CTE9000 / CTU9000...CS Series
OEM stainless steel submersible pressure transducers
FEATURES · 0...100 to 0...2000 mbar, 0...1.5 to 0...30 psi gage · For corrosive media · 4...20 mA output · Field interchangeable
MEDIA COMPATIBILITY
Wetted materials: stainless steel 1.4404 (316L), PUR, PE, ABS, Loctite 603 Housing: protection class IP 68 (according to DIN EN 60529) respectively NEMA 6P1
SPECIFICATIONS9,10
Maximum ratings Supply voltage (reverse polarity protection)2 Temperature limits Storage Operating Compensated Vibration (5 to 500 Hz) Mechanical shock Proof pressure3 CTEM9100..., CTEM9200... CTU9001..., CTU9003... all others 9...32 V
ELECTRICAL CONNECTION
+Vs
-40 to 70°C -25 to 70°C 0 to 70°C 10 gRMS 50 g
+9 to 32 V red sensor -Vs black bare
V RL GND
600 mbar 9 psi 2 x rated pressure
August 2008 / 594
1/4
www.sensortechnics.com
CTE9000 / CTU9000...CS Series
OEM stainless steel submersible pressure transducers
INDIVIDUAL PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 15 V, RL = 100 Ω, tamb = 25°C)
Characteristics
Operating pressure CTEM9100... CTEM9200... CTEM9400... CTEM9600... CTEM91K0... CTEM91K6... CTEM92K0... CTU9001... CTU9003... CTU9005... CTU9010... CTU9015... CTU9020... CTU9030...
Min.
0 0 0 0 0 0 0 0 0 0 0 0 0 0 3.8 15.8
Typ.
Max.
100 200 400 600 1000 1600 2000 1.5 3 5 10 15 20 30 4.2 16.2 0.08 0.05 0.05
Unit
mbar
p si
Zero pressure offset Full scale span
7
4.0 16.0 0.04 0.02 0.02 0.03 0.03 ±0.2 ±0.1 ±0.04 ±0.1 1
mA
Thermal effects (0 to 70°C)4
Offset Span
0.1 bar / 1.5 psi all others
%FSO/°C
Thermal effects (-25 to 0°C) Non-linearity (BSL) and hysteresis Repeatability Output noise (0 < f < 1 kHz) Long term stability6 Response time (10 to 90 %) Power supply rejection Power consumption (IL = 20 mA)
5
Offset Span
±0.5 %FSO
ms %FSO/V mW
Offset Span
0.05 0.08 260
Specification notes (for all devices):
1. The package is an all-sealed housing. For proper function the gage port is vented to the atmosphere through the connecting cable. Thus the vent tube of the cable end must have access to the ambient pressure. 2. The minimum supply voltage is directly proportional to the load resistance seen by the transmitter. For more details see the load limitation diagrams. 3. Proof pressure is the maximum pressure which may be applied without causing damage to the sensing element. 4. Thermal effects tested and guaranteed from 0 to 70°C relative to 25°C. All specifications shown are relative to 25°C. 5. Non-linearity refers to the B est S traight L ine fit measured for offset, full scale span and 1/2 full scale span. 6. Long term stability is the change in output after one year or 1 million pressure cycles. 7. Span is the arithmetic difference in transmitter output signal measured at zero pressure and the maximum operating pressure. 8. Tests are in accordance with EN61000-6-2, April 1999. 9. CE-labelling is in accordance with 89/336/EEC. 10. The pressure transmitters must not be used as safety accessories according to article 1, 2.1.3 of the directive 97/23/EC.
August 2008 / 594
2/4
www.sensortechnics.com
CTE9000 / CTU9000...CS Series
OEM stainless steel submersible pressure transducers
ELECTROMAGNETIC CAPABILITY8
Test conditions
Radiated, radio frequency electromagnetic field immunity (RFI) Electrical fast transient / burst immunity (EFT) Electrostatic discharge immunity test (ESD) Immunity to conducted disturbances induced by radio-frequency fields EN61000-4-3: Grade 3 10 V/m, 80 MHz to 1000 MHz 80 % AMC (1 kHz) Grade 3 ±2 kV Grade 4 ±8 kV, contact discharge Grade 3 0,15 to 80 MHz 10 V, 80 % AMC (1 kHz)
Criterion
A
Interference
很抱歉,暂时无法提供与“CTEM9010G4CXS”相匹配的价格&库存,您可以联系我们找货