Rogowski Current Probes
User Manual
T3RC0300-UM
T3RC0600-HF
T3RC3000-HF
T3RC3000-LF
T3RC6000-LF
INTRODUCTION ........................................................................................3
TECHNICAL SPECIFICATION .......................................................................3
STANDARDS AND APPROVALS ..................................................................5
ROGOWSKI PROBES OVERVIEW ................................................................6
T3RC0300-UM HANDLING INSTRUCTIONS .................................................7
T3RC0300-UM SAFETY AND PRE-USE CHECKS ............................................8
T3RC0600-HF AND T3RC3000-HF HANDLING INSTRUCTIONS ......................9
T3RC0600-HF AND T3RC3000-HF SAFETY AND PRE-USE CHECKS ...............10
T3RC3000-LF AND T3RC6000-LF HANDLING INSTRUCTIONS .....................11
T3RC3000-LF AND T3RC3000-LF SAFETY AND PRE-USE CHECKS ................11
ADDITIONAL SAFETY AND PRE-USE CHECKS FOR ALL PROBES...................12
POWER OPTIONS .....................................................................................13
REPLACING THE BATTERIES ......................................................................13
SWITCHING ON ....................................................................................... 13
OBTAINING THE BEST MEASUREMENT .................................................... 14
CLEANING AND DECONTAMINATION ...................................................... 15
WARRANTY ............................................................................................. 15
RETURNING YOUR ROGOWSKI PROBE FOR RECALIBRATION OR REPAIR . 15
2
INTRODUCTION
The Teledyne Test Tools Rogowski Current Waveform Transducers combines high
bandwidth performance with minimal disruption to the circuit under test. These instructions
must be followed whenever the unit is used. They are intended to help you obtain the best
and safest performance from the transducer.
Throughout these instructions there are a number of warnings which must be observed to
ensure safe operation of the Teledyne Test Tools Rogowski probe. These warnings are
identified by the following symbol:
The intended use is for the measurement of AC current. Teledyne Test Tools accepts no
responsibility for any damage resulting from careless use, or failure to observe these
instructions.
TECHNICAL SPECIFICATION
Peak Current
Refer to ID Label
Peak Output Voltage
±6V (corresponding to ±Peak Current Rating)
Output Cable
0.5m BNC to BNC 50Ω cable
In order to maintain safety, the BNC terminal must be connected to
equipment that is separated from hazardous voltages by at least
double insulation.
Output Load
≥ 100kΩ (for rated accuracy – recommend DC1MΩ on scope)
= 50Ω (The Integrator can drive a 50Ω load if necessary)
NB. A load of 50Ω will reduce the Rogowski Probe sensitivity by
half its normal value and limit the peak output voltage to ±2V.
Bandwidth (-3dB)
T3RC0300-UM: 9.2 Hz - 30 MHz
T3RC0600-HF: 12 Hz - 30 MHz
T3RC3000-HF: 3 Hz - 23 MHz
T3RC3000-LF: 0.11 Hz - 6.5 MHz
T3RC6000-LF: 0.055 Hz - 6.5 MHz
Operating Temp. Range
T3RC0300-UM
T3RC0600-HF
T3RC3000-HF
0oC to +40oC (Integrator electronics)
Operating Temp. Range
T3RC3000-LF
T3RC6000-LF
0oC to +40oC (Integrator electronics)
- 40oC to +125oC (Coil and cable)
- 20oC to +100oC (Coil and cable)
3
Accuracy (typ.)
(5% to 100% Peak Current)
Calibration
Nominally ±0.2% with the conductor central in the Rogowski loop.
See the calibration certificate for further details.
Positional Accuracy
T3RC0300-UM
T3RC0600-HF
T3RC3000-HF
±2% variation of accuracy with conductor position in the loop
(see ‘Obtaining The Best Measurement’)
T3RC3000-LF
T3RC6000-LF
±1% variation of accuracy with conductor position in the loop
(see ‘Obtaining The Best Measurement’)
Linearity
±0.05% of reading
Noise (Max)
T3RC0300-UM:
T3RC0600-HF:
T3RC3000-HF:
T3RC3000-LF:
T3RC6000-LF:
2.5 mV rms
1.7 mV rms
1.4 mV rms
2.5 mV rms
2.5 mV rms
Oscilloscope
Attenuation Ratio (for
correct Oscilloscope
vertical scaling)
T3RC0300-UM:
T3RC0600-HF:
T3RC3000-HF:
T3RC3000-LF:
T3RC6000-LF:
50
100
500
500
1000
Coil Insulation
T3RC0300-UM:
T3RC0600-HF:
T3RC3000-HF:
T3RC3000-LF:
T3RC6000-LF:
1.2 kV peak
5 kV peak
5 kV peak
10 kV peak
10 kV peak
Absolute di/dt Ratings
T3RC0300-UM:
T3RC0600-HF:
T3RC3000-HF:
T3RC3000-LF:
T3RC6000-LF:
20 kA/us
40 kA/us
80 kA/us
11 kA/us
11 kA/us
Environmental
Indoor use / Altitude up to 2000m
Installation Category II, Pollution Degree 2.
Max. Rel. Humidity: 80% up to 31oC
Decreasing Linearly to 50% at 40oC
Overvoltage Transient
T3RC0300-UM:
T3RC0600-HF:
T3RC3000-HF:
T3RC3000-LF:
T3RC6000-LF:
4
Cat II 600V
Cat III 600V
Cat III 600V
Cat III 1000V
Cat III 1000V
Operation with External Adaptor
DC Supply Voltage
DC Quiescent Current
DC Socket Type
12Vdc (+/-10%)
60mA @12VDC – using standard alkaline battery
130mA @ 12VDC – using NiMH Rechargeable battery
1.3mm jack socket – Centre Positive Tip
The unit should only be powered by the supplied mains adaptor.
Adaptor Supply Voltage
100 – 240VAC
Adaptor Supply Frequency
50 - 60Hz
Adaptor Supply Current
450mA Max @ 100VAC Input
Supply Voltage Fluctuation
up to 10%
Operation with Battery
Using 4 x 1.5V AA alkaline Battery Life: 25hrs typ.
Battery Supply
Using 4 x 1.2V NiMH rechargeable batteries
Battery Life: 10hrs typ. (based on a 2000mAh cell)
Typical Recharge Time: 40 hrs
STANDARDS AND APPROVALS
EMC Approvals
EN61326-1:2013
FCC Title 47 (CFR:2010, Part 15b)
Safety Approvals
EN61010-1:2010
EN61010-2-032:2012
Quality System
IS09001:2015
5
Teledyne Test Tools Rogowski Probes Overview
1. Rogowski Coil (Probe)
2. Probe ‘Free End’
3. Ferrule
4. Connecting Cable
5. DC Power Socket
6. DC Power Status LED
7. Enclosure
8. BNC Output Socket
9. 4 x AA batteries
10.Status Indicator LED
GREEN = ON
RED = Battery LOW
11.Power Button
6
T3RC0300-UM HANDLING INSTRUCTIONS
Every effort has been made to ensure that the T3RC0300-UM is as robust as possible. To
achieve such an extremely thin Rogowski coil the T3RC0300-UM is necessarily delicate.
To prolong the life of the coil please observe the following handling instructions.
When fully inserted the ferrule will grip the coil.
OPEN IN THIS
DIRECTION
When clipping or unclipping the coil, hold the ferrule as shown
Never force the free end of the coil into the ferrule.
This may damage the coil insulation.
Take care not to put any force onto the cable attached to the
coil. This may damage the coil.
When not in use return the coil to its protective case.
The minimum bend radius of this coil is 10mm. This is the minimum radius that the
coil can be bent without causing damage or shortening its life.
7
T3RC0300-UM SAFETY AND PRE-USE CHECKS
The integrity of the insulation around the Rogowski coil itself must be VISUALLY
INSPECTED before use, and the T3RC0300-UM must NOT BE USED if there are
signs of damage.
When bending the flexible coil around a conductor, avoid tight bends and sharp edges
that could damage the coil.
The voltage rating (safe PEAK working voltage) is clearly labelled on the cable
adjacent to the coil. For the T3RC0300-UM Ultra Mini this is 1.2kV peak. Never use at
voltages greater than this value.
Voltage ratings are only valid if the ‘free end’ of the coil is fully inserted into the socket,
and remains fully inserted during use. The coil has a friction fit, the coil is fully inserted
when the user feels they can ease the coil free end into the ferrule no further.
The voltage ratings are appropriate for intermittent use of the T3RC0300-UM
as a test instrument and not for continuous use in a permanent installation.
The ratings are derived from the following standard test: Coils rated for 1.2kV peak
are flash tested for one minute at 3.0kVrms using a 50Hz sine-wave voltage.
For permanent installation the coil should be situated such that corona, which will
eventually damage the coil insulation, cannot occur. For information regarding
permanent installation of Teledyne Test Tools Rogowski coils on high voltage
equipment please consult Teledyne Test Tools.
The user must ensure that the absolute maximum rated di/dt is not exceeded.
THE INTEGRATOR
The T3RC0300-UM must only be used with oscilloscopes or monitoring equipment
which have their BNC INPUTS PROPERLY GROUNDED.
8
T3RC0600-HF AND T3RC3000-HF HANDLING INSTRUCTIONS
To achieve such an extremely thin Rogowski coil, the T3RC0600-HF and T3RC3000-HF are
necessarily delicate.
Every effort has been made to ensure that the T3RC0600-HF and T3RC3000-HF are as robust as
possible. To prolong the life of the coil please observe the following handling instructions.
Releasing The Coil
The free end of the coil is
unclipped by applying gentle
pressure in the direction shown.
Correct Coil Insertion
The coil is correctly positioned when the
free end of the coil is fully inserted into
the ferrule.
When correctly inserted, the ferrule will
grip the coil and hold it securely.
Never force the free end of the coil into the ferrule.
This may damage the coil insulation.
Take care not to put any force onto the cable attached to the coil.
This may damage the coil.
When not in use return the Rogowski coil to its protective case.
The minimum bend radius of this coil is 14mm. This is the minimum radius that the coil
can be bent without risk of damaging the coil or shortening its life.
9
T3RC0600-HF and T3RC3000-HF SAFETY AND PRE-USE
CHECKS
The integrity of the insulation around the Rogowski coil itself must be VISUALLY
INSPECTED before use, and the T3RC0600-HF and T3RC3000-HF must NOT
BE USED if there are signs of damage.
When bending the flexible coil around a conductor, avoid tight bends and sharp edges
that could damage the coil.
The voltage rating (safe PEAK working voltage) is clearly labelled on the cable
adjacent to the coil. For the T3RC0600-HF and T3RC3000-HF this is 5kV peak. Never
use at voltages greater than this value.
Voltage ratings are only valid if the ‘free end’ of the coil is fully inserted into the socket,
and remains fully inserted during use. The coil has a friction fit, the coil is fully inserted
when the user feels they can ease the coil free end into the ferrule no further.
The voltage ratings are appropriate for intermittent use of the T3RC0600-HF
and T3RC3000-HF as a test instrument and not for continuous use in a
permanent installation.
The ratings are derived from the following standard test: Coils rated for 5kV peak are
flash tested for one minute at 8kVrms using a 50Hz sine-wave voltage.
For permanent installation the coil should be situated such that corona, which will
eventually damage the coil insulation, cannot occur. For information regarding
permanent installation of Teledyne Test Tools Rogowski coils on high voltage
equipment please consult Teledyne Test Tools.
The user must ensure that the absolute maximum rated di/dt is not exceeded
THE INTEGRATOR
The T3RC0600-HF and T3RC3000-HF must only be used with oscilloscopes or
monitoring equipment which have their BNC INPUTS PROPERLY GROUNDED.
10
T3RC3000-LF AND T3RC6000-LF HANDLING INSTRUCTIONS
The coil free end is fully inserted in the ferrule when the user feels the free end of the coil engage
with the internal click-in mechanism inside the ferrule.
Visual indication that the coil mechanism is fully inserted is provided by the black coil marker
situated near the end of the coil, as shown below:
Once the free end of the coil is fully engaged in the ferrule, turn the ferrule lock nut to lock the coil
end in position.
To remove the probe from the device under test first loosen the ferrule lock nut then remove the
coil end from the ferrule.
The locking mechanism must not be actuated at temperatures below 0 C.
Care must be taken not to over-tighten the locking nut.
When not in use return the Rogowski probe to its protective case.
T3RC3000-LF and T3RC6000-LF SAFETY AND PRE-USE
CHECKS
The integrity of the insulation around the Rogowski coil itself must be VISUALLY
INSPECTED before use, and the T3RC3000-LF and T36000-LF must NOT BE
USED if there are signs of damage.
When bending the flexible coil around a conductor, avoid tight bends and sharp edges
that could damage the coil.
11
The voltage rating (safe PEAK working voltage) is clearly labelled on the cable
adjacent to the coil. For the T3RC3000-LF and T36000-LF this is 10kV peak. Never
use at voltages greater than this value.
Voltage ratings are only valid if the ‘free end’ of the coil is fully inserted into the socket,
and remains fully inserted during use i.e. right up to the black indicator line on the free
end of the coil.
The voltage ratings are appropriate for intermittent use of the T3RC3000-LF
and T36000-LF as a test instrument and not for continuous use in a
permanent installation.
The ratings are derived from the following standard test: Coils rated for 10kV peak
are flash tested for one minute at 15kVrms using a 50Hz sine-wave voltage.
For permanent installation the coil should be situated such that corona, which will
eventually damage the coil insulation, cannot occur. For information regarding
permanent installation of Teledyne Test Tools Rogowski coils on high voltage
equipment please consult Teledyne Test Tools.
The user must ensure that the absolute maximum rated di/dt is not exceeded
THE INTEGRATOR
The Rogowski probe must only be used with oscilloscopes or monitoring equipment
which have their BNC INPUTS PROPERLY GROUNDED.
SAFETY AND PRE-USE CHECKS FOR ALL TELEDYNE TEST
TOOLS ROGOWSKI PROBES
The Teledyne Test Tools Rogowski probes do not provide protection of
the user from electric shock whilst applying or removing the probe from
HAZARDOUS LIVE conductors. Additional protective means are
necessary to avoid electric shock from HAZARDOUS LIVE conductors
which cannot be de-energised.
If HAZARDOUS LIVE voltages are present and accessible in the
installation, appropriate protective equipment must be used.
The Rogowski coil must not be used if the coil outer layer is damaged or in
any way compromised.
The Teledyne Test Tools Rogowski probes must only be used in
conjunction with properly earthed test equipment.
If the Teledyne Test Tools Rogowski probes are used in a manner not
specified by the manufacturer the protection provided by the equipment
may be impaired.
12
POWER OPTIONS
The Teledyne Test Tools Rogowski probes can be powered by an external DC source or by battery.
External Supply
The Teledyne Test Tools Rogowski probes can be powered by an external DC supply. The DC
voltage must be 12V(±10%) using a centre positive connector. When the DC supply is present
a RED indicating LED adjacent to the socket is illuminated.
Standard Alkaline Batteries
Four fully charged AA alkaline batteries provide approx. 25 hours operation. Healthy batteries
are indicated by the GREEN LED. If the LED is RED the batteries are depleted and must be
replaced.
When the DC voltage is present the batteries are inoperative.
REPLACING THE BATTERIES
Low battery is indicated by a red LED located near the power switch. To replace the batteries,
switch-off the Rogowski probe integrator, remove the battery cover and replace with appropriate
batteries for your Teledyne Test Tools Rogowski probe type.
It is recommended that the circuit under test is de-energised or the probe is removed from the
test circuit before replacing the batteries.
SWITCHING ON
Before installing the Rogowski probe and taking a measurement refer to SAFETY AND
PRE-USE CHECKS and HANDLING INSTRUCTIONS to ensure safe operation of
your Rogowski probe.
Your Rogowski probe must only be used with oscilloscopes or monitoring equipment
which have their BNC inputs properly grounded.
The user is advised to de-energise the circuit under test first. If the probe is being fitted
to a live circuit the user must ensure the use of appropriate personal protective
equipment.
1. Connect the BNC output of the transducer to your grounded oscilloscope or current
monitoring equipment.
2. Having carried out the VISUAL INSPECTION of the Rogowski coil, unclip the coil and wrap
it around the de-energised conductor under test.
13
3. Insert the ‘free end’ of the coil fully inside the ferrule.
4. Re-energise the conductor.
5. Use the push button to switch ON by pressing and releasing the button, and to turn OFF
by depressing the button fully; the LED indicates that the transducer is ON when the LED
is GREEN.
6. After switch-on the Teledyne Test Tools Rogowski probes, they may require a settling period
of up to 2 minutes before providing a correct current measurement. The settling period is
dependent on warm-up time and the low frequency bandwidth.
OBTAINING THE BEST MEASUREMENT
The Rogowski coil should be positioned so that the conductor under test is encircled by the coil
but is not adjacent to the cable attachment (see picture below). The diagram shows the direction
a positive current should pass through the coil loop in order to obtain a positive output voltage.
The Teledyne Test Tools Rogowski probe has been calibrated with the conductor near the
centre (position A), and this is the ideal position for accuracy.
For the best high frequency performance, the centre of the current should lie on the line shown
A to B, where B is half way around the circumference of the coil.
B
I current
into page
A
KEEP CONDUCTORS
AWAY FROM
SHADED AREA
SENSITIVITY TO EXTERNAL CURRENTS AND VOLTAGES
The sensitivity of the Teledyne Test Tools Rogowski probes to currents that are outside the closed
loop of the Rogowski coil is very small, provided that the external currents are less than the
current rating of the Teledyne Test Tools Rogowski probes or that such currents are relatively
distant from the coil. In the vicinity of a multi-turn conductor the effects are far stronger than from
a single conductor carrying the same current, and such positions should be avoided.
14
If there is a surface with a high voltage very close to the coil and the voltage is subject to high
rates of change or high frequency oscillations in the MHz range, then measurement error can
arise due to capacitive coupling to the coil.
To check for any unwanted response to adjacent fields, it is recommended to quantify the output
of the Teledyne Test Tools Rogowski probes when close to (but not encircling) the conductor under
test. This will reveal the magnitude of any unwanted responses to currents close to, but outside,
the coil.
OUTPUT CABLES
Your Teledyne Test Tools Rogowski probe has been supplied with a 50ohm BNC to BNC output
cable. It is possible to use longer cables and these should be of a 50ohm single screened
co-axial cable type.
Teledyne Test Tools does not consider the use of extension cables to be problematic from the
noise viewpoint but consideration should be given to the routing of very long cables.
CLEANING AND DECONTAMINATION
To prevent surface contamination the Rogowski coil and cable should be inspected regularly.
The Rogowski coil must be disconnected from any external circuit and turned OFF prior to
cleaning. To clean the coil and cable use a mild detergent and a damp cloth to remove any
contamination, wipe dry with clean cloth to remove the detergent and dry thoroughly before
placing back into service.
Teledyne Test Tools accepts no responsibility for the use of any other cleaning solvents or
cleaning methods.
WARRANTY
The Rogowski probe is guaranteed to be free from defects due to materials and workmanship
for 1 year.
In the event of a defect where the transducer has not been misused the Rogowski probe should
be returned to Teledyne Test Tools with freight charges to be paid by the customer. Correction
shall be in the form of repair or replacement.
RETURNING YOUR ROGOWSKI PROBE FOR RECALIBRATION OR
REPAIR
To return the Rogowski probe to Teledyne Test Tools for repair or recalibration please contact
Teledyne LeCroy in advance for shipping instructions (refer to information on the back page for
contact details and our service procedure).
15
ABOUT TELEDYNE TEST TOOLS
Company Profile
Teledyne LeCroy is a leading provider of oscilloscopes, protocol
analyzers and related test and measurement solutions that
enable companies across a wide range of industries to design
and test electronic devices of all types. Since our founding in
1964, we have focused on creating products that improve
productivity by helping engineers resolve design issues faster
and more effectively. Oscilloscopes are tools used by
designers and engineers to measure and analyze complex
electronic signals in order to develop high-performance
systems and to validate electronic designs in order to improve
time to market.
The Teledyne Test Tools brand extends the Teledyne LeCroy
product portfolio with a comprehensive range of test
equipment solutions. This new range of products delivers
a broad range of quality test solutions that enable engineers to
rapidly validate product and design and reduce time-tomarket. Designers, engineers and educators rely on Teledyne
Test Tools solutions to meet their most challenging needs for
testing, education and electronics validation.
Teledyne LeCroy
(US Headquarters)
Location and Facilities
700 Chestnut Ridge Road
Chestnut Ridge, NY. USA 10977-6499
Headquartered in Chestnut Ridge, New York, Teledyne
Test Tools and Teledyne LeCroy has sales, service and
development subsidiaries in the US and throughout
Europe and Asia. Teledyne Test Tools and Teledyne LeCroy
products are employed across a wide variety of industries,
including semiconductor, computer, consumer electronics,
education, military/aerospace, automotive/industrial, and
telecommunications.
Phone:
800-553-2769 or 845-425-2000
Fax Sales:
845-578-5985
Phone Support: 1-800-553-2769
Email Sales:
contact.corp@teledynelecroy.com
Email Support: support@teledynelecroy.com
Web Site:
http://teledynelecroy.com
World wide support contacts can be found at:
h�ps://teledynelecroy.com/support/contact
Distributed by:
World wide instrument service can be found at:
h�ps://teledynelecroy.com/support/service.aspx
RoHS and WEEE informa�on can be found at:
h�ps://teledynelecroy.com/support/rohs.aspx
teledynelecroy.com
© 2018 Teledyne Test Tools is a brand and trademark of Teledyne LeCroy Inc. All rights reserved. Specifications,prices,availability and delivery subject to change without notice.
Product brand or brand names are trademarks or requested trademarks of their respective holders.
T3 stands for Teledyne Test Tools.
30Apr2019