For gas
Digital Pressure Sensor
DP-101(A) / 102(A) SERIES Ver. 2
Conforming to
EMC Directive
Recognition
Certification
[DP-101(A)/102(A) only]
Pressure Sensor Evolutions
To Come
2013.11
panasonic.net/id/pidsx/global
Achieved further efficiency with
4 upgrades, keeping the same
operability
Note: The upgrade will be implemented from production in October 2013, based on stock status.
Upgrade 1
Superior visibility
Improved visibility in Digital Display
Improvements to the digital display deliver a wide viewing
angle along with increased clarity. The display pressure
range and set pressure range have also been increased.
Old DP-100 series
New DP-100 series
Upgrade 2
Long-distance transmission of analog output
Addition of analog current output capability to multifunctional models
Users can now select either voltage output or current output as analog output according to their
application.
Upgrade 3
Reduced environmental impact
14
% lower
after update
14% lower power consumption
(during normal operation)
Thanks to a redesign of its circuitry, power consumption of
the low-power-consumption DP-100 series during normal
operation has been reduced by 14%. The display is shut
off entirely during ECO / FULL mode operation for power
savings of up to 50% compared to normal operation, and
display brightness is lowered during ECO / STD mode
operation for power savings of up to 30% compared to
normal operation.
Old DP-100 series New DP-100 series
840 mW (normal operation) 720 mW (normal operation)
50
Up to
% lower
in ECO mode
Normal operation
720 mW
ECO / FULL operation
360 mW
Upgrade 4
Enhanced power circuitry
Addition of a reverse polarity protection circuit to the
transistor output circuit
To prevent from breakage due to miswiring.
1
A new global standard
Dual Display
Direct setting
“Current value” and “threshold value”
can be checked at the same time!
Current value
Threshold value
ect setting
Dir
Works like a dial
control type sensor
Main display
Comparative Output 1 operation indicator
Comparative Output 2 operation indicator
Work as analog voltage output operation
indicator in Multi-function type
Mode selection key
Dual display allows direct
setting of threshold value
Decrement key
Increment key
Previous model with single display
Threshold value setting method
Equipped with a 30 mm 1.181 in square
compact-sized dual display. The current
value and the threshold value can be
checked at the same time, so the
threshold value can be set and checked
smoothly without switcing to another
screen mode. ON / OFF operations still
continue while the threshold values are
being set, so setting to the same
sensitivity as dial control-type sensors is
possible. Key lock function is equipped
as well.
Sub display
-91.8
Take a note on current value
Operate MODE key
Adjust threshold value Return to RUN operation
DP-100 with dual display
Simply press a button
Man-hours reduced by
during RUN operation
3/
4
3-color display (Red, Green, Orange)
Readable digital display!
The main display changes color in line with changes in the
status of output ON / OFF operation, and it also changes color
while setting is in progress. The sensor status can therefore be
understood easily, and operating errors can be reduced.
Alphanumeric indication in 12 segments is used. This improved
visual checking.
OFF: Green (or Red)
ON: Red (or Green)
SET: Orange
Main
Display
During normal operation
During setting
2
Copy function helps operation to be accurate and quick
Copy function reduces man-hours and human error
Sensors can be connected to a master sensor one by one, and a copy of the setting details for the master sensor can be transmitted as data
to other sensors. If making the same settings for multiple sensors, this prevents setting errors among other sensors and in addition, when
machinery design are changed, there would be less change in work orders.
Setting details can be copied.
Copying via wiring
Master sensor
Slave sensor
Power supply
+V
+V
0V
0V
Comparative
output 1
Comparative
output 1
Comparative
output 2
Comparative
output 2
Details transmitted
Details received
Detail
s cop
ied
Note: Settings cannot be copied from the new version (Ver. 2) to the old version.
However, settings can be copied from the old version to the new version (Ver. 2).
Advantage
1
Setting man-hours are reduced and sensor setting time is shortened.
4th equipment:
5 minutes
approx.
3rd equipment:
5 minutes
approx.
2nd equipment:
5 minutes
approx.
1st equipment:
Advantage
2
Number of pressure sensor: 1 unit per 1 equipment, 4 equipment per 1 line
Reduction
of lead time
20 seconds
approx.
5 minutes
approx.
The setting details for the master sensor are
retained, so continuous copying is possible.
For 2nd and subsequent lines, the data
transmission time only is required.
4th equipment:
3rd equipment:
2nd equipment:
1st equipment:
Master
sensor
5 minutes
approx.
20 seconds
approx.
4th equipment:
3rd equipment:
2nd equipment:
1st equipment:
Sensor setting time:
Sensor setting time:
Sensor setting time:
20 minutes approx.
5 minutes 20 seconds approx.
20 seconds approx.
Previously
DP-100, 1st line
DP-100, 2nd line
Human operating error is reduced.
Because all details are copied automatically, problems as a result of human error can be
prevented.
Instruction manuals can be updated easily when changes are made to equipment design.
3
Setting is smooth and easy
The sensor’s setting operation mode has a 3-level
configuration to suit the frequency of use.
The setting levels are clearly separated into “RUN mode” for operation
settings that are carried out daily, “MENU SETTING mode” for basic settings,
and “PRO mode” for special and detailed setting. These make setting
operations easy to understand and easy to carry out.
RUN mode
Settings such as threshold value
adjustment and key lock operation
can be carried out while the
sensor is operating.
RUN mode
Simple setting
MENU SETTING mode
MENU SETTING mode
Basic settings such as output
mode setting and NO / NC
switching can be carried out.
PRO mode
Special and
detailed setting
PRO mode
High-level function settings such
as hysteresis adjustment and the
copy function can be carried out.
Displayed in orange while setting is in progress
The display appears in red and green during RUN operation, but it changes to orange while setting is in progress, so that the sensor status
can be viewed at a glance.
RUN operation
RUN mode
While setting
MENU SETTING mode
PRO mode
or
Red or green when output is ON / OFF
Orange while setting is in progress
Default settings that can be used straight away
Buttons with good clicking touch
Easy-to-use default settings are provided for applications that are
used frequently by pressure sensors. The default settings for low
pressure types are ideal for suction confirmation applications, and
those for high pressure types are ideal for checking reference
pressure.
The buttons have a good clicking touch,
allowing smooth setting.
Low pressure type
Comparative output 1
Hysterisys 3digits
ON
High pressure type
Comparative output 1
Ideal for
checking
suction
Hysterisys 3digits
Ideal for
checking
reference
pressure
OFF
–100 kPa
The clicking feeling is
transmitted even through gloves.
–50 kPa
0 kPa
0 MPa
0.5 MPa
1 MPa
Reset function
If a problem ever occurs with the sensor settings, they can be reset
to the default settings.
4
Full range of performance and functions in a compact body
All lineup models are compound pressure
types
High performance accomplished
No sensor settings are required to switch between positive pressure
and negative pressure, so that the number of registered part
numbers can be decreased.
Low pressure type
The low pressure type displays measurements in 0.1 kPa at a
resolution of 1/2,000 and has a response time of 2.5 ms (variable up
to 5,000 ms), ±0.5 % F.S. temperature characteristics and ±0.1 %
F.S. repeatability, achieving high detection performance.
Ideal for applications such
as suction. No misoperations due to vacuum
breakdown occur.
– 100 kPa
For low
pressure
For high
pressure
0 kPa
DP-101Ƒ
Resolution: 1/2,000
Response time: 2.5 ms
Temperature characteristics: ±0.5 % F.S.
Repeatability: ±0.1 % F.S.
100 kPa
Ideal for applications such
as checking reference
pressure. Can also be used
for simple suction.
Displays measurements in 0.1 kPa
1 MPa
DP-102Ƒ
Equipped with independent dual output
Standard type
Equipped with two independent comparative outputs, and separate
sensing modes can be selected for each of them. Since there are
two comparative outputs, one of the comparative outputs can even
be used for alarm output. In addition, output, which is not being
used, can be disabled.
Three output modes are suitable for a wide
range of applications
1
– 100 kPa
Comparative output
0
ON
OFF
H: Fixed
Notes: 1) Hysteresis can be fixed to one of eight different levels.
2) “
” appears in the sub display for comparative output 1,
and “
” appears for comparative output 2.
2
Vacuum breakdown checking
P-1
0 kPa
Comparative output 1 ON
Hysteresis
mode
OFF
Hysteresis mode
This mode is used for setting comparative output hysteresis to
the desired level and for carrying out ON / OFF control.
P-2
3
Lo-1
Hi-1
P-2
H: 1 digit or more
” appears in the sub display for comparative
” or “
” appears for comparative output 2.
Window comparator mode
This mode is used for setting comparative output ON and OFF
at pressures within the setting range.
Pressure
ON
Comparative output 2
EASY mode
OFF
H (Hysteresis)
Note: “
” or “
output 1, and “
Reference
pressure
checking
Alarm output
Hi
Lo
0
ON
Comparative output
OFF
Ɣ Reference pressure alarm output is possible during
reference pressure checking!
0 kPa
H (Hysteresis)
P
Suction checking
Pressure
ON
Comparative output 2
EASY mode
OFF
Pressure
Ɣ Vacuum breakdown can also be notified during
suction applications!
ON
Comparative output 1
EASY mode
OFF
EASY mode
This mode is used for comparative output ON / OFF control.
H (Hysteresis)
Hi
Lo
0
ON
Comparative output
OFF
H (Hysteresis)
H: Fixed
Notes: 1) Hysteresis can be fixed to one of eight different levels.
” or “
” appears in the sub display for comparative
2) “
output 1, and “
” or “
” appears for comparative output 2.
5
Equipped with auto-reference / remote zero-adjustment functions,
More precise pressure management is achieved with a minimum of effort
If the reference pressure of the
device changes, two functions
are selectable. One is
auto-reference function, which
partially shift the comparative
output judgment level by the
amount that the reference
pressure shifts. The other is
remote zero-adjustment
function, which can reset the
display value to zero via
external input. These functions
are ideal for places where the
reference pressure fluctuates
wildly, or where fine settings are
required.
Multi-function type
Without auto-reference and remote zero-adjustment functions
Comparative output: Window comparator mode
Hi-1 30, Lo-1 25
Fixed set value
40
NG?
30
Leak threshold level
25
20
OK
NG?
15
Variation in the filling pressure
Trial 1
Trial 2
Trial 3
Pressure
0
Time
Because the threshold level is fixed for conventional pressure sensors,
changes in the reference pressure result in wrong decisions.
With auto-reference function applied
With remote zero-adjustment function applied
Comparative output: Window comparator mode
Hi-1 0, Lo-1 – 5
Comparative output: Window comparator mode
Hi-1 0, Lo-1 – 5
Sets the absolute threshold level
Sets the absolute threshold level
Auto-reference input
Remote zero-adjustment input
The display is forced to “0”, and only the
filling pressure drop range is displayed.
The display remains at “30” and
only the threshold level is changed.
30
OK
25
20
Threshold level after applying
auto-reference input
15
30
0
25
–5
OK
20
Threshold level after applying
remote zero-adjustment input
15
Threshold level before applying
remote zero-adjustment input
Threshold level before applying
auto-reference input
Trial 1
Pressure 0
Trial 1
Pressure 0
Auto-reference input value
–5
Time
When auto-reference input is applied, the reference pressure “30” is
added to the threshold level. If the reference pressure changes to “20” or
“40”, the auto-reference input compensates for this every time by changing
the threshold level, so any variation in the filling pressure can be ignored.
–5
Displayed when remote zeroadjustment input is applied
Time
When remote zero-adjustment input is applied, the reference pressure is forced to “0”.
If the reference pressure changes to “20” or “40”, the remote zero-adjustment input
adjusts the reference pressure to “0” every time the reference pressure changes, so
any variation in the filling pressure can be ignored.
Peak hold and Bottom hold functions
Energy-saving design! Equipped with an ECO mode
The peak values and bottom values for fluctuating pressures can be
displayed using the dual display.
This mode lowers the display luminance to cut power consumption
by approximately 30 %. The displays can also be turned off
completely to achieve a power saving of approximately 40 %.
Power consumption
Normal
Approx.
Power consumption
Approx.
ECO-STD 30 % less
ECO-FULL 40 % less
Current consumption for
24 V power supply: 25 mA or less
Current consumption for
24 V power supply: 20 mA or less
Blinks alternately
Current consumption for
24 V power supply: 35 mA or less
6
Other useful functions
Sub display can be customized
Setting details can be recognized at a glance
The sub display can be set to indicate any other desired values or
letters apart from the threshold value. This eliminates the need for
tasks such as affixing a label to the device to indicate the normal
pressure value.
The DP-100 setting details appear in the digital display. Because
the settings are in numeric form that can be easily understood, it is
useful such as when receiving technical support by telephone.
What are the check
codes for the setting?
The code on the main display is 3205.
The sub display shows 7011.
2nd digit: Setting status for
comparative output 2
3rd digit: Threshold value setting
1st digit: Setting status for
comparative output 1
4th digit: Display color setting
Indicates desired values and letters
5th digit: Response time
setting
8th digit: ECO mode setting
7th digit: Display refresh rate setting
6th digit: Display unit setting
Cable can be connected with one-touch
Connector attached cable (2m 6.562 ft), as an accessory, can be
connected easily with one-touch connection.
Ɣ Types without connector attached
cable are also available
DP-10Ƒ-J
Commercially-available connectors can be used for cable
connections. Cables in required length can be used, so this
contributes to reduction in waste of unwanted cables.
Environmentally
friendly
Commercially-available
parts can be used!
Connection cable *
* Options: 1 m 3.281 ft / 3 m 9.843 ft / 5 m 16.404 ft
types are also available.
Ɣ M8 plug-in connector types are also available (Only for Europe)
Mating cable
CN-24A-C2 (cable length 2 m 6.562 ft)
CN-24A-C5 (cable length 5 m 16.404 ft)
M8 connector
G 1/8 male thread
M5 female thread
7
DP-11Ƒ-E-P-J
Installation is also easy!
Tight installation to panels is possible
An exclusive mounting bracket that is suitable for 1 to 6 mm 0.039
to 0.236 in panel thickness is available.
Tight installation is possible
Neat and
space
saving
Front protection
cover available
MS-DP1-3
A single mounting hole!
Ɣ An exclusive mounting bracket that supports tight installation is available
Space savings can also be achieved even when an L-shaped mounting bracket is used.
MS-DP1-1
Tight installation is possible
MS-DP1-5
Ceiling mounting
Floor mounting
Positioning bosses for easier
mounting bracket installation
Rear mounting
Short pressure port type is lightweight and takes up little space
Space saving!
Light weight of 30 g! *
Compact size with a depth of only 30 mm 1.181 in, so that it can
easily fit into narrow spaces.
10 g lighter than standard types. This reduces the loads on movable
parts such as robot arms.
Short pressure
port type DP-10Ƒ-M
Standard type
M5 female thread+R1 8 male thread
Elbow joint
* Excluding cables with connector attached
Short pressure port type
Shortened to
12.5 mm 0.492 in !
Light weight
30g approx.
30 mm
1.181 in
Ideal for clean environments!
Stainless steel (SUS303) which does not rust or generate gas is
used as the port material.
M5 female thread [Material: Stainless steel (SUS)]
* The illustration shows connection using an elbow joint. The elbow joint is sold separately.
8
DP-100
Flat installation on the wall by shifting the direction
For short pressure port type
of the pressure port
Rc1/8 conversion bushing is available. Compatible
For short pressure port type
with previous model
By mounting the flat attachment
to DP-10Ƒ-M(-P), pressure port
and cable can now be pulled out
in downward, left or right
directions. Flat mounting on
surfaces such as the wall is
made possible.
By equipping the push-in converter with DP-10Ƒ-M(-P), pressure
port can be converted from M5 female thread to Rc1/8 female thread.
Bore diameter conversion to the DP2 / DP3 series is possible.
Previous model DP2 / DP3
series can be switched over to
DP-100 series.
Previous model
Flat attachment
06'3)Ƒ
20 mm
0.787 in pitch
Model No.
MS-DP1-FM
M5 female thread
Rc1/8 female thread
20 mm
0.787 in pitch
Conversion bushing
MS-DP1-7
Pressure port
M5 female thread
MS-DP1-FR
Rc1/8 female thread
MS-DP1-FN
NPT1/8 female thread
MS-DP1-FE
G1/8 female thread
3 possible directions to mount
ORDER GUIDE
Type
Asia
Standard
Multi-function
M8 plug-in connector type
Europe
Multi-function
Standard
Multi-function
North America
Standard pressure port type
Standard
Appearance
Model No.
For low pressure
í100.0 to +100.0 NPa
DP-101
For high pressure
í0.100 to +1.000 MPa
DP-102
For low pressure
í100.0 to +100.0 NPa
DP-101A
For high pressure
í0.100 to +1.000 MPa
DP-102A
For low pressure
í100.0 to +100.0 NPa
DP-101-E-P
For high pressure
í0.100 to +1.000 MPa
DP-102-E-P
For low pressure
í100.0 to +100.0 NPa
For high pressure
í0.100 to +1.000 MPa
For low pressure
í100.0 to +100.0 NPa
DP-111-E-P-J
For high pressure
í0.100 to +1.000 MPa
DP-112-E-P-J
For low pressure
í100.0 to +100.0 NPa
For high pressure
í0.100 to +1.000 MPa
For low pressure
í100.0 to +100.0 NPa
Standard
For high pressure
For low pressure
* CN-14A-C2
Connector attached
cable 2 m 6.562 ft
is attached.
Excluding M8 plug-in
connector type
(
(
Multi-function
For high pressure
For low pressure
í0.100 to +1.000 MPa
)
í100.0 to +100.0 NPa
)
í0.100 to +1.000 MPa
í100.0 to +100.0 NPa
Standard
For high pressure
í0.100 to +1.000 MPa
Asia
Short pressure port type
Rated pressure range
For low pressure
í100.0 to +100.0 NPa
Multi-function
For high pressure
í0.100 to +1.000 MPa
Pressure port
Comparative output
M5 female thread
+
NPN open-collector transistor
R 1/8
male thread
M5 female thread
+
PNP open-collector transistor
G 1/8
DP-101A-E-P
male thread
DP-102A-E-P
M5 female thread
+
PNP open-collector transistor
G 1/8
DP-111A-E-P-J
male thread
DP-112A-E-P-J
DP-101-N
NPN open-collector transistor
DP-101-N-P
PNP open-collector transistor
NPN open-collector transistor
M5 female thread
PNP open-collector transistor
+
NPT 1/8
NPN open-collector transistor
DP-101A-N
male thread
PNP open-collector transistor
DP-101A-N-P
DP-102-N
DP-102-N-P
DP-102A-N
NPN open-collector transistor
DP-102A-N-P
PNP open-collector transistor
DP-101-M
NPN open-collector transistor
DP-101-M-P
PNP open-collector transistor
DP-102-M
NPN open-collector transistor
DP-102-M-P
M5 female thread
PNP open-collector transistor
DP-101A-M
NPN open-collector transistor
DP-101A-M-P
PNP open-collector transistor
DP-102A-M
NPN open-collector transistor
DP-102A-M-P
PNP open-collector transistor
Type without connector attached cable
Type without connector attached cable CN-14A-C2 is available. When ordering this type, suffix “-J” to the end of Model No.
(e.g.) Type without connector attached cable of DP-101-N is “DP-101-N-J”
Accessory
CN-14A-C2
(Connector attached cable 2 m 6.562 ft)
9
DP-100
OPTIONS
Connector attached cable
Designation
Model No.
Description
&1$&Ƒ
&1$5&Ƒ
CN-14A-C1
CN-14A-C2 (Note)
CN-14A-C3
CN-14A-C5
CN-14A-R-C1
CN-14A-R-C2
CN-14A-R-C3
CN-14A-R-C5
Length: 1 m 3.281 ft
M8 connector
attached cable
CN-24A-C2
Length: 2 m 6.562 ft
CN-24A-C5
For M8 plug-in connector type
The connector on one end
Length: 5 m 16.404 ft Cable outer diameter: ø4 mm ø0.157 in
Connector
CN-14A
Set of 10 housings and 40 contacts
Sensor
mounting
bracNet
MS-DP1-1
Allows sensors to be installed on the flooring or ceiling.
Multiple sensors can also be mounted closely.
MS-DP1-5
Allows sensors to be installed on the wall. Multiple sensors
can also be mounted closely.
MS-DP1-2
Allows installation to panels with thicNness of 1 to 6 mm 0.039
to 0.236 in. Multiple sensors can also be mounted closely.
MS-DP1-4
Allows replacement from DP2 / DP3 series to DP-100 series. For newly designed
set-up, please use panel mounting bracNet MS-DP1-2 for panel mounting.
Front protection
cover
MS-DP1-3
Protects the adjustment surfaces of sensors.
(Can be attached when using the panel mounting bracNet)
Conversion
bushing
MS-DP1-7
By equipping with DP-10Ƒ-M(-P), pressure port can be converted to
Rc1/8 female thread. Replacement from DP2 / DP3 series is possible.
Flat
attachment
MS-DP1-FM
MS-DP1-FR
MS-DP1-FN
MS-DP1-FE
Connector
attached cable
Connector
attached cable
(Flexible cable)
Panel mounting
bracNet
Length: 2 m 6.562 ft
Length: 3 m 9.843 ft
0.2 mm2 4-core cabtyre cable with
connector on one end
Cable outer diameter: ø3.7 mm ø0.146 in
Length: 5 m 16.404 ft
Length: 1 m 3.281 ft
Length: 2 m 6.562 ft
Length: 3 m 9.843 ft
0.2 mm2 4-core flexible cabtyre cable
with connector on one end
Cable outer diameter: ø3.7 mm ø0.146 in
Length: 5 m 16.404 ft
M8 connector attached cable
&1$&Ƒ
31.4
1.236
ø9
ø0.354
(Unit: mm in)
M5 female thread
Pressure port and cable can now be
pulled out in downward, left or right
NPT1/8 female thread directions. Flat mounting on surfaces
such as the wall is made possible.
G1/8 female thread
Rc1/8 female thread
Sensor mounting bracket
MS-DP1-1
M3 (length 6 mm 0.236 in)
screws with washers
Accessory for
MS-DP1-1
Note: The connector attached cable CN-14A-C2 is supplied with the DP-100 series.
(Excluding M8 plug-in connector type).
(
MS-DP1-2
MS-DP1-3
MS-DP1-4
DP2 / DP3
Front protection cover
DPX-04 (optional)
can be installed on
MS-DP1-4.
)
Sensor mounting
bracNet
MS-DP1-1
Panel mounting bracket, Front protection cover
DP
2-2
0
ñ101
.3kP
a
OUT
1
OUT
2
0-AD
J
Front protection cover
MS-DP1-3
MOD
E
Remove
Remove
MS-DP1-5
Insert
Panel mounting bracNet
MS-DP1-2
M3 (length 6 mm 0.236 in)
screws with washers
Accessory for
MS-DP1-5
Mounting holes
for DP2 / DP3 series
can be used as is.
DP-100
(
)
Panel mounting bracNet MS-DP1-4
Sensor mounting
bracNet
MS-DP1-5
Flat attachment
MS-DP1-FM
ø4
ø0.157
MS-DP1-FR
MS-DP1-FN
MS-DP1-FE
Net weight: MS-DP1-FM 15g approx.
MS-DP1-FR/FN/FE 25g approx.
Two M3 (length 8 mm 0.315 in) screws,
two M4 (length 20 mm 0.787 in) screws are attached.
Conversion bushing
MS-DP1-7
M5 female
thread
Rc1/8 female
thread
Recommended connector
Contact: SPHD-001T-P0.5, Housing: PAP-04V-S
(Manufactured by J.S.T. Mfg. Co., Ltd.)
Note: Contact the manufacturer for details of the recommended products.
Recommended crimping tool
Model No.: YC-610R
(Manufactured by J.S.T. Mfg. Co., Ltd.)
Conversion bushing
MS-DP1-7
Note: Contact the manufacturer for details of the recommended products.
10
DP-100
SPECIFICATIONS
Standard
Item
Model No.
Type
For low pressure
DP-101(-M)(-P)
DP-101-E-P
DP-111-E-P-J
DP-101-N(-P)
Asia (Note 2)
Europe
M8 plug-in connector type
North America (Note 2)
Multi-function
For low pressure
For high pressure
DP-101A(-M)(-P)
DP-102A(-M)(-P)
DP-101A-E-P
DP-102A-E-P
DP-111A-E-P-J
DP-112A-E-P-J
DP-101A-N(-P)
DP-102A-N(-P)
For high pressure
DP-102(-M)(-P)
DP-102-E-P
DP-112-E-P-J
DP-102-N(-P)
Type of pressure
Gauge pressure
í100.0 to +100.0 kPa
í101.0 to +101.0 kPa
í1.030 to +1.030 kgf/cm2
í1.010 to +1.010 bar
í14.64 to +14.64 psi
í757 to +757 mmHg
í29.8 to 29.8 inHg
Rated pressure range
Set pressure range
500 kPa
Pressure withstandability
í0.100 to +1.000 MPa
í0.101 to +1.010 MPa
í101 to +1,010 kPa
í1.03 to +10.30 kgf/cm2
í1.01 to +10.10 bar
í14.6 to +146.4 psi
í100.0 to +100.0 kPa
í101.0 to +101.0 kPa
í1.030 to +1.030 kgf/cm2
í1.010 to +1.010 bar
í14.64 to +14.64 psi
í757 to +757 mmHg
í29.8 to 29.8 inHg
1.5 MPa
í0.100 to +1.000 MPa
í0.101 to +1.010 MPa
í101 to +1,010 kPa
í1.03 to +10.30 kgf/cm2
í1.01 to +10.10 bar
í14.6 to +146.4 psi
500 kPa
1.5 MPa
Applicable fluid
Non-corrosive gas
Selectable unit
For low pressure: kPa, kgf/cm2, bar, psi, mmHg, inHg, For high pressure: MPa, kPa, kgf/cm2, bar, psi
12 to 24 V DC ±10 % Ripple P-P 10 % or less
Supply voltage
Power consumption
Comparative output
Normal operation: 720 mW or less (Current consumption 30 mA or less at 24 V supply voltage)
ECO mode: 480 mW or less at STD (Current consumption 20 mA or less at 24 V supply voltage)
360 mW or less at FULL (Current consumption 15 mA or less at 24 V supply voltage)
NPN open-collector transistor
PNP open-collector transistor
Ma[imum source current: 100 mA
Ma[imum sink current: 100 mA
Applied voltage: 30 V DC or less (between comparative output and +V)
Applied voltage: 30 V DC or less (between comparative output and 0 V)
Residual voltage: 2 V or less (at 100 mA sink current)
Residual voltage: 2 V or less (at 100 mA source current)
Output operation / Output modes
NO / NC (selectable by key operation) / EASY mode / Hysteresis mode / Window comparator mode
Hysteresis
Minimum 1 digit (variable) (however, 2 digits when using psi unit)
Repeatability
±0.1 % F.S. (within ±2 digits)
Response time
±0.2 % F.S. (within ±2 digits)
±0.1 % F.S. (within ±2 digits)
±0.2 % F.S. (within ±2 digits)
2.5 ms, 5 ms, 10 ms, 25 ms, 50 ms, 100 ms, 250 ms, 500 ms, 1,000 ms, 5,000 ms, selectable by key operation
Short-circuit protection
Incorporated
ON voltage: 0.4 V DC or less
OFF voltage: 5 to 30 V DC, or open
,nput impeGance: 0 Nȍ appro[.
Input time: 1 ms or more
E[ternal input (Note 3)
Auto-reference function /
Remote zero-adjustment
function
ON voltage: 5 V to +V DC
OFF voltage: 0.6 V DC or less, or open
Input impeGance: 10 Nȍ appro[.
Input time: 1 ms or more
Analog voltage output (Note 3)
Output voltage: 1 to 5 V DC
Zero point: within 3 V ±5 % F.S.
Span: within 4 V ±5 % F.S.
Linearity: within ±1 % F.S.
Output impeGance: 1 Nȍ appro[.
Output voltage: 0.6 to 5 V
Zero point: within 1 V ±5 % F.S.
Span: within 4.4 V ±5 % F.S.
Linearity: within ±1 % F.S.
Output impeGance: 1 Nȍ appro[.
Analog current output (Note 3)
Output current: 4 to 20 mA
Zero point: 12 mA ±5 % F.S.
Span: 16 mA ±5 % F.S.
Linearity: within ±1 % F.S.
LoaG resistance: 250 ȍ (ma[.)
Output current: 2.4 to 20 mA
Zero point: 4 mA ±5 % F.S.
Span: 17.6 mA ±5 % F.S.
Linearity: within ±1 % F.S.
LoaG resistance: 250 ȍ (ma[.)
Display
4 digits + 4 digits 3-color LCD display (Display refresh rate: 250 ms, 500 ms, 1,000 ms, selectable by key operation)
í101.0 to +101.0 kPa
í101.0 to +101.0 kPa
í0.101 to +1.010 MPa
í0.101 to +1.010 MPa
í1.030 to +1.030 kgf/cm2
í101 to +1,010 kPa
í1.030 to +1.030 kgf/cm2
í101 to +1,010 kPa
í1.010 to +1.010 bar
í1.010 to +1.010 bar
í1.03 to +10.30 kgf/cm2
í1.03 to +10.30 kgf/cm2
í14.64 to +14.64 psi
í14.64 to +14.64 psi
í1.01 to +10.10 bar
í1.01 to +10.10 bar
í757 to +757 mmHg
í757 to +757 mmHg
í14.6 to +146.4 psi
í14.6 to +146.4 psi
í29.8 to 29.8 inHg
í29.8 to 29.8 inHg
Displayable pressure range
Environmental resistance
Indicator
(
Orange LED
Comparative output 1 operation indicator, comparative output 2 operation indicator:
Lights up when each comparative output is ON
Protection
Ambient temperature
Ambient humidity
Voltage withstandability
Orange LED
Comparative output 1 operation indicator: Lights up when comparative output is ON,
Analog voltage output operation indicator: Lights up when setting
)(
)
IP40 (IEC)
í10 to +50 °C +14 to +122 °F, Storage: í10 to +60 °C +14 to +140 °F
35 to 85 % RH (No dew condensation or icing allowed), Storage: 35 to 85 % RH
1,000 V AC for one min. between all supply terminals connected together and enclosure
Insulation resistance
50Mȍ or more with 500 V DC megger between all supply terminals connected together and enclosure
Vibration resistance
10 to 500 Hz frequency, 3 mm 0.118 in amplitude, in X, Y and Z directions for two hours each (when panel is mounted: 10 to 150 Hz frequency, 0.75 mm 0.030 in amplitude, in X, Y and Z directions for two hours each)
Shock resistance
100 m/s2 acceleration (10 G appro[.) in X, Y and Z directions for three times each
Temperature characteristics
Within ±0.5 % F.S. (at +20 °C +68 °F)
Pressure port
Asia: M5 female thread + R (PT) 1/8 male thread >e[cluding DP-Ƒ-M(-P)], Europe: M5 female thread + G 1/8 male thread, North America: M5 female thread + NPT 1/8 male thread
Within ±1 % F.S. (at +20 °C +68 °F)
Within ±0.5 % F.S. (at +20 °C +68 °F)
Within ±1 % F.S. (at +20 °C +68 °F)
Material
Enclosure: PBT (glass fiber reinforced), LCD display: Acrylic, Pressure port: Stainless steel (SUS303) , Mounting threaded part: Brass (nickel plated), Switch part: Silicone rubber
Connecting method / Cable length Connector / Total length up to 100 m 328.084 ft (less than 30 m 98.425 ft when conforming to CE marking) is possible with 0.3 mm2, or more, cable
Weight
Accessories
Net weight: 40 g appro[. (DP-10Ƒ-M(-P): 30 g appro[.), Gross weight: 135 g appro[. (DP-10Ƒ-M(-P):125 g appro[.)
CN-14A-C2 (Connector attached cable 2 m 6.562 ft): 1pc. (e[cluding M8 plug-in connector type)
Notes: 1) Where measurement conditions have not been specified precisely, the conditions used were an ambient temperature of +20 °C +68 °F.
2) Model Nos. of Asia type having “-M´ are short pressure port type. Model Nos. of Asia and North America types having the suffi[ “-P” are PNP output type.
3) Cannot be used at the same time.
11
DP-100
I/O CIRCUIT AND WIRING DIAGRAMS
DP-10Ƒ
NPN output type
Terminal No.
I/O circuit diagram
Standard type
D2
2
Tr1
ZD1
D3
3
5V
(Brown) +V
(Black)
Load
Comparative output 1
Load
100 mA ma[.
(White) Comparative output 2
+
í
12 to 24 V DC
±10 %
Sensor circuit
D1
Sensor circuit
Color code of connector attached cable
Color code of connector attached cable
1
Tr2
Multi-function type
Terminal No.
Tr1
Tr2
3
D2
1kȍ
2
Internal circuit
(Blue) 0 V
Internal circuit
Users’ circuit
Load
(Black) Comparative output 1
+
í
12 to 24 V DC
±10 %
100 mA ma[.
4
4
(White) Analog output
or E[ternal input
ZD1
100 mA ma[.
ZD2
1
D1
D3
(Brown) +V
(Blue) 0 V
Users’ circuit
Symbols D1, D2 : Reverse supply polarity protection diode
ZD1 : Surge absorption zener diode
Tr1 : PNP input transistor
Tr2 : NPN output transistor
Symbols D : Reverse supply polarity protection diode
ZD1, ZD2 : Surge absorption zener diode
Tr1, Tr2 : NPN output transistor
Terminal arrangement diagram
Terminal
Designation
1
+V
Comparative output 1
2
2
3
DP-10Ƒ-P
4
3
Standard type: Comparative output 2
Multi-function type: Analog output or E[ternal input
4
0V
PNP output type
Terminal No.
I/O circuit diagram
Multi-function type
Terminal No.
Standard type
Color code of connector attached cable
1
Sensor circuit
Tr1
100 mA ma[.
ZD2
Tr2
2
(Black) Comparative output 1
100 mA ma[.
Load
3
(White) Comparative output 2
Load
(Blue) 0 V
4
Internal circuit
(Brown) +V
D1
D
ZD1
Color code of connector attached cable
1
(Brown) +V
Users’ circuit
Symbols D : Reverse supply polarity protection diode
ZD1, ZD2 : Surge absorption zener diode
Tr1, Tr2 : PNP output transistor
+
í
12 to 24 V DC
±10 %
Sensor circuit
1
ZD1
Tr1
100 mA ma[.
2
D3
1kȍ
Tr2
3
D2
4
Internal circuit
(Black) Comparative output 1
(White) Analog output
Load
or E[ternal input
+
í
12 to 24 V DC
±10 %
(Blue) 0 V
Users’ circuit
Symbols D1, D2 : Reverse supply polarity protection diode
ZD1: Surge absorption zener diode
Tr1 : PNP output transistor
Tr2 : NPN input transistor
Terminal arrangement diagram
Terminal
1
2
3
4
Designation
1
+V
2
Comparative output 1
3
Standard type: Comparative output 2
Multi-function type: Analog output or E[ternal input
4
0V
12
DP-100
I/O CIRCUIT AND WIRING DIAGRAMS
'3Ƒ(3- PNP output type
I/O circuit diagram
Color code of connection cable
Standard type
(Brown) +V
1
ZD1
100 mA max.
Tr1
4
ZD2
Tr2
+
(Black) Comparative output 1
100 mA max.
Load
2
(White) Comparative output 2
Load
(Blue) 0 V
í
12 to 24 V DC
±10 %
ZD1
Tr1
100 mA max.
4
1k
í
Tr2
D2
12 to 24 V DC
±10 %
(White) Analog voltage output Load
2
or External input
Internal circuit
Users’ circuit
+
(Black) Comparative output 1
3
3
Internal circuit
Sensor circuit
1
D1
(Brown) +V
D
Sensor circuit
Terminal No.
Color code of connection cable
Multi-function type
Terminal No.
(Blue) 0 V
Users’ circuit
Symbols D1, D2 : Reverse supply polarity protection diode
ZD1: Surge absorption zener diode
Tr1 : PNP output transistor
Tr2 : NPN input transistor
Symbols D : Reverse supply polarity protection diode
ZD1, ZD2 : Surge absorption zener diode
Tr1, Tr2 : PNP output transistor
Terminal arrangement diagram
1
2
4
3
Terminal
Designation
1
+V
2
Standard type: Comparative output 2
Multi-function type: Analog output or External input
0V
Comparative output 1
3
4
PRECAUTIONS FOR PROPER USE
Never use this product as a sensing device for
personnel protection.
In case of using sensing devices for personnel
protection, use products which meet laws and standards,
such as OSHA, ANSI or IEC etc., for personnel
protection applicable in each region or country.
The DP-100 series is designed for use with non-corrosive
gas. It cannot be used with liquid or corrosive gas.
Mounting
MS-DP1-1/MS-DP1-5 sensor mounting brackets are available
separately, and it should be used for mounting. When tightening
the sensor to the sensor mounting bracket, use a tightening torque
of 0.5 N·m or less.
M3 (length 6 mm 0.236 in)
screws with washers
Accessory for MS-DP1-1
and MS-DP1-5
(
)
Wiring
Make sure that the power supply is off while wiring.
Verify that the supply voltage variation is within the rating.
If power is supplied from a commercial switching regulator, ensure
that the frame ground (F.G.) terminal of the power supply is
connected to an actual ground.
In case noise generating equipment (switching regulator, inverter
motor, etc.) is used in the vicinity of this sensor, connect the frame
ground (F.G.) terminal of the equipment to an actual ground.
Do not run the wires together with high-voltage lines or power lines or put
them in the same raceway. This can cause malfunction due to induction.
Incorrect wiring will cause problems with operation.
Connection
Do not apply stress directly to the
connection cable leader or to the
connector.
Connector
attached cable
[CN-14A(-R)-CƑ]
Sensor mounting bracket
MS-DP1-1 (Optional)
The MS-DP1-2 panel mounting bracket (optional) and the MS-DP1-3
front protection cover (optional) are also available.
Front protection cover
MS-DP1-3 (optional)
Conditions in use for CE conformity
The DP-100 series is a CE conformity product complying with
EMC Directive. The harmonized standard with regard to immunity
that applies to this product is EN 61000-6-2 and the following
condition must be met to conform to that standard.
Condition
The line to connect with this sensor should be less than
30 m 98.425 ft.
13
Panel mounting bracket
MS-DP1-2 (optional)
DP-100
PRECAUTIONS FOR PROPER USE
Mounting
Piping
The MS-DP1-4 panel mounting bracket is available when
switching from the DP2 / DP3 series.
If connecting a commercially-available
coupling to the pressure port, attach a
12 mm 0.472 in spanner (14 mm 0.551 in
spanner for DP-100-E type) to the hexagonal
section of the pressure port to secure it, and
tighten at a torque of 9.8 N·m or less. If it is
tightened using excessive torque, it may
damage the coupling or the pressure port.
In addition, wrap sealing tape around the
coupling when connecting it to prevent leaks.
DP2 / DP3
DP
2-2
0
ñ10
1.3k
Pa
Front protective cover
DPX-04 (Optional)
can be installed on
MS-DP1-4.
OUT
1
Remove
OUT
$'
-
2
MO
DE
Remove
Insert
DP-100
Mounting holes
for DP2 / DP3 series
can be used as is.
Panel mounting bracket MS-DP1-4
If connecting a commercially-available joint to the pressure port of
the DP-10Ƒ-M, hold the main unit in your hand to steady it, and
tighten to a torque of 1.0 N·m or less. If it is tightened to an
excessive torque, the joint or the main unit may become
damaged.
If connecting a commercially-available joint to the pressure port of
the MS-DP1-7, tighten to a torque of 9.8 N·m or less.
An conversion bushing is available for when using the DP-10Ƒ-M short
pressure port type. It can be used to switch between this model and the
DP2 / DP3 series. When connecting to the pressure port, use a tightening
torque of 1.0 N·m or less.
M5 female thread
Rc 1 8 female thread
Conversion bushing MS-DP1-7
The MS-DP1-FƑ flat attachment is available.
If using the MS-DP1-FƑ flat attachment (optional), install by
following the procedures given below.
①Decide the direction of this product to mount with the sensor.
12 mm 0.472 in
spanner
14 mm 0.551 in
spanner
The tightening torque should be 1 N·m or less when connecting a
coupling to the pressure port of MS-DP1-FM.
Pressure port
Coupling (Purchase separately.)
Pressure port
faces downward
Pressure port
faces leftward
Pressure port
faces rightward
Note: It is not possible to mount this product such
that the pressure port faces upward.
When connecting the coupling to the pressure port of
MS-DP1-FR/FE/FN, hold the pressure port with a 14 mm 0.551 in
spanner and make sure that the tightening torque is 9.8 N·m or
less.
In addition, in order to prevent any leakage, wind a sealing tape
on the coupling when connecting.
Pressure port
faces upward
②Mount this product with the M3 female threads of the sensor by
using the attached M3 (length 8 mm 0.315 in) screws. The
tightening torque should be 0.5 N·m or less.
M3 female thread
14 mm 0.551 in
spanner
M3 (length 8 mm 0.315 in)
screw (Accessory)
M3 female thread
Note: Do not tighten the pressure port by holding the product with the
spanner. It may cause the product breakage.
③Mount this product with the mounting surface by using the
attached M4 (length 20 mm 0.787 in) screws. The tightening
torque should be 1.2 N·m or less.
Flat attachment
Groove
Connector attached cable
M4 (length 20 mm 0.787 in)
screw (Accessory)
Note: Take care that if the cable with connector is sticking out of the side
groove of this product when mounting, the cable may disconnected.
Make sure to mount MS-DP1-FƑ with the sensor properly. If it is
not mounted properly, air leakage may occur.
Take care that the excessive mounting and dismounting of this
product may cause deterioration of the O-ring.
If you touch the O-ring of MS-DP1-FƑ, or any scratch or dust, etc.
is attached to it, air leakage may occur and the sensing
performance may deteriorate.
Take sufficient care when using and storing MS-DP1-FƑ.
14
DP-100
PRECAUTIONS FOR PROPER USE
Others
PRO mode
This product has been developed / produced for industrial use.
Use within the rated pressure range.
Do not apply pressure exceeding the pressure withstandability
value. The diaphragm will get damaged and correct operation shall
not be maintained.
Do not use during the initial transient time (0.5 sec. approx.) after
the power supply is switched on.
Avoid dust, dirt, and steam.
Take care that the sensor does not come in direct contact with
water, oil, grease, or organic solvents, such as, thinner, etc.
Do not insert wires, etc., into the pressure port. The diaphragm will
get damaged and correct operation shall not be maintained.
Do not operate the keys with pointed or sharp objects.
If the mode selection key is pressed and held for 5 seconds in
RUN mode, the mode will switch to PRO mode.
If the mode selection key is pressed while a setting is being made,
the mode will switch to RUN mode. In this case, the settings that
have been changed will be entered.
RUN mode
7KLVLVWKHQRUPDORSHUDWLQJPRGH
Setting item
Description
Threshold value
setting
The threshold values for ON / OFF operation can be changed directly
by pressing the increment key (UP) and the decrement key (DOWN).
Zero-adjustment
function
This forces the pressure value display to be reset to zero when
the pressure port is open on the atmospheric pressure side.
Key lock function
Stops key operations from being accepted.
Peak hold /
bottom hold function
Displays the peak value and bottom value for fluctuating
pressure. The peak value appears in the main display,
and the bottom value appears in the sub display.
Setting item
Description
Sub display switching
Changes the information in the sub display
during RUN mode operation to the desired
alphanumeric display.
Display refresh rate
switching
Changes the display refresh rate for the
pressure value displayed in the main display.
Hysteresis fix value
switching
Sets the hysteresis for EASY mode and window
comparator mode. (8 steps)
Linked display color
switching
(standard type only)
Allows the display color for the main display to
be switched in line with the output operation for
comparative output 1 or comparative output 2.
ECO mode setting
Allows power consumption to be reduced by
dimming the display or turning it off.
Setting check code
Allows the setting details to be checked via
codes.
Setting copy mode
Allows the setting details for the master sensor
to be copied to slave sensors.
Reset setting
Resets the settings to the factory settings.
Table of codes
If the mode selection key is pressed and held for 2 seconds in
RUN mode, the mode will switch to MENU SETTING mode.
If the mode selection key is pressed while a setting is being made,
the mode will switch to RUN mode. In this case, the settings that
have been changed will be entered.
Setting item
Comparative output 1
output mode setting
Code
2nd digit
MENU SETTING mode
1st digit
NO
NC
Sets normally open (NO) or normally closed (NC).
Response time setting
Sets the response time.
The response time can be selected from 2.5 ms,
5 ms, 10 ms, 25 ms, 50 ms, 100 ms, 250 ms,
500 ms, 1,000 ms and 5,000 ms.
Display color switching
for main display
Allows the color for the main display to be changed.
The colors can be set to ‘red / green’ or ‘green / red’
to correspond to ON / OFF output, or it can be fixed
at ‘red’ or ‘green’ all the time.
Unit switching
Pressure unit can be changed.
15
NO
NO
NC
NC
NO
Hysteresis
Hysteresis
Window
comparator
Code
NO / NC switching
OFF
EASY
Comparative output 2 output mode
Sets the output mode for comparative output 2.
setting (standard type only)
Allows switching between analog voltage output /
analog current output, and auto-reference input /
remote zero-adjust-ment input.
OFF
EASY
Sets the output mode for comparative output 1.
Multi-function 3rd digit
type
4th digit
Standard
type only
Comparative output 1 NO / NC Comparative output 2 NO / NC Analog voltage output / Threshold Display color for Display
output mode
switching output mode
switching External input
value display main display color linking
Description
Analog output / external
input switching
(multi-function type only)
Standard type
NO
NC
—
—
—
—
5th digit
Response time
2.5 ms
5 ms
10 ms
25 ms
50 ms
100 ms
250 ms
500 ms
1,000 ms
5,000 ms
Window
comparator
—
Analog voltage
P-1, Lo-1
output
Red
when ON
Autoreference Hi-1
Comparative
output 1
Remote
zero-adjustment P-2, Lo-2 Green
when ON
Analog current
Hi-2
output
Comparative
output 1
NC
—
ADJ.
NO
—
—
NC
—
—
—
—
—
6th digit
Unit switching
MPa
kPa
kgf/cm2
bar
psi
mmHg
inchHg
—
—
—
7th digit
Display refresh rate
250 ms
500 ms
1,000 ms
—
—
—
—
—
—
—
Comparative
output 2
Comparative
output 2
Comparative
Always output 1
red
Comparative
output 2
Comparative
Always output 1
green Comparative
output 2
8th digit
ECO mode
OFF
STD
FULL
—
—
—
—
—
—
—
DP-100
DIMENSIONS (Unit: mm in)
DP-10Ƒ
The CAD data in the dimensions can be downloaded from the website.
Sensor
42.5
1.673
30
1.181
25.5
1.004
20
0.787
7.5 9.5
0.295 0.374
12
0.472
20
0.787
30
1.181
R (PT) 1/ 8
male thread
(North America: NPT 1/ 8)
1.5
0.059
Connector
M5 female thread
M3 female thread, 4 0.157 deep
Europe
Europe
10
7
0.394 0.276
14
0.551
G 1/ 8
male thread
DP-10Ƒ-M(-P) Sensor
30
1.181
30
1.181
25.5
1.004
20
0.787
4.5
0.177
20
0.787
ø8.3
ø0.327
30
1.181
1.5
0.059
connector
M5 female thread,
4 0.157 deep
M3 female thread,
4 0.157 deep
DP-11Ƒ-E-P-- Sensor
30
1.181
47.5
1.870
25.5
1.004
20
0.787
7
0.276
9.5
0.374
30
1.181
20
0.787
M8 connector
14
0.551
4.5
0.177
G 1/8 male thread
1.5
0.059
15
0.591
7
0.276
M5 female thread
M3 female thread,
4 0.157 deep
16
DP-100
DIMENSIONS (Unit: mm in)
MS-DP1-1
The CAD data in the dimensions can be downloaded from the website.
Sensor mounting bracket (Optional)
ø2.3
ø0.091
1
0.039
22
0.866
30
1.181
20
0.787
13
0.512
14.5
0.571
4.2
0.165
2-ø3.5 ø0.138
25.5
1.004
42.5
1.673
9.5
0.374
7.5
0.295
5.3
(0.209
)
5.5
0.217
20
0.787
45
1.772
44
1.732
30
1.181 45
1.772
12
0.472
30
1.181
15
0.591
Sensor mounting bracket (Optional)
1.5
0.059
30
1.181
Material: Cold rolled carbon steel (SPCC)
(Uni-chrome plated)
Two M3 (length 6 mm 0.236 in) screws with washers are attached.
MS-DP1-5
2-R2.1 R0.083
9.5
0.374
2-R2.1 R0.083
R13
20
0.787 R0.512
t2
t 0.079
Mounting drawing with DP-1Ƒ
22
0.866
9.5
14.5
0.571 0.374
4.2
0.165
Assembly dimensions
t2
t 0.079
Assembly dimensions
Mounting drawing with DP-1Ƒ-M
R6.75
R0.266
22.75
0.896
29.5
1.161
20
0.787
4.5
0.177
16
0.630
ø4.5
ø0.177
28
1.102
8.5
0.335
30
1.181
4.5
0.177
44.5
30 1.752
1.181
20
0.787
13.5
0.531
20
0.787
2-ø3.5 ø0.138
49.5
(1.949
)
ø4.5
ø0.177
20
0.787
29.5
1.161
30
1.181
Material: Cold rolled carbon steel (SPCC)
(Uni-chrome plated)
Two M3 (length 6 mm 0.236 in) screws with washers are attached.
MS-DP1-2
MS-DP1-3
ø8.3
ø0.327
20
49.5
0.787 28.5 1.949
1.122
t 1.5
t 0.059
1.5
0.059
28
1.102
53.5
2.106
4.5
0.177
t 1.5
t0.059
Panel mounting bracket (Optional), Front protection cover (Optional)
Assembly dimensions
Mounting drawing with '3Ƒ
Panel cut-out dimensions
When 1 unit is installed
0
31í
0
1.220 í0.016
0
39.3
( 1.547
)
34.5
1.358
(
34.5 20
1.358 0.787
7.2
0.283
11
0.433
18.3 7.5
0.720 0.295
Front
protection
cover
)
50
1.969
33.4
1.315
When “n” units are installed horizontally in series When “n” units are
installed vertically in series
0
31í
0
1.220 í0.016
31 1.220 Xn + 3.5 0.138 X (n-1)
12
0.472
0
31í
0
1.220 í0.016
55 2.165
or more
M5 female thread,
4 0.157 deep
(52 2.047 max)
M3 female thread,
4 0.157 deep
View A
Material: POM (Panel mounting bracket)
Polycarbonate (Front protection cover)
31 1.220 Xn + 3.5 0.138 X (n-1)
Panel thickness dimension
1 to 6 mm 0.039 to 0.236 in
Connector
33.4
1.315
17
9.5
0.374
R(PT) 1/ 8
male thread
8.7
0.343
31í
0
1.220 í0.016
55 2.165
or more
Note: The panel thickness should be
1 to 6 mm 0.039 to 0.236 in.
Note: The panel thickness should be
1 to 6 mm 0.039 to 0.236 in.
DP-100
DIMENSIONS (Unit: mm in)
MS-DP1-4
The CAD data in the dimensions can be downloaded from the website.
Panel mounting bracket (Optional)
Assembly dimensions
Panel cut-out dimensions
Mounting drawing with DP-10Ƒ
(
42.7
1.681
40
1.575
30
1.181
)
(
9.7
0.382
36
1.417
26.5
1.043
6.5
0.256
3
0.118
)
Front protection cover (DPX-04)
Panel thickness dimension 1 to 3.2 mm
0.039 to 0.126 in
MS-DP1-7
36 +0.5
0
1.417+0.020
0
Connector
R 1/ 8
male thread
80
3.150
12
0.472
42.7
1.681
40
30
1.575 1.181
56
2.205
Panel mounting
bracket body
8
0.315
Panel mounting
bracket
36 +0.5
0
1.417+0.020
0
M5 female
thread
Spacer
Note: The panel tickness should be 1 to 32 mm 0.039 to 1.260 in.
Material: Panel mounting bracket body Nylon 6
Panel mounting bracket Stainless steel (SUS304)
Spacer Cold rolled carbon steel (SPCC)(Uni-chrome plated)
Conversion bushing (Optional)
15.8
0.622
14
0.551
3.5
0.138
(
7.8
0.307
0.5
0.020
11 0.433
M5 male thread
Gasket
)
8
0.315
10 0.394
Rc1/8 female thread
Material: Brass (Nickel plated)
Weight: 10 g approx.
MS-DP1-FM Flat attachment (Optional)
MS-DP1-FR/FN/FE Flat attachment (Optional)
Assembly dimensions
Assembly dimensions
Mounting drawing with DP-10Ƒ-M
(
)
44.5
1.752
24
0.945
(
Mounting drawing with DP-10Ƒ-M
52.5
( 2.067
)
24
(0.945
)
37
1.457
11.5
0.453
)
30
1.181
44.5
1.752
11
0.433
24
0.945
6.5 0.256
5.5 0.217 13.5
0.531
24
0.945
M5 female thread
20
0.787
24
0.945
30
1.181
17.5
0.689
Material: Enclosure Polybutylene
terephthalate (PBT)
(Glass fiber reinforced)
Pressure port Stainless steel
(SUS303)
8.5 0.335
O-ring Hydrogenated Nitrile
Butadiene Rubber
(H-NBR)
Weight: 15 g approx. (flat attachment only)
Two M3 (length 8 mm 0.315 in) screws,
two M4 (length 20 mm 0.787 in) screws are
attached.
2-ø4.3
ø0.169
Pressure port
G1/8 female thread (Note)
11.5
0.453
11
0.433
30
1.181
52.5
2.067
6.5 0.256
5.5 0.217
2-ø4.3 ø0.169
37
1.457
14
0.551
20
0.787
24
0.945
30
1.181
Product distinguishing mark
MS-DP1-FR: 0 line
MS-DP1-FN: 1 line
MS-DP1-FE: 2 line
17.5
0.689
8.5
0.335
Material: Enclosure Polybutylene terephthalate (PBT)
(Glass fiber reinforced)
Pressure port Stainless steel (SUS303)
O-ring Hydrogenated Nitrile
Butadiene Rubber (H-NBR)
Weight: 25 g approx. (flat attachment only)
Note: MS-DP1-FR has a
Two M3 (length 8 mm 0.315 in) screws, two M4
Rc1/ 8 female thread. (length 20 mm 0.787 in) screws are attached.
MS-DP1-FN has a
NPT1/ 8 female thread.
18
DIMENSIONS (Unit: mm in)
The CAD data in the dimensions can be downloaded from the website.
CN-14A(-R)-CƑ Connector attached cable (Optional, CN-14A-C2 is attached to the sensor)
L
35
(1.378
)
Model No.
50
(1.969
)
(0.3158 )
(0.3158 )
ø3.7 ø0.146 cable
Cable length L (mm in)
CN-14A(-R)-C1
1,000 39.370
CN-14A(-R)-C2
2,000 78.740
CN-14A(-R)-C3
3,000 118.110
CN-14A(-R)-C5
5,000 196.850
Please contact ..........
2431-1 Ushiyama-cho, Kasugai-shi, Aichi, 486-0901, Japan
Global Sales Department
■Telephone: +81-568-33-7861 ■Facsimile: +81-568-33-8591
panasonic.net/id/pidsx/global
All Rights Reserved ©
No. CE-DP100V2-5 November, 2013
Specifications are subject to change without notice.
2013
Printed in Japan