1.3 GHz and 4 GHz Multiplexers
1.3 GHz and 4 GHz Multiplexers
NI SCXI-1190, NI PXI-2590, NI SCXI-1191, NI PXI-2591
• 4x1 unterminated multiplexer • 50 characteristic impedance • High-bandwidth electromechanical relays • Excellent insertion loss, VSWR, and isolation specifications • Fully software programmable
Operating Systems
• Windows 2000/NT/XP
Recommended Software
• LabVIEW • LabVIEW Real-Time Module (PXI Modules only) • LabWindows/CVI • Measurement Studio • NI Switch Executive
SCXI-1190 (quad 4x1), PXI-2590 (single 4x1)
• 1.3 GHz bandwidth • SMB direct connectivity • 1 A at 24 VDC capacity
Other Compatible Software
• Visual Basic • C/C++
SCXI-1191 (quad 4x1), PXI-2591 (single 4x1)
• 4 GHz bandwidth • SMA direct connectivity • 330 mA at 30 VDC capacity
Driver Software (included)
• NI-SWITCH
Compliance
• CE
Overview
The National Instruments PXI-2590 and PXI-2591 are 4x1 highfrequency unterminated multiplexing switches. The NI SCXI-1190 and NI SCXI-1191 perform the same functions as the PXI-2590 and PXI-2591 but also offer a quad 4x1 multiplexing configuration. The switches are capable of switching signals from DC to 1.3 GHz (SCXI-1190 and PXI-2590) and DC to 4 GHz (SCXI-1191 and PXI-2591). The characteristic impedance of the channels is 50 . For the SCXI-1190 and the PXI-2590, the maximum voltage rating of the switches is 24 Vrms with a maximum current of 1 A. For the SCXI-1191 and the PXI-2591, the maximum voltage rating is 30 Vrms with a maximum current of 330 mA. All are well suited for applications that require the routing of high-frequency signals inside automated test equipment (ATE) systems because they can switch signals with a very low insertion loss. In addition, the excellent voltage standing-wave ratio (VSWR) and isolation parameters make these modules the perfect choice in a system geared to highfrequency applications as shown in Table 1.
Ordering Information
NI SCXI-1190 ................................................................776572-90 NI PXI-2590 ..................................................................777987-01 NI SCXI-1191 ................................................................776572-91 NI PXI-2591 ..................................................................778339-01
Includes switch module and NI-SWITCH driver software.
For information on extended warranty and value added services, see page 20.
See page 499 for accessory and cable information.
BUY ONLINE!
Visit ni.com/products and enter pxi2590, pxi2591, scxi1190, and/or scxi1191.
Module SCXI-1190, PXI-2590 (@ 1.3 GHz) SCXI-1190, PXI-2590 (@ 2 GHz) SCXI-1191, PXI-2591 (@ 4 GHz)
Insertion Loss < 1.5 dB < 3 dB < 0.9 dB
VSWR < 1.5 < 2.5 < 1.5
Switches
Table 1. Parameters of High-Frequency Switching Modules
Extended Features and Specifications
National Instruments switch modules are built with a number of core features that are covered in detail in the Switch Overview section.
For additional information about the PXI-2590, PXI-2591, SCXI-1190, and SCXI-1191, including software, certifications and compliance, relay control, etc., please see page 20. For detailed specifications, please see page 504.
492
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com
Switch Specifications
Specifications
Specifications (continued)
2 1.9 1.8 1.7 1.6 VSWR 1.5 8x1 1.4 1.3 1.2 1.1 0 0 100 200 300 400 500 600 Frequency (MHz) 700 800 900 1000 16x1 32x1
SCXI-1190, PXI-2590
Input Characteristics
Number of channels per bank.......................... Number of banks SCXI-1190 .................................................. PXI-2590..................................................... Input voltage Channel-to-earth........................................ Maximum switching voltage AC ............................................................ DC ............................................................ Maximum switching capacity per channel AC ............................................................ DC ............................................................ Maximum switching power per channel ......... Contact resistance (initial) ............................... Contact material............................................... 4 4 1 24 VDC 24 Vrms 24 VDC 1 Arms at 24 Vrms 1 A at 24 VDC 24 W 100 mΩ maximum Gold-clad silver
Input Characteristics
Maximum switching voltage............................ Maximum switching current ............................ Maximum carry current.................................... Simultaneous channels at maximum current PXI-2593 ................................................... SCXI-1193 .................................................. DC path resistance Initial ......................................................... End of life .................................................. 150 V, CAT I 0.5 A (per channel) 1 A (per channel) 2, maximum 4, maximum