REVISION
THESE DRAWINGS ARE THE PROPERTY OF QUALTEK
ELECTRONICS AND NOT TO BE REPRODUCED FOR
OR COMMUNICATED TO A THIRD PARTY WITHOUT
THE EXPRESS WRITTEN PERMISSION OF AN OFFICER
OF QUALTEK ELECTRONICS.
REV.
B
C
APPROVED
DP
LY
1-2.
CURRENT RATING
A. @ 115 VAC........................................6 A RMS MAX
B. @ 250 VAC........................................6 A RMS MAX
1-3.
OPERATING FREQUENCY......................50/60Hz
1-4.
LEAKAGE CURRENT: EACH LINE TO GROUND
A. @ 115 VAC 60Hz............................2.0 uA MAX
B. @ 250 VAC 50Hz.............................5.0 uA MAX
1-5.
HIPOT RATING (FOR ONE MINUTE)
A. LINE TO GROUND...........................2250 VDC
B. LINE TO LINE.....................................1450 VDC
1-6.
MINIMUM INSERTION LOSS IN dB: (AT 50 OHM SYSTEM)
29.0 ± 0.5
0
40.0 ± 0.1
60.2 ± 0.5
0
M3(2X) PANEL CUTOUT R2.5MAX (4X)
.01
.05
.10
.15
.50
1.0
5.0
10
30
COM. MODE dB
/
12
18
20
25
25
20
13
6
DIF. MODE dB
/
4
4
2
28
50
50
50
25
6.3x0.8(5X)
82.5 ± 0.8
88.0 ± 1.0
FREQ. MHz
59.9 ± 0.3
2. ELECTRICAL SCHEMATIC
DATE
06-01-06
09-13-16
RoHS Compliancy
Updated Dim & Fuse Changed to 5x20mm Size Only
ELECTRICAL CHARACTERISTICS:
1-1.
VOLTAGE RATING...................................115/250VAC
1. PHYSICAL DIMENSIONS
65.0 ± 0.5
63.5 ± 0.5
50.0 ± 0.5
40.0 ± 0.5
28.4 ± 0.5
DESCRIPTION
28.2 ± 0.8
L:
C1:
C2:
R:
2.5mH
0.22uF
0.1uF
1MΩ
1-7.
OPERATING TEMPERATURE.................-25° TO +100° C
(DERATING ABOVE 40˚C)
1-8.
SAFETY:
1.9.
FOR USE WITH UL 544 HEALTH CARE EQUIPMENT
1-10.
FUSE SIZE 5x20mm (FUSES NOT INCLUDED)
1-11.
RoHS COMPLIANT
Dim
range
(mm)
Tolerance
+/(mm)
0-1
0.15
1-6
0.20
6-18
0.30
18-50
0.50
CONTROLLED
Qualtek Electronics Corp.
Part Number:
Description:
PPC DIVISION
865-06/017
EMI MEDICAL FILTER
UNIT: mm
REV. C
50-120
0.70
Drawn / Date
Checked / Date
Approved / Date
120-250
0.90
250-500
1.20
B.ZHANG
01-13-03
01-13-03
JJH / 01-13-03
很抱歉,暂时无法提供与“865-06/017”相匹配的价格&库存,您可以联系我们找货
免费人工找货