T3DMM6-5 /T3DMM6-5-SC Data Sheet
6.5 Digit Digital Multimeter with optional Scanner capability
Broad Measurement Range
DC: 1000 Volts
AC: 750 Volts
Current: 10A
Tools for Improved Debugging
●●
●●
●●
●●
●●
●●
Wide range of measurements – DC/AC voltage and
Current, Resistance, Capacitance, Frequency, Period,
Temperature, and more.
More application coverage from a single Digital
multimeter.
True-RMS measurements – All AC Voltage and
Current ranges give True-RMS readings.
Excellent accuracy regardless of the waveform shape.
Low level measurement, high sensitivity ranges –
Voltage ranges as low as 200 mV full scale, DC Current
200 µA, AC Current 200 µA full scale.
High sensitivity ranges give greater accuracy of
small signal measurements.
Advanced measurement features – Min, Max,
Average, Standard Deviation dBm/dB, Pass/fail,
Histogram, Trend, Relative measurements.
Advanced features for today’s measurement needs.
Built-in cold terminal thermocouple compensation –
4.3 inch (10.92 cm) color TFT-LCD 480 x 272 display.
Accurate Temperature measurements.
USB Device, USB Host and LAN support
Remote control your measurements.
Key Specifications
DC Voltage
200 mV to 1000 V
DC Current
200 µA to 10 A
True RMS AC Voltage
200 mV to 750 V
True RMS AC Current
200 µA to 10 A
2 /4 Wire Resistance
200 Ohms to 100 MOhms
Connectivity
USB Device, LAN
Remote Control
SCPI, LabView Driver
PRODUCT OVERVIEW
Teledyne Test Tools T3DMM6-5 is a 6½ digit digital multimeter incorporating the latest 4.3 inch (10.92cm) dualdisplay technology which can be configured to show data histograms, Data fluctuation Trends, Bar Graph, Statistics
or the traditional Number mode, all in an easy to use interface.
A great feature of the Teledyne Test Tools T3DMM6-5 is it’s ability to made highly accurate True RMS AC Voltage and
Current measurements, meaning no loss of accuracy even when measuring complex voltage and current waveforms.
The T3DMM6-5 is especially well suited for the needs of High precision multifunctional environments, as well as
supporting a full range of automatic measurements.
Main Functions
Application fields
Basic Measurement Function
●●
Research Laboratory
●●
DC Voltage: 200 mV ~ 1000 V
●●
Development Laboratory
●●
DC Current: 200 μA ~ 10 A
●●
Repair and Maintenance
●●
AC Voltage: True-RMS, 200 mV ~ 750 V
●●
Calibration Laboratory
●●
AC Current: True-RMS, 200 µA ~ 10 A
●●
Automatic Production Test
●●
2/4-Wire Resistance: 200 Ω ~ 100 MΩ
●●
General bench-top use
●●
Capacitance: 2 nF ~ 100 mF
●●
Continuity Test: Range is fixed at 2 kΩ
●●
Diode Test: Adjustable range is 0 ~ 4 V.
●●
Real 6½ digit (2,200,000 count) readings resolution
●●
Frequency Measurement: 3 Hz ~ 1 MHz
●●
Up to 10,000 rdgs/s measurement speed
●●
Period Measurement: 1 μs ~ 333 ms
●●
True-RMS AC Voltage and AC Current measuring
●●
Temperature: Support for TC and RTD sensor
●●
Max, Min, Average, Standard Deviation,
dBm/dB, Relative Measurement, Pass/Fail
Histogram, Trend Chart, Bar Graph
●●
●●
Main Features
●●
User-friendly Design
●●
●●
4.3” TFT-LCD, 480*272
●●
●●
Dual display, Chinese and English Menu
●●
Built-in front panel accessible help system
●●
File management (support for U-disc and local storage)
2
1 Gb flash memory for mass storage configuration files
and data files
Built-in cold terminal compensation for thermocouple
T3DMM6-5-SC: Includes Scanner with 16 inputs
(12 general purpose, 4 current channels)
Standard interface: USB Device, USB Host, LAN
USB & LAN remote interfaces support common SCPI
command set. Compatible with other popular DMMs
on the market.
SPECIAL FEATURES
Dual Display
Bar Chart
Statistics
Histogram
Trend Chart
Hold Measurement
dBm Hold Measurement
Rear Panel Interfaces
The Scanner Module is only included with the
T3DMM6-5-SC DMM, it is not included with the
T3DMM6-5
3
SPECIFICATIONS
DC Characteristics
Function
DC Voltage
DC Current
Resistance 6)
Diode Test 7)
Continuity Test
Range 2)
200.0000 mV
2.000000 V
20.00000 V
200.0000 V
1000.000 V4)
200.0000 μA
2.000000 mA
20.00000 mA
200.0000 mA
2.000000 A
10.00000 A5)
200.0000 Ω
2.000000 KΩ
20.00000 KΩ
200.0000 KΩ
1.000000 MΩ
10.00000 MΩ
100.0000 MΩ
0 ~ 2V
2 ~ 4V
2000.0 Ω
Accuracy ± (% of reading + % of range) 1)
Test Current
Or Shunt
Voltage
< 0.03V
< 0.25V
< 0.07 V
< 0.7V
< 0.12 V
< 0.6 V
1 mA
1 mA
100 μA
10 μA
2 μA
200 nA
200 nA || 10 MΩ
1 mA
1 mA
1 mA
24 Hour3)
TCAL °C ± 1°C
90 day
TCAL °C ± 5°C
1 Year
TCAL °C ± 5°C
0.0020 + 0.0015
0.0015 + 0.0004
0.0020 + 0.0003
0.0020 + 0.0005
0.0020 + 0.0005
0.009 + 0.010
0.007 + 0.001
0.006 + 0.008
0.009 + 0.001
0.045 + 0.015
0.090 + 0.002
0.0030 + 0.0031
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0010
0.015 + 0.001
0.300 + 0.010
0.002 + 0.009
0.002 + 0.010
0.002 + 0.010
0.0030 + 0.0020
0.0020 + 0.0004
0.0030 + 0.0004
0.0040 + 0.0004
0.0040 + 0.0008
0.040 + 0.005
0.030 + 0.001
0.030 + 0.005
0.030 + 0.001
0.080 + 0.005
0.120 + 0.005
0.008 + 0.005
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.010 + 0.001
0.030 + 0.001
0.800 + 0.010
0.008 + 0.020
0.008 + 0.020
0.008 + 0.020
0.0040 + 0.0023
0.0035 + 0.0006
0.0040 + 0.0004
0.0050 + 0.0005
0.0055 + 0.0008
0.050 + 0.005
0.050 + 0.002
0.050 + 0.005
0.050 + 0.002
0.100 + 0.012
0.150 + 0.005
0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.012 + 0.001
0.040 + 0.001
0.800 + 0.010
0.010 + 0.020
0.010 + 0.020
0.010 + 0.020
Temperature
coefficient 0 °C to
(TCAL °C - 5 °C)
(TCAL °C + 5 °C) to
50 °C
0.0005 + 0.0003
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0020 + 0.0026
0.0020 + 0.0001
0.0020 + 0.0015
0.0020 + 0.0001
0.0050 + 0.0008
0.0050 + 0.0018
0.0006 + 0.0006
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0005
0.1500 + 0.0002
0.0010 + 0.0020
0.0010 + 0.0020
0.0010 + 0.0020
Remarks:
1)
Specifications are for 90-minute warm-up and 100 NPLC integration time. For integration time 7A DC or 7A AC RMS, use 30 seconds ON followed by 30 seconds OFF.
6)
Specifications are for 4-wire resistance measurement or 2-wire resistance measurement using REL operation. Without REL operation, add 0.2 Ω additional error in 2-wire
resistance measurement.
7)
Accuracy specifications for the voltage measured at the input terminal only. 1 mA test current is typical. Variation in the current source will create some variation in the
voltage drop across a diode junction. Adjustable voltage range: 0 ~ 4V.
4
SPECIFICATIONS
Performance Versus Integration Time – 50 Hz (60 Hz) Power-line Frequency
Integration
Time
Number of
Power line
Cycles5)
(NPLC)
0.005(0.006)
0.05 (0.06)
0.5 (0.6)
1
10
100
Resolution1)
NMRR2)
Readings/s3)
RMS Noise Adder 4) (% of Range)
(ppm Range) (dB)
50 Hz
60 Hz
DCV 20 V
2.7
1.6
1
0.22
0.08
0.035
10000
1000
100
50
5
0.5
10000
1000
100
60
6
0.6
0.0006
0.0004
0.0003
0
0
0
0
0
0
60
60
60
DCV 2 V
200 V
Resistance
2 KΩ 20 KΩ
0.0008
0.0005
0.0003
0.0001
0
0
DCV 1000 V
DCI 2 mA
200 mA
0.0015
0.0008
0.0006
0.0002
0
0
DCV 200 mV
Resistance
200 Ω
DCI 10 A
0.0040
0.0025
0.0025
0.0005
0.0002
0
Remarks:
1)
Typical value. Resolution is defined as the typical 20V range RMS noise.
2)
Normal mode rejection ratio for power-line frequency ± 0.1%. For power-line frequency ± 1%, subtract 20 dB. For ± 3 %, subtract 30 dB.
3)
Maximum rate for DCV, DCI, 2-wire resistance and 4-wire resistance functions.
4)
The basic DC accuracy specifications include RMS noise at 100 NPLC. For 3Hz ac filter and sine wave input.
2)
10 % over range on all ranges except ACV 750 V and ACI 10 A ranges.
3)
Relative to calibration standards.
4)
Specifications are for sine wave input > 5 % of range. For inputs within 1% and 5 % of range and 3 Hz
> 20 Hz
> 200 Hz
10 Hz – 20 Hz
0
0.74
–
20 Hz – 40 Hz
0
0.22
–
40 Hz – 100 Hz
0
0.06
0.73
100 Hz – 200 Hz
0
0.01
0.22
200 Hz – 1 kHz
0
0
0.18
> 1 kHz
0
0
0
Temperature
coefficient 0 °C to
(TCAL °C - 5 °C)
(TCAL °C + 5 °C) to
50 °C
0.200 + 0.005
0.100 + 0.005
0.015 + 0.005
0.030 + 0.005
0.100 + 0.005
0.035 + 0.005
0.015 + 0.005
0.030 + 0.005
0.200 + 0.005
0.100 + 0.005
0.015 + 0.005
0.030 + 0.005
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006
0.030 + 0.006
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006
0.030 + 0.006
0.100 + 0.008
0.035 + 0.008
0.015 + 0.008
Additional Crest Factor Errors (non-sine wave)7)
Crest Factor
error (% of reading)
1 – 2
2 – 3
3 – 4
4 – 5
0.05
0.2
0.4
0.5
Remarks:
1)
Specifications are for 90-minute warm-up, > 3 Hz ac filter and sine wave input.
2)
10 % over range on all ranges except ACV 750 V and ACI 10 A ranges.
3)
Relative to calibration standards.
4)
Specifications are for sine wave input > 5 % of range. For inputs within 1% and 5 % of range and DC 7A or AC RMS 7A, use 30 seconds ON followed by 30 seconds OFF.
7)
For frequency blow 100 Hz, the specification of the low frequency filters is for sine wave only.
8)
Specifications are for sine wave input > 5 % of range. For inputs within 1% to 5 % of range, add 0.1% of range additional error.
Specifications are typical values for 200 µA, 2 mA, 2 A and 10 A ranges when frequency > 1 kHz.
7
SPECIFICATIONS
Frequency and Period Characteristics
Function
Frequency,
Period
Range
200 mV to
750 V
Accuracy± (% of Reading) 1) 2)
Frequency
Range
24 Hour3)
TCAL °C ± 1°C
90 day
TCAL °C ± 5°C
1 Year
TCAL °C ± 5°C
3 Hz – 5 Hz
5 Hz – 10 Hz
10 Hz – 40 Hz
40 Hz –300 KHz
300 KHz – 1 MHz
0.07
0.04
0.02
0.005
0.005
0.07
0.04
0.02
0.006
0.006
0.07
0.04
0.02
0.007
0.007
Frequency
3 Hz– 5Hz
5 Hz– 10 Hz
10 Hz–40 Hz
40 Hz–100 Hz
100 Hz–300 Hz
300Hz–1 KHz
> 1 K Hz
1 s (0.1 ppm)
0
0
0
0
0
0
0
0.1 s (1 ppm)
0.12
0.17
0.20
0.06
0.03
0.01
0
Gate Time (Resolution)
0.01 s (10 ppm)
0.12
0.17
0.20
0.21
0.21
0.07
0.02
Temperature
coefficient
0 °C to (TCAL °C - 5 °C)
(TCAL °C + 5 °C) to 50 °C
0.005
0.005
0.001
0.001
0.001
0.001 s (100 ppm)
0.12
0.17
0.20
0.21
0.21
0.07
0.02
Remarks:
1)
Specifications are for 90 minutes warm-up, using 1s gate time.
2)
For frequency ≤ 300 kHz, the specification is the 10 % to 110 % of range of the AC input voltage. For frequency > 300 kHz, the specification is the 20 % to 110 % of range of
the AC input voltage. The maximum input is limited to 750 V rms or 8 × 107 Volts-Hz (whichever is less). The 200 mV range is full range input or input that is larger than the
full range. For 20 mV to 200 mV, multiply % of reading error × 10.
3)
Relative to calibration standards.
Capacitance Characteristic
Accuracy± (% of Reading + % of Range) 1)
Function
Range 2)
Test Current
1 Year
TCAL °C ± 5°C
Capacitance
2.0000 nF
20.000 nF
200.00 nF
2.0000 μF
20.000 μF
200.00 μF
2.0000 mF
20.000 mF
100.00 mF
10 μA
10 μA
100 μA
100 μA
1 mA
1 mA
1 mA
1 mA
1 mA
2 + 2.4
1 + 0.1
1 + 0.1
1 + 0.1
1 + 0.1
1 + 0.1
1 + 0.1
1 + 0.1
3 + 0.1
Remarks:
1)
Specifications are for 90 minutes warm-up and using REL operation. Additional errors may be caused by non-film capacitors.
2)
Specifications are the 1% to 110 % of range on 2 nF range and 10 % to 110 % of range on all other ranges.
8
Temperature
coefficient
0 °C to (TCAL °C - 5 °C)
(TCAL °C + 5 °C) to 50 °C
0.05 + 0.06
0.05 + 0.01
0.01 + 0.01
0.01 + 0.01
0.01 + 0.01
0.01 + 0.01
0.01 + 0.01
0.01 + 0.01
0.05 + 0.02
Temperature Characteristics
Function
Temperature
Probe
Type
Accuracy± (% of Reading) 1)
Type
α = 0.00385
RTD 2)
(R0 is 49 Ω to 2.1 kΩ)
B
E
J
K
Thermocouple3)
N
R
S
T
Optimum
Range
1 Year
TCAL °C ± 5°C
-200 °C ~ 660 °C
0.16 °C
Temperature
coefficient
0 °C to (TCAL °C - 5 °C)
(TCAL °C + 5 °C) to 50 °C
0.01°C
0 °C ~ 1820 °C
-270 °C ~ 1000 °C
-210 °C ~ 1200 °C
-270 °C ~ 1370 °C
-270 °C ~ 1300 °C
-270 °C ~ 1760 °C
-270 °C ~ 1760 °C
-270 °C ~ 400 °C
0.76 °C
0.5 °C
0.5 °C
0.5 °C
0.5 °C
0.5 °C
0.6 °C
0.5 °C
0.14 °C
0.02 °C
0.02 °C
0.03 °C
0.04 °C
0.09 °C
0.11 °C
0.03 °C
Remarks:
1)
Specifications are for 90 minutes warm-up. Exclusive of sensor error.
2)
Specification is for 4WR sensor measurement or 2WR measurement using REL operation.
3)
Relative to cold junction temperature, accuracy is based on ITS-90. Built-in cold junction temperature refers to the temperature inside the banana jack
and its accuracy is ± 2.5 .
Measurement Rate
Function
DC Voltage
DC Current
2-wire Resistance
4-wire Resistance
AC Voltage
AC Current
Frequency and Period1)
Capacitance2)
Measurement rate 3)
Setting
0.005 (0.006) NPLC
0.05 (0.06) NPLC
0.5 (0.5) NPLC
1 NPLC
10 NPLC
100 NPLC
3 Hz AC Filter
20 Hz
200 Hz
1 s Gate time
0.1 s
0.01 s
0.001 s
100 mF Range
Integration
100 (100) µs
1 (1) ms
4 (4) ms
20 (16.7) ms
200 (167) ms
2 (1.67) s
Readings/s 50 Hz (60 Hz)
10000 (10000)
1000 (1000)
100 (100)
50 (60)
5 (6)
0.5 (0.6)
0.5
2
50
1
10
100
500
0.5
Remarks:
1)
20 V range, 1 kHz input.
2)
The measurement period changes with the capacitance under test.
3)
Auto zero off, auto range off.
9
MEASURING METHOD AND OTHER CHARACTERISTICS
DC Voltage
Input Resistance
200 mV, 2 V, 20 V ranges: Selectable 10 MΩ or > 10 GΩ
(For these ranges, input beyond ±26V are clamped through 106 kΩ (typical)
200 V and 1000 V ranges; 10 MΩ ± 1 %
Input Offset Current
50 pA, 25 °C, typical
Input Protection
1000 V
CMRR (common mode rejection ratio) 140 dB for 1 kΩ unbalance in LO lead, ± 500 VDC peak maximum
Resistance
Measurement Method
Open-circuit Voltage
Max. Lead Resistance (4-wire)
Offset Compensation
Input Protection
Selectable 4-wire or 2-wire resistance
Current source referenced to LO input
Limited to 300 Vrms (or > 5 Arms) will cause self-heating of signal-conditioning components and these error are included in the
instrument specifications. Internal temperature changes due to self-heating may cause additional error on lower ac voltage ranges.
The additional error will be lower than 0.02 % of reading and will generally dissipate within a few minutes as the signal-conditioning
components return to normal operating temperature.
Frequency and Period
Measurement Method
Input Impedance
Input Protection
Measurement Considerations
Settling Time Considerations
Reciprocal-counting technique, AC-coupled input using the AC voltage function.
1 MΩ ± 2% in parallel with < 150 pF capacitance on any range
750 V rms on all ranges
All frequency counters are susceptible to error when measuring low-voltage,
low-frequency signals. Shielding inputs from external noise is recommended.
Errors will occur when attempting to measure the frequency or period of an input following
a dc offset voltage change. Make sure the RC input circuit at the input terminals has fully
settled (about 1 s) before making accurate measurements.
Capacitance Measurement
Measurement Method
Connection Type
Input Protection
Measurement considerations
Apply constant current into the capacitance, and measure the rate of voltage change.
2-wire
1000 V on all ranges
Since small capacitance measurements are susceptible to the external noise, shielding
inputs from external noise pickup is critical for minimizing measurement errors.
Temperature Measurement
Measurement Method
Measurement considerations
Support for TC and RTD types of sensor
The built-in cold junction temperature tracks the temperature inside the banana jack.
The change of the temperature in the banana jack may cause additional error. When using
the built-in cold junction compensation, connect the sensor terminal of the thermocouple
to the banana jack and allow it to warm up for at least 3 minutes to minimize errors.
11
MEASURING METHOD AND OTHER CHARACTERISTICS
Triggering and Storage
Trigger
Time Base Resolution
Trigger Delay
Reading Sensitivity
Single Trigger Samples
External Trigger Input
VMC Output
Pre-trigger or Post-trigger, Internal Trigger or External Trigger,
Rising Edge Trigger or Falling Edge Trigger
40 µs, 0.01 % Accuracy
0 to 1000s
0.01 %,0.1 %,1 %or10 %reading
1 to 599999999
Level: TTL compatible (High level when input terminal is disconnected)
Trigger: Selectable rising edge or falling edge
InputImpedance:≥30KΩ//500pF
Delay: 100000 operations, at 1 A 30V DC (at 0.5 Hz)
> 100000 operations, at 0.3 A 125 V AC (at 0.5 Hz)
75 mΩ (maximum at 6 VDC, 1 A)
5 ms maximum on/off (channel to channel )
Minimum 1 GΩ ( 500 VDC )
Clamp terminal, #24 AWG wire size
Remarks: Please do not use the scanner capability to measure signals that exceed the specifications
as this can damage or reduce the life of the scanner and cause electric shock or personal injury.
Channel Capabilities
Item
No. of wires
No. of channels
DCV, ACV
DCI, ACI 1)
2W Resistance
4W Resistance
Capacitance
Diode/Continuity
Period/Frequency
Temp (Thermocouple)
Temp (RTD)
2 wires (H , L)
2 wires (H , L)
2 wires (H , L)
4 wires (Input H , L + sense H, L)
2 wires (H , L)
2 wires (H , L)
2 wires (H , L)
2 wires (H , L)
2 wires (H , L)
12 (CH1 ~ CH12)
4 (CH13 ~ CH16) (2A Range Only )
12 (CH1 ~ CH12)
6 pairs (CH1 [input] & CH7[sense], 2&8, •••, 6&12)
12 (CH1 ~ CH12)
12 (CH1 ~ CH12)
12 (CH1 ~ CH12)
12 (CH1 ~ CH12)
12 (CH1 ~ CH12)
Remarks:
1)
For continuous current < 2.2 A, Accuracy ± (% 3 (reading) + 0.02 % (range)).
13
GENERAL SPECIFICATIONS
Power Supply
AC 100 V ~ 120 V
45 Hz – 66 Hz
AC 200 V ~ 240 V
45 Hz – 66 Hz
Detect the power-line frequency automatically at power-on, 400 Hz defaults to 50 Hz
Power Consumption
25VA max
Mechanical
Dimension
Weight
(length × width × height): 345.45 mm × 260.29 mm × 107.21 mm
3.377 Kg (Net weight)
Other characteristics
Display Screen
Working Environment
EMC
Safety
Remote Interface
Programming Language
Warm Up Time
14
4.3 ”TFT-LCD with resolution 480*272
Full accuracy for 0°C to 50 °C
Full accuracy to 40°C , 80 % R.H., Non-condensing
Storage Temperature – 20 °C to 70 °C
Shock and Vibration: conforming to MIL-T-28800E, III, 5 level (only for sine)
Height above sea level: up to 3000 meters
Conforming to EMC (2004/108/EC) and EN 61326-1:2013
IEC 61010-1; EN 61010-1; UL 61010-1; CAN/CSA-C22.2 No. 61010-1
Measurement CAT I 1000 V/CAT II 600 V
10/100Mbit LAN, USB2.0 Full Speed Device, Host
Standard SCPI
90 minutes
Ordering Information
Product Name
Models
Standard Accessories
Teledyne Test Tools T3DMM6-5 Digital Multimeter
T3DMM6-5
6.5 Digit Digital Multimeter
T3DMM6-5-SC 6.5 Digit Digital Multimeter including Scanner capability
Power Cord that fits the destination country
Two Test Leads, Two Alligator Clips
USB Cable
Quick Start Guide
Warranty Card
15
ABOUT TELEDYNE TEST TOOLS
Company Profile
Teledyne LeCroy is a leading provider of oscilloscopes, protocol
analyzers and related test and measurement solutions that
enable companies across a wide range of industries to design
and test electronic devices of all types. Since our founding
in 1964, we have focused on creating products that improve
productivity by helping engineers resolve design issues
faster and more effectively. Oscilloscopes are tools used by
designers and engineers to measure and analyze complex
electronic signals in order to develop high-performance
systems and to validate electronic designs in order to improve
time to market.
The Teledyne Test Tools brand extends the Teledyne LeCroy
product portfolio with a comprehensive range of test
equipment solutions. This new range of products delivers
a broad range of quality test solutions that enable engineers
to rapidly validate product and design and reduce time-tomarket. Designers, engineers and educators rely on Teledyne
Test Tools solutions to meet their most challenging needs for
testing, education and electronics validation.
Location and Facilities
Headquartered in Chestnut Ridge, New York, Teledyne
Test Tools and Teledyne LeCroy has sales, service and
development subsidiaries in the US and throughout
Europe and Asia. Teledyne Test Tools and Teledyne LeCroy
products are employed across a wide variety of industries,
including semiconductor, computer, consumer electronics,
education, military/aerospace, automotive/industrial, and
telecommunications.
Distributed by:
Teledyne LeCroy
(US Headquarters)
700 Chestnut Ridge Road
Chestnut Ridge, NY. USA 10977-6499
Phone:
Fax Sales:
Phone Support:
Email Sales:
Email Support:
Web Site:
800-553-2769 or 845-425-2000
845-578-5985
1-800-553-2769
contact.corp@teledynelecroy.com
support@teledynelecroy.com
http://teledynelecroy.com/
Teledyne LeCroy
(European Headquarters)
Teledyne LeCroy GmbH
Im Breitspiel 11c
D-69126 Heidelberg, Germany
Phone:
Fax:
Phone Service:
Phone Support:
Email Sales:
Email Service:
Email Support:
Web Site:
+49 6221 82700
+49 6221 834655
+49 6221 8270 85
+49 6221 8270 28
contact.gmbh@teledynelecroy.com
service.gmbh@teledynelecroy.com
tlc.t3.appsupport.eu@teledyne.com
http://teledynelecroy.com/germany
teledynelecroy.com
© 2018 Teledyne Test Tools is a brand and trademark of Teledyne LeCroy Inc. All rights reserved. Specifications, prices, availability and delivery subject to change without notice.
Product brand or brand names are trademarks or requested trademarks of their respective holders.
T3 stands for Teledyne Test Tools.