T3SA3100 / T3SA3200 Data Sheet
2.1 GHz and 3.2 GHz Spectrum Analyzers
Broad Measurement Range
Frequency Range: 9 kHz to 2.1 GHz / 3.2 GHz
Tools for Improved Debugging
●●
Frequency Range from 9 kHz to 3.2 GHz.
More application coverage from a single Spectrum
Analyzer.
●●
-161 dBm/Hz Displayed Average Noise Level (Typ.)
View and measure very small signals.
-98 dBc/Hz @10 kHz Offset Phase Noise
(1 GHz, Typ.)
Improved specification gives more accurate
measurement results.
Optional Tracking Generator
Optional EMI Pre-compliance Test Kit
Optional Reflection Measurement Kit
Make fast and simple transmission measurements
Make EMI Receiver measurements to CISPR 16-1-1
Measure VSWR, Return Loss, Reflection Coefficient
●●
Built-in switchable pre-amplifier.
Integrated pre-amplifier allows higher sensitivity
measurements.
●●
10.1 inch (25.65 cm) color WVGA 1024 x 600 display.
Clear and flexible display aids ease of use.
●●
USB Device, USB Host and LAN support.
Remote control your measurements.
●●
●●
Key Specification
Model
T3SA3200
T3SA3100
Frequency Range
9 kHz ~ 3.2 GHz
9 kHz ~ 2.1 GHz
Resolution Bandwidth
1 Hz ~ 1 MHz, in 1-3-10 sequence
1 Hz ~ 1 MHz, in 1-3-10 sequence
Displayed Average Noise Level
-161 dBm/Hz, Normalize to 1 Hz (typ.)
-161 dBm/Hz, Normalize to 1 Hz (typ.)
Phase Noise
< -98 dBc/Hz@1 GHz, 10 kHz offset
< -98 dBc/Hz@1 GHz, 10 kHz offset
Amplitude Precision
< 0.7 dB
< 0.7 dB
PRODUCT OVERVIEW
Teledyne Test Tools T3SA3000 Spectrum Analyzer range consists of models with frequency ranges
from 9 kHz to 2.1 GHz or 9 kHz to 3.2 GHz. The small footprint and easy user interface is augmented
by a high performance specification with many advanced measurement functions and capabilities.
The high performance Spectrum Analysis capability can
be enhanced further with options to extend it’s measure
ment capability.
●●
●●
●●
●●
Optional Tracking Generator: Adds tracking generator
capabilities to your Spectrum Analyzer to make
transmission measurements on back planes, cables,
filters, amplifiers, etc.
Optional EMI Pre-compliance test kit: Add EMI Receiver
Measurements following CISPR 16-1-1.
Optional Reflection Measurement Kit: Adds VSWR,
Return Loss and Reflection Coefficient measurements.
Optional Advanced Measurement kit: Adds channel
power, adjacent channel power ratio, time domain
power, occupied bandwidth, third order intercept, etc,
to further enhance the measurement capability of your
spectrum analyzer.
Teledyne Test Tools spectrum analyzers offer comprehensive measurement capabilities even in the base units. The
enhancement options support the user when conducting
more complex measurements and make daily measurement tasks easier and faster.
Typical Applications
●●
Research Laboratory
●●
Development Laboratory
●●
Repair and Maintenance
●●
Calibration Laboratory
●●
Automatic Production Test
●●
General bench-top use
2
User-friendly Design
●●
10.1 inch (25.65 cm) 1024*600 display
●●
Intuitive, easy to use menu system
●●
“Preset” and “Auto Tune” for quick set up
●●
Built-in front panel accessible help system
●●
File management (support for U-disc and local storage)
●●
Lightweight, small footprint, easy to transport
Features and Benefits
●●
Frequency Range from 9 kHz up to 3.2 GHz
●●
-161 dBm/Hz Displayed Average Noise Level (Typ.)
●●
-98 dBc/Hz @10 kHz Offset Phase Noise (1 GHz, Typ.)
●●
Total Amplitude Accuracy < 0.7 dB
●●
1 Hz Minimum Resolution Bandwidth (RBW)
●●
All-Digital IF Technology
●●
Standard Preamplifier
●●
Up to 3.2 GHz Tracking Generator Kit (Opt.)
●●
Reflection Measurement Kit (Opt.)
●●
Advanced Measurement Kit (Opt.)
●●
EMI Pre-compliance Test Kit (Opt.)
●●
10.1 lnch WVGA (1024 x 600) Display
DESIGN FEATURES
Supports four independent traces
and cursors
-151 dBm Displayed Average Noise Level
(RBW = 10 Hz)
1 Hz Minimum Resolution Bandwidth (RBW)
Zero span and demodulation capabilities
Phase noise -98 dBc/Hz @ 1 GHz,
offset 10 kHz
On-screen VSWR/Return Loss
measurements with the reflection
measurement option (T3SA3000-RFM)
EMI filter and Quasi-Peak detector
following CISPR 16-1-1 ( T3SA3000-EMI)
Advanced measurement kit (T3SA3000-ADM)
includes on-screen ACPR measurements
3
SPECIFICATIONS
Specifications are valid under the following conditions: The instrument
is within the calibration period, has been stored between 0 and
50°C for at least 2 hours prior to use, and has been powered on and
warmed up for at least 40 minutes. The specifications include the
measurement uncertainty, unless otherwise noted.
Specifications: All products are guaranteed to meet published
specifications when operating in temperatures from 5 to 45 °C, unless
otherwise noted.
Typical: Performance deemed typical implies that 80 percent of the
measurement results will meet the typical published performance with
a 95th percentile confidence level at room temperature (approximately
25 °C). Typical performance is not warranted and does not include
measurement uncertainty.
Nominal: The expected performance or design attribute.
Frequency Characteristic
T3SA3200
T3SA3100
9 kHz – 3.2 GHz
1 Hz
9 kHz – 2.1 GHz
1 Hz
0 Hz, 100 Hz to 3.2 GHz
± Span / (number of sweep points - 1)
0 Hz, 100 Hz to 2.1 GHz
Frequency
Frequency range
Frequency resolution
Frequency Span
Range
Accuracy
Internal Reference Source
Reference frequency
Frequency reference accuracy
Initial calibration accuracy
Temperature stability
Frequency aging rate
10.000000 MHz
± [(time since last adjustment × frequency aging rate) + temperature stability + calibration accuracy]
< 1 ppm
< 1 ppm/year, 0 °C ~ 50 °C
< 0.5 ppm/first year, 3.0 ppm/20 years
Marker
Marker resolution
Marker uncertainty
Frequency counter resolution
Frequency counter uncertainty
Span / (number of sweep points - 1)
± [frequency indication × frequency reference uncertainty + 1 % × span + 10 % × resolution
bandwidth + marker resolution]
1 Hz
± [frequency indication × frequency reference accuracy + counter resolution]
Bandwidths
Resolution bandwidth (-3 dB)
Resolution filter shape factor
RBW uncertainty
Video bandwidth (-3 dB)
VBW uncertainty
1 Hz ~ 1 MHz 1), in 1-3-10 sequence
< 4.8:1 (60 dB:3 dB), Gaussian-like
< 5 %
1 Hz ~ 3 MHz, in 1-3-10 sequence
< 5%
The DANL with RBW set to 1 or 3 Hz will be similar to 10 Hz.
1)
4
SPECIFICATIONS
Amplitude Characteristic
T3SA3200
T3SA3100
Amplitude and Level
Measurement range
DANL to +10 dBm, 100 kHz ~ 1 MHz, preamplifier off DANL to +20 dBm, 1 MHz ~ 3.2 GHz,
preamplifier off
Reference level
-100 dBm to +30 dBm, 1 dB steps Preamplifier
20 dB (nom.), 9 kHz ~ 3.2 GHz Input attenuation
0 ~ 51 dB, 1 dB steps
Maximum input DC voltage
± 50 VDC
Maximum average RF power
30 dBm, 3 minutes, fc ≥ 10 MHz, attenuation > 20 dBm, preamp off Maximum
damage level
33 dBm, fc ≥ 10 MHz, attenuation > 20 dBm, preamp off
Displayed Average Noise Level (DANL)
20 °C ~ 30 °C, attenuation = 0 dB, sample detector, trace average > 50
Preamp off
RBW = 10 Hz
Normalization to 1 Hz
9 kHz ~ 100 kHz
-100 dBm (nom.)
-110 dBm (nom.)
100 kHz ~ 1 MHz
-97 dBm, -101 dBm (typ.)
-107 dBm, -111 dBm (typ.)
1 MHz ~ 10 MHz
-122 dBm, -126 dBm (typ.)
-132 dBm, -136 dBm (typ.)
10 MHz ~ 200 MHz
-127 dBm,-131 dBm (typ.)
-137 dBm, -141 dBm (typ.)
200 MHz ~ 2.1 GHz
-125 dBm, -129 dBm (typ.)
-135 dBm, -139 dBm (typ.)
2.1 GHz ~ 3.2 GHz
-116 dBm, -122 dBm (typ.)
-126 dBm, -132 dBm (typ.)
Preamp on
9 kHz ~ 100 kHz
-107 dBm (nom.)
-117 dBm (nom.)
100 kHz ~ 1 MHz
-122 dBm, -127 dBm (typ.)
-132 dBm, -137 dBm (typ.)
1 MHz ~ 10 MHz
-138 dBm, -144 dBm (typ.)
-148 dBm, -154 dBm (typ.)
10 MHz ~ 200 MHz
-146 dBm, -151 dBm (typ.)
-156 dBm, -161 dBm (typ.)
200 MHz ~ 2.1 GHz
-145 dBm, -148 dBm (typ.)
-155 dBm, -158 dBm (typ.)
2.1 GHz ~ 3.2 GHz
-135 dBm, -139 dBm (typ.)
-145 dBm, -149 dBm (typ.)
Phase Noise
20 °C ~ 30 °C, f c= 1 GHz
Phase Noise