APM-40 SERIES
AC/DC Power Module 40W, Industrial & Medical Safety
FEATURES
►Fully encapsulated Plastic Case
►3 Mounting Versions:
- PCB Mounting with Solder Pins
- Chassis Mounting with Screw Terminals
- DIN-Rail Mounting
► Package Dimension 88.9x63.5x30.0 mm (PCB Version)
►Universal Input 85-264VAC, 47-440 Hz
►Protection Class ll
►Extended Operating Temp.Range -40°C to +60°C at full Load
►LED Output Indicator (Chassis Version Models)
►Industrial Safety to UL/cUL/IEC/EN 60950-1 and UL508
►Medical Safety Approval to UL/cUL/IEC/EN 60601-1 3rd Edition
►Over Load and Over Temperature Protection
►3 Year Product Warranty
PRODUCT OVERVIEW
The new MINMAX APM-40 series is a range of fully encapsulated AC/DC power modules. These high performance products feature an extended operating
temperature range of -40°C to +80°C. Universal input voltage 85-264VAC and UL/IEC/EN safety approvals including medical safety and UL508 listing
qualify these power supplies modules for applications in products with worldwide markets.EMI-filter meets EN55022, class B and FCC,part15, class B.
The APM-40 series power modules provide an economical solution for many space critical applications in commercial, medical and industrial electronic
equipment.
Model Selection Guide
Model Number PCB Mounting
(For model with Chassis
Mounting, add suffix C)
Output
Voltage
VDC
5
12
15
24
±12
±15
APM-40S05
APM-40S12
APM-40S15
APM-40S24
APM-40D12
APM-40D15
Output Current
Input Current
Max.
115VAC, 60Hz
230VAC, 50Hz
@Max. Load
mA
8000
3330
2660
1660
±1660
±1330
mA(typ.)
716
689
680
687
687
680
Max. capacitive
Load
@Max. Load, 115VAC
µF
8000
3900
3900
680
1500#
1000#
# For each output
429
414
408
413
413
408
Efficiency
(typ.)
%
81
84
85
84
84
85
Input Specifications
Parameter
Model
AC Voltage Input Range
Input Frequency Range
All Models
DC Voltage Input Range
No-Load Power Consumption
Inrush Current (Cold Start at 25℃)
E-mail:sales@minmax.com.tw
2014/04/03 REV:0
Typ.
Max.
Unit
85
---
264
VAC
47
---
440
Hz
120
---
370
VDC
---
---
0.3
W
115VAC
---
---
30
A
230VAC
---
---
60
A
Tel:886-6-2923150
Page 1 of 5
Min.
APM-40 SERIES
AC/DC Power Module 40W, Industrial & Medical Safety
Output Specifications
Parameter
Conditions
Output Voltage Accuracy
Line Regulation
Min.
Typ.
--±2.0
--±0.5
Single Output Model
--±1.0
Dual Output Models
--±2.0
No minimum Load Requirement
5VDC Output Models
1.5
--0-20 MHz Bandwidth
Other Output Models
1.0
--Zener diode clamp
120
----±0.02
----85VAC, Hiccup Mode, auto-recovery
105
--(long term overload condition may cause damage)
Hiccup mode, indefinite (automatic recovery)
Load Regulation
Min.Load
Ripple & Noise
Over Voltage Protection
Temperature Coefficient
Overshoot
Current Limitation
Short Circuit Protection
Max.
Unit
---------
%
%
%
%
1.8
1.3
----5
%VPP of Vo
%VPP of Vo
% of Vo
%/℃
%
---
% Inom.
General Specifications
Parameter
Conditions
I/O Isolation Voltage (reinforced)
Leakage Current
I/O Isolation Resistance
Switching Frequency
Min.
500 VDC
115VAC, 60Hz
230VAC, 50Hz
MIL-HDBK-217F@25℃, Ground Benign
Hold-up Time
MTBF (calculated)
Protection Class II
Safety Approvals
Typ.
Max.
Unit
4000
------80
--1000
------130
----25
----80
--200,000
----According IEC/EN 60536
IEC/EN 60950-1, 60601-1 3rd, 2XMOPP
cUL/UL 60950-1, 60601-1 3rd, 2XMOPP, UL 508 listed
VACrms
µA
MΩ
KHz
ms
ms
Hours
EMC Specifications
Parameter
Conducted and radiated EMI
ESD
Radiated immunity
Fast transient
Surge
Conducted immunity
PFMF
Dips
Interruptions
Standards & Level
Performance
EN55011, EN55022, FCC part 15
EN61000-4-2 air ± 8kV , Contact ± 4kV
EN61000-4-3 10V/m
EN61000-4-4 ±2kV
EN61000-4-5 ±1kV
EN61000-4-6 10Vrms
EN61000-4-8 30A/m
EN61000-4-11 30% 10ms
EN61000-4-11 >95% 5000ms
Class B
A
A
A
A
A
A
A
B
Environmental Specifications
Parameter
Temperature Range (operational)
Power Derating
Storage Temperature Range
Over Temperature Protection
Humidity (non condensing)
Cooling
E-mail:sales@minmax.com.tw
2014/04/03 REV:0
Conditions
Min.
Max.
Ambient
Above +60℃
-40℃
+80℃
1.5W / ℃
+95℃
-40℃
Shutdown at 90℃ (automatic recovery at approx. 67℃)
--Free-Air convection
Tel:886-6-2923150
Page 2 of 5
95% rel. H
APM-40 SERIES
AC/DC Power Module 40W, Industrial & Medical Safety
Notes
1
2
3
4
5
6
7
8
9
10
This product is not designed for use in critical life support systems, equipment used in hazardous environment, nuclear control systems or other
such applications which necessitate specific safety and regulatory standards other the ones listed in this datasheet.
Specifications typical at Ta=+25℃, resistive load, 115VAC, 60Hz input voltage, after warm-up time rated output current unless otherwise noted.
Ripple & Noise measured with a 0.1µF/50V MLCC and a 1µF/50V Aluminum electrolytic.
Safety approvals cover frequency 47-63 Hz.
We recommend to protect the converter by a slow blow fuse in the input supply line.
Other input and output voltage may be available, please contact factory.
To order the module with chassis mount package, please add a suffix C (e.g. APM-40S05C).
Part number for DIN-Rail mounting bracket: AC-DIN-02
That “natural convection” is about 20LFM but is not equal to still air (0 LFM).
Specifications are subject to change without notice.
Package Specifications PCB Mounting
Mechanical Dimensions
Pin Connections
5
6
15.9 [0.63]
12.7
[0.50]
15.24
[0.60]
4
0.7
[0.03]
3
81.3 [3.20]
31.6 [1.24]
Pin
Single Output
Dual Output
1
AC (N)
AC (N)
2
AC (L)
AC (L)
3
+Vout
+Vout
4
No Pin
No Pin
5
-Vout
Common
6
No Pin
No Pin
7
NC
-Vout
88.9 [3.50]
7
12.7
[0.50]
1.0
[ 0.04]
15.24
[0.60]
30.0
[1.18]
6.0
[0.24]
NC: No Connection
Bottom View
2
1
5.0
[0.20]
Mounting M4.0 Thread
►All dimensions in mm
►Tolerance: ±0.5
55.88 [2.20]
►Pin
63.5 [2.50]
Physical Characteristics
Case Size
:
88.9x63.5x30.0mm (3.50x2.50x1.18 inches)
Case Material
:
Plastic resin (flammability to UL 94V-0 rated)
Pin Material
:
Copper Alloy with Gold Plate Over Nickel Subplate
Weight
:
310g
E-mail:sales@minmax.com.tw
2014/04/03 REV:0
Tel:886-6-2923150
Page 3 of 5
(inches)
(±0.02)
diameter ⇔ 1.0 ±0.1 (0.04±0.004)
APM-40 SERIES
AC/DC Power Module 40W, Industrial & Medical Safety
Package Specifications Chassis Mounting (order code suffix C)
Mechanical Dimensions
Connections
Indication LED
50.0 [1.97]
7
6
5
4
3
2
Top view
63.8 [2.51]
4 x 3.5
1
DCON
POWER "GOOD" INDICATOR
Pin
Single Output
Dual Output
1
AC (N)
AC (N)
2
AC (L)
AC (L)
3
+Vout
+Vout
4
NC
NC
5
-Vout
Common
6
NC
NC
7
NC
-Vout
100.0 [3.94]
112.0 [4.41]
NC: No Connection
92.0 [3.62]
16.8 [0.66]
5.0 [0.20]
34.1 [1.34]
10.0 [0.39]
Physical Characteristics
Case Size
:
112.0x63.8x34.1mm (4.41x2.51x1.34 inches)
Case Material
:
Plastic resin (flammability to UL 94V-0 rated)
Weight
:
320g
Package Specifications with DIN Rail Mounting Bracket
Mechanical Dimensions
7
6
5
4
3
Indication LED
2
Top view
1
63.8 [2.51]
3.5
50.0 [1.97]
4x
DCON
POWER "GOOD" INDICATOR
100.0 [3.94]
112.0 [4.41]
34.1 [1.34]
11.0 [0.43]
3.0 [0.12]
5.0
[0.20]
10.0 [0.39]
16.8 [0.66]
92.0 [3.62]
Physical Characteristics
Case Size
:
112.0x63.8x34.1mm (4.41x2.51x1.34 inches)
Case Material
:
Plastic resin (flammability to UL 94V-0 rated)
Weight
:
374g
E-mail:sales@minmax.com.tw
2014/04/03 REV:0
Tel:886-6-2923150
Page 4 of 5
►All dimensions in mm
►Tolerance: ±0.5
(inches)
(±0.02)
APM-40 SERIES
AC/DC Power Module 40W, Industrial & Medical Safety
DIN-Rail Mounting Bracket (Order code for Kit : AC-DIN-02)
Minmax Technology Co., Ltd.
18, Sin Sin Road, An-Ping Industrial District, Tainan 702, Taiwan
Tel: 886-6-2923150 Fax: 886-6-2923149 E-mail: sales@minmax.com.tw
2014/04/03 REV:0
Page 5 of 5
很抱歉,暂时无法提供与“APM-40D12”相匹配的价格&库存,您可以联系我们找货
免费人工找货