https://www.phoenixcontact.com/us/products/2810395
Temperature measuring transducer - MINI MCR-SL-PT100-LP-NCSP - 2810395
Please be informed that the data shown in this PDF Document is generated from our Online Catalog. Please find the complete data in the user's
documentation. Our General Terms of Use for Downloads are valid
(http://phoenixcontact.com/download)
Configurable loop-powered temperature transducer for Pt 100 temperature sensors, configured via DIP switches,
with spring-cage connection, not pre-configured. Replacement part: 2810308 MINI MCR-SL-PT100-LP-NC.
Your advantages
2, 3 or 4-wire Pt 100 sensors
Highly-compact loop-powered temperature transducer for electrical isolation, conversion, amplification, and filtering of Pt 100 signals to create
standard signals
Does not require additional auxiliary voltage
Error indication via diagnostic LED and analog signal
2-way isolation
Temperature measuring range of -150°C to +300°C
Supplied by an output loop
Input signals can be configured via DIP switches
Key Commercial Data
Packing unit
1 pc
GTIN
GTIN
4046356142472
Weight per Piece (excluding packing)
80.000 g
Custom tariff number
85437090
Country of origin
Germany
06/27/2020 Page 1 / 9
https://www.phoenixcontact.com/us/products/2810395
Temperature measuring transducer - MINI MCR-SL-PT100-LP-NCSP - 2810395
Technical data
Note
Utilization restriction
EMC: class A product, see manufacturer's declaration in the download
area
Dimensions
Width
6.2 mm
Height
93.1 mm
Depth
102.5 mm
Ambient conditions
Ambient temperature (operation)
-20 °C ... 65 °C
Ambient temperature (storage/transport)
-40 °C ... 85 °C
Maximum altitude
≤ 2000 m
Permissible humidity (operation)
5 % ... 95 % (non-condensing)
Degree of protection
IP20
Noise immunity
EN 61000-6-2 When being exposed to interference, there may be minimal
deviations.
Input data
Configurable/programmable
Yes, unconfigured
Sensor types (RTD) that can be used
Pt 100 (IEC 60751/EN 60751)
Sensor input current
1 mA (constant)
Connection technology
2, 3, 4-wire
Output data
Number of outputs
1
Configurable/programmable
Yes, unconfigured
Current output signal
4 mA ... 20 mA
20 mA ... 4 mA
Max. output current
23 mA (output limit)
Load/output load current output
(Usupply - 12 V) / 22 mA
Ripple
< 20 mVPP (at 500 Ω)
Power supply
Designation
Loop-powered
Supply voltage range
12 V DC ... 30 V DC
Max. current consumption
< 4.5 mA (without signal current)
Power consumption
< 150 mW (without signal current)
Connection data
06/27/2020 Page 2 / 9
https://www.phoenixcontact.com/us/products/2810395
Temperature measuring transducer - MINI MCR-SL-PT100-LP-NCSP - 2810395
Technical data
Connection data
Connection method
Spring-cage connection
Stripping length
8 mm
Conductor cross section solid
0.2 mm² ... 2.5 mm²
Conductor cross section flexible
0.2 mm² ... 2.5 mm²
Conductor cross section AWG
24 ... 12
General
Transmission error in the set measuring range
((90 K / set measuring range [K]) + 0.05)%
Transmission error in the full measuring range
≤ 0,25 %
Maximum temperature coefficient
< 0.02 %/K
Linearity error
< 0.05 % (for full measuring range)
Electrical isolation
Basic insulation according to EN 61010
Overvoltage category
II
Degree of pollution
2
Rated insulation voltage
30 V AC
Test voltage, input/output/supply
1.5 kV (50 Hz, 1 min.)
Electromagnetic compatibility
Conformance with EMC directive
Noise emission
EN 61000-6-4
Noise immunity
EN 61000-6-2 When being exposed to interference, there may be minimal
deviations.
Color
green
Housing material
PBT
Mounting position
any
Conformance
CE-compliant
ATEX
II 3 G Ex nA IIC T4 Gc X
UL, USA/Canada
UL 508 Recognized
Class I, Div. 2, Groups A, B, C, D T4
Fire protection for rail vehicles (DIN EN 45545-2) R22
HL 1 - HL 2
Fire protection for rail vehicles (DIN EN 45545-2) R23
HL 1 - HL 2
Fire protection for rail vehicles (DIN EN 45545-2) R24
HL 1 - HL 2
EMC data
Designation
Electromagnetic RF field
Standards/regulations
EN 61000-4-3
Typical deviation from the measuring range final value
5%
Designation
Fast transients (burst)
Standards/regulations
EN 61000-4-4
06/27/2020 Page 3 / 9
https://www.phoenixcontact.com/us/products/2810395
Temperature measuring transducer - MINI MCR-SL-PT100-LP-NCSP - 2810395
Technical data
EMC data
Typical deviation from the measuring range final value
5%
Designation
Conducted interferences
Standards/regulations
EN 61000-4-6
Typical deviation from the measuring range final value
5%
Standards and Regulations
Electromagnetic compatibility
Conformance with EMC directive
Noise emission
EN 61000-6-4
Connection in acc. with standard
CUL
Standards/regulations
EN 61000-4-2
Designation
Electromagnetic RF field
Standards/regulations
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
Designation
Conducted interferences
Standards/regulations
EN 61000-4-6
Electrical isolation
Basic insulation according to EN 61010
Conformance
CE-compliant
ATEX
II 3 G Ex nA IIC T4 Gc X
UL, USA/Canada
UL 508 Recognized
Class I, Div. 2, Groups A, B, C, D T4
Environmental Product Compliance
REACh SVHC
Lead 7439-92-1
China RoHS
Environmentally Friendly Use Period = 50 years
For details about hazardous substances go to tab “Downloads”, Category
“Manufacturer's declaration”
Drawings
06/27/2020 Page 4 / 9
https://www.phoenixcontact.com/us/products/2810395
Temperature measuring transducer - MINI MCR-SL-PT100-LP-NCSP - 2810395
Application drawing
06/27/2020 Page 5 / 9
https://www.phoenixcontact.com/us/products/2810395
Temperature measuring transducer - MINI MCR-SL-PT100-LP-NCSP - 2810395
Block diagram
06/27/2020 Page 6 / 9
https://www.phoenixcontact.com/us/products/2810395
Temperature measuring transducer - MINI MCR-SL-PT100-LP-NCSP - 2810395
Dimensional drawing
Classifications
eCl@ss
eCl@ss 4.0
27200200
eCl@ss 4.1
27200200
eCl@ss 5.0
27200200
eCl@ss 5.1
27200200
eCl@ss 6.0
27200200
eCl@ss 7.0
27200206
eCl@ss 8.0
27200206
eCl@ss 9.0
27210129
06/27/2020 Page 7 / 9
https://www.phoenixcontact.com/us/products/2810395
Temperature measuring transducer - MINI MCR-SL-PT100-LP-NCSP - 2810395
Classifications
ETIM
ETIM 2.0
EC001446
ETIM 3.0
EC001446
ETIM 4.0
EC001446
ETIM 5.0
EC001446
ETIM 6.0
EC002919
ETIM 7.0
EC002919
UNSPSC
UNSPSC 6.01
30211506
UNSPSC 7.0901
39121008
UNSPSC 11
39121008
UNSPSC 12.01
39121008
UNSPSC 13.2
41112105
UNSPSC 19.0
41112105
Approvals
Approvals
Approvals
UL Recognized / cUL Recognized / cULus Recognized
Ex Approvals
ATEX / UL Listed / cUL Listed / EAC Ex / cULus Listed
Approval details
UL Recognized
http://database.ul.com/cgi-bin/XYV/template/LISEXT/1FRAME/index.htm
FILE E 238705
cUL Recognized
http://database.ul.com/cgi-bin/XYV/template/LISEXT/1FRAME/index.htm
FILE E 238705
06/27/2020 Page 8 / 9
https://www.phoenixcontact.com/us/products/2810395
Temperature measuring transducer - MINI MCR-SL-PT100-LP-NCSP - 2810395
Approvals
cULus Recognized
Phoenix Contact 2020 © - all rights reserved
http://www.phoenixcontact.com