R P T- 1 6 0
160W Triple Output Medical Type
series
■ Features :
Universal AC input / Full range
Built-in active PFC function
Protections: Short circuit / Overload / Over voltage / Over temperature
5"x3" compact size
Free air convection for 100W and 145W with 20.5 CFM forced air
Medical safety approved (2 x MOPP between primary to secondary)
With power good and fail signal output
No load power consumption under 0.75W by PS-ON control (G model)
Standby 5V@0.8A with fan, @0.6A without fan (G model)
Suitable for BF application with appropriate system consideration
3 years warranty
RPT G
G: With 5Vsb & no load power
consumption < 0.75 W
Blank: Basic function (without 5Vsb)
160A
EN 60601-1
IEC 60601-1
8
SPECIFICATION
RPT□-160A
MODEL
OUTPUT
RPT□-160B
RPT□-160C
RPT□-160D
OUTPUT NUMBER
CH1
CH2
CH3
CH1
CH2
CH3
CH1
CH2
CH3
CH1
CH2
DC VOLTAGE
5V
12V
-5V
5V
12V
-12V
5V
15V
-15V
5V
12V
24V
RATED CURRENT (20.5CFM) 14A
5.5A
1A
14A
5A
1A
14A
3.6A
1A
11A
5A
1.2A
CURRENT RANGE (convection) 0.6 ~ 9A
0.2 ~ 3.8A 0.1 ~ 0.6A 0.6 ~ 9A
0.2 ~ 3.4A 0.1 ~ 0.8A 0.6 ~ 9A
0.1 ~ 2.6A 0.1 ~ 0.8A 0.3 ~ 8A
CURRENT RANGE (20.5CFM) 0.6 ~ 14A 0.2 ~ 5.5A 0.1 ~ 1A
0.6 ~ 14A 0.2 ~ 5A
RATED POWER (convection) Note.7 98.6W
98.4W
99W
98.2W
RATED POWER (20.5CFM) Note.8 145W
146W
143W
147.8W
0.1 ~ 1A
CH3
0.2 ~ 2.6A 0.15 ~ 1A
0.6 ~ 14A 0.1 ~ 3.6A 0.1 ~ 1A 0.3 ~ 11A 0.2 ~ 5A
0.15 ~ 1.2A
RIPPLE & NOISE (max.) Note.2 100mVp-p 120mVp-p 120mVp-p 100mVp-p 120mVp-p 120mVp-p 100mVp-p 150mVp-p 150mVp-p 100mVp-p 120mVp-p 200mVp-p
VOLTAGE ADJ. RANGE
CH1:5 ~ 5.5V
VOLTAGE TOLERANCE Note.3 ±2.0% ±5.0%
-5,+7%
±2.0%
±5.0%
-4,+5%
±2.0%
±4.0%
±8.0%
±2.0%
±5.0%
+7,-5%
LINE REGULATION
±0.5%
±1.0%
±1.0%
±0.5%
±1.0%
±1.0%
±0.5%
±1.0%
±1.0%
±0.5%
±1.0%
±1.0%
LOAD REGULATION
±1.5%
±3.0%
-5,+6%
±1.5%
±3.0%
-4,+5%
±2.0%
±3.0%
±8.0%
±1.5%
±3.0%
-3,+4%
SETUP, RISE TIME
1800ms, 30ms/230VAC
16ms/230VAC/115VAC at full load
HOLD UP TIME (Typ.)
VOLTAGE RANGE
INPUT
Note.6 90 ~ 264VAC
47 ~ 63Hz
POWER FACTOR (Typ.)
EFFICIENCY (Typ.)
PF>0.93/230VAC
84%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.8A/115VAC
0.9A/230VAC
COLD START 35A/115VAC
70A/230VAC
PROTECTION OVER VOLTAGE
OVER TEMPERATURE
WORKING HUMIDITY
SAFETY STANDARDS
ISOLATION LEVEL
SAFETY & WITHSTAND VOLTAGE
EMC
ISOLATION RESISTANCE
EMC EMISSION
EMC IMMUNITY
MTBF
NOTE
84%
83%
83%
105 ~ 135% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
CH1: 5.75 ~ 6.75V
Protection type : Shut down o/p voltage, re-power on to recover
TSW1: Shut down o/p voltage, recovers automatically after temperature goes down
TSW2: Shut down o/p voltage, re-power on to recover
5VSB : 5V@0.6A without fan, 0.8A with fan 20.5CFM ; tolerance ± 2%, ripple : 50mVp-p(max.)
5V STANDBY (G model)
PS-ON INPUT SIGNAL (G model) Power on: PS-ON = "Hi" or " > 2 ~ 5V" ; Power off: PS-ON = "Low" or " < 0 ~ 0.5V"
PF>1ms
POWER GOOD / POWER FAIL 500ms>PG>10ms
-20 ~ +70℃ (Refer to "Derating Curve")
WORKING TEMP.
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
OTHERS
PF>0.98/115VAC at full load
Note.9 Earth leakage current < 200 μA/264VAC , Touch current < 100 μA/264VAC
OVERLOAD
(Note 4)
127 ~ 370VDC
FREQUENCY RANGE
LEAKAGE CURRENT
FUNCTION
3500ms, 30ms/115VAC at full load
DIMENSION
PACKING
20 ~ 90% RH non-condensing
-40 ~ +85℃, 10 ~ 95% RH
±0.03%/℃ (0 ~ 50℃)
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1, TUV EN60601-1 approved
Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
I/P-O/P:4KVAC
I/P-FG:2KVAC
O/P-FG:1.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
Compliance to EN55011 (CISPR11), EN55032 (CISPR32) Class B, EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
191.4K hrs min. MIL-HDBK-217F (25℃)
127*76.2*34.6mm (L*W*H)
0.33Kg; 36pcs/12.9Kg/0.79CUFT
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on
a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to
perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com)
5. HS1,HS2 & HS3 can not be shorted.
6. Derating may be needed under low input voltages. Please check the derating curve for more details.
7. The rated power includes 5Vsb @ 0.6A.
8. The rated power includes 5Vsb @ 0.8A.
9. Touch current was measured from primary input to DC output.
File Name:RPT-160-SPEC 2017-07-14
R P T- 1 6 0
160W Triple Output Medical Type
Mechanical Specification
Unit:mm
FAN
AC Input Connector (CN1) : JST B3P-VH or equivalent
5cm
Pin No.
1
2
3
5
20.5CFM min.
SVR1
M1
76.2
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
DC Output Connector (CN2) : JST B8P-VH or equivalent
1
2
66.2
Assignment
AC/L
No Pin
AC/N
CN3
HS1
1
2
3
HS3
FS2
Pin No.
1,2,3,4
5,6
7
8
1
2
3
4
5
6
7
8
HS2
FS1
CN1
series
CN2
M2
5
4 -ψ
3.5
117
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Power Good Connector(CN3):JST B2B-XH or equivalent
CN901
LED
CN901
Assignment
COM
CH1
CH2
CH3
Pin No.
Status
Mating Housing
Terminal
1
2
PG
GND
JST XHP
or equivalent
JST SXH-001T-P0.6
or equivalent
1234
5VSB Connector(CN901) : JST B-XH or equivalent
127
Pin No.
1
2,4
3
PS/ON
34.6
5mA(max.)
GND
Assignment
PS/ON
GND
5VSB
Mating Housing
Terminal
JST XHP
or equivalent
JST SXH-001T
or equivalent
3 max.
1.HS1,HS2,HS3 can not be shorted
2.M1 and M2 are Safety ground and should all be grounded.
Block Diagram
POWER SWITCHING
PWM CONTROL
RECTIFIERS
&
PFC
EMI
FILTER
I/P
FG
POWER
SWITCHING
O.L.P.
fosc :100KHz
+5VSB
DETECTION
CIRCUIT
RECTIFIERS
&
FILTER
RECTIFIERS
&
FILTER
RECTIFIERS
&
FILTER
RECTIFIERS
&
FILTER
PG
DETECTION
CIRCUIT
CONTROL
+5VSB
V3
+V2
+V1
COM
PG
O.V.P.
PS/ON
CONTROL
Derating Curve
PS/ON
Output Derating VS Input Voltage
100
100
With FAN
90
80
80
70
Without FAN
LOAD (%)
LOAD (%)
60
40
20
-20
0
10
20
30
40
50
AMBIENT TEMPERATURE (℃)
60
70
(HORIZONTAL)
60
50
40
90
95
100
115
120
140
160
180
200
220
240
264
INPUT VOLTAGE (VAC) 60Hz
File Name:RPT-160-SPEC 2017-07-14