USB / Ethernet
FCPM-6000RC
Integrated Frequency Counter & Power Meter
50Ω
-30 dBm to +20 dBm, 1 MHz to 6000 MHz
The Big Deal
• Automatically synchronized
power & frequency measurements
• USB and Ethernet control
• Includes GUI with measurement applications
software, simplifying complex measurements
• Measurement speed 30ms
Applications
•
•
•
•
CASE STYLE: JL2029
Included Accessories
Description
Model No.
Production testing systems
Field testing & remote location monitoring
automatic, scheduled data collection
Evaluate high-power, multi-port devices with built-in virtual
couplers/attenuators & other software tools.
Software Package
USB-RJ45-CBL-7+
2.6 ft. USB cable
Qty.
1
RoHS Compliant
See our web site for RoHS Compliance
methodologies and qualifications
Product Overview
Mini-Circuits’ FCPM-6000RC Integrated Frequency Counter & Power Meter is a compact (5.00 x 2.66 x 1.36”) precision test
device controlled via USB or Ethernet (HTTP and Telnet protocols) or operated as standalone test instrument. It simplifies test
setups by enabling synchronized frequency and power measurements from a single device. The unit features an LCD display
allowing convenient readings directly off the measurement head, while our user-friendly GUI software enables easy remote test
management via USB or Ethernet.
Full software support is provided, including our user-friendly GUI application for Windows and a full API and programming
instructions for both Windows and Linux environments (32-bit and 64-bit systems). The latest version of the full software package can
be downloaded from https://www.minicircuits.com/softwaredownload/fcpm.html at any time.
Key Features
Feature
Advantages
USB & Ethernet control
USB HID and Ethernet (HTTP / Telnet) interfaces provide easy compatibility with a wide range of
software setups and programming environments
‘Measurement Application’ GUI
software built-in
Automated measurement setups which allow the user to perform measurements on RF components
such as Couplers, Filters, Amplifiers, etc., display numerical data and graphs, and analyze the data.
LCD character display
LCD character display providing the frequency and power readings, as well as information on the
unit’s status.
Operates with external or internal
reference
The FCPM-6000RC can operate either synchronized with an external 10 MHz reference signal or
independently, using its internal reference depending on user preference.
No calibration required before taking
measurement
The FCPM-6000RC does not require any reference signal for calibration, reducing test time and
simplifying operation.
5V power supply
Powered via USB plug from PC, AC adapter or from commercially available Power Over Ethernet (PoE)
splitter with 5V output.
Trademarks: Windows is a registered trademark of Microsoft Corporation in the United States and other countries. Linux is a registered trademark of Linus Torvalds. Mac is a registered trademark of Apple Corporation. Pentium is a registered trademark of Intel Corporation. Neither Mini-Circuits nor the Mini-Circuits FCPM6000RC are affiliated with or endorsed by the owners of the above referenced trademarks
Rev. D
Mini-Circuits and the Mini-Circuits logo are registered trademarks of Scientific Components Corporation.
Mini-Circuits
®
www.minicircuits.com P.O. Box 350166, Brooklyn, NY 11235-0003 (718) 934-4500 sales@minicircuits.com
M169105
EDR-11250
FCPM-6000RC
RAV
180902
Page 1 of 8
Integrated Frequency Counter & Power Meter
FCPM-6000RC
Electrical Specifications (CW) 1, -30 dBm to +20 dBm, 1 to 6000 MHz
Parameter
Dynamic range 2,3,4
@ 25ºC
@ 0ºC to 50ºC
VSWR
@ -30 to +5 dBm 5,6
Uncertainty of
Power Measurement
@ 25ºC
@ +5 to +15 dBm
@ +15 to +20 dBm
@ -30 to +5 dBm 5,6
Uncertainty of
Power Measurement
@ 0ºC to 50ºC
@ +5 to +15 dBm
@ +15 to +20 dBm
Freq. Range
(MHz)
Min.
Typ.
Max.
Units
1 - 5700
-30
−
+20
5700 - 6000
-28
−
+20
1 - 6000
-27
−
+20
1 - 6000
−
1.05
1.25
:1
1 - 3000
−
± 0.10
± 0.30
dB
3000 - 6000
−
± 0.15
± 0.30
dB
1 - 3000
−
± 0.15
± 0.30
dB
3000 - 6000
−
± 0.15
± 0.30
dB
1 - 3000
−
± 0.20
± 0.40
dB
3000 - 6000
−
± 0.20
± 0.40
dB
1 - 3000
−
± 0.25
−
dB
3000 - 6000
−
± 0.25
−
dB
1 - 3000
−
± 0.20
−
dB
3000 - 6000
−
± 0.20
−
dB
1 - 3000
−
± 0.35
−
dB
dB
dBm
3000 - 6000
−
± 0.30
−
Power Linearity @ 25ºC
1 - 6000
−
± 3.0
−
%
Power Resolution
1 - 6000
0.01
−
−
dB
1 - 40 MHz
−
1
−
Frequency Resolution
Frequency Uncertainty
(@1 sec measurement
sample time) 4,7
Averaging Range
Frequency Aging
Using int. Reference
Measurement Speed
(for power meas.)
@ Low Noise Mode
@ Faster Mode
40-190 MHz
−
10
−
190-6000 MHz
−
100
−
Hz
1 - 40 MHz
−
±2
−
40-190 MHz
−
±20
−
190-1400 MHz
−
±200
−
1400 - 6000 MHz
−
±800
−
1 - 6000
1
−
999
−
−
−
−
±1.0
ppm/year
−
100
−
−
30
−
1 - 6000
Hz
msec
Sample time (for frequency meas.) 8
−
100
1000
3000
msec
Input Impedance
−
−
50
−
Ω
Reference In Frequency
−
−
10
−
MHz
Reference In Impedance
−
−
50
−
Ω
Reference In Power
−
-5
−
+10
dBm
1 - 6000
4.5
5
5.5
V
Current (via USB port, in USB control)
1 - 6000
−
370
500
mA
Current (via USB port, in Ethernet control)
1 - 6000
−
380
500
mA
Supply Voltage
via USB port
Ethernet communication
Supports both Telnet and HTTP protocols over TCP/IP with dynamic(DHCP) or static IP
1
All specifications apply to continuous wave (CW) signals.
continuous safe operational power limit: +23 dBm. Performance is guaranteed up to +20 dBm.
3 The FCPM-6000RC can operate down to -32 dBm, however performance is guaranteed only in the range specified in the table.
4 Minimum power for Frequency measurement at 190-240 MHz may degrade by up to 3 dB due to measurement band switching.
5 When using Faster mode at high frequencies below -20dBm, use of averaging is recommended to prevent noise errors.
6 When using Faster mode power reading below -20dBm, uncertainty value may increase by up to 0.2 dB relative to Low noise mode power reading.
7 Accuracy shown using external 10 MHz reference synchronized to test signal. Using Internal Reference adds 2 ppm of tested frequency to the accuracy
values shown.
8 Software function set by user, default option 1000 mec.
2 Maximum
Mini-Circuits
®
www.minicircuits.com P.O. Box 350166, Brooklyn, NY 11235-0003 (718) 934-4500 sales@minicircuits.com
Page 2 of 8
Integrated Frequency Counter & Power Meter
Absolute Maximum Ratings
Parameter
Connections
Reference Input
Ratings
Operating Temperature
Storage Temperature
6V
15 V
DC Voltage @ Reference Input
RF Power(CW) @ Reference Input
Power & Control
-30°C to 70°C
DC Voltage @ RF port
(BNC-Female)
Signal Input
0°C to 50°C
VUSB Max.
RF Power(CW) @ RF port
FCPM-6000RC
(N-Typ-Male)
(Push-Pull connector)
+25 dBm
25 V
+13 dBm
Permanent damage may occur if any of these limits are exceeded. Operating in the range
between operating power limits and absolute maximum ratings for extended periods of time may
result in reduced life and reliability.
Mini-Circuits
®
www.minicircuits.com P.O. Box 350166, Brooklyn, NY 11235-0003 (718) 934-4500 sales@minicircuits.com
Page 3 of 8
FCPM-6000RC
Integrated Frequency Counter & Power Meter
Typical Performance Curves
VSWR
Uncertainty of Power Measurement @ +25°C
0.2
1.15
0°C
1.12
0.1
UUncertainty (dB)
+25°C
+50°C
VSWR (:1)
1.09
1.06
1.03
0.0
-0.1
-30 dBm
-10 dBm
0 dBm
+5 dBm
+10 dBm
+15 dBm
+20 dBm
-0.2
-0.3
-0.4
1.00
0
1000
2000
3000
4000
5000
0
6000
1000
Frequency (MHz)
2000
3000
4000
Frequency (MHz)
5000
6000
Linearity @ +25OC
8%
-30 to 0 dBm
6%
0 to +20 dBm
Linearity (%)
4%
2%
0%
-2%
-4%
0
1000
2000
3000
4000
5000
6000
Frequency (MHz)
Typ. Input Power Threshhold for Frequency
Measurement @ +25°C
-27
Uncertainty of Frequency Measurement @ +25°C
500
Uncertainty (Hz)
Power In (dBm)
-30 dBm
-20 dBm
-10 dBm
0 dBm
+5 dBm
+15 dBm
+20 dBm
300
-30
-33
-36
100
-100
-300
-39
-500
-42
0
1000
2000
3000
4000
Frequency (MHz)
5000
0
6000
Mini-Circuits
1000
2000
3000
4000
Frequency (MHz)
5000
6000
®
www.minicircuits.com P.O. Box 350166, Brooklyn, NY 11235-0003 (718) 934-4500 sales@minicircuits.com
Page 4 of 8
FCPM-6000RC
Integrated Frequency Counter & Power Meter
Outline Drawing (JL2029)
Note 9
Indicates Grey per
Pantone 427C
Indicates Cutout
P/N: B16-10-174+ REV: ORx1
Outline Dimensions ( inch
mm )
9
A
B
C
D
E
F
G
H
J
K
L
WT.
GRAMS
5.00
(127.0)
2.66
(67.6)
1.36
(34.5)
1.31
(33.3)
0.50
(12.7)
2.300
(58.42)
0.18
(4.6)
4.000
(101.60)
0.50
(12.7)
81
(2057)
20
(508)
500
FCPM-6000RC to be used with the supplied control cable only.
Mini-Circuits
®
www.minicircuits.com P.O. Box 350166, Brooklyn, NY 11235-0003 (718) 934-4500 sales@minicircuits.com
Page 5 of 8
FCPM-6000RC
Integrated Frequency Counter & Power Meter
Connection diagrams
Connection diagram for USB control
RJ45 connector - Not in use for USB control
Push-Pull
connector
Connect to
D.U.T.
USB-RJ45-CBL-7+
USB Type A plug Connect to computer USB socket
Reference input
(connect to external
reference if needed)
Connection diagram for Ethernet control, using power adapter
RJ45 connector - Connected to LAN
network port
USB Type A plug Connect to USB-AC/DC-5
power adaptor
Push-Pull
connector
Connect to
D.U.T.
USB-RJ45-CBL-7+
USB-AC/DC-5
Reference input
(connect to external
reference if needed)
Connection diagram for Ethernet control, using PoE system
Push-Pull
connector
Connect to
D.U.T.
USB-RJ45-CBL-7+
Reference input
(connect to external
reference if needed)
Note: Commercially available PoE
splitter not supplied by Mini-Circuits
RJ45 connector for LAN data
PoE Splitter
RJ45 connector Connected to PoE/LAN port
USB Type A plug for +5V power only
Mini-Circuits
®
www.minicircuits.com P.O. Box 350166, Brooklyn, NY 11235-0003 (718) 934-4500 sales@minicircuits.com
Page 6 of 8
Integrated Frequency Counter & Power Meter
FCPM-6000RC
Software & Documentation Download:
•
Mini-Circuits’ full software and support package including user guide, Windows GUI, DLL files, programming manual and
examples can be downloaded free of charge from https://www.minicircuits.com/softwaredownload/fcpm.html.
Please contact testsolutions@minicircuits.com for support
•
Minimum System Requirements
Parameter
Requirements
Interface
USB HID or HTTP Get/Post or Telnet protocols
System requirements
GUI:
Windows 32 & 64 bit systems from Windows 98 up to Windows 10
USB API (ActiveX & .Net)
Windows 32 & 64 bit systems with ActiveX or .Net support from Windows 98 up to Windows 10
USB direct programming support
Linux, Windows systems from Windows 98 up to Windows 10
HTTP or Telnet
Any computer with a network port and Ethernet-TCP/IP (HTTP or Telnet protocols) support
Hardware
Pentium ® II or higher, RAM 256 MB
Y control cable for USB
and Ethernet (supplied)
Model to be used with the supplied control cable only
Graphical User Interface (GUI) for Windows
Key Features:
•
•
•
•
•
•
•
•
Automatically synchronized power and frequency
measurements.
Relative and Average power measurements
Setting measurement speed for power and frequency
independantly.
Freq. & Power measurment data recording
Measurement application tools
online graphical display of power measurement
USB, HTTP or Telnet control of FCPM
Setting Ethernet configuration
Application Programming Interface (API)
Windows Support:
• API DLL files exposing the full power sensor functionality
• ActiveX COM DLL file for creation of 32-bit programs
• .Net library DLL file for creation of 32 / 64-bit programs
• HTTP Get/Post and Telnet protocols use SCPI commands to provide full control.
• Supported by most common programming environments (refer to application note AN-49-001 for summary of tested
environments)
Linux Support:
• Full power sensor control in a Linux environment is achieved by way of USB interrupt commands.
Mini-Circuits
®
www.minicircuits.com P.O. Box 350166, Brooklyn, NY 11235-0003 (718) 934-4500 sales@minicircuits.com
Page 7 of 8
Integrated Frequency Counter & Power Meter
Ordering Information
Model
FCPM-6000RC
Description
FCPM-6000RC
Integrated Frequency Counter & Power Meter
Included Accessories
Part No.
Description
USB-RJ45-CBL-7+
6.6 ft (2 m) “Y” data cable with USB Type-A and RJ45
plug connectors 9
9 FCPM-6000RC
to be used with the supplied control cable only.
Optional Accessories
Description
USB-AC/DC-5
AC/DC 5VDC Power Adapter with US, EU, IL, UK, AUS, and China power plugs 10,11
USB-RJ45-CBL-7+ (spare)
6.6 ft (2 m) “Y” data cable with USB Type-A and RJ45 plug connectors
NF-SM50+
N-Type Female to SMA Male Adapter (For mating with SMA devices).
NF-SF50+
N-Type Female to SMA Female Adapter
NF-BM50+
N-Type Female to BNC Male Adapter.
10
11
Power plugs for other countries are also available, if you need a power plug for a country not listed in the table please contact testsolutions@minicircuits.
com for support.
The USB-AC/DC-5 may be used to provide the 5VDC power input via USB port if operating with Ethernet control. Not required if using USB control.
Calibration
Description
CAL-FCPM-6000RC
Calibration Service
Click Here
Additional Notes
A. Performance and quality attributes and conditions not expressly stated in this specification document are intended to be excluded and do not form a part of this
specification document.
B. Electrical specifications and performance data contained in this specification document are based on Mini-Circuit’s applicable established test performance criteria and
measurement instructions.
C. The parts covered by this specification document are subject to Mini-Circuits standard limited warranty and terms and conditions (collectively, “Standard Terms”);
Purchasers of this part are entitled to the rights and benefits contained therein. For a full statement of the Standard Terms and the exclusive rights and remedies
thereunder, please visit Mini-Circuits’ website at www.minicircuits.com/MCLStore/terms.jsp
Mini-Circuits
®
www.minicircuits.com P.O. Box 350166, Brooklyn, NY 11235-0003 (718) 934-4500 sales@minicircuits.com
Page 8 of 8