5 GHz 4x1 and Quad 4x1 Multiplexers
NI PXI-2595, NI SCXI-1195
• • • • 5 GHz bandwidth 50 Ω characteristic impedance SMA direct connectivity Excellent insertion loss, VSWR, and isolation specifications • Onboard relay counting • Fully software programmable
Operating Systems
• Windows 2000/NT/XP • Linux®
Recommended Software
• • • • • NI Switch Executive LabVIEW LabVIEW Real-Time LabWindows/CVI Measurement Studio
NI PXI-2595
• 4x1 multiplexer configuration • 1-slot 3U PXI module
Other Compatible Software
• Visual Basic • C/C++
NI SCXI-1195
• Quad 4x1 multiplexer configuration • 1-slot 4U SCXI module
Driver/Services Software (included)
• NI-SWITCH • NI-DAQmx
Overview
The National Instruments PXI-2595 and SCXI-1195 are 4x1 highfrequency, unterminated multiplexing switches. The NI SCXI-1195 performs the same functions as the PXI-2595 but in a higher density, quad 4x1 multiplexing configuration. The PXI-2595 and SCXI-1195 are capable of switching signals from DC to 5 GHz. The characteristic impedance of these modules is 50 Ω. The PXI-2595 and SCXI-1195 have a maximum voltage rating of 30 Vrms with a maximum switching current of 500 mA. They are well-suited for applications that require the routing of high-frequency signals inside automated test equipment (ATE) systems because they can switch signals with very low insertion loss. In addition, the excellent voltage standing-wave ratio (VSWR) and isolation parameters make these modules the perfect choice in any system that is geared toward high-frequency applications
Software
All National Instruments PXI and SCXI switch modules are shipped with NI-SWITCH, an IVI-compliant driver offering complete functionality for all switch modules. For additional assistance in configuring, programming, and managing higher-channel-count switching systems, NI Switch Executive software offers an easy-to-use, intelligent switch management and visual routing environment. Use the NI-SWITCH soft front panel for simple relay operations or debugging switch code/execution.
Ordering Information
NI PXI-2595............................................................................778572-95 NI SCXI-1195 ........................................................................776572-95
Includes NI-SWITCH and NI-DAQmx driver software.
Relay Count Tracking
The PXI-2595 and SCXI-1195 count relay closures on each of their RF relays. Relay counts are incremented each time a relay is actuated. The counts, stored on board the module, are retrievable programmatically, and you can use them for predictive maintenance to reduce unexpected system downtime.
Accessories SMA male-male cable (semirigid) 0.15 cm ..............................................................................763443-01 0.45 cm ..............................................................................763444-01 NI Switch Executive Development System ............................................................778546-01 Deployment Engine ..............................................................778548-00
BUY NOW!
For complete product specifications, pricing, and accessory information, call (800) 813 3693 (U.S.) or go to ni.com/switches.
Specifications
All specifications are subject to change without notice. Visit ni.com/manuals for the most current specifications.
Physical Characteristics
Relay type ........................................... I/O connectors PXI-2595.......................................... SCXI-1195 ....................................... Dimensions (W by H by D) PXI-2595.......................................... electromechanical, latching 5 SMA jacks 20 SMA jacks
Configuration
PXI-2595 .............................................. 4x1 multiplexer SCXI-1195............................................ Quad 4x1 multiplexer
Input Characteristics
All input characteristics are DC, ACrms, or a combination unless otherwise specified. Maximum switching voltage .............. 30 V Maximum switching current............... 0.5 A Maximum carry current....................... 0.5 A Maximum RF power............................ 10 W Note: When operating the SCXI-1195 at ambient temperatures >30 °C, a load derating may apply to the total power that the module can handle. DC path resistance Initial...............................................