EME05 Series
5W
AC-DC POWER SUPPLIES
CONVECTION
COOLED
The EME05 is a series of open frame and encapsulated AC-DC single output power supplies
designed for medical applications. The series provides two mechanical options including
open frame and encapsulated PCB mount. With approvals to world-wide medical safety
standards, compliance with class B for both conducted and radiated emissions and a 130%,
30s peak load capability, these class II isolation parts benefit system designers with easy
integration into the latest healthcare products and applications.
Features
Applications
• Compact Size
• Medical Approvals (2 x MOPP)
• Single Outputs from 3.3 to 48VDC
Healthcare
• Open Frame & Encapsulated PCB Mount
Home
Healthcare
Medical
Diagnostic
Dimensions
• Class II
• Peak Load Capability
EME05:
• No External Components Required
1.50 x 1.00 x 0.60” (36.1 x 25.4 x 15.2 mm)
• 3 Year Warranty
EME05-P:
1.40 x 0.94 x 0.69” (35.6 x 23.7 x 17.6 mm)
Models & Ratings
Model Number(2)
Output Voltage
Output Current
Nominal
Peak(1)
Output Power
EME05US03
3.3VDC
1510mA
1960mA
5W
EME05US05
5.0VDC
1000mA
1300mA
5W
EME05US09
9.0VDC
555mA
722mA
5W
EME05US12
12.0VDC
416mA
541mA
5W
EME05US15
15.0VDC
333mA
433mA
5W
EME05US24
24.0VDC
208mA
270mA
5W
EME05US36
36.0VDC
138mA
180mA
5W
EME05US48
48.0VDC
104mA
135mA
5W
Notes:
1. Peak load lasting 400
Mean Time Between Failure
W/in3
khrs
0.03 (14)
Weight
2 x MOPP
MIL-HDBK-217F, +25°C GB
Open frame versions (-P)
lb (g)
0.053 (24)
Encapsulated version
Environmental
Characteristic
Minimum
Typical
Maximum
Units
Operating Temperature
-25
+70
°C
Storage Temperature
-40
+85
°C
95
%RH
Notes & Conditions
Derate linearly from 100% at +50°C to 50% at +70°C
Convection-cooled
Cooling
Humidity
3048
Operating Altitude
m
Non-condensing
100%
90%
Shock
IEC68-2-27, 30 g, 11 ms half sine, 3 times in each of 6 axes
Vibration
IEC68-2-6, 2 g, 10Hz to 500kHz, 10 mins/cycle, 60 mins
each cycle
70%
Efficiency
80%
EME05US12-P
115Vac input
230Vac input
60%
50%
Efficiency Graphs
40%
10
20
30
40
50
60
70
80
90
100
Load (%)
EME05US12
EME05US24
100%
100%
90%
EME05US24-P
90%
80%
70%
115Vac input
230Vac input
60%
50%
40%
Efficiency
Efficiency
80%
70%
115Vac input
230Vac input
60%
50%
10
20
30
40
50
60
70
80
90
40%
100
Load (%)
10
20
30
40
50
60
70
80
90
100
Load (%)
100%
90%
Efficiency
80%
70%
115Vac input
230Vac input
60%
50%
40%
10
20
30
40
50
60
70
80
90
100
Load (%)
3
EME05 Series
AC-DC POWER SUPPLIES
Safety Approvals
Certification
Standard
Notes & Conditions
CB
IEC60601-1
UL
ANSI/AAMI ES60601-1:2005 & CSA C22.2, No.60601-1:08
TUV
EN60601-1
CE
Meets all applicable directives
UKCA
Meets all applicable legislation
Means of Protection
Primary to Secondary
Medical
Category
2 x MOPP (Means of Patient Protection)
IEC60601-1 Ed 3.1
EMC: Emissions
Phenomenon
Standard
Test Level
Conducted
EN55011
Class B
Radiated
EN55011
Class B
Harmonic Current
EN61000-3-2
Class A
Voltage Flicker
EN61000-3-3
Notes & Conditions
EMC: Immunity
Phenomenon
Test Level
Criteria
IEC60601-1-2
Ed 4.0: 2014
A
ESD Immunity
EN61000-4-2
±6kV contact,
±15kV air discharge
A
Radiated Immunity
EN61000-4-3
10 V/m
A
EFT
EN61000-4-4
3
A
Surge
EN61000-4-5
2
A
Conducted
EN61000-4-6
10Vrms
A
Magnetic Fields
EN61000-4-8
30A/m
A
70% UT (70 VAC) for 500ms
A
40% UT (40 VAC) for 100ms
A
Medical Device EMC
Standard
EN60601-1-2
(100 VAC)
Dips and Interruptions
EN60601-1-2
(240 VAC)