4000 pF maximum (See chart below)
2.00 pF/inch × 72 in = 144 pF The total capacitance is enough to meet the 50 pF minimum span of the electronics.
These charts are meant as an application aid; actual values may differ slightly. Always provide a 10% margin of error to ensure satisfactory performance. These curves represent the probe located in the center of the vessel. If the probe is near one wall of a large vessel, multiply the distance from the vessel wall by 2 (to develop a diameter), choose the closest curve in the chart to your application, and then multiply the resultant pF value × 78%. This will account for the probe not being totally surrounded by the ground reference.
RANGE OF CAPACITANCE ADJUSTMENT
Level Kotron Electronics Model 80-8032 Model 810 Model 811 Point Models 80/81 Narrow Wide High Range Wide Low Range Models 822/832/842 Model 82CE Continuous Model 804 Model 801 Model 805 Zero set point 0 to 350 pF 0 to 500 pF 0 to 1000 pF 0 to 3000 pF 0 to 3000 pF 0 to 3000 pF 0 to 50,000 pF 0 to 1000 pF 0 to 50,000 pF 0 to 50,000 pF 0 to 10,000 pF Span/Differential 0.5 pF, fixed 0.5 pF, fixed 0.5 pF to 700 pF 0.5 pF, fixed 4 to 1500 pF 2 to 1500 pF 0.5 to 50,000 pF 50 to 4000 pF 5 to 50,000 pF, 0.5 pF (setpoint min.) 5 to 50,000 pF, 0.5 pF (setpoint min.) 5 pF to 10,000 pF
6
CAPACITANCE
GAIN GRAPHS
PICOFARAD
8XA-1AXX-XXX or 8XA-4AXX-XXX 8XA-2AXX-XXX or 8XA-5AXX-XXX pF per inch cm 10 4.0
9 8 7 6 5 4 3 2 1 0 3.6 3.2 2.8 2.4 2.0 1.6 1.2 0.8 0.4 0 10 20 30
(PF)
CONTINUED
Capacitance gain for Teflon® and Halar® coated probes
Tank diameter inches (cm) 4 12 24 48 96 (10) (30) (60) (120) (240) Stillwell
Capacitance gain for Kynar coated probes
8XA-3XXX-XXX or 8XA-6XXX-XXX
pF per inch cm 40 16.0 40 50 60 70 80 Dielectric value of media
®
Stillwell Tank diameter inches (cm)
30 12.0
20
8.0
4 12 24 48 96
(10) (30) (60) (120) (240)
10
4.0
00 10 20 30 40 50 60 Dielectric value of media 70 80
Capacitance gain chart for bare probes
8XB-XXXX-XXX or 8XC-XXXX-XXX
pF per inch cm 20 8.0 18 7.2 Stillwell
16 6.4 14 5.6 12 4.8 10 8 6 4 2 0 4.0 3.2 2.4 1.6 0.8 0 1 2 3 4 5 6 7 Dielectric value of media 8 9 10 Tank diameter inches (cm) 4 (10) 12 (30) 24 (60) 48 (120) 96 (240)
7
CAPACITANCE
GAIN GRAPHS
PICOFARAD
(PF)
CONTINUED
Capacitance gain chart for bare flexible probe
8X2-AA1A-XXX 0.19" (5 mm) probe diameter
pF per inch cm 4 1.6
Tank diameter inches (cm) 4 (10) 12 (30)
3
1.2
2
0.8
24 (60) 48 (120) 96 (240)
1
0.4
0
0 1 2 3 4 5 6 Dielectric value of media 7 8
Capacitance gain chart for standard flexible probe
8X1-5A1A-XXX 0.19" (5 mm) probe diameter
pF per inch cm 7 2.8 6 5 4 3 2 1 0 2.4 2.0 1.6 1.2 0.8 0.4 0 10 20 30 40 50 60 Dielectric value of media 70 80 Tank diameter inches (cm) 4 (10) 12 (30) 24 (60) 48 (120) 96 (240)
8
PRESSURE
TEMPERATURE
CURVE
6000
Teflon Halar
5000
Kynar HTHP
4000
Guarded Insulated Flexible
Pressure: psig
3000
2000
1000
0 -100
0
100
200
300
400
500
600
700
800
900
1000
Temperature ° F
PROBE
QUICK
SELECTION
GUIDE
The most popular Kotron probes, handling the majority of applications, are outlined below. Rigid probe lengths are available in inches (second digit A) or centimeters (second digit C). Flexible probe lengths are available in feet (second digit A) or meters (second digit C). Selection charts for all Kotron probes are located on pages 10 through 18. Insulated rigid
3
⁄4" NPT, carbon steel rod, Teflon® insulation and seal
8AA-1A1A-XXX 8CA-1A1A-XXX 8AB-AA1A-XXX 8CB-AA1A-XXX 8AC-AA1A-XXX 8CC-AA1A-XXX 8AA-1A2B-XXX 8CA-1A2B-XXX 8AA-1A1E-XXX 8CA-1A1E-XXX 8A2-AA1A-XXX 8C2-AA1A-XXX 8A1-5A1A-XXX 8C1-5A1A-XXX
Bare rigid
3
⁄4" NPT, 316 stainless steel rod, Teflon® seal
Bare high temperature rigid
3
⁄4" NPT, 316 stainless steel rod, ceramic seal
Stillwell
1" NPT, carbon steel rod, 316 stainless steel stillwell, Teflon® insulation and seal
3
Reference wire rigid
⁄4" NPT, carbon steel rod, Teflon® insulation and seal
Bare flexible
3
⁄4" NPT, 316 stainless steel cable, ceramic seal
Insulated flexible
3
⁄4" NPT, 316 stainless steel cable, Halar® insulation and seal
9
INSULATED
MODEL NUMBER
RIGID
PROBES
Models available for quick shipment, usually within one week after factory receipt of a purchase order, through the Expedite Ship Plan (ESP).
INSULATED RIGID PROBES (for conductive or non conductive media) A complete measuring system consists of a: 1. Kotron electronics (switch or transmitter) 2. Kotron probe 3. Optional heat extension for process temperatures >200° F (95° C): P/N 089-6593-001 (see page 5)
BASIC MODEL NUMBER - INSULATED RIGID PROBE
8AA 8CA FM/CSA/ATEX, probe length in inches FM/CSA/ATEX, probe length in centimeters
ROD, SEAL, AND INSULATION MATERIAL
1A 2A 3A 4A 5A 6A
Carbon steel rod with Teflon® (TFE) insulation max +400° F @ 200 psig/max 3000 psig @ +100° F (max +200° C @ 13.8 bar/max. 205 bar @ +40° C) Carbon steel rod with Halar® (ECTFE) insulation max +200° F @ 50 psig/max 3000 psig @ +100° F (max +95° C @ 3.5 bar/max. 205 bar @ +40° C) Carbon steel rod with Kynar® (PVDF) insulation max +200° F @ 200 psig/max 3000 psig @ +150° F (max +95° C @ 13.8 bar/max. 205 bar @ +65° C) 316/316L SS rod with Teflon® (TFE) insulation max +400° F @ 200 psig/max 3000 psig @ +100° F (max +200° C @ 13.8 bar/max. 205 bar @ +40° C) 316/316L SS rod with Halar® (ECTFE) insulation max +200° F @ 50 psig/max 3000 psig @ +100° F (max +95° C @ 3.5 bar/max. 205 bar @ +40° C) 316/316L SS rod with Kynar® (PVDF) insulation max +200° F @ 200 psig/max 3000 psig @ +150° F (max +95° C @ 13.8 bar/max. 205 bar @ +65° C)
Temperature at electronics should not exceed 160° F (70° C). See pages 17 and 18 for other materials of construction. Minimum temperature -40° F (-40° C)
PROCESS CONNECTION - 316L SS (flange socket welded to probe)
1 2 E 4 5 6 7 8 9 ⁄4" NPT (not available with Configuration Style code B) 1" NPT G1 (1" BSP) thread 1" 150 lbs ANSI RF flange (not available with Configuration Style code B) 11⁄2" 150 lbs ANSI RF flange 2" 150 lbs ANSI RF flange 3" 150 lbs ANSI RF flange 4" 150 lbs ANSI RF flange 1" 300 lbs ANSI RF flange (not available with Configuration Style code B)
Process connections continue on page 11.
3
8 10
A
A
INSULATED
MODEL NUMBER
RIGID
PROBES
PROCESS CONNECTION - 316L SS (flange socket welded to probe)
A B C D H J K L M N S T U V Y Z 11⁄2" 300 lbs ANSI RF flange 2" 300 lbs ANSI RF flange 3" 300 lbs ANSI RF flange 4" 300 lbs ANSI RF flange DN 25 PN 16 DIN 2527 Form B DN 40 PN 16 DIN 2527 Form B DN 50 PN 16 DIN 2527 Form B DN 25 PN 25/40 DIN 2527 Form B DN 40 PN 25/40 DIN 2527 Form B DN 50 PN 25/40 DIN 2527 Form B 1/11⁄2" Sanitary flange, 16 AMP-3A 2" Sanitary flange, 16 AMP-3A 3" Sanitary flange, 16 AMP-3A 4" Sanitary flange, 16 AMP-3A DN 25 Sanitary flange DIN 11851 DN 50 Sanitary flange DIN 11851
flange (not available with Configuration Style code B) flange flange flange (not available with Configuration Style code B) flange flange (available with Configuration Style code A and D only) (available with Configuration Style code A and D only) (available with Configuration Style code A and D only) (available with Configuration Style code A and D only) (available with Configuration Style code A and D only) (available with Configuration Style code A and D only)
CONFIGURATION STYLE
A B C D E Insulated rigid probe Insulated rigid probe with 1" (25 mm) 316 SS diameter stillwell Insulated rigid probe with 6" (15 cm) 316/316L SS inactive sheath Insulated rigid bent probe 90° (specify L1 and L2) (Minimum L1 length is 6.5 inches (16.5 cm); probe insertion length is L1 + L2 – 1") Insulated rigid probe with 316/316L SS reference wire
INSERTION LENGTH specify length in 1 inch (1 cm) increments (specify in digit 2)
6 to 234 inches (15 to 595 centimeters) Example: 6 inches = 006 / 60 centimeters = 060 ESP available 6 to 234 inches in 1 inch increments ( 15 to 595 centimeters in 1 cm increment)
8
A
A 11
BARE
MODEL
RIGID
NUMBER
PROBES
A complete measuring system consists of a: 1. Kotron electronics (switch or transmitter) 2. Kotron probe 3. Optional heat extension for process temperatures >200° F (95° C): P/N 089-6593-001 (see page 5)
BASIC MODEL NUMBER - BARE PROBES
8AB 8CB FM/CSA/ATEX, probe length in inches FM/CSA/ATEX, probe length in centimeters
Bare probes cannot be installed in a hazardous area unless used with an intrinsically safe electronic circuit.
ROD AND SEAL MATERIAL
AA
316 SS rod with Teflon® seal max +400° F @ 200 psig/max 3000 psig @ +100° F (max +200° C @ 13.8 bar/max. 205 bar @ +40° C)
Temperature at electronics should not exceed +160° F (+70° C). See pages 17 and 18 for other materials of construction. Minimum temperature -40° F (-40° C)
PROCESS CONNECTION - 316L SS (flange socket welded to probe)
1 2 E 4 5 6 7 8 9 A B C D H J K L M N ⁄4" NPT (not available with Configuration Style code B) 1" NPT G1 (1" BSP) thread 1" 150 lbs ANSI RF flange (not available with Configuration Style code B) 11⁄2" 150 lbs ANSI RF flange 2" 150 lbs ANSI RF flange 3" 150 lbs ANSI RF flange 4" 150 lbs ANSI RF flange 1" 300 lbs ANSI RF flange (not available with Configuration Style code B) 11⁄2" 300 lbs ANSI RF flange 2" 300 lbs ANSI RF flange 3" 300 lbs ANSI RF flange 4" 300 lbs ANSI RF flange DN 25 PN 16 DIN 2527 Form B flange (not available with Configuration Style code B) DN 40 PN 16 DIN 2527 Form B flange DN 50 PN 16 DIN 2527 Form B flange DN 25 PN 25/40 DIN 2527 Form B flange (not available with Configuration Style code B) DN 40 PN 25/40 DIN 2527 Form B flange DN 50 PN 25/40 DIN 2527 Form B flange
3
CONFIGURATION STYLE
A B D Bare rigid probe Bare rigid probe with 1" (25 mm) 316 SS diameter stillwell (1" process connection minimum) Bare rigid bent probe 90° (specify L1 and L2) (Minimum L1 length is 6.5 inches (16.5 cm); probe insertion length is L1 + L2 – 1")
INSERTION LENGTH specify length in 1 inch (1 cm) increments (specify in digit 2)
6 to 234 inches (15 to 595 centimeters) Example: 6 inches = 006 / 60 centimeters = 060
8 12
B
HIGH BARE
MODEL
TEMPERATURE/HIGH RIGID PROBES
NUMBER
PRESSURE
A complete measuring system consists of a: 1. Kotron electronics (switch or transmitter) 2. Kotron probe 3. Optional heat extension for process temperatures >200° F (95° C): P/N 089-6593-001 (see page 5) BASIC MODEL NUMBER - HIGH TEMPERATURE/HIGH PRESSURE BARE PROBE
8AC 8CC FM/CSA/ATEX, probe length in inches FM/CSA/ATEX, probe length in centimeters
Bare probes cannot be installed in a hazardous area unless used with an intrinsically safe electronic circuit.
ROD AND SEAL MATERIAL
AA
316 SS rod with ceramic seal max +1000° F @ 500 psig/max 5000 psig @ +100° F (max +535° C @ 35 bar/max. 345 bar @ +40° C)
Temperature at electronics should not exceed +160° F (+70° C). See pages 17 and 18 for other materials of construction. Minimum temperature -40° F (-40° C)
PROCESS CONNECTION - 316L SS (flange socket welded to probe)
1 2 E 4 5 6 7 8 9 A B C D H J K L M N ⁄4" NPT (not available with Configuration Style code B) 1" NPT G1 (1" BSP) thread 1" 150 lbs ANSI RF flange (not available with Configuration Style code B) 11⁄2" 150 lbs ANSI RF flange 2" 150 lbs ANSI RF flange 3" 150 lbs ANSI RF flange 4" 150 lbs ANSI RF flange 1" 300 lbs ANSI RF flange (not available with Configuration Style code B) 11⁄2" 300 lbs ANSI RF flange 2" 300 lbs ANSI RF flange 3" 300 lbs ANSI RF flange 4" 300 lbs ANSI RF flange DN 25 PN 16 DIN 2527 Form B flange (not available with Configuration Style code B) DN 40 PN 16 DIN 2527 Form B flange DN 50 PN 16 DIN 2527 Form B flange DN 25 PN 25/40 DIN 2527 Form B flange (not available with Configuration Style code B) DN 40 PN 25/40 DIN 2527 Form B flange DN 50 PN 25/40 DIN 2527 Form B flange
3
CONFIGURATION STYLE
A B D High temperature/high pressure probe High temperature/high pressure probe with 1" (25 mm) 316 SS diameter stillwell (1" process connection minimum) High temperature/high pressure bent probe 90° (specify L1 and L2) (Minimum L1 length is 10.5 inches (27 cm); probe insertion length is L1 + L2 – 1")
INSERTION LENGTH specify length in 1 inch (1 cm) increments (specify in digit 2)
6 to 234 inches (15 to 595 centimeters) Example: 6 inches = 006 / 60 centimeters = 060
8
C
AA 13
GUARDED
MODEL
RIGID
PROBES
NUMBER
GUARDED RIGID PROBES (for use with 811 and 822 only) A complete measuring system consists of a: 1. Kotron electronics 2. Kotron probe
BASIC MODEL NUMBER
8AD 8CD FM/CSA, probe length in inches FM/CSA, probe length in centimeters
ROD, SEAL, AND INSULATION MATERIAL
AA
316/316L SS rod with Ryton® guard insulation max +400° F @ 250 psig/max 3500 psig @ +100° F (max +200° C @ 17 bar/max 240 bar @ +40° C)
Temperature at electronics should not exceed +160° F (+70° C). Minimum temperature -40° F (-40° C)
PROCESS CONNECTION, 316/316L SS
1 E
3
⁄4" NPT G1 (1" BSP) thread
CONFIGURATION STYLE
A 316 SS guarded probe
INSERTION LENGTH (specify in digit 2)
0 0 0 0 1 3 4 9 8 6 5 2 18" for 3⁄4" NPT - 19" for G1 (1" BSP) thread 36" for 3⁄4" NPT - 37" for G1 (1" BSP) thread 45 cm for 3⁄4" NPT - 48 cm for G1 (1" BSP) thread 92 cm for 3⁄4" NPT - 95 cm for G1 (1" BSP) thread
8 14
D
AA
A
0
INSULATED
MODEL
FLEXIBLE
PROBES
Models available for quick shipment, usually within one week after factory receipt of a purchase order, through the Expedite Ship Plan (ESP).
NUMBER
INSULATED FLEXIBLE PROBES (for conductive and non conductive media) A complete measuring system consists of a: 1. Kotron electronics (switch or transmitter) 2. Kotron probe 3. Optional 316 SS Anchor assembly: P/N 032-8814-001 316 SS Weight: P/N 004-4355-001 Kynar® Insulated Weight: P/N 032-8902-001 BASIC MODEL NUMBER - INSULATED FLEXIBLE PROBE
8A1 8C1 FM/CSA/ATEX, insulated flexible probe, probe length in feet FM/CSA/ATEX, insulated flexible probe, probe length in meters
Flexible probes cannot be installed in a hazardous area unless used with an intrinsically safe electronic circuit. Minimum temperature -40° F (-40° C)
CABLE AND SEAL MATERIAL
5A
316 SS cable with Halar® (ECTFE) insulation max +200° F @ 50 psig/max 100 psig @ +160° F (max +95° C @ 3.8 bar/max 7 bar @ +70° C)
Temperature at electronics should not exceed +160° F (+70° C).
PROCESS CONNECTION, 316 SS
1
3
⁄4" NPT
Threaded flanges may be ordered separately.
CONFIGURATION STYLE
A Insulated flexible probe
INSERTION LENGTH specify length in 1 foot (1 meter) increments (specify in digit 2)
10 to 150 feet (3 to 45 meters) Example: 10 feet = 010 / 3 meters = 003 ESP available 10 to 150 feet in 1 foot increments (3 to 45 meters in 1 meter increments)
8
5A1A 15
BARE
MODEL
FLEXIBLE
NUMBER
PROBES
BARE FLEXIBLE PROBES (for conductive and non conductive media) A complete measuring system consists of a: 1. Kotron electronics (switch or transmitter) 2. Kotron probe 3. Optional heat extension for process temperatures >200° F (95° C): P/N 089-6593-001 (see page 5) BASIC MODEL NUMBER - BARE FLEXIBLE PROBE
8A2 8C2
Bare probes cannot be installed in a hazardous area unless used with an intrinsically safe electronic circuit. Minimum temperature -40° F (-40° C)
FM/CSA/ATEX, bare flexible probe, probe length in feet FM/CSA/ATEX, bare flexible probe, probe length in meters
CABLE AND SEAL MATERIAL
AA
316 SS cable, ceramic seal max +1000° F @ 500 psig/max 5000 psig @ +100° F (max +535° C @ 35 bar/max 345 bar @ +40° C)
Temperature at electronics should not exceed +160° F (+70° C).
PROCESS CONNECTION, 316 SS
1
3
Threaded flanges may be ordered separately.
⁄4" NPT
CONFIGURATION STYLE
A Bare flexible probe
INSERTION LENGTH specify length in 1 foot (1 meter) increments (specify in digit 2)
10 to 150 feet (3 to 45 meters) Example: 10 feet = 010 / 3 meters = 003
8
2
AA1A
SPECIALTY
MODEL
PROBES
NUMBER
BASIC MODEL NUMBER - HASTELLOY® C PROBE
8AB-BB1 8CB-BB1
FM/CSA/ATEX, bare rigid Hastelloy C probe with 3⁄4" NPT process connection and Teflon seal, probe length in inches FM/CSA/ATEX, bare rigid Hastelloy C probe with 3⁄4" NPT process connection and Teflon seal, probe length in centimeters
Bare probes cannot be installed in a hazardous area unless used with an intrinsically safe electronic circuit. Temperature at electronics should not exceed +160° F (+70° C). Max +400° F @ 200 psig/max 3000 psig @ +100° F (max +200° C @ 13.8 bar/max 205 bar @ +40° C) Minimum temperature -40° F (-40° C)
CONFIGURATION STYLE
A D Standard probe Bent probe 90° (specify L1 and L2) (Minimum L1 length is 6.5 inches (16.5 cm); probe insertion length is L1 + L2 – 1")
INSERTION LENGTH specify length in 1 inch (1 cm) increments (specify in digit 2)
6 to 234 inches (15 to 595 centimeters) Example: 6 inches = 006 / 60 centimeters = 060
8 16
B
BB1
SPECIALTY
MODEL
PROBES
CONTINUED
NUMBER
SPECIALTY PROBES (for conductive and non conductive media) A complete measuring system consists of a: 1. Kotron electronics (switch or transmitter) 2. Kotron probe 3. Optional heat extension for process temperatures >200° F (95° C): P/N 089-6593-001 (see page 5)
BASIC MODEL NUMBER - REFERENCE ROD PROBES
041-50 041-60 FM/CSA, length in inches ATEX, length in centimeters
Temperature at electronics should not exceed +160° F (+70° C).
ROD MATERIAL, SEAL, AND PROCESS CONNECTION
78
Minimum temperature -40° F (-40° C)
Carbon steel rod with Halar® (ECTFE) insulation, 3" 150 lb carbon steel faced flange max +285° F @ 230 psig/max 285 psig @ +70° F (max +140° C @ 15.8 bar/max 20 bar @ +21° C)
INSERTION LENGTH specify length in 1 inch (1 cm) increments (specify in digit 4)
6 to 234 inches (15 to 595 centimeters) Example: 6 inches = 006 / 60 centimeters = 060
41
078
BASIC MODEL NUMBER - PROBES WITH FACED FLANGE PROCESS CONNECTION
041-50 041-60 CSA, insulated rigid plastic probe, length in inches ATEX, insulated rigid plastic probe, length in centimeters
Temperature at electronics should not exceed +160° F (+70° C).
ROD MATERIAL, SEAL, AND FLANGED PROCESS CONNECTION
6 6 6 6 6 6 0 1 2 3 4 5 Carbon Carbon Carbon Carbon Carbon Carbon steel steel steel steel steel steel rod rod rod rod rod rod with with with with with with Halar® (ECTFE) insulation, 11⁄2"150 Halar (ECTFE) insulation, 11⁄2" 300 Halar (ECTFE) insulation, 2" 150 Halar (ECTFE) insulation, 2" 300 Halar (ECTFE) insulation, 3" 150 Halar (ECTFE) insulation, 3" 300 lb lb lb lb lb lb
carbon carbon carbon carbon carbon carbon
steel steel steel steel steel steel
flange flange flange flange flange flange
with with with with with with
Halar® face Halar face Halar face Halar face Halar face Halar face
INSERTION LENGTH specify length in 1 inch (1 cm) increments (specify in digit 4)
6 to 234 inches (15 to 595 centimeters) Example: 6 inches = 006 / 60 centimeters = 060
Minimum temperature -40° F (-40° C) Max +285° F @ 250 psig/max 285 psig @ +100° F (max +140° C @ 17.2 bar/max 19.6 bar @ +38° C). Max +285° F @ 300 psig/max 740 psig @ +100° F (max +140° C @ 20.7 bar/max 51 bar @ +38° C).
41
0
17
SPECIALTY
MODEL
PROBES
CONTINUED
NUMBER
SPECIALTY PROBES (for conductive and non conductive media) A complete measuring system consists of a: 1. Kotron electronics (switch or transmitter) 2. Kotron probe 3. Optional heat extension for process temperatures >200° F (95° C): P/N 089-6593-001 (see page 5)
Basic Model Number - STILLWELL PROBE (316 SS PROCESS CONNECTION AND STILLWELL)
041-5027 041-6027
FM/CSA/ATEX, 3⁄4" NPT, carbon steel rod with TFE insulation, probe length in inches FM/CSA/ATEX, 3⁄4" NPT, carbon steel rod with TFE insulation, probe length in centimeters
Temperature at electronics should not exceed +160° F (+70° C). Max +300° F @ 500 psig/max 1000 psig @ +70° F (max +150° C @ 34.5 bar/max 70 bar @ +21.1° C) Minimum temperature -40° F (-40° C)
INSERTION LENGTH specify length in 1 inch (1 cm) increments (specify in digit 4)
6 to 234 inches (15 to 595 centimeters) Example: 6 inches = 006 / 60 centimeters = 060
41
027
18
DIMENSIONS
INCHES (MM)
PROBES
Optional Mounting Flange
Optional Mounting Flange
Optional Mounting Flange
3/4" NPT 3/4" NPT Optional Anchor Assembly Insertion Length
3/4" NPT 0.1875 (5) 0.25 (6) 2.95 (75)
Insertion Length
4.87 (124) 5 (127)
0.375 (10)
0.5 (13) - Bare Probe 0.625 (16) - Insulated Probe
Optional Weight 1.32 (34) 0.875 (22) 0.79 (20)
Guarded probe
Rigid probe
Flexible probe
CONNECTIONS
3/4" NPT 150 Grit (32 micro inch) Finish
16 AMP Solid End Cap Type Fitting 1/32" Radius TFE Insulation .625" (16) O.D.
Insertion Length Insertion Length
Insertion Length
Insertion Length
Threaded 3⁄4"-1" NPT
Threaded G1 (1" BSP)
Sanitary Flange
Welded flange ANSI / DIN
19
QUALITY
The quality assurance system in place at Magnetrol guarantees the highest level of quality throughout the company. Magnetrol is committed to providing full customer satisfaction both in quality products and quality service. Magnetrol’s quality assurance system is registered to ISO 9001 affirming its commitment to known international quality standards providing the strongest assurance of product/service quality available.
ESP
Several Kotron Sensing Probes are available for quick shipment, usually within one week after factory receipt of a purchase order, through the Expedite Ship Plan (ESP). Models covered by ESP service are color coded in the selection data charts. To take advantage of ESP, simply match the color coded model number codes (standard dimensions apply). ESP service may not apply to orders of ten units or more. Contact your local representative for lead times on larger volume orders, as well as other products and options.
Expedite Ship Plan
WARRANTY
All Magnetrol electronic level and flow controls are warranted free of defects in materials or workmanship for one full year from the date of original factory shipment. If returned within the warranty period; and, upon factory inspection of the control, the cause of the claim is determined to be covered under the warranty; then, Magnetrol will repair or replace the control at no cost to the purchaser (or owner) other than transportation. Magnetrol shall not be liable for misapplication, labor claims, direct or consequential damage or expense arising from the installation or use of equipment. There are no other warranties expressed or implied, except special written warranties covering some Magnetrol products.
5300 Belmont Road • Downers Grove, Illinois 60515-4499 • 630-969-4000 • Fax 630-969-9489 • www.magnetrol.com 145 Jardin Drive, Units 1 & 2 • Concord, Ontario Canada L4K 1X7 • 905-738-9600 • Fax 905-738-1306 Heikensstraat 6 • B 9240 Zele, Belgium • 052 45.11.11 • Fax 052 45.09.93 Regent Business Ctr., Jubilee Rd. • Burgess Hill, Sussex RH15 9TL U.K. • 01444-871313 • Fax 01444-871317
Copyright © 2008 Magnetrol International, Incorporated. All rights reserved. Printed in the USA.
Performance specifications are effective with date of issue and are subject to change without notice.
Magnetrol & Magnetrol logotype, Magneseal® and Kotron® are registered trademarks of Magnetrol International Hastelloy® is a registered trademark of Haynes International, Inc. Teflon® is a registered trademark of DuPont. Halar® is a registered trademark of Allied Chemical Corp. Kynar® is a registered trademark of Arkema, Inc. Ryton® is a registered trademark of Chevron Phillips Chemical Company
BULLETIN: 50-125.13 EFFECTIVE: November 2008 SUPERSEDES: March 2004