Triple-Output Programmable DC Power Supply
NI PXI-4110 NEW!
• 3 independent DC power supplies • 0 to 6 VDC, 1 A, nonisolated • 0 to +20 VDC, 1 A, isolated • 0 to -20 VDC, 1 A, isolated • 16-bit voltage setpoint and current limit • 16-bit voltage/current readback measurements • 20 mA and 1 A current modes • 2 power source options • Internal (PXI backplane) – 9 W output • Auxiliary – full 46 W output • Isolated channels can be combined for 0 to +40 VDC operation
Operating Systems
• Windows Vista/XP/2000
Recommended Software
• • • • • LabVIEW LabVIEW Real-Time Module LabWindows™/CVI Measurement Studio LabVIEW SignalExpress
Other Compatible Software
• Microsoft Visual Basic • C/C++
Driver Software (included)
• NI-DCPower
Overview
The National Instruments PXI-4110 is a programmable, triple-output precision DC power supply in a single-slot, 3U PXI module. The NI PXI-4110 has two isolated channels, one from 0 to +20 V and the other from 0 to -20 V, and a single nonisolated 0 to 6 V supply, all capable of sourcing up to 1 A per channel. The PXI-4110 has 16-bit resolution for programming the voltage setpoint and current limit and for using the voltage and current readback measurement functionality. The versatile supply rails and high accuracy make the PXI-4110 an excellent general-purpose, single-quadrant power supply for design validation and manufacturing test applications.
Internal/External Supply Options
You can power the PXI-4110 either internally from the PXI backplane or externally through the NI APS-4100, a front-panel-connected auxiliary DC supply. Using internal power reduces the number of connections required on the front panel but also limits the available output power because of per-slot PXI power restrictions. When internally powered, the nonisolated, 0 to 6 V channel can be operated at its full 1 A current range, but the isolated channels are limited to 100 mA. When externally powered, all channels can be operated at full power of 1 A per channel for a total maximum output power of 46 W.
Linear Supply with Switching Preregulation
The PXI-4110 uses a combination of switching and linear regulation to provide excellent output power and accuracy in the 3U PXI module. On each channel, input power (coming from either the PXI backplane or the APS-4100) is regulated to within a certain percentage of the desired output power. This preregulation stage is governed by an intelligent PID algorithm implemented on board the module, ensuring the amount of power passed to the second (linear) stage is at the most efficient level, given the desired output. After additional filtering, you can use traditional linear regulation techniques and amplification to further regulate the signal and source the final voltage or current. Because the output is linearly regulated, it has very quick load response and high precision – even at levels as low as 0 V. Also, because the linear regulation occurs on the preregulated signal, the power dissipation is relatively small and easily cooled in a PXI slot.
Power Supply with Precision Source Capability
The PXI-4110 has the ability to source both voltage and current from each of its three outputs. As a voltage source, it can be programmed in 120 µV steps on the +6 V channel and 400 µV steps on each of the 20 V channels. As a current source, it can be programmed in 20 µA steps on each channel in the 1 A current range. Additionally, you can set each of the 20 V channels to a 20 mA current range for 400 nA programming resolution. You can use this impressive level of current resolution in traditional power supply applications or in many applications that typically require a separate precision source measure unit.
Triple-Output Programmable DC Power Supply
Preregulation Control Power Amplifier Current Sense (20 mA/1 A)
Transformer
Rectifier and Filter
Load Linear Regulation (Isolated Channels Only) Current Feedback Voltage Feedback Auxiliary Power (Isolated Channels Only)
+
The PXI-4110 operates with only nominal temperature increases internally due to the intelligent PID control of the output devices. If an overtemperature condition occurs in the PXI chassis due to fan failure or intake blockage, the output channels are shut down and a warning is issued. This type of condition requires user software intervention to reset, thus preventing the module from damage at excessive temperatures.
Software
NI-DCPower, an IVI-compliant instrument driver, offers complete programmatic control of the PXI-4110. You can use an available test panel to quickly troubleshoot or debug power supply operation and take advantage of the DCPower Express VI for an intuitive, configurationbased method of programming in the National Instruments LabVIEW graphical development environment.
Internal Power
Power Switch
–
Architecture of Single Isolated Channel on the PXI-4110
Extensive Protection Features
In addition to the standard voltage and current limiting functionality of the PXI-4110, several other features are included to protect the supply and the load. Each output is protected against a reverse-polarity voltage application as well as excessive voltages – up to 15 V above the maximum channel voltage. Output fuses provide additional protection to prevent catastrophic failure as a last line of defense. The operating voltage range for the auxiliary power input is 11 to 15.5 V. If voltages outside these limits are detected, the module shuts down until an input voltage within range is applied. If an input in excess of 20 V is applied, the input crowbar protection turns on, protecting the input solid-state switching devices (and preregulator power supply) from overvoltage damage.
Ordering Information
NI PXI-4110............................................................................779647-10
Includes NI-DCPower and the DCPower Test Panel.
NI APS-4100 ..........................................................................779671-01 NI PXI-4110, NI APS-4110 Bundle ........................................779647-11
BUY NOW!
For complete product specifications, pricing, and accessory information, call 800 813 3693 (U.S.) or go to ni.com/modularinstruments.
BUY ONLINE a t ni.com o r CALL 800 813 3693 (U.S.)
2
Triple-Output Programmable DC Power Supply
Specifications
Specifications subject to change without notice. For the most current and complete specifcations, visit ni.com/modularinstruments.
Voltage Output Speed8
Auxiliary Power Internal Power Rise Time9 Fall Time10 Rise Time9 Fall Time10 Channels Full Load No Load Full Load No Load Full Load No Load Full Load No Load 0
很抱歉,暂时无法提供与“NIPXI-4110”相匹配的价格&库存,您可以联系我们找货
免费人工找货