TRUMPower
DESCRIPTION
30-48 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES TMC40 SERIES
T he TMC40 series of compact, open PCB constructed, AC-DC switching power supplies are capable of delivering 30-48 watts of continuous output power at convection cooling. They operate at 90-264 VAC input voltage without the need of voltage selection, and are suited for medical, information technology and industrial applications. Approval to both EN60601-1 and EN60950-1 safety standards improves design-in time and reduces end equipment compliance costs.
FEATURES
M edical and ITE approvals C ompact size 2” x4” x1.18” S ingle, dual and triple outputs F ull-range input 90-264 VAC L ow earth leakage current L evel B emissions R oHS compliant
SAFETY STANDARD APPROVALS
U L 2601-1 CSA C22.2 No. 601.1
T ÜV EN60601-1
INPUT SPECIFICATIONS
Input voltage: Input frequency: Input current: Earth Leakage current: 90-264 VAC 47-63 Hz 0.9 A (rms) for 100 VAC 0.5 A (rms) for 240 VAC 150 µA max. @ 264 VAC, 63 Hz
GENERAL SPECIFICATIONS
Switching frequency: Efficiency: Hold-up time: Line regulation: 62±5 KHz 80-88% typical except TMC40-T31-3 and TMC40-T31-5 at 75% typical 12 msec minimum at 110 VAC ±0.5% maximum at full2±5 KHz 6 load 25 A @ 115 VAC, or 50 A @ 230 VAC, except 80-88% typical at 25 C cold start TMC40-T31-3 and 4000 VAC from input to output TMC40-T31-5 at 75% 1500 VAC from input to ground typical 500 VAC from output to ground MTBF: EMC Performance EN55011 / EN55022: FCC: VCCI: EN61000-3-2: EN61000-3-3: EN61000-4-2: EN61000-4-3: EN61000-4-4: EN61000-4-5: EN61000-4-6: EN61000-4-8: EN61000-4-11: Class B conducted, class B radiated Class B conducted, class B radiated Class B conducted, class B radiated Harmonic distortion, class A and D Line flicker ESD, ±8 KV air and ±6 KV contact Radiated immunity, 3 V/m Fast transient/burst, ±2 KV Surge, ±1 KV diff., ±2 KV com. Conducted immunity, 3 Vrms Magnetic field immunity, 3 A/m Voltage dips, 30% reduction for 500 ms 60% reduction for 100 ms >95% reduction for 10 ms 400,000 hours at full load at 25 C ambient, calculated per MIL-HDBK-217F
o o
OUTPUT SPECIFICATIONS
Output voltage/current: Maximum output power: Ripple and noise: Overvoltage protection: Overcurrent protection: Temperature coefficient: Transient response: See rating chart See rating chart 2% peak to peak on 5.0 V model, 1% peak to peak on other models. Provided on output #1 only; set at 112–132% of its nominal output voltage All outputs protected to short circuit conditions All outputs ±0.04% /°C maximum Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 µs after a 25% step load change
Inrush current: Withstand voltage:
ENVIRONMENTAL SPECIFICATIONS
Operating temperature: Storage temperature: Relative humidity: Derating: -10 C to +70 C -40 C to +85 C 5% to 95% non-condensing o Derate from 100% to +50 C linearly to o 50% at +70 C
o o o o
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com
UNIVERSAL INPUT
OUTPUT VOLTAGE/CURRENT RATING CHART
Output #1 P/N T MC40-S05 T MC40-S12 TMC40-S15 T MC40-S24 TMC40-S48 T MC40-D23 TMC40-D25 T MC40-T31 TMC40-T31-3 T MC40-T31-5 TMC40-T32 T MC40-T39 Vnom. 5V 12 V 15 V 24 V 48 V +5 V +5 V +5 V + 3.3 V +5 V +5 V +5 V Imin. 0A 0A 0A 0A 0A 0.5 A 0.5 A 0.5 A 0.8 A 0.5 A 0.5 A 0.5 A Imax. 8.0 A 3.5 A 3.0 A 2.0 A 1.0 A 6A 6A 6A 6A 6A 6A 6A Tol. 2% 2% 2% 2% 2% 3% 3% 3% 3% 3% 3% 3% +12 V +24 V +12 V +5 V +3.3 V +15 V +24 V Vnom.
TMC40 MEDICAL & ITE SERIES
Maximum Output Tol. Power 40 W 42 W 45 W 48 W 48 W 40 W 40 W 0.3 A 0.3 A 0.3 A 0.3 A 0.3 A 4% 4% 4% 4% 4% 40 W 30 W 30 W 40 W 40 W
Output #2 Imin. (N/A) (N/A) (N/A) (N/A) (N/A) 0.1 A 0.1 A 0.1 A 0.1 A 0A 0.1 A 0.1 A 2A 1A 2A 2A 1.5 A 1.5 A 1A 5% 5% 5% 5% 5% 5% 5% - 12 V +12 V +12 V - 15 V - 12 V Imax. Tol. Vnom.
Output #3 Imin. (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) 0A 0A 0A 0A 0A Imax.
N OTE: 1. Safety approvals are for PCB form only. To order unit with cover fitted, add suffix “C”.
MECHANICAL SPECIFICATIONS
NOTES: 1. 2. 3. Dimensions shown in inches [mm] Tolerance 0.02 [0.5] maximum Input connector mates with Molex housing 09-50-3031 and Molex 2878 series crimp terminal. Output connector mates with Moles housing 09-50-3061 and Moles 2878 series crimp terminal. Ground tab is 0.25 [6.35] x 0.032 [0.8] To ensure compliance with level B emissions, connect the two “*” marked mounting holes with metallic standoffs to chassis. Weight: 205 grams (0.45 lbs.) approx.
4.
5. 6.
7.
PIN CHART
PIN 1 MODEL TMC40-S05 TMC40-S24 T MC40-D23 T MC40-T31 TMC40-S12 TMC40-S48 TMC40-D25 TMC40-T32 TMC40-T39 TMC40-S15 2 3 4 5 6
+V1
+V1
RTN
RTN
N.C
N.C
+V1 +V1 +V1
+V1 +V1 +V1
RTN RTN RTN
RTN RTN RTN
N.C -V3 +V3
+V2 +V2 +V2
T MC40-T31-3 ,TMC40-T31-5
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com
很抱歉,暂时无法提供与“TMC40-D25”相匹配的价格&库存,您可以联系我们找货
免费人工找货