ISM02 Series
DC-DC Converter
XP Power
2 Watt
• International medical safety approvals
• Single & dual unregulated outputs
• Compact SMD package
• 4kV VAC reinforced isolation
• 1 x MOPP at 300 VAC
• 2 μA patient leakage current
• Operating temperature -25 °C to +105 °C
• Full load to 60°C convection
• Tape & reel package available
Dimensions:
• 3 year warranty
ISM02:
0.94 x 0.71 x 0.36” (24.0 x 18.1 x 9.3 mm)
The ISM02 provides a compact cost effective SMD solution for
healthcare applications providing 1 MOPP, 2 MOOP and IEC60601-1
compliance. Features include a high operating temperature, low leakage
current, reinforced insulation and short circuit protection.
Models & Ratings
Input voltage
5V
(4.5-5.5V)
12V
(10.8-13.2V)
24V
(21.6-26.4V)
Output voltage
Output current
Max
Min
Input current(1)
No load
Full load
Maximum
capacitive load(2)
Efficiency(6)
Model number(3)
5V
400 mA
8.0 mA
605 mA
330 μF
72%
ISM0205S05
12 V
165 mA
3.0 mA
600 mA
330 μF
72%
ISM0205S12
15 V
133 mA
2.5 mA
605 mA
330 μF
72%
ISM0205S15
±12 V
±83 mA
±1.5 mA
555 mA
±100 μF
72%
ISM0205D12
±15 V
±66 mA
±1.0 mA
540 mA
±100 μF
73%
ISM0205D15
5V
400 mA
8.0 mA
255 mA
330 μF
70%
ISM0212S05
12 V
165 mA
3.0 mA
250 mA
330 μF
72%
ISM0212S12
15 V
133 mA
2.5 mA
250 mA
330 μF
72%
ISM0212S15
±12 V
±83 mA
±1.5 mA
225 mA
±100 μF
74%
ISM0212D12
±15 V
±66 mA
±1.0 mA
220 mA
±100 μF
75%
ISM0212D15
5V
400 mA
8.0 mA
125 mA
330 μF
68%
ISM0224S05
12 V
165 mA
3.0 mA
125 mA
330 μF
68%
ISM0224S12
15 V
133 mA
2.5 mA
125 mA
330 μF
68%
ISM0224S15
±12 V
±83 mA
±1.5 mA
110 mA
±100 μF
74%
ISM0224D12
±15 V
±66 mA
±1.0 mA
110 mA
±100 μF
75%
ISM0224D15
90 mA
40 mA
30 mA
Notes
1. Input currents measured at nominal input voltage.
2. Maximum capacitive load is per output.
3. For optional water washable version add suffix ‘-P’.
1
4. Add suffix ‘-TR’ for tape and reel. MOQ 200 pcs.
5. Standard tube quantity 10 pcs.
6. Measured at maximum load.
www.xppower.com
ISM02 Series
DC-DC Converter
XP Power
Input
Characteristic
Input Voltage Range
Input Filter
Minimum
Typical
4.5
10.8
21.6
Internal Pi type filter
Maximum
5.5
13.2
26.4
9
18
30
Input Surge
Units
VDC
VDC for 1 s
Notes & Conditions
5 V nominal
12 V nominal
24 V nominal
5 V nominal
12 V models
24 V models
Output
Characteristic
Output Voltage
Initial Set Accuracy
Output Voltage Balance
Minimum Load
Line Regulation
Load Regulation
Minimum
5
Typical
Maximum
30
±4.0
±1.0
±1.5
10
Units
VDC
%
%
A
%
%
±5.0
%
5
% deviation
Ripple & Noise
150
mV pk-pk
Short Circuit Protection
Maximum Capacitive Load
Temperature Coefficient
0.5
s
0.02
%/ ˚C
Maximum
Units
%
VAC
VDC
μA
Ω
pF
kHz
W/in3
MHrs
±2.0
±0.1
±1.2
Cross Regulation
Transient Response
3
Notes & Conditions
See Models and Ratings table
For dual output with balanced loads
See Models and Ratings table
From 20% to 100% load. 5V output version is 12% max
On dual output models when one load is varied between
25% and 100% and other is fixed at 100%
Recovery within 1% in less than 250 μs for a 25% load
change.
20 MHz bandwidth. Measured using 0.47 μF ceramic
capacitor.
With auto recovery
See Models and Ratings table
General
Characteristic
Minimum
Efficiency
Isolation: Input to Output
Leakage Current
Isolation Resistance
Isolation Capacitance
Switching Frequency
Power Density
Mean Time Between Failure
Typical
72
4000
6000
2
109
15
80
50
20
100
12.2
2
Notes & Conditions
See Models and Ratings table
60 s, reinforced isolation at 300 VAC, 1 x MOPP/2 x MOOP
1s
240 VAC, 60 Hz
At 500 VDC
MIL-HDBK-217F, +25 °C GB
Lead-Free Reflow Solder Process
IPC/JEDEC J-STD-020D.1
Case Material
Pin Material
Weight
Moisture Sensitivity Level
Non conductive black plastic. UL94V-0 rated. Not suitable for water washing.
Solder-coated phosphor bronze
lb (g)
0.01 (3.75)
IPC/JEDEC J-STD-020D.1
Level 2
Environmental
Characteristic
Maximum
Units
Notes & Conditions
Operating Temperature
Minimum
-40
Typical
+105
°C
See derating curve
Storage Temperature
-50
+125
°C
Case Temperature
Humidity
+105
°C
95
%RH
Cooling
Non-condensing
Natural convection
Safety Approvals
Safety Agency
Safety Standard
Notes & Conditions
CB Medical
IEC60601-1
UL Medical
ANSI AMMI ES60601-1
EN Medical
EN60601-1
UL ITE
UL/cUL60950-1, UL/cUL62368-1
CE
Meets all applicable directives
UKCA
Meets all applicable legislation
www.xppower.com
2
ISM02 Series
DC-DC Converter
XP Power
EMC: Emissions
Phenomenon
Standard
Test Level
Conducted
EN55011
Class A
Notes & Conditions
See Application Notes
EMC: Immunity
Test Level
Criteria
ESD
Phenomenon
EN61000-4-2
Standard
±15 kV air discharge, ±8 kV contact
A
Radiated
EFT/Burst
EN61000-4-3
EN61000-4-4
10 V/m
±2 kV
A
A
Surge
EN61000-4-5
±1 kV
A
Conducted
EN61000-4-6
6kV rms
A
Magnetic Fields
EN61000-4-8
30 A/m
A
Notes & Conditions
Mechanical Details
Connecting Pin Pattern
0.94 (24.0)
0.70 (17.8)
7 8
0.77 (19.5)
1
0.54 (13.7)
10 9
0.65 (16.6)
16
0.71 (18.1)
0.10
(2.5)
0.54 (13.7)
0.60 (15.2)
0.12
(3.1)
0.04 (1.0)
0.06
(1.6)
0.10 (2.5)
0.35 (9.0)
Pin Connections
S
0.36±0.02
(9.25±0.5)
SEATING
PLANE
0~
4
0.01 (0.25)
0.11
(2.9)
0.15 S
Pin
1
7
8
9
10
16
Single
-Vin
No Connection
No Connection
+Vout
-Vout
+Vin
Dual
-Vin
No Connection
Common
+Vout
-Vout
+Vin
Application Notes
Derating Curve
EMC Circuit for Class A
100
16
Output Power (%)
Convection
400 LFM forced air
9
C1
ISM02
10
1
0
-40
-25
60
75
C1
2.2µF/150V
105
Ambient Temperature (°C)
L
88μH
Notes
1. All dimensions are in inches (mm)
2. Weight: 0.01 lbs (3.75 g) approx.
3
3. Tolerance: X.XX±0.01 (X.X±0.25)
X.XXX±0.005 (X.XX±0.13)
4. Pin Tolerance: ±0.002 (±0.05)
www.xppower.com
01 Dec 2021
很抱歉,暂时无法提供与“ISM0205S12”相匹配的价格&库存,您可以联系我们找货
免费人工找货- 国内价格 香港价格
- 1+187.008971+23.19838
- 5+181.412185+22.50410
- 10+175.8143610+21.80969
- 25+168.2794425+20.87499
- 100+157.08546100+19.48638
- 500+140.20296500+17.39212