0
登录后你可以
  • 下载海量资料
  • 学习在线课程
  • 观看技术视频
  • 写文章/发帖/加入社区
创作中心
发布
  • 发文章

  • 发资料

  • 发帖

  • 提问

  • 发视频

创作活动
PXI-2530

PXI-2530

  • 厂商:

    NI

  • 封装:

  • 描述:

    PXI-2530 - High-Density Multiconfiguration Multiplexer/Matrix - National Instruments Corporation

  • 数据手册
  • 价格&库存
PXI-2530 数据手册
High-Density Multiconfiguration Multiplexer/Matrix NI PXI-2530, NI SCXI™-1130 • 14 multiplexer/matrix configurations • Scanning speed up to 900 operations/s • Maximum switch capacity • Up to 60 VDC, 30 Vrms • Up to 400 mA • 32,000-step scan list for deterministic scanning • Fully software programmable • Tight synchronization with instruments through hardware triggers NEW Operating Systems • Windows 2000/NT/XP Recommended Software • LabVIEW™ • LabWindows™/CVI™ • Measurement Studio • NI Switch Executive Other Compatible Software • Visual Basic • C/C++ NI PXI-2530 • 128x1 (1-wire), 64x1 (2-wire), 32x1 (4-wire), or 8 banks of 16x1 (1-wire) • 4x32 (1-wire), 8x16 (1-wire), 4x16 (2-wire) matrix configurations Driver Software (included) • NI-SWITCH NI SCXI-1130 • 256x1 (1-wire), 128x1 (2-wire), 64x1 (4-wire), or 16 banks of 16x1 (1-wire) • 4x64 (1-wire), 8x32 (1-wire), 4x32 (2-wire) matrix configurations Overview The National Instruments PXI-2530 and SCXI-1130 are high-density, multiconfiguration multiplexer/matrix switches. With 14 possible configurations and multiple expansion options, these modules offer an excellent solution for large or complex-channel-count systems. The NI PXI-2530 and NI SCXI-1130 incorporate reed relays into the PXI and SCXI platforms for long-life functionality. At up to 10 times the speed of traditional armature relays, the PXI-2530 and SCXI-1130 are an ideal front-end for high-speed measurement devices, such as the NI PXI-4070 61⁄2-digit FlexDMM. Multiplexer You can configure the PXI-2530 and SCXI-1130 in several different multiplexer configurations. The 1 and 2-wire modes are designed for measuring/sourcing voltage or 2-wire resistance measurements. For improved accuracy of resistance measurements, configure the PXI-2530 and SCXI-1130 in 4-wire mode. Matrix Configured as a matrix, the PXI-2530 or SCXI-1130 becomes a general-purpose signal router. With the TB-2631, the PXI-2530 forms a 4x32 matrix in 1-wire mode and a 4x16 matrix in 2-wire mode. The TB-2632 is also available to configure the PXI-2530 into an 8x16 1-wire matrix. Using the SCXI-1378 terminal block, the SCXI-1130 can form a 4x64 1-wire or a 4x32 2-wire configuration matrix. With the SCXI-1379, an 8x32 1-wire matrix is available. Through software, you can control the matrix to connect any row(s) to any column(s). You can connect the rows and columns of multiple modules to expand an NI matrix system and meet the needs of high-channelcount systems. With the SCXI-1130, you can use expansion cables and plugs to externally connect the rows and columns of the matrix modules. For example, to create a 4x128 matrix, connect two SCXI-1378 terminal blocks together with a matrix expansion cable to double the number of columns. Modes of Operation The PXI-2530 and SCXI-1130 are completely software configurable to operate as a multiplexer, matrix, or independent switch bank. For matrix and multiplexer hardware options, see Table 1. Accessory Type TB-2630 Multiplexer Configuration 128x1 (1-wire), 64x1 (2-wire), 32x1 (4-wire) or 8 banks of 16x1 (1-wire) TB-2631 Matrix 4x32 (1-wire) and 4x16 (2-wire) TB-2632 Matrix 8x16 (1-wire) SCXI-1377 Multiplexer 256x1 (1-wire),128x1 (2-wire), 64x1 (4-wire) or 16 banks of 16x1 (1-wire) SCXI-1378 Matrix 4x64 (1-wire) and 4x32 (2-wire) SCXI-1379 Matrix 8x32 (1-wire) Connectivity Ribbon cable headers Screw terminals Screw terminals Spring terminals NI PXI-2530 NI SCXI-1130 Screw terminals Screw terminals Table 1. Matrix and Multiplexer Hardware Options High-Density Multiconfiguration Multiplexer/Matrix Independent When the PXI-2530 and SCXI-1130 are configured with software as independent switch banks, you can control the state of any relay on the module. Use this mode for custom applications, where the banks of 16x1 multiplexers can be combined to form other multiplexer sizes or operate independently. Signal Connections Terminal blocks provide screw terminals, spring terminals, or ribbon cable terminals for easy connections. The terminal blocks for the SCXI-1130 are twice as deep as traditional SCXI terminal blocks, such as the SCXI-1300 and SCXI-1303. Software Automatic Scanning The PXI-2530 and SCXI-1130 are able to maximize throughput in automated test applications by the use of scanning. Scanning improves throughput by downloading a list of up to 32,000 connections to the switch modules and cycling through the list using an event (trigger) without any interruption from the host processor. Scanning is most efficiently accomplished by mating the PXI-2530 or SCXI-1130 with an instrument, such as the NI PXI-4070 FlexDMM, that issues a trigger after each measurement. All National Instruments PXI and SCXI switch modules are shipped with NI-SWITCH, an IVI-compliant driver offering complete functionality for all switch modules. For additional assistance in configuring, programming, and managing higher-channel-count switching systems, NI Switch Executive software offers an easy-touse, intelligent switch management and visual routing environment. Ordering Information NI PXI-2530 ..................................................................778660-01 NI SCXI-1130 ................................................................778661-01 Includes NI-SWITCH driver software. SCXI Module Control Every SCXI switch system requires an external switch controller. The switch controller uses the digital communications bus on the SCXI chassis to control the switch circuitry. An NI 4021 Switch Controller, available for both PCI and PXI, is the recommended SCXI switch controller. Accessories NI TB-2630....................................................................778733-01 NI TB-2631....................................................................778734-01 NI TB-2632....................................................................778735-01 NI SCXI-1377 ................................................................778716-01 NI SCXI-1378 ................................................................778731-01 NI SCXI-1379 ................................................................778732-01 NI PCI-4021 ..................................................................778277-01 NI PXI-4021 ..................................................................778278-01 BUY ONLINE! Visit ni.com/info and enter pxi2530 and/or scxi1130. 2 National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com High-Density Multiconfiguration Multiplexer/Matrix Specifications PXI-2530 Multiplexer sizes ....................................... 128x1 (1-wire) 64x1 (2-wire) 32x1 (4-wire) 16x1 (1-wire) (8) Matrix sizes ............................................... 4x32 (1-wire) 8x16 (1-wire) 4x16 (2-wire) SCXI-1130 Multiplexer sizes ....................................... 256x1 (1-wire) 128x1 (2-wire) 64x1 (4-wire) 16x1 (1-wire) (16) Matrix sizes ............................................... 4x64 (1-wire) 8x32 (1-wire) 4x32 (2-wire) Physical Relay types ....................................................... Contact material............................................... I/O connectors .................................................. Dimensions PXI-2530..................................................... SCXI-1130 .................................................. reed rhodium 176-pin docking station plug 2 x 10 x 16 cm (.8 x 3.9 x 6.4 in.) 3.0 x 17.2 x 20.3 cm (1.2 x 6.9 x 8.0 in.) Environment Operating temperature..................................... Storage temperature ........................................ Relative humidity ............................................. Pollution degree ............................................... Indoor use only 0 to 55 ˚C -40 to 70 ˚C 5 to 85% noncondensing 2 Safety This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: • IEC 61010-1, EN 61010-1 • UL 3111-1, UL 61010B-1 • CAN/CSA C22.2 No. 1010.1 CE Compliance This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows: Low-Voltage Directive (safety) .................. 73/23/EEC Electromagnetic Compatibility Directive (EMC) ................................... 89/336/EEC Input Characteristics Maximum switching voltage............................ 60 VDC, 30 Vrms, CAT I (channel-to-channel and channel-to-ground) Maximum current ............................................. 0.4 A (per terminal or internal path) Maximum switching power ............................. 10 W (per channel, resistive) Typical DC path resistance (channel-to-common) Initial ...................................................
PXI-2530 价格&库存

很抱歉,暂时无法提供与“PXI-2530”相匹配的价格&库存,您可以联系我们找货

免费人工找货